Neurolutions Upper Extremity Rehabilitation System
DEN200046 · Neurolutions, Inc. · QOL · Apr 23, 2021 · Neurology
Device Facts
Record ID
DEN200046
Device Name
Neurolutions Upper Extremity Rehabilitation System
Applicant
Neurolutions, Inc.
Product Code
QOL · Neurology
Decision Date
Apr 23, 2021
Decision
DENG
Submission Type
Direct
Regulation
21 CFR 890.5420
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
DEN200046 · Apr 23, 2021
Neurolutions Upper Extremity Rehabilitation System
Neurolutions, Inc.
Published clinical literature (2000-2020) reporting standard of care rehabilitation outcomes for chronic stroke patients
A meta-analysis of published literature was conducted to establish a historical control for standard rehabilitation care, allowing for comparison of Fugl-Meyer Upper Extremity (FM-UE) assessment changes against the investigational device's clinical study results.
Meta-analysis; Historical control; Standard of care; Chronic stroke
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Meta-analysis of chronic stroke standard of care literature; Meta-analysis of published literature; Follow-up/Duration: Not applicable; Study Period: 2000 to June 2020
Chronic stroke patients; Sample Size: 10 patients (subset of literature); Number of Sites: Not applicable
Not applicable for this study
Fugl-Meyer Upper Extremity Assessment (FM-UE)
Indications for Use
The Neurolutions IpsiHand Upper Extremity Rehabilitation System is indicated for use in chronic stroke patients (≥ 6 months post-stroke) age 18 or older undergoing stroke rehabilitation, to facilitate muscle re-education and for maintaining or increasing range of motion in the upper extremity.
Device Story
System uses non-invasive EEG sensors in a biometric headset to detect goal-oriented brain activity from the unaffected hemisphere; signals processed by a Windows tablet; software translates intentions into motor commands for a wearable handpiece. Handpiece uses a linear actuator to drive a 3-finger pincer grip (one degree of freedom) to open/close the impaired hand. Operates in three modes: BCI ('thought') mode, volitional mode, and continuous passive motion (CPM) mode. Used in clinic or home settings under prescription. Provides physical therapy exercises to facilitate muscle re-education and improve range of motion in chronic stroke survivors. Benefits include improved motor function (ARAT/UEFM scores); helps patients perform exercises otherwise impossible due to impairment.
Clinical Evidence
Evidence from three prospective, non-randomized, self-controlled clinical studies (ORS-0008, ORS-0012, ORS-0013) and a meta-analysis. Total cohort demonstrated statistically significant improvements in motor function (ARAT, UEFM, Motricity Index, grip strength). In ORS-0012 (n=17), mean UEFM improvement was statistically significant (p < 0.0001) at 12 weeks. Meta-analysis of standard care showed no clinically meaningful improvement in chronic stroke patients, supporting the device's therapeutic effect. Adverse events were minor (fatigue, discomfort, skin redness).
Technological Characteristics
System components: Biometric headset (dry EEG sensors), Windows tablet, motor-driven handpiece. Materials: ABS D100, silver/silver chloride, polycarbonate, foam padding. Connectivity: Wi-Fi 802.11 b/g/n, Bluetooth 2.1. Power: Rechargeable lithium-ion battery. Operation: Linear actuator for 3-finger pincer grip. Biocompatibility: ISO 10993-1. Electrical safety: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11.
Indications for Use
Indicated for chronic stroke patients (≥ 6 months post-stroke) age 18+ undergoing rehabilitation to facilitate muscle re-education and maintain/increase upper extremity range of motion. Contraindicated in patients with severe spasticity/rigid contractures preventing proper fit, or skull defects (craniotomy/craniectomy).
Regulatory Classification
Identification
An EEG-driven upper extremity powered exerciser is a non-invasive prescription device intended for rehabilitation by driving movement or exercise of an impaired upper extremity in response to the detection of purpose oriented electrical activity produced by the patient's brain.
Special Controls
In combination with the general controls of the FD&C Act, the electroencephalography (EEG)driven powered exerciser is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must capture any adverse events observed during clinical use and must demonstrate that the EEG signal can be translated into intended motion.
(2) Software verification, validation, and hazard analysis must be performed.
(3) Performance data must demonstrate the electromagnetic compatibility, electrical safety, battery safety, and wireless compatibility of the device.
