K241601 · Cpmt Laser (Canadian Pioneer Medical Technology Corporation) · NGX · Feb 25, 2025 · Physical Medicine
Device Facts
Record ID
K241601
Device Name
EMS (FlexPulse, MagnaCore, Magnetika)
Applicant
Cpmt Laser (Canadian Pioneer Medical Technology Corporation)
Product Code
NGX · Physical Medicine
Decision Date
Feb 25, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is indicated to be used for: · Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen. - · Strengthening, Toning and Firming of buttocks and thigh
Device Story
Electromagnetic Stimulator Device (FlexPulse, MagnaCore, MagnetiKa) is a non-invasive, powered muscle stimulator. System comprises main unit (power supply, control unit with LCD, cooling unit) and handpiece tools containing electromagnetic induction coils and cooling fans. Device generates electromagnetic field interacting with human tissues to initiate nerve action potentials, resulting in muscle contraction. Operated in clinical settings; hands-free application via fixation belts. Two outputs allow simultaneous treatment. Healthcare providers select parameters (intensity, time) via touch screen. Output affects muscle conditioning; benefits include improved tone and strength in abdomen, buttocks, and thighs.
Clinical Evidence
Bench testing only. Device performance verified against design specifications and safety standards including IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), IEC 60601-2-10 (nerve/muscle stimulators), IEC 60601-1-6 (usability), ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation/sensitization). Software verification and validation testing performed per FDA guidance.
Technological Characteristics
Powered muscle stimulator; magnetic field energy source. 15.6-inch touch screen interface. Magnetic field intensity 0.5-1.8T. Symmetrical biphasic sine wave pulse; 1-150 Hz repetition rate; 270±20% μs pulse duration. Two outputs. Biocompatibility per ISO 10993-5/10. Standards: IEC 60601-1, 60601-1-2, 60601-1-6, 60601-2-10.
Indications for Use
Indicated for improvement of abdominal tone, strengthening of abdominal muscles, development of firmer abdomen, and strengthening, toning, and firming of buttocks and thighs.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
February 25, 2025
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.
CPMT LASER (Canadian Pioneer Medical Technology Corporation) Dr. Rashid Reza Mir Sayah Managing Director 460 Garyray Drive, North York Toronto, ON M9L1P8 Canada
Re: K241601
Trade/Device Name: EMS (FlexPulse, MagnaCore, Magnetika) Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: June 4, 2024 Received: June 4, 2024
Dear Dr. Sayah:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}------------------------------------------------
# Sincerely, Tushar Bansal -S
for Heather Dean, PhD
Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K241601
Device Name
Electromagnetic Stimulator Device (FlexPulse, MagnaCore, MagnetiKa)
Indications for Use (Describe)
The device is indicated to be used for:
· Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.
- · Strengthening, Toning and Firming of buttocks and thigh
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) summary
# Submitter:
CPMT Laser (Canadian Pioneer Medical Technology Corporation) 460 Garyray Drive #1, North York, Toronto, ON, M9L 1P8, Canada Contact Person: Dr. Rashid Reza Mir Sayah Phone: (437) 772-7788 Email: canadianpioneer@yahoo.com Date Prepared: February 10, 2025
## Subject Device:
Trade/Device Name: Electromagnetic Stimulator Device Models: FlexPulse, MagnaCore, MagnetiKa Regulation Number: 21CFR 890.5850 Regulation Name: Stimulator, Muscle, Powered, For Muscle Conditioning Regulatory Class: II Product code: NGX 510(k) number: K241601 Manufacturer: CPMT Laser (Canadian Pioneer Medical Technology Corporation)
#### Predicate Device:
Trade/Device Name: BTL 799-2 Regulation Number: 21CFR 890.5850 Regulation Name: Stimulator, Muscle, Powered, For Muscle Conditioning Regulatory Class: II Product code: NGX 510(k) number: K180813 Manufacturer: BTL Industries, Inc.
#### Device Description:
The Electromagnetic Stimulator Device, with models FlexPulse, MagnaCore, and MagnetiKa, consists of a main unit, handpiece tools, and a power cable. The main unit has a power supply unit, a control unit, and a cooling unit. The control unit has a control element and a liquid crystal display. The handpiece tools have electromagnetic induction coils and cooling fans, it is a non-invasive device.
