THOR VR SINGLE DIODE LASER TREATMENT PROBE, MODEL 200 MW SINGLE VR DIODE PROBE

K070024 · Thor International · ILY · May 16, 2007 · Physical Medicine

Device Facts

Record IDK070024
Device NameTHOR VR SINGLE DIODE LASER TREATMENT PROBE, MODEL 200 MW SINGLE VR DIODE PROBE
ApplicantThor International
Product CodeILY · Physical Medicine
Decision DateMay 16, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The THOR VR Single Diode Laser Treatment Probe is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, the temporary increase in local blood circulation and/or temporary relaxation of muscles.

Device Story

THOR VR Single Diode Laser Treatment Probe is a non-invasive, hand-held therapeutic device; delivers continuous infrared energy to skin surface. System consists of control unit housing electronics and handheld treatment probe. Device elevates tissue temperature to provide therapeutic heat. Used in clinical or home settings; operated by healthcare providers or patients. Output is infrared energy; clinical effect is temporary pain relief, increased circulation, and muscle relaxation.

Clinical Evidence

Bench testing only. Functional performance and electrical safety testing conducted in accordance with applicable standards for infrared lamp medical devices.

Technological Characteristics

Infrared lamp (21 CFR 890.5500); utilizes infrared and visible laser diodes for topical heating; hand-held probe form factor; non-invasive; continuous heat therapy; electrical energy source.

Indications for Use

Indicated for patients requiring topical heating for temporary relief of minor muscle and joint pain, stiffness, minor arthritis pain, muscle spasm, increased local blood circulation, and muscle relaxation.

Regulatory Classification

Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # APPENDIX B MAY 16 2007 ## 510(k) PREMARKET NOTIFICATION SUMMARY (per 21 CFR 807.92) # THOR VR Single Diode Laser Treatment Probe - Applicant: I. THOR International Ltd. 18a East Street Chesham HP5 1HQ United Kingdom December 21, 2006 Date Prepared: ### II. Device Name THOR VR Single Diode Laser Treatment Probe Proprietary Name: Therapeutic VR Single Diode Laser Treatment Common / Usual Name: Probe Infrared Lamp (21 CFR 890.5500) Classification Name: Product Code: ILY ### Intended Use of the Device III. The THOR VR Single Diode Laser Treatment Probe is intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for the temporary relief of minor muscle and joint pain and stiffness, minor arthritis pain, or muscle spasm, the temporary increase in local blood circulation and/or temporary relaxation of muscles. ### Predicate Devices IV. Predicate devices to the THOR VR Diode Probe are the THOR Infrared and Visible Diode Probes (K033923). These devices were cleared for introduction into interstate commerce via the FDA's 510(k) Notification process. ## V. Description of the Device The THOR VR Single Diode Laser Treatment Probe is non-invasive, easy to use, hand-held therapeutic device providing continuous heat therapy. The System is comprised of a Control Unit that houses the electronics and controls the {1}------------------------------------------------ Ko 70024 handheld treatment probe, which delivers infrared energy. The probes are intended to be placed directly on the skin to provide heating. ### Summary of the technical characteristics of the THOR VR Single Diode VI. Laser Treatment Probe to the referenced predicate devices The THOR VR Single Diode Laser Treatment Probe and the aforementioned predicate devices are infrared lamps as defined in 21 CFR 890.5500. These devices utilize infrared and visible laser diodes to generate topical heating for the purpose of elevating tissue temperatures for temporary relief of muscle and joint pain. #### Testing VII. Testing of the THOR VR Single Diode Laser Treatment Probe will include functional performance testing and electrical safety testing in accordance with all applicable standards for this type medical device. ## VIII. Conclusions Pursuant to the testing and comparison to the predicate devices, the THOR VR Single Diode Laser Treatment Probe has the same intended uses, with similar functional and performance characteristics. The System is designed to comply with the generally accepted therapeutic heat performance specifications by producing a level of tissue temperature reported in literature and accepted by the Federal Food and Drug Administration. {2}------------------------------------------------ Public Health Service Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with its wings spread, symbolizing protection and service. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" in a circular arrangement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 16 2007 Thor International, Ltd. % Texas Applied Biomedical Services Ms. M. Joyce Heinrich Regulatory Consultant 12101 Cullen Boulevard, Suite A Houston, Texas 77047 Re: K070024 Trade/Device Name: THOR VR Single Diode Laser Treatment Probe Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: II Product Code: ILY Dated: March 22, 2007 Received: March 27, 2007 Dear Ms. Heinrich: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Ms. M. Joyce Heinrich forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark M Wilkerson Mark N. Melkerson Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### APPENDIX C # STATEMENT OF INDICATIONS FOR USE K070024 510(k) Number (if known): ### Device Name: THOR VR Single Diode Laser Treatment Probe ## Indications for Use: The THOR VR Single Diode Laser Treatment Probe is intended to emit energy in The infrared spectrum to provide topical heating for the purpose of elevating the infrared Specaram to premporary relief of minor muscle and joint pain and tissue temperature for the componsible spasm, the temporary increase in local blood circulation and/or temporary relaxation of muscles. Prescription Use: ____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR Over the Counter Use: (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODpE) Mark N Millican (Division Sign-Of (Division of General. Restorative, and Neurological L 510(k) Number K070024 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...