(4) The device components that contact the patient must be demonstrated to be biocompatible.
(5) Performance data must validate the reprocessing instructions for the reusable components of the device.
(6) Labeling must include:
(i) Instructions on fitting the device to the patient;
(ii) Information on how the device operates and the typical sensations experienced during treatment; and
(iii) Reprocessing instructions.
Submission Summary (Full Text)
{0}------------------------------------------------
## DE NOVO CLASSIFICATION REQUEST FOR NEUROLUTIONS IPSIHAND UPPER EXTREMITY REHABILITATION SYSTEM
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Electroencephalography (EEG)-driven upper extremity powered exerciser. An EEG-driven upper extremity powered exerciser is a non-invasive prescription device intended for rehabilitation by driving movement or exercise of an impaired upper extremity in response to the detection of purpose oriented electrical activity produced by the patient's brain.
NEW REGULATION NUMBER: 21 CFR 890.5420
CLASSIFICATION: Class II
PRODUCT CODE: QOL
#### BACKGROUND
DEVICE NAME: Neurolutions IpsiHand Upper Extremity Rehabilitation System
SUBMISSION NUMBER: DEN200046
DATE DE NOVO RECEIVED: July 23, 2020
#### SPONSOR INFORMATION:
Neurolutions, Inc. 1101 Pacific Ave, Suite 300 San Cruz, California 95060
#### INDICATIONS FOR USE
The Neurolutions IpsiHand Upper Extremity Rehabilitation System is indicated as follows:
The Neurolutions IpsiHand Upper Extremity Rehabilitation System is indicated for use in chronic stroke patients (≥ 6 months post-stroke) age 18 or older undergoing stroke rehabilitation, to facilitate muscle re-education and for maintaining or increasing range of motion in the upper extremity.
#### LIMITATIONS
The sale, distribution, and use of the Neurolutions System are restricted to prescription use in accordance with 21 CFR 801.109.
{1}------------------------------------------------
The Neurolutions System is contraindicated for use in patients having any of the following conditions:
- . Severe spasticity or rigid contractures in the wrist and/or digits that would prevent the Neurolutions Handpiece from being properly fit or positioned for use.
- . Skull defects due to craniotomy or craniectomy.
The safety and effectiveness of Neurolutions System therapy has not been evaluated in the following patient populations:
- Patients with Dementia, or who are too cognitively impaired to understand tasks
- Patients with severe, receptive aphasia who have difficulty understanding written or spoken language, or who are unable to follow written instructions
- Patients with severe unilateral visual inattention (neglect) that would visually limit use of the Tablet
The Neurolutions System should be used with caution in patients with nerve or sensory impairment that may limit or interfere with the patient's ability to sense pain in response to potential pressure points on the Neurolutions Handpiece.
As with use of any medical device, there are risks associated with treatment using the Neurolutions System. Some patients may experience any of the following adverse events during use of the device.
- · Fatigue, discomfort and/or headache associated with device use
- · Skin redness or pressure point(s) associated with the mechanical hand exerciser
Durability testing has not been completed beyond 6-months, persistence of benefits beyond 6-months post device use are currently unknown.
# PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
# DEVICE DESCRIPTION
The Neurolutions IpsiHand Upper Extremity Rehabilitation System (a.k.a. Neurolutions System or IpsiHand System) detects goal-oriented brain activity using non-invasive EEG electrodes to allow a stroke patient to perform therapeutic exercises that they would otherwise not be able to perform, due to their impairment.
{2}------------------------------------------------
The Neurolutions System shown in the figure below, consists of the Neurolutions Handpiece, a Tablet computer, and a Biometric Headset.
Image /page/2/Picture/1 description: A person is wearing a hand rehabilitation device. The device is blue and black and is attached to the person's hand and wrist with straps. The device appears to be designed to help the person move their fingers and hand. The person's hand is resting on a black surface.
Neurolutions Handpiece
Image /page/2/Picture/3 description: The image shows a white tablet with a Windows operating system displayed on the screen. The tablet is positioned at a slight angle, showcasing the screen and the thin profile of the device. The Windows interface is visible, featuring the start menu and various application icons, suggesting the tablet is in active use or standby mode.
Tablet
Image /page/2/Picture/5 description: The image shows a black headset with multiple sensors attached to it, placed on a transparent head model. The headset has a central circular pad on top, with several smaller circular sensors extending outwards and downwards. The sensors appear to be positioned to make contact with the head. The transparent head model is featureless, allowing the headset and sensors to be clearly visible.