The device generates an electromagnetic field that interacts with human body tissues, aiding in muscle stimulation to strengthen and tone the abdomen, buttocks, and thighs. With two outputs, it allows simultaneous treatment using two applicators.
{5}------------------------------------------------
#### Indications for Use:
The device is indicated to be used for:
- Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.
- Strengthening, Toning and Firming of buttocks and thighs.
## Comparison of Technological Characteristics with the Predicate Device:
#### Substantial equivalence discussion
The comparison and discussion between the subject device and the predicate device is listed in the table below:
| | Subject Device | Predicate Device | Reference<br>Device | Comparison |
|----------------------|------------------------------|------------------------------|---------------------------------|-------------------------|
| 510(K) | K241601 | K180813 | K190456 | - |
| Manufacturer | CPMT Laser | BTL Industries,<br>Inc. | BTL Industries,<br>Inc. | - |
| Trade Name | EMS | BTL 799-2 | BTL 799-2L | - |
| Regulation<br>number | 21 CFR 890.5850 | 21 CFR<br>890.5850 | 21 CFR<br>890.5850 | Identical (the<br>same) |
| Regulation<br>Name | Powered muscle<br>stimulator | Powered muscle<br>stimulator | Powered<br>muscle<br>stimulator | Identical (the<br>same) |
| Product code | NGX | NGX | NGX | Identical (the<br>same) |
| Class | II | II | II | |
#### Table: Substantial equivalence comparison
{6}------------------------------------------------
| Indications for<br>use/Intended<br>use | The device is<br>indicated to be<br>used for: | The device is<br>indicated to be<br>used for: | The device is<br>indicated to be<br>used for: | |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| | • Improvement of<br>abdominal tone,<br>strengthening of<br>the abdominal<br>muscles,<br>development of<br>firmer abdomen.<br>• Strengthening,<br>Toning and<br>Firming of buttocks<br>and thighs. | • Improvement of<br>abdominal tone,<br>strengthening of<br>the abdominal<br>muscles,<br>development of<br>firmer abdomen.<br>• Strengthening,<br>Toning and<br>Firming of<br>buttocks and<br>thighs. | • Improvement<br>of abdominal<br>tone,<br>strengthening<br>of the<br>abdominal<br>muscles,<br>development of<br>firmer<br>abdomen.<br>• Strengthening,<br>Toning<br>and Firming<br>of buttocks<br>and<br>thighs. | Identical |
| Primary Function | Muscle stimulation | Muscle stimulation | Muscle stimulation | Identical |
|------------------------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------|------------------------------------------------------------------------------------|-----------|
| Principal of Action | Initiating action potential of nerves results in muscle contraction | Initiating action potential of nerves results in muscle contraction | Initiating action potential of nerves results in muscle contraction | Identical |
| User Interface | Touch screen | Touch screen | Touch screen | Identical |
| Touch Screen size | 15.6 inches | 15.6 inches | 15.6 inches | Identical |
| Firmware Controlled | Yes | Yes | Yes | Identical |
| Type of Energy | Magnetic field | Magnetic field | Magnetic field | Identical |
| Magnetic Field Intensity (on the coil surface) | 0.5-1.8T,±20% | 299-6 applicator:<br>0.5–1.8 T | BTL 299-6 applicator: 0.5 - 1.8 T ,±20%<br>BTL 299-7 applicator: 0.7 - 2.0 T ,±20% | Identical |
| Number of Outputs | 2 | 2 | 2 | Identical |
{7}------------------------------------------------
| Maximum Magnetic Field Intensity at Applicator Center Surface | 1.0T±20% | BTL 299-6 applicator: 1.154 T±20%; | BTL 299-6 applicator: 1.154 T±20%;<br>BTL 299-7 applicator: 1.173 T±20%. | The maximum field intensity at the center surface of the applicator is 1.0T, in comparison to 1.154T for the predicate device. Nonetheless, the magnetic field intensity falls within the acceptable range of the predicate device. |
|---------------------------------------------------------------|--------------------------------|------------------------------------|------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Number of Magnetic Coils in the Applicator | 1 | 1 | 1 | Identical |
