K072064 · Therapeutic Clinical Technologies, Inc. · ITH · Apr 21, 2008 · Physical Medicine
Device Facts
Record ID
K072064
Device Name
GRAVILAX TRACTION SYSTEM
Applicant
Therapeutic Clinical Technologies, Inc.
Product Code
ITH · Physical Medicine
Decision Date
Apr 21, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The GraviLax Traction System is intended for use in exerting therapeutic pulling forces on the patient's body. The device utilizes a cylindrical hot tub in which the patient is suspended vertically by a shoulder harness and which contains salt water whose salinity is adjusted to a specific gravity of approximately 1.2 providing an increased natural buoyancy to the user's body. Decompressive force (unloading due to distraction and positioning) to the sacral, lumbar and thoracic spine is achieved by application of suitable weights to the patient's ankles. This device is for prescription use only and is limited to use by qualified personnel in a suitably equipped treatment facility.
Device Story
GraviLax Traction System provides spinal decompression via vertical suspension in a saline-filled cylindrical hot tub. Device utilizes buoyancy from high-salinity water (specific gravity ~1.2) to support patient weight while shoulder harness maintains vertical position. Decompressive force applied to sacral, lumbar, and thoracic spine via ankle weights. Used in clinical treatment facilities by qualified personnel. Benefits include spinal unloading through distraction and positioning. System is purely mechanical; no electronic or software components.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Powered traction equipment; cylindrical hot tub design; saline solution (specific gravity 1.2); shoulder harness; ankle weights. Mechanical operation; no software or electronic sensing components.
Indications for Use
Indicated for patients requiring therapeutic traction/decompression of the sacral, lumbar, and thoracic spine. Prescription use only; for use by qualified personnel in clinical treatment facilities.
Regulatory Classification
Identification
Powered traction equipment consists of powered devices intended for medical purposes for use in conjunction with traction accessories, such as belts and harnesses, to exert therapeutic pulling forces on the patient's body.
K101889 — DECOMPRESSION OF CHOICE AND NEURO DECOMPRESSION DEVICE · Pivotal Health Solutions · Aug 27, 2010
K243775 — True Non-Surgical Spinal Decompression System (DRX9000-SL) · Excite Medical of Tampa Bay, LLC · Jul 18, 2025
K130249 — INTEGRITY SPINAL CARE SYSTEM 3.0 · Integrity Life Sciences · Feb 6, 2014
K031862 — SPINERX-LDM · Spinerx Technology · Oct 31, 2003
K992545 — TM-300 TRACTION SYSTEM · Ito Co., Ltd. · Sep 1, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
APR 2 1 ZUUR
## 510(k) Summary
Daniel J. Manelli Submitted by: Manelli & Fisher, P.L.L.C. 5335 Wisconsin Ave., NW (Suite 440) Washington, DC 20015
Telephone: 202-885-5548
On behalf of Therapeutic Clinical Technologies, Inc. 2655 South Rainbow Blvd. (Ste. 300) Las Vegas, Nevada 89146
510(k) Submission: GraviLax Traction Device Date: March 27, 2008
Description:
The device is a powered traction device pursuant to 21 CFR 890.5900 (FDA product code ITH) class 2. The device is substantially equivalent to various other traction devices which achieve their effect through the force of gravity and/or mechanical means, including the following:
HangUps InvertAlign; STL International Inc. K991835; BAX-D Genesis System; VAD-D Medical Technology Services, LLC. K053503; Digit-Trac 930 Traction System; Ever Prosperous Instrument, Inc. K052453; Extentrac Elite; Advanced Back Technologies, Inc. K031996; and DRS System; Professional Distribution Systems, Inc. K981822.
Indications for Use: The GraviLax Traction System is intended for use in exerting therapeutic pulling force on the patient's body. The device utilizes a cylindrical hot tub in which the patient is supported vertically by a shoulder harness and which contains salt water whose salinity is adjusted to a specific gravity of approximately 1.2 providing an increased natural buoyancy to the user's body. Decompressive force (unloading due to distraction and positioning) to the sacral, lumbar and thoracic spine is achieved by application of suitable weights to the patient's ankles. This device is for prescription use only and is limited to use by qualified personnel in a suitably equipped treatment facility.
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The text is in all capital letters and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Therapeutic Clinical Technologies % Manelli & Fisher, P.L.L.C. Mr. Daniel J. Manelli 5335 Wisconsin Avenue, Suite 440 Washington, DC 20016
APR 2 1 2008
K072064 Re: Trade/Device Name: GraviLax Traction System Regulation Number: 21 CFR 890.5900 Regulation Name: Power traction equipment Regulatory Class: Class II Product Code: ITH Dated: March 27, 2008 Received: March 28, 2008
Dear Mr. Manelli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic Free (110) the elevice, subject to the general controls provisions of the Act. The I ou mayy more servisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can may be our of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 Comply will an the Pet 270 cart 801); good manufacturing practice requirements as set OI It I art 607), moeing (21 CFR Part 820); and if applicable, the electronic forth in the quality by the (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. Daniel J. Manelli
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M Milliman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Exhibit A
## Indications for Use
510(k) Number (if known): K072064
Device Name: GraviLax Traction System
Indications for Use:
The GraviLax Traction System is intended for use in exerting therapeutic pulling forces on the patient's body. The device utilizes a cylindrical hot tub in which the patient is suspended vertically by a shoulder harness and which contains salt water whose salinity is adjusted to a specific gravity of approximately 1.2 providing an increased natural buoyancy to the user's body. Decompressive force (unloading due to distraction and positioning) to the sacral, lumbar and thoracic spine is achieved by application of suitable weights to the patient's ankles This device is for prescription use only and is limited to use by qualified personnel in a suitably equipped treatment facility.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR (21 CFR 801 Subpart C)
Over-The-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Neil R.P. Ogden for mxm
Division of General, Restorative, and Neurological Devices
**510(k) Number** K072064
Page 1 of 1
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.