Biometric Headset
Using a standard Windows Tablet as the patient interface, the System translates brain signals from the Biometric Headset into movement of the motor-driven Handpiece worn over the patient's hand and wrist. The motion of the Handpiece, in turn, opens and closes the patient's impaired hand. The combined action of these System components allows the stroke patient to perform physical therapy exercises that they would otherwise not be able to perform, due to their impairment. The Biometric Headset, which contains dry electroencephalographic (EEG) sensors, receives electrical signals from the motor or pre-motor cortex, predominantly of the unaffected hemisphere of the patient's brain, and in doing so, detects the patient's intentions to move their affected hand. These intentions to move are translated into motor movements of the Handpiece using software that drives a linear actuator contained inside of the Handpiece. The Handpiece opens and closes the hand using a 3-finger pincer grip mechanism designed to emulate the movement of grasping an object using the two forefingers and thumb in a grasping motion (one degree of freedom).
The device functions as powered exercise equipment for the patient's hand using three separate modes: the main mode of therapy is referred to as the brain-computer interface (BCI) or 'thought' mode in which the patient's hand is opened or closed by the powered orthosis based on brain signals received from the Biometric headset; the second mode is referred to as a 'volitional' mode in which a patient actively opens and closes their hand with the System enabling independent range of motion; the third is a continuous passive motion (CPM) mode in which the System simply moves the patient's hand passively through a comfortable range of motion in a repetitive fashion.
The Neurolutions System is designed for use in clinic or home settings as part of prescribed rehabilitation therapy.
# SUMMARY OF NONCLINICAL/BENCH STUDIES
# BIOCOMPATIBILITY/MATERIALS
Only the Biometric Headset and Neurolutions System Handpiece are considered patientcontacting components of the System. The sponsor tested (6) samples each of the loop
{3}------------------------------------------------
straps, orthosis padding, headset top foam padding and headset ear pad covers, which are patient-contacting throughout the therapy. However, the orthosis housing, EEG electrode components, plastic components of the finger tray and thumb rest, chargers, and the polycarbonate components of the biometric headset were not tested, as these components are in contact with the skin for a brief period and represent a low risk. The tablet was also not tested for biocompatibility, as it is an off-the-shelf consumer electronics device that is in contact with the patient only when initiating or updating the therapy program, thus posing a low risk for biocompatibility issues.
The biocompatibility evaluation for the Biometric Headset and Neurolutions System Handpiece was conducted in accordance with the International Standard ISO 10993-1: 2009 "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing Within a Risk Management Process". The Biometric Headset and Neurolutions System Handpiece are categorized as a surface device in limited (< 24 hours) contact of the skin. Assessment of the device included the following tests:
- Cytotoxicity (ISO 10993-5:2009) ●
- . Sensitization Test (ISO 10993-10:2010)
- Intracutaneous Reactivity (ISO 10993-10:2010) ●
The inner rings passed sensitization and irritation testing but demonstrated some cytotoxicity. It appears likely that the Silver/Silver Chloride (manufactured from (b) (4) fine silver) that is deposited on the Acrylonitrile Butadiene Styrene (ABS D100) base material for the inner sensor electrode is the cause for the failure. This appears to be a low risk and was determined to be acceptable. All other components met the prespecified acceptance criteria for all tests.
#### SHELF LIFE/REPROCESSING/STERILITY
The Neurolutions System is not provided sterile and is not intended to be sterilized prior to use. The device consists of the following reusable components: a handpiece, a tablet, and a headset. Cleaning and disinfection methods were validated in accordance with AAMI TIR12:2010 (Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A guide for device manufacturers). Testing was performed by Nelson Labs to demonstrate low-level disinfection of the Neurolutions System when used by different patients in a clinic setting (multiple patient use), and during factory reprocessing performed by Neurolutions, Inc. Validated cleaning and disinfection instructions have been provided for each of the three components.
The risk of contamination, particularly when the device is used between multiple patients in clinic use, is also ameliorated by the use of a legally marketed disposable protective cover that is worn as a barrier between the patient's skin and the Neurolutions Handpiece.
