ZX2 Laser System

K150466 · Lzr7, Inc. · ILY · Jun 8, 2015 · Physical Medicine

Device Facts

Record IDK150466
Device NameZX2 Laser System
ApplicantLzr7, Inc.
Product CodeILY · Physical Medicine
Decision DateJun 8, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ZX2 Laser System is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm; the temporary increase in local blood circulation; and / or the temporary relaxation of muscle.

Device Story

Non-invasive, low-energy infrared therapeutic laser system; comprises AC control unit and handheld laser probes (808 nm, 658 nm, 904 nm diodes). Control unit features dual independent probe outputs, user interface, key-activated switch, and emergency stop. Used in clinical settings by practitioners to provide topical heating; elevates tissue temperature to 40-45°C. Heat sensor provides automatic shut-off at >44°C. Benefits include temporary pain relief, increased local blood circulation, and muscle relaxation.

Clinical Evidence

Bench testing only. Temperature vs. time studies conducted on three human subjects (knee, shoulder, neck, low back) demonstrated the device achieves and maintains a therapeutic heat range of 40-45°C within 3-4 minutes. Safety testing confirmed compliance with EN 60601-1, EN 60601-1-2, and EN 60601-2-22.

Technological Characteristics

Infrared lamp (21 CFR 890.5500); GaAlAs, GaAs, and GaInP semiconductor laser diodes (658 nm, 808 nm, 904 nm). AC-powered control unit with dual independent probe outputs. Includes integrated heat sensor for automatic shut-off at 44°C. Compliant with EN 60601-1, EN 60601-1-2, and EN 60601-2-22 standards.

Indications for Use

Indicated for patients requiring temporary relief of minor muscle/joint pain, stiffness, minor arthritis pain, or muscle spasm; and temporary increase in local blood circulation or muscle relaxation.