| Type of Operation | Continuous | Continuous | Continuous | Identical |
| Pulse Repletion Rate | 1 – 150 Hz | 1 – 150 Hz | 1 – 150 Hz | Identical |
| Pulse Duration | $270 \pm 20\% \mu s$ | $280 \pm 20\% \mu s$ | BTL 299-6<br>applicator: $280 \pm 20\% \mu s$<br>BTL 299-7<br>applicator: $190 \pm 20\% \mu s$ | The subject device has a slightly shorter pulse width compared to the predicate device but falls within the range between the predicate and reference devices. |
| Pulse Amplitude | 0 – 100%, step 1% | 0 – 100% | Not publicly available | Identical |
| Selection of<br>parameters<br>(Intensity, Time) | Yes | Yes | Yes | Identical |
| Shape of<br>Stimulation<br>Pulse | Symmetrical Biphasic Sine Wave | Symmetrical Biphasic Sine Wave | Symmetrical Biphasic Sine Wave | Identical |
| Induced Current<br>in the Tissues | 20.6mA | 28-30mA | Not Publicly Available | The current induced in the tissue by the subject device is lower than that of the predicate device, making it safer while achieving the same effect. |
{8}------------------------------------------------
| Therapy Time | Up to 60 min | Up to 60 min | Up to 60 min | Identical |
|---------------|-----------------------------------------------------|-----------------------------------------------------|-----------------------------------------------------|-----------|
| Application | Hands-free,<br>applicator fixed by<br>fixation belt | Hands-free,<br>applicator fixed by<br>fixation belt | Hands-free,<br>applicator fixed by<br>fixation belt | Identical |
| Energy Source | 100–240 VAC, 50–60 Hz | 100–240 VAC, 50–60 Hz | 100–240 VAC, 50–60 Hz | Identical |
{9}------------------------------------------------
| System<br>Dimensions<br>(WxHxD) | 525x433x1279mm/<br>640×560×1390mm | 500×1380×580 mm<br>(20x55x23 in) | 500×1380×580 mm<br>(20x55x23 in) | The varying<br>dimensions do<br>not impact the<br>safety or<br>effectiveness of<br>the device. |
|---------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Environmental<br>Specifications | For indoor use<br>only | For indoor use<br>only | For indoor use<br>only | Identical |
| Ambient<br>Temperature | -20°C ~ +55°C | -10°C to +55°C | -10°C to +55°C | This variation<br>does not affect<br>the safety or<br>effectiveness of<br>the device. |
| Conformance<br>Standard | IEC 60601-1; IEC<br>60601-1-2; IEC<br>60601-1-6; IEC<br>60601-2-10 | IEC 60601-1;<br>IEC 60601-1-2;<br>IEC 60601-1-6;<br>IEC 60601-2-10 | IEC 60601-1;<br>IEC 60601-1-2;<br>IEC 60601-1-6;<br>IEC 60601-2-10 | Identical |
| Biocompatibility | ISO 10993-5 ISO<br>10993-10 | ISO 10993-5 ISO<br>10993-10 | ISO 10993-5<br>ISO 10993-10 | Identical |
# Summary of Non-Clinical Tests:
Non clinical tests were conducted to verify that the subject devices met all design specifications and was substantially equivalent (SE) to the predicate device. The test
{10}------------------------------------------------
results demonstrated that the proposed device complies with the following:
● ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010((R)2012 (Consolidated Text) Medical Electrical Equipment -- Part 1: General Requirements for Basic Safety and essential performance
● IEC 60601-1-2:2014 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
IEC60601-2-10:2016, Medical electrical equipment -- Part 2-10: Particular requirements for the basic safety and essential performance of nerve and muscle stimulators
● IEC 60601-1-6: 2013, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
Software Verification and Validation Testing:
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software was considered as a "moderate" level of concern.
#### Conclusion
The EMS device is considered to have substantial equivalence to its predicate device. The intended uses of the proposed device are included within the scope of the predicate device. All three models fall under the umbrella of the predicate device.
Non-clinical testing has shown that the device is equally safe, effective, and is substantially equivalent to the legally marketed device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.