# ELECTROMAGNETIC CAPABILITY & ELECTRICAL SAFETY
{4}------------------------------------------------
The Neurolutions System was tested according to the following FDA-recognized consensus standards:
- IEC 60601-1:2005 (Modified to be equivalent to (AAMI/ANSI ES60601-. 1:2005/(R)2012 and C1:2009/(R)2012 and, A2:2010/(R)2012) "Medical Electrical Equipment; Part 1: General requirements for basic safety and essential performance." Results demonstrated that the device is compliant to this standard.
- IEC 60601-1-2:2014 "Medical Electrical Equipment Part 1-2: General ● Requirements for Basic Safety and Essential Performance-Collateral Standard: Electromagnetic disturbances - Requirements and Tests."
- IEC 60601-1-11:2015 "Medical electrical equipment: General requirements for . basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment."
## Wireless
The Neurolutions System includes the following wireless technologies operating in the ISM band and has been evaluated against ANSI/AAMI/IEC 60601-1-2 (4th edition) for immunity and emissions appropriate for the healthcare and home environment.
- Wi-Fi 802.11 b/g/n 2.4GHz
- Bluetooth 2.1 2.4GHz
Although the sponsor provided wireless compatibility testing. they did not provide wireless coexistence testing. Thus, the sponsor added a General Precaution warning in the labeling explaining that the device may be susceptible to interference from common radio frequency (RF) devices and added a section in the troubleshooting to address such issues.
#### Battery testing
Evaluation of the product to IEC 60601-1:2005 + A1:2012 included review of 3rd party testing of sub-components such as the rechargeable lithium ion Handpiece battery to IEC 62133 Edition 2.0 2012-12.
#### SOFTWARE
The sponsor provided documentation acceptable for software and firmware with a "Moderate" Level of Concern (LoC), as outlined in the FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." issued May 11, 2005. The primary risk to patients is delay of treatment due to software/firmware malfunction or failure.
#### SUMMARY OF CLINICAL INFORMATION
A total of three separate clinical investigations were performed to evaluate the safety and effectiveness of the Neurolutions System. The first study (QRS-0008) was completed and the results published in Bundy et al. 2017, Contralesional Brain-Computer Interface Control of a Powered Exoskeleton for Motor Recovery in Chronic Stroke Survivors. The second study (ORS-0012) was completed, and the third study (QRS-0013) was on-going at the time of submission,
{5}------------------------------------------------
with preliminary analyses provided. Additionally, a meta-analysis was performed to provide further interpretation of the outcomes of these clinical investigations which utilized subject selfcontrols.
# A. Treatment of Chronic Stroke with IpsiHand, Clinical Study (ORS-0008)
This feasibility study tested whether a powered hand orthosis driven by a brain-computer interface (BCI), using neural activity from the unaffected cortical hemisphere, could affect motor recovery in chronic hemiparetic stroke survivors. The study was a prospective, non-randomized, self-controlled study performed in two phases at one investigational site. Phase 1 was conducted to determine the feasibility of recording ipsilateral (contralesional) motor commands and to demonstrate the ability to use the signals to control a computer, and Phase 2 was intended to determine if the BCI-system can be used to impact rehabilitation was assessed by examining changes in commonly used functional motor scores.
Ten chronic (≥ 6 months) hemiparetic (Modified Ashworth Scale of 1+ or less of elbow flexion in the affected upper extremity) stroke survivors utilized the BCI IpsiHand System at home for 12 weeks. During this time, data was collected before, during, and upon completion of use of the Neurolutions IpsiHand System.
Subjects had a statistically significant improvement in the Action Research Arm Test (ARAT) of a mean 660 points (from (b) (4) points represents the Minimal Clinically Important Difference (MCID) in chronic stroke survivors. Six of the 10 subjects showed a difference exceeding the ARAT MCID. Secondary outcome measures including Canadian Occupational Performance Measure (COPM), Motricity Index, and grip strength also significantly improved.
No patient injury or adverse events have been reported in any subject; and only minor fatigue was reported in one (1) subject and was self-limiting.
# B. IpsiHand Device Use in Stroke Patients to Assess Functional Motor Outcomes, Clinical Study (ORS-0012)
The study was a prospective, non-randomized, self-controlled study performed in two phases at two (2) investigational sites. The protocol allowed for enrollment of up to 20 adult (at least 18 vears of age) stroke survivors who sustained a cerebrovascular accident (CVA) six (6) or more months prior to the date of screening and who had moderate functional impairment of the right or left upper extremity as evidenced by screening assessments.
