K203710 · Storz Medical AG · IPF · May 3, 2021 · Physical Medicine
Device Facts
Record ID
K203710
Device Name
Storz Medical MAGNETOLITH Muscle Stimulator
Applicant
Storz Medical AG
Product Code
IPF · Physical Medicine
Decision Date
May 3, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MAGNETOLITH is indicated for - Relaxation of muscle spasms, - Prevention or retardation of disuse atrophy, - Increasing local blood circulation, - Muscle re-education, - Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, - Maintaining or increasing range of motion.
Device Story
MAGNETOLITH delivers high-energy electromagnetic pulses to stimulate muscle tissue via magnetic induction field principle; non-invasive; no direct skin contact required. System comprises control unit, handpiece applicator, applicator holder, mains cable, and trolley. Applicator uses treatment coil to generate electromagnetic induction, creating voltage across electrical conductors in changing magnetic field; water-cooled applicator prevents tissue heating. Used in clinical settings by healthcare professionals. Output consists of electromagnetic pulses; healthcare provider monitors therapy via touch screen interface. Device facilitates muscle contraction to achieve therapeutic goals like spasm relaxation and atrophy prevention.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed device meets design specifications. Software validated as moderate level of concern. Compliance with IEC 60601-1:2012, IEC 60601-1-2:2014, and IEC 60601-1-6:2012 demonstrated. No clinical data provided.
Technological Characteristics
Powered muscle stimulator using electromagnetic induction. Components: control unit, water-cooled handpiece applicator, trolley. Energy source: 100-240 VAC, 50/60 Hz. Magnetic field intensity: 0.4T (surface), 0.08T (coil center). Pulse duration: 125μs ± 20%. Pulse repetition rate: 1-10Hz. Connectivity: touch screen interface. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62304, ISO 14971, ISO 13485.
Indications for Use
Indicated for patients requiring muscle spasm relaxation, prevention/retardation of disuse atrophy, increased local blood circulation, muscle re-education, post-surgical calf muscle stimulation to prevent venous thrombosis, and maintenance/increase of range of motion.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
K160992 — HPM-6000 · BTL Industries, Inc. · Oct 21, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo, which features a stylized eagle. To the right of the eagle is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
May 3, 2021
Storz Medical AG % Michael Dayton President & CEO Biomed Research, Inc. 3959 Van Dyke Road Suite 245 Lutz, Florida 33558
Re: K203710
Trade/Device Name: Storz Medical MAGNETOLITH Muscle Stimulator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, NGX Dated: February 1, 2021 Received: February 3, 2021
Dear Michael Dayton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning vour device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 53 1 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-ind-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Patrick Antkowiak -S
Patrick Antkowiak, PhD Assistant Director (Acting) DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K203710
Device Name
MAGNETOLITH Muscle Stimulator
Indications for Use (Describe)
The MAGNETOLITH Muscle Stimulator is indicated for:
-Relaxation of muscle spasms,
-Prevention or retardation of disuse atrophy,
-Increase local blood circulation.
-Muscle re-education,
-Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis,
-Maintaining or increasing range of motion.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
# Storz Medical AG
### MAGNETOLITH Muscle Stimulator
#### 1. SPONSOR
Storz Medical AG Lohstampfestr. 8 CH-8274 Tägerwilen Switzerland
Contact Person: Pavel Novak, Ph.D. Telephone: 41 (0) 71 677 45 13 FAX: 41 (0) 71 677 45 05
Date Prepared: May 3, 2021
#### DEVICE NAME 2.
Proprietary Name: MAGNETOLITH Regulation Name: Powered Muscle Stimulator
| Regulation Number: | 21 CFR 890.5850 |
|--------------------|-----------------|
| Product Code: | IPF, NGX |
| Regulatory Class: | II |
#### 3. PREDICATE DEVICES
Equivalence is claimed to the following predicate devices: HPM-6000 (K160992, primary), and Bemer Classic & Pro (K151834, secondary).
