Nova HD+

K220938 · Aura Wellness, LLC · NGX · Aug 22, 2023 · Physical Medicine

Device Facts

Record IDK220938
Device NameNova HD+
ApplicantAura Wellness, LLC
Product CodeNGX · Physical Medicine
Decision DateAug 22, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5850
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Nova HD+ is intended for use by adults for the stimulation of healthy muscles in order to improve or facilitate muscle performance.

Device Story

Nova HD+ is a noninvasive powered muscle stimulator using electrically generated magnetic fields to stimulate healthy muscles. System comprises a main control unit and a large loop coil applicator. Operated by user or clinician in home or clinical settings. User selects treatment duration (1-20 minutes) and intensity (levels 1-10) via control unit interface. Device generates magnetic pulses to facilitate muscle performance. Firmware manages user interface and pulse generator parameters. Benefits include improved muscle performance through noninvasive magnetic field induction.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including signal waveform output measurements, magnetic flux output (mT) across all intensity levels (1-10), and electrical/EMC safety testing per IEC 60601-1, 60601-1-2, 60601-1-11, and 60601-1-6. Biocompatibility testing was performed per ISO 10993-1, 10993-5, and 10993-10.

Technological Characteristics

Powered muscle stimulator; non-invasive magnetic field induction. Main unit dimensions: 9.7" x 10.64" x 3.57". Large loop coil applicator (147.2 in²). Peak frequency: 1.58 kHz; Max flux density: 1.8 mT. Power: 81.6W max. Microprocessor-controlled. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 62366-1, IEC 62304. Biocompatibility: ISO 10993-1, 5, 10.

Indications for Use

Indicated for adults for stimulation of healthy muscles to improve or facilitate muscle performance. No specific disease state or contraindications listed.