The goal of the study was to assess whether a motor-activity based BCI system (Neurolutions IpsiHand System) would provide rehabilitative benefits for stroke survivors. Forty-two subjects were evaluated against the inclusion criteria, and those 31 who met those criteria were assigned to Group 1 or Group 2. The study was carried out in two phases. In Phase 1, each participant (Group 1 and Group 2) completed up to three (3) sessions for recording EEG signals. by wearing a research-grade EEG Headset with EEG electrodes in place. The signals were recorded with BCI software. If the participant's EEG signals were adequate for controlling a BCI
{6}------------------------------------------------
mediated hand orthosis, the participant continued on to Phase 2. Of the 31 enrolled subjects, 7 did not meets the Phase 1 EEG criteria and 24 progressed to Phase 2, which evaluated whether a BCI system can be used to impact stroke rehabilitation. In Phase 2, each subject (Group 1 and Group 2) was issued a BCI-mediated hand orthosis and advised to use the device daily at home, a minimum of five out of seven consecutive days per week for a total of 12 weeks. Each subject had a motor assessment at 4 weeks, 8 weeks, and 12 weeks of device use. Based on progress achieved during the first 12 weeks of device use (if a subject achieved a 2-point improvement on UEFM from 8 to 12 weeks), clinical discretion, and other variables, some subjects agreed to continue using the Neurolutions System at home for up to 36 weeks. Seventeen subjects completed the study protocol, and 12 completed the 6-month durability visit.
All eligible subjects underwent the same assessments, except that Group 1 subjects completed functional MRI imaging (fMRI) to obtain structural, diffusion tensor imaging (DTI), resting state (rs) and task-based (tb) fMRI at the following specific time-points during Phase 2:
- · Prior to using home BCI therapy to establish a baseline
- · 12 weeks of home BCI therapy device use
All subjects were followed for an initial period of 12 weeks. Subjects who achieved motor progress, as evidenced by results of their motor assessments throughout their participation, were allowed to continue using the device at home for up to 36 weeks, or until progress plateaued (plateau was indicated by 'less than 2 points of Fugl Meyer change at the 12-week visit'). A total of eight subjects who qualified to continue with the study past 12 weeks, declined to continue for a number of reasons including personal/scheduling conflicts. or a desire to return to previous therapy.
The Fugl-Meyer - Upper Extremity (UEFM) was administered at baseline. during use of the device, and post treatment to assess motor impairment and functional movement of the affected upper extremity as related to the impact of BCI therapy.
From baseline to 12 weeks, patients (n=17) demonstrated a mean improvement of (b) (9) with a standard deviation (SD) of (b) (4) (two sided, one-sample t-test, p-value < 0.0001).
Twelve of the 17 subjects who completed at least 12-weeks of device use also completed a durability visit 6-months post device use. The mean improvement was (0)(4) points on the ARAT assessment at 12 weeks with a SD of (b) (4) (two-sided, one-sample t-test, p-value (b) (4)). The mean change from baseline to completion was (b) (4) (n=17). The mean change in motricity at 12 weeks for the study analysis (n=17) was (6) (4) points with a SD of (b) (4) (p <.0001). The mean increase in gross grasp was (b) (4) pounds with a SD of (b) (4)
change in AMAT scores was (b) (4) points (SD (b) (4) and p < 0001). . The mean
The only adverse events reported were one case of minor fatigue, discomfort (including headache) reported with device use following heart surgery and one case of skin redness indicative of a pressure point on the patient's thumb, which was resolved with re-emphasized training on thumb positioning.
{7}------------------------------------------------
## C. Chronic Stroke Rehabilitation with Contralesional Brain Computer Interface, Preliminary Clinical Study Report (ORS-0013)
The third clinical investigation evaluating the Neurolutions System is currently being conducted with the approval of the Washington University IRB, located in St. Louis, MO. The clinical study (preliminary report provided in QRS-0013, Rev. B) is a prospective, non-randomized, selfcontrolled crossover study performed in two phases at one investigational site (St. Louis, MO).
Phase 1 of the study investigated the feasibility of recording electroencephalogram (EEG) signals from the affected and unaffected brain hemispheres and was intended to demonstrate the possibility of using these signals to control the motor output of a robotic hand orthosis (Handpiece). Phase 2 was the home use portion of the study, which assessed if the BCI system could be used to impact stroke rehabilitation. Subjects were enrolled into either Group 1 (additional fMRI assessment and range of motion (ROM) exercises, crossover group) or Group 2 (no fMRI or ROM).