#### 4. DEVICE DESCRIPTION
The MAGNETOLITH delivers electromagnetic pulses to stimulate muscle tissues by applying the Magnetic Induction Field Principle (high-energy magnetic pulses) directly over (but not in direct contact with) the patient in order to achieve the same powered muscle stimulation without having to place adhesive electrode pads directly onto the patient's body. Key components of the device are the Control Unit, Handpiece Applicator, Applicator holder, a Mains cable, and a Trolley. These device components are provided non-sterile and are intended to be reusable.
The principle of operation of the MAGNETOLITHis by the application of a treatment coil to generate electromagnetic induction. i.e., the production of voltage across an
{4}------------------------------------------------
electrical conductor in a changing magnetic field. The duration of the individual pulses is short and therefore the system does not increase temperature within the tissues. In addition, the MAGNETOLITH Applicator is cooled by circulating water.
#### 5. INTENDED USE
The MAGNETOLITH is indicated for
- · Relaxation of muscle spasms,
- Prevention or retardation of disuse atrophy, .
- · Increasing local blood circulation,
- · Muscle re-education,
- Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis, .
- Maintaining or increasing range of motion. .
#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The technological characteristics of the MAGNETOLITH device and the predicate device is substantially equivalent in that each device is designed to non-invasively use electromagnetic induction to initiate action potential of nerves resulting in muscle contraction.
#### 7. PERFORMANCE TESTING
Verification and validation testing was performed and demonstrated that the MAGNETOLITH meets the design specifications and is safe and effective for its intended use. All tests required by the verification and validation plan were completed and passed.
The MAGNETOLITH software was validated and demonstrated to be of a moderate level of concern; while hazard analysis / risk management was performed and demonstrated that all risks are mitigated to an acceptable level. While the MAGNETOLITH has cleaning and sanitation requirements, there are no body contacting components and therefore biocompatibility testing was not performed. The MAGNETOLITH was tested and demonstrated to conform to the general safety requirements of IEC 60601-1:2012; IEC 60601-6:2013; and electromagnetic compatibility requirements of IEC 60601-1-2:2014.
No performance standards applicable to this device have been adopted under Section 514 of the Act. The MAGNETOLITH complies with the applicable requirements of the following international consensus standards:
- ISO 14971:2000/A1:2007: Medical devices: Application of risk management to . medical devices.
{5}------------------------------------------------
- IEC 60601-1:2012, (Ed. 3.1): Medical electrical equipment Part 1: General ● requirements for basic safety and essential performance.
- IEC 60601-1-2: 2014(4th Ed.): Medical electrical equipment Part 1-2: General . requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
- . IEC 60601-1-6: 2012 (3.1 Ed.): Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
- IEC 62304:2015: Medical Device Software - Software Life CycleProcess.
- ISO 13485:2016: Medical devices Ouality management systems -. Requirements for regulatory purposes.
The performance testing demonstrated that the MAGNETOLITH Muscle Stimulator is substantially equivalent to the predicate devices and that it is as safe, as effective, and performs as well as or better than the predicate devices. The table below compares the characteristics of the MAGNETOLITH to the predicate devices.