Regulatory Classification

Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square. August 22, 2023 Aura Wellness, LLC Scott Blomberg Director of Regulatory Affairs 11530 Electron Drive Louisville, Kentucky 40299 Re: K220938 Trade/Device Name: Nova HD+ Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: August 21, 2023 Received: August 21, 2023 Dear Scott Blomberg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Heather L. Dean -S Heather Dean, Ph.D Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K220938 Device Name Nova HD+ Indications for Use (Describe) The Nova HD+ is intended for use by adults for the stimulation of healthy muscles in order to improve or facilitate muscle performance. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for AURA. The logo consists of a colorful circular design above the word "AURA" in a simple, sans-serif font. The circular design is made up of several small circles in various colors, including shades of blue, green, purple, orange, and red. The circles are arranged in a pattern that radiates outward from the center, creating a visually appealing and modern design. ## 510(k) SUMMARY K220938 ## (as required by 21CFR 807.92) #### I Submitter Aura Wellness, LLC 11530 Electron Drive Louisville, KY 40299 Phone: 502-714-1993 Fax: 502-369-5226 | Submitter Contact: | Scott Blomberg | |-----------------------------|--------------------| | Submission Correspondent: | scott@aurawell.com | | Establishment Registration: | 10081462 | | Submission Date: | August 21, 2023 | #### II Device | Proprietary or Trade Name: | Nova HD+ | |----------------------------|------------------------------------------------------| | Common/Usual Name: | Powered Muscle Stimulator | | Classification Name: | Stimulator, Muscle, Powered, For Muscle Conditioning | | Regulation: | 890.5850 | | Regulatory Class: | 2 | | Product Code: | NGX | | Device Panel: | Neurological and Physical Medicine Devices | | III Predicate Device | Bemer Classic Set and Bemer Pro-Set (K210174) | #### IV Device Description The Nova HD+ system is a noninvasive physical medicine device that uses electrically generated magnetic fields to stimulate muscles in order to improve and facilitate muscle performance. The Nova HD+ is noninvasive, fully reusable (no disposable components such as electrodes) and have configurations allowing both home care and clinical professional use. In addition to the main control unit, the Nova HD+ includes a large loop coil applicator (coils) which allow the operator to administer optimal treatment to the desired anatomical part of the body. The device contains firmware that controls the user interface, the operator controls the magnetic pulse generator intensity and duration of the treatment. #### V Indication for Use The Nova HD+ is intended for use by adults for the stimulation of healthy muscles in order to improve or facilitate muscle performance. Environments of use: Over the Counter (OTC), both home care and clinical professional use Nova HD+ Traditional 510(k) {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows a logo for a company or organization called "AURA". The logo consists of the word "AURA" in a simple, sans-serif font, positioned below a colorful, circular design. The circular design is made up of many small circles in various colors, including shades of blue, green, purple, and orange. The overall design is modern and visually appealing. #### VI Comparison of Technological Characteristics and Performance with the Predicate Table 1 below provides the summarized substantial equivalence comparison of general intended uses/actions, specific indications for use, equivalence of key clinical and technical features between subject and predicate device, along with a full listing of technical and conformance specifications. ## Table 1: Comparison of Subject vs. Predicate | | | Nova HD+ | BEMER Pro Set | |----------------------------------------|---------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | Subject Device<br>K220938 | Predicate Device<br>K210174 | | Classification Code | Identical | Primary: Powered Muscle<br>Stimulator NGX 890.5850 | Same | | Indication for Use | Similar | The Nova HD+ is intended for<br>the stimulation of healthy muscles<br>in order to improve or facilitate<br>muscle performance. | The BEMER therapy is<br>indicated:<br>• To temporarily increase local<br>blood circulation in healthy leg<br>muscles.<br>• To stimulate healthy muscles in<br>order to improve and facilitate<br>muscle performance | | Primary Mode of<br>Action | Identical | Non-invasive tissue stimulation<br>via magnetic field induction | Same | | Therapy Timer | Similar | 1-20 minutes | 8-20 minutes | | Model (System) | Similar | Nova HD+ | Pro Set | | Weight | Similar | System:<br>5.6 pounds (2.5 kg) | System:<br>2.9 pounds (1.3 kg)<br>[B.BOX Classic]<br>System:<br>3.1 pounds (1.4 kg)<br>[B.BOX Professional] | | Dimensions (Main<br>Unit): | Similar | 9.7" x 10.64" x 3.57" | Public information not available | | Size of Therapy Area | Similar | 147.2 in² (Large Loop) | Public information not available | | Magnetic Field Peak<br>Frequency | Different but<br>does not raise<br>any safety<br>concerns | 1.58 kHz | Public information not available. | | Average Flux<br>Density (Applicator) | Different but<br>does not raise<br>any safety<br>concerns | 1.8 mT (max level) | 150 µT (max level) | | Power Consumption | Different but<br>does not raise<br>safety or<br>effectiveness<br>concerns | 81.6 Watts maximum | 30 Watts maximum | | Input Power | Similar | 100-130 Vac, 60 Hz, 0.68 A | 100-240 Vac, 50-60 Hz, 0.6 A; | | | | Nova HD+ | BEMER Pro Set | | | | Subject Device<br>K220938 | Predicate Device<br>K210174 | | Number of Output<br>Channels | Similar | 1 | 2 | | Microprocessor<br>controlled | Identical | Yes | Same | | Therapy Amplitude<br>Control | Identical | Power level 1-10 controlled by<br>operator | Power level 1-10 controlled by<br>operator | | User-controlled<br>therapy start/stop | Identical | Yes | Yes | | Compliance with<br>voluntary standards | Identical | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-6<br>IEC 62366-1<br>IEC 60601-1-11 | Same | | Firmware developed<br>per IEC 62304 | Identical | Yes | Same | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows a logo for a company called AURA. The logo consists of the word "AURA" in a simple, sans-serif font, with the letters in a light gray color. Above the word "AURA" is a circular design made up of many small, colored dots. The dots are arranged in concentric circles, with the colors ranging from green and blue on the outside to purple, orange, and yellow in the center. #### VII Equivalence Rationale ## Principle of Operation The principle of operation for the Nova HD+ is identical to the predicate, Bemer Classic Set and Bemer Pro-Set (K210174). The Nova HD+ and the predicate systems deliver therapeutic massage to the patient through the use of low-level magnetic fields administered directly to the body. ## Use Mode The use of the Nova HD+ is essentially the same as the predicate device. Both units have main control units have interchangeable treatment applicators (coils) for administering the therapy to desired area of the controls on the control unit allow the user to select the time and intensity of the therapy and allow for adjustment of the intensity during the session. ## Non-clinical Testing Performance testing demonstrated that the subject treatment applicator performed equivalent to the predicate applicator. Testing included: Performance testing involved multiple measurements of: - Signal waveform output (volts) generated from Nova HD+ with its applicators. ● - Magnetic flux output (mT) generated from the applicator at all signal intensity input levels 1-10. ## Software - All software features documented and tested in accordance with IEC 62304 for 'Moderate' level of concern. Electrical / EMC evaluated per - IEC 60601-1 Medical Electrical Equipment Part 1: General Requirements for Basic ● Safety And Essential Performance - IEC 60601-1-2 Medical Electrical Equipment Part 1-2: General Requirements For ● Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for AURA. The logo consists of a colorful circular design made up of many small circles in shades of green, blue, purple, and orange. Below the circular design, the word "AURA" is written in a simple, sans-serif font. - . IEC 60601-1-11 - Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety And Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in The Home Healthcare Environment - IEC 60601-1-6 - Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability - . IEC 62366-1 - Medical devices - Part 1: Application of usability engineering to medical devices ## Biocompatibility - ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management ● process . ISO 10993-5 - Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity - ISO 10993-10 Biological Evaluation of Medical Devices Part 10: Tests for Irritation And . Skin Sensitization ## Clinical Testing Not applicable. ## CONCLUSIONS The analysis of the differences between Nova HD+ and the predicate device does not raise new questions of safety and effectiveness for the subject device performance test results, Aura Wellness determines that the Nova HD+ system performs within its design specifications and is equivalent to the predicate device. The information in this 510(k) submission demonstrates that the Nova HD+ system is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...