Group 1 subjects underwent 12 weeks each of device use and the ROM Home Exercise Program (24 weeks total). For the first 12 weeks, half of the subjects used the Neurolutions IpsiHand System before the ROM home exercise program, and half of the subjects used the Neurolutions IpsiHand System after the ROM home exercise program. For the second 12 weeks, the subjects switched the treatment regimen regarding use of the investigational device before or after the ROM home exercise program.
The preliminary report presented results from 13 chronic hemiparetic stroke survivors who utilized the Neurolutions System in BCI-mode at home for 12- weeks. Three of the subjects are actively participating in the crossover portion (second treatment regimen) of the study. During the study, the clinical specialist collected data before, during, and upon completion of device use.
Based on this preliminary statistical analysis, from baseline to 12 weeks, on the patients who completed 12 weeks of therapy (n=12) demonstrated a mean improvement of (0) (4) with a SD of (b) (4) (two-sided, one sample t-test, p < 0.0001)
The results of Phase 1 of the study support the feasibility of recording EEG signals from the unaffected (contralesional) hemisphere and suggest that these signals can be used to control the motor output of a robotic hand orthosis (Handpiece). The results of Phase 2 of the study suggest that brain-controlled movement of the affected hand, through use of the Neurolutions InsiHand System, can positively impact rehabilitation by improving motor function.
To interpret the therapy outcomes of the Neurolutions IpsiHand System relative to a control intervention, an independent meta-analysis including studies published from 2000 to June 2020 was conducted of serial changes in terms of upper limb motor function for chronic stroke patients treated with standard rehabilitation care (as measured by Fugl-Meyer Upper Extremity Assessment, FM-UE). Analysis shows the effect size computed from serial differences, in terms of FM-UE assessment from a total 10 patients, ranges from (b) (4) to (0) (8) with an average change in FM-UE of (b) (4). Although small, this estimated average improvement can be hypothesized to originate from the training effect, or placebo effect, as some chronic stroke
{8}------------------------------------------------
patients treated with regular rehabilitation care will exhibit an improvement after periods of inactivity. Importantly, this small improvement is far below the Minimal Clinically Important Difference (MCID) scores for the FM-UE, which has been defined as an increase of 5.25 points from baseline in those patients in the chronic phase of stroke. The findings of this meta-analysis conclude that chronic stroke survivors do not demonstrate a clinically meaningful improvement in their motor function when receiving standard of care alone, as evidenced by the Fugl-Meyer assessment.
To date, there has been no patient injury or adverse events reported in any subject.
## Pediatric Extrapolation
The Neurolutions IpsiHand Upper Extremity Rehabilitation System is indicated for patients age 18 and older. For medical devices, the FD&C Act defines patients before their 22nd birthday as pediatric patients. In this De Novo request, complete data from patients between 28-79 (mean age 57.9) were used to support the use of the device in adult patients. It was appropriate to indicate the device for individuals 18 and older because patients aged 18 to 21 do not carry additional differences or risks relative to the patient population studied, incidence of stroke and hemiparesis is present in the transitional adolescent population (pediatric sub-population) of 18-21 years of age, and this device has a likely benefit for this group.
## LABELING
The labeling (User Manual) meets the requirements of 21 CFR Part 801.109 for prescription devices.
The labeling for providers and patients includes information describing the data collected with regard to the durability of the effects of device use. Namely, providers and patients are made aware that 6-month durability data were collected in a small cohort and the permanency of these gains beyond that window remain unknown.
Information in the labeling for providers and patients also describes the data collected with regard to the proportion of users that attained a clinically meaningful increased in the Fugl-Meyer with device use. Namely, 33.3% of patients who used the device as directed did not achieve a clinically meaningful increase in Fugl-Meyer score during the 12-week trial of device use. It is important that providers and patients be made aware that not all patients will experience a clinically meaningful benefit from device use.
The labeling includes instructions explaining how users can navigate the software application to use the different modes of the device (Brain-Computer Interface, Volitional, and Continuous Passive Motion). Instructions for use includes information on fitting the device to the patient, how the device operates, and the typical sensations experienced during treatment.