| 510(k) number<br>Device name<br>Company name | K160992<br>HPM-6000<br>BTL Industries, Inc. | K151834<br>Bemer Classic & Pro<br>Bemer Int'l AG | K203710<br>MAGNETOLITH<br>Storz Medical AG | Comparison |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Code and<br>Regulation | Physical Medicine<br>21 CFR 890.5850<br>IPF - Stimulator,<br>Muscle, Powered | Physical Medicine<br>21 CFR 890.5850<br>NGX - Stimulator, Muscle,<br>Powered | Physical Medicine<br>21 CFR 890.5850<br>IPF - Stimulator,<br>Muscle, Powered<br>NGX - Stimulator, Muscle,<br>Powered | Same |
| Intended Use /<br>Indications for Use | The HPM-6000 device is intended<br>to be used under medical<br>supervision for adjunctive therapy<br>for the treatment of medical<br>diseases and conditions.<br>The HPM-6000 device is indicated<br>for use in stimulating<br>neuromuscular tissue for bulk<br>muscle excitation in the legs or<br>arms for rehabilitative purposes.<br>Indications for Use for Muscle<br>Stimulators:<br>-Relaxation of muscle spasms<br>-Prevention or retardation of disuse<br>atrophy<br>-Increasing local blood circulation<br>-Muscle re-education<br>-Immediate post- surgical<br>stimulation of calf muscles to<br>prevent venous thrombosis<br>-Maintaining or increasing range<br>of motion | Indications for Use:<br>-To temporarily increase local<br>blood circulation in healthy leg<br>muscles<br>-To stimulate healthy muscles in<br>order to improve and facilitate<br>muscle performance | Indications for Use:<br>-Relaxation of muscle spasms<br>-Prevention or retardation of<br>disuse atrophy<br>-Increasing local blood<br>circulation<br>-Muscle re-education<br>-Immediate post- surgical<br>stimulation of calf muscles to<br>prevent venous thrombosis<br>-Maintaining or increasing range<br>of motion<br>-To stimulate healthy muscles in<br>order to improve and facilitate<br>muscle performance | Similar |
#### 8. SUBSTANTIAL EQUIVALENCE COMPARISON
{6}------------------------------------------------
| Principle of Action | Non- invasive tissue stimulation<br>via magnetic field induction | Non- invasive tissue stimulation<br>via magnetic field induction | Non- invasive tissue stimulation<br>via magnetic field induction | Same |
|--------------------------------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------|------------------------------------------------------------------|----------|
| Clinical Use | Prescription use | Over the counter use | Prescription use | Same |
| Electrical<br>Protection | Class II, BF | unknown | Class I, B | Similar¹ |
| User<br>Interface | Touch screen | Touch screen | Touch Screen | Same |
| Firmware<br>Controlled | Yes | Yes | Yes | Same |
| Type of<br>Energy | Magnetic field | Magnetic field | Magnetic field | Same |
| Number<br>of outputs | 1 | 2 | 1 | Same |
| Number of<br>Magnetic Coils in<br>the Applicator | 1 | unknown | 1 | Same |
| Magnetic Field<br>Intensity | BTL 299-1 applicator: 0.5 - 1.8 T<br>BTL 299-2 applicator: 0.7 - 2.5 T | 35-100 μT | 0.4T ±20% at surface<br>0.08T ±20% at center of coil | Similar |
| Pulse Repetition<br>Rate | 1 - 150 Hz | 10-30 Hz | 1-10Hz | Similar |
| Pulse Duration | BTL 299-1 applicator: 280μs ±<br>20%<br>BTL 299-2 applicator: 280μs ±<br>20% | 10-33μs | 125μs ± 20% | Similar |
| Therapy Time | Up to 60 min | 20 minutes | 10-20 minutes | Similar |
| Energy Source | 100 - 240 VAC,<br>50-60 Hz | 100 - 240 VAC,<br>50/60 Hz | 100 - 240 VAC,<br>50/60 Hz | Same |
| System Dimensions<br>(W×H×D) | 500×970×580 mm (20×38×23 in) | 320 x 320 x 70 mm | 454 x 187 x 460 mm | Similar |
| Ambient<br>Temperature | 10° - 30°C | 10° - 30°C | 10° - 30°C | Same |
| Relative Humidity | 30 - 75% | unknown | 5-55% | Similar |
| Environmental<br>Specifications | For indoor use only | For indoor use only | For indoor use only | Same |
'Electrical Protection: The electrical security measures in devices are slightly different, but both comply with theIEC60601-1 standard for basic safety and essential performance.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.