Regulatory Classification

Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing to the right, with flowing lines suggesting movement or connection. The profiles are arranged in a way that they appear to be overlapping. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the design. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 8, 2015 LZR7 Incorporated % Ms. M. Joyce Heinrich Texas Applied Biomedical Services 12101 Cullen Boulevard, Suite A Houston, Texas 77047 Re: K150466 Trade/Device Name: ZX2 Laser System Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: ILY Dated: May 8, 2015 Received: May 11, 2015 Dear Ms. Heinrich: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150466 Device Name ZX2 Laser System ### Indications for Use (Describe) The ZX2 Laser System is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffits pain, or muscle spasm; the temporary increase in local blood circulation; and / or the temporary relaxation of muscle. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|---------------------------------------------| | X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Exhibit 2. 510(k) Summary (as per 21 CFR 807.92) I. | GENERAL INFORMATION | | |-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Generic Name: | Infrared Laser System | | Trade Name: | ZX2 LASER SYSTEM | | Device Classification: | Class II, Performance Standards -<br>21CFR Part 890.5500 - Infrared Lamp | | Product Code: | ILY | | Sponsor Name and Address: | LZR 7, Inc.<br>1980 Broadway<br>Placerville, CA | | Applicant Name and Address: | M. Joyce Heinrich<br>Texas Applied Biomedical Services, Inc.<br>12101 Cullen Blvd, Suite A<br>Houston, Texas 77047<br>Telephone:<br>713 / 734-4433<br>Facsimile:<br>866 - 362-3968<br>Email:<br>jheinrich4@comcast.net | 510(k) Number: K150466 #### II. Device Description The ZX2 Laser System is a non-invasive, low energy, infrared therapeutic medical laser. The ZX2 Laser incorporates an AC Control Unit and a variety of hand-held laser probes. The Control Unit has two (2) completely independent probe outputs which allow use of the device by two operators at once and/or treatment using two probes at once. The Control Unit houses the power supply, electronics and user interface panel. The Control Unit is connected to a suitable grounded outlet. The Control Unit is equipped with a key which when inserted into the switch the Unit can be activated. The ZX2 has an emergency stop button that, when pressed, will deactivate the device. The device houses a split screen panel on the face of the control unit and dual control buttons for use with two independent probes. {4}------------------------------------------------ The hand-held probes that are optional with the ZX2 Laser System include: - 1. 750 mW Deep Probe 808 nm GaAlAs Semiconductor laser - 2. 1.5 W Deep Probe 808 nm GaAlAs Semiconductor laser - 3. 6.1 W Multi Probe 808 nm GaAlAs Semiconductor laser / 658 nm Ga Aa InP - 4. SP4 Probe 904 nm Ga As Semiconductor Laser Pulsed The above laser probes incorporate laser diode/s at 658nm, 808nm and 904nm diodes, classifying it under 21 CFR 890.5500, ILY infrared lamp for adjunctive use in pain therapy. #### III. Indications for Use The ZX2 Laser System is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm; the temporary increase in local blood circulation; and / or the temporary relaxation of muscle. #### IV. Predicate Devices The ZX2 Laser is substantially equivalent to other infrared therapeutic devices that are currently in commercial distribution. The primary predicate is Apollo Physical Therapy Products International's Apollo IR Heat Lamp System (K060134). #### Summary of the Technical Characteristics of the ZX2 Laser as Related to V. the Referenced Predicate Device. The ZX2 Laser System and the aforementioned predicate devices are infrared lamps as defined in 21 CFR 890.5500. These devices utilize infrared laser diodes, and visible laser diodes to generate topical heating for the purpose of elevating tissue temperatures for temporary relief of muscle and joint pain. #### VI. Electrical and Safety Testing Electrical safety and functional performance testing were conducted on the ZX2 Laser demonstrating that the device is compliant with FDA recognized consensus standards. These standards include the following international standards: - 1. EN 60601-1: 2006 Ed. 3 Medical Electrical Equipment, Part 1, General Requirements for Safety - 2. EN 60601-1-2:2012 Medical Electrical Equipment, General Requirement for Safety. Electromagnetic Compatibility - 3. EN 60601-2-22:1996 Medical Laser Equipment {5}------------------------------------------------ #### Non-clinical Testing VII. ## THERAPEUTIC HEAT TEMPERATURE DATA The ZX2 Laser System is capable of achieving therapeutic heat temperature range of 40 - 45 degrees centigrade as accepted by the FDA. An increase in topical heating of the tissue level by at least 5° centigrade was reached within 3 - 4 minute(s) demonstrated in the bench testing that was conducted on the treatment head. - The therapeutic temperature range was maintained until the device was turned off. The achieved therapeutic temperature range was maintained for the 10 minute testing time. The spot size for the devices used in the Temperature vs. Time studies was 9.35 cm2. The use of light energy to generate heat for therapeutic use has been well documented and is generally accepted alternative treatment modality for the temporary relief of pain and tissue repair. The ZX2 Laser System emits infrared light energy to provide topical heating for the purpose of elevating tissue temperature to safely increase local blood flow and circulation where heat is applied. These devices have been used in Europe for over 10 years demonstrating clinical efficacy and benefit with no reported adverse effects. The average ambient temperature at the beginning of the testing was approximately 19 degrees centigrade. The temperature versus time testing was performed on various body areas: i.e., knee, shoulder, neck and low back and on three (3) different individuals. The data were analyzed and an average of the temperature for each body area is summarized. The pre-exposed topical skin temperature ranged from 31 to 35 degrees centigrade. The topical temperature during exposure following brief stabilization time ranged from 27 to 33 degrees centigrade. These data demonstrate the ZX2 Laser System meets the generally accepted topical temperature range for therapeutic heat of 40 - 45 degrees centigrade during the recommended treatment time of 10 minutes. Representative results of the Temperature vs. Time tests are summarized in Tables 1 - 3 below. To ensure the topical skin temperature does not exceed the maximum acceptable therapeutic heat temperature, the ZX2 incorporates a heat sensor that has an automatic shut off once the device exceeds a temperature of 44 degrees centigrade. {6}------------------------------------------------ | Table 1. | | | | |--------------------------------------------|-----------|-------------------------------|-----------| | Sex: M Age: 52 Ethnicity: White | | | | | | | Room Temperature: end = 19 °C | | | Room Temperature: start = 19 °C | | | | | Device: 6.1W Multi - Average Power: 6100mW | | Warming time: 2 minutes | | | | | | | | Low Back | | Knee | | | Time (min) | Temp (°C) | Time (min) | Temp (°C) | | initial | 35 | initial | 32 | | 1 | 33 | 1 | 29 | | 2 | 36 | 2 | 33 | | 3 | 40 | 3 | 36 | | 4 | 41 | 4 | 39 | | 5 | 43 | 5 | 40 | | 6 | 45 | 6 | 42 | | 7 | 45 | 7 | 44 | | 8 | 46 | 8 | 44 | | 9 | 46 | 9 | 45 | | 10 | 47 | 10 | 46 | | Mean (Active): | | Mean (Active): | | | | | | | | | | | | | | | | | | anie | | |------|--| |------|--| | Sex: F Age: 44 Ethnicity: White | | Room Temperature: end = 21 °C | | |--------------------------------------------|-----------|-------------------------------|-----------| | Room Temperature: start = 21 °C | | Warming time: 2 minutes | | | Device: 6,1W Multi - Average Power: 6100mW | | | | | Neck | | Knee | | | Time (min) | Temp (°C) | Time (min) | Temp (°C) | | initial | 34 | initial | 31 | | 1 | 33 | 1 | 27 | | 2 | 35 | 2 | 31 | | 3 | 39 | 3 | 35 | | 4 | 41 | 4 | 40 | | 5 | 42 | 5 | 41 | | 6 | 44 | 6 | 42 | | 7 | 45 | 7 | 43 | | 8 | 45 | 8 | 44 | | 9 | 46 | 9 | 44 | | 10 | 46 | 10 | 45 | | Mean (Active): | | Mean (Active): | | {7}------------------------------------------------ | Table 3. | | | | |--------------------------------------------|-----------|-------------------------------|-----------| | Sex: M Age: 41 Ethnicity: White | | Room Temperature: end = 17 °C | | | Room Temperature: start = 16 °C | | Warming time: 2 minutes | | | Device: 6,1W Multi - Average Power: 6100mW | | | | | Shoulder | | Low Back | | | Time (min) | Temp (°C) | Time (min) | Temp (°C) | | initial | 32 | initial | 31 | | 1 | 31 | 1 | 31 | | 2 | 35 | 2 | 34 | | 3 | 38 | 3 | 38 | | 4 | 41 | 4 | 40 | | 5 | 42 | 5 | 42 | | 6 | 43 | 6 | 44 | | 7 | 45 | 7 | 45 | | 8 | 44 | 8 | 45 | | 9 | 45 | 9 | 46 | | 10 | 46 | 10 | 46 | | Mean (Active): | | Mean (Active): | | {8}------------------------------------------------ # VIII. Conclusions By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device may have the same intended use and different technological characteristics if they can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regards its safety and effectiveness as compared to the predicate device. The ZX2 Laser has the same intended uses, technical, functional and performance characteristics as the predicate device listed above. The ZX2 Laser is designed to comply with applicable performance standards promulgated by Federal Food and Drug Administration.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...