The labeling also outlines the cleaning for home use and clinical use.
# RISKS TO HEALTH
{9}------------------------------------------------
The table below identifies the risks to health that may be associated with use of the electroencephalography (EEG)-driven powered exerciser and the measures necessary to mitigate these risks.
| Identified Risks to Health | Mitigation Measures |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Device provides ineffective treatment,<br>leading to worsening condition | Clinical performance testing<br>Software verification, validation, and hazard analysis<br>Wireless compatibility testing |
| Unintended motion leading to injury | Software verification, validation, and hazard analysis |
| Thermal injury including burns and<br>shock | Electromagnetic compatibility (EMC) testing<br>Electrical safety testing<br>Battery safety testing<br>Labeling |
| Adverse tissue reaction | Biocompatibility evaluation<br>Labeling |
| Cross contamination, leading to<br>infection or adverse tissue reaction | Reprocessing validation<br>Labeling |
| Pain or discomfort including:<br>• Headache<br>• Fatigue<br>• Skin redness | Labeling<br>Clinical performance testing |
#### SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the electroencephalography (EEG)driven powered exerciser is subject to the following special controls:
- (1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must capture any adverse events observed during clinical use and must demonstrate that the EEG signal can be translated into intended motion.
- (2) Software verification, validation, and hazard analysis must be performed.
- (3) Performance data must demonstrate the electromagnetic compatibility, electrical safety, battery safety, and wireless compatibility of the device.
- (4) The device components that contact the patient must be demonstrated to be biocompatible.
- (5) Performance data must validate the reprocessing instructions for the reusable components of the device.
- (6) Labeling must include:
- (i) Instructions on fitting the device to the patient;
- (ii) Information on how the device operates and the typical sensations experienced during treatment; and
- (iii) Reprocessing instructions.
#### BENEFIT-RISK DETERMINATION
{10}------------------------------------------------
The risks of the device are based on data collected in the clinical studies described above. Namely, the studies included adverse events such as pain or discomfort due to headache and fatigue as well as adverse tissue reaction if the device was not positioned properly. Known risks for skin contacting devices include issues of biocompatibility, and when intended for multipatient use, there is a known risk of cross contamination. With devices of this type, malfunction due to electromagnetic interference or software faults can also pose risks such as thermal injury. unintended motion, or inability to treat due to device failure. None of the above are serious and all can be resolved quickly and easily by discontinuing use or adjusting the device.
The probable benefits of the device are also based on data collected in the clinical studies as described above. Improvements include improved upper extremity function as measured by the Action Research Arm Test (ARAT) and the upper extremity Fugl-Mever (UEFM) score, results that were statistically significant and clinically meaningful. There were also statistically significant improvements in Motricity Index and grip strength and a self-scored subjective measure of each subject's ability to use their affected arm in functional tasks. Improvements were observed in pinch strength, Modified Ashworth Scale and Active Range of Motion (AROM).
Sources of uncertainty in the benefits included the effect of concomitant use of Botox by some study participants, small sample sizes with an open label study design that does not control for the placebo effect, and limited follow up time to establish durability of treatment effect. The first was addressed using subgroup analysis showing that when the data were combined across studies there was a larger mean UEFM improvement in Botox patients ((b) (4) pts) than in non-Botox patients (0) (4) pts). The small sample size and lack of placebo control was addressed by pooling data and citing historical controls. Due to the limited follow up time, the sponsor includes in the labeling information on the durability of the treatment effect.
# PATIENT PERSPECTIVES
This submission did not include specific information on patient perspectives for this device.
# BENEFIT/RISK CONCLUSION
In conclusion, given the available information above, for the following indication statement:
The Neurolutions IpsiHand Upper Extremity Rehabilitation System is indicated for use in chronic stroke patients (> 6 months post-stroke) age 18 or older undergoing stroke rehabilitation, to facilitate muscle re-education and for maintaining or increasing range of motion in the upper extremity.
The probable benefits outweigh the probable risks for the Neurolutions IpsiHand Upper Extremity Rehabilitation System. The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.
# CONCLUSION
{11}------------------------------------------------
The De Novo request for the Neurolutions IpsiHand Upper Extremity Rehabilitation System is granted and the device is classified as follows:
Product Code: QOL Device Type: Electroencephalography (EEG)-driven upper extremity powered exerciser Regulation Number: 21 CFR 890.5420 Class: II
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.