DECOMPRESSION OF CHOICE AND NEURO DECOMPRESSION DEVICE
K101889 · Pivotal Health Solutions · ITH · Aug 27, 2010 · Physical Medicine
Device Facts
Record ID
K101889
Device Name
DECOMPRESSION OF CHOICE AND NEURO DECOMPRESSION DEVICE
Applicant
Pivotal Health Solutions
Product Code
ITH · Physical Medicine
Decision Date
Aug 27, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5900
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Decompression' of Choice (DOC) and AAAR 2K10 is a non invasive method of applying distractive forces to the spine through controlled tensions. It is designed to apply decompression' forces to intervertebral discs. The Decompression' of Choice (DOC) and AAAR 2K10 may be used for back pain, neck pain, hemiated discs, protruding discs, degenerative disc disease, sciatica and posterior facet syndromes.
Device Story
Non-surgical spinal traction/decompression system; applies controlled distractive forces to spine to reduce disc bulges/herniations and promote healing. System comprises base, elevation, cervical, thoracic, and pelvic sections. Input: user-defined treatment protocols (static, intermittent, cyclic traction). Operation: gear-motor/linear actuator controlled by user interface; load cell provides real-time feedback to adjust forces for patient posture/resistance. Output: controlled mechanical traction force. Used in clinical settings by healthcare personnel. Benefits: non-invasive decompression of intervertebral disc space; reduction of pressure on spinal structures.
Clinical Evidence
Bench testing only. Performance testing conducted on cervical and lumbar sections using a calibrated scale to simulate a patient. Measured force output during manual distraction confirmed accuracy of applied traction forces against target poundage. Compliance with IEC 60601-1, IEC 60601-2-46, IEC 60601-1-4, UL 60601-1, and IEC 60601-1-2 standards verified.
Technological Characteristics
Structural steel frame (welded/bolted, rust-protected coating); cushions (plywood, foam, vinyl). Power: electric power supply with linear actuator. Sensing: load cell for force feedback. Connectivity: PIC interface. Traction modes: static, intermittent, cyclic (ramp up/down). Safety: cervical limit 50 lbs, lumbar limit 200 lbs. Standards: IEC 60601-1, IEC 60601-2-46, IEC 60601-1-4, UL 60601-1, IEC 60601-1-2.
Indications for Use
Indicated for patients suffering from back pain, neck pain, herniated discs, protruding discs, degenerative disc disease, sciatica, and posterior facet syndromes requiring non-surgical spinal decompression via distractive forces.
Regulatory Classification
Identification
Powered traction equipment consists of powered devices intended for medical purposes for use in conjunction with traction accessories, such as belts and harnesses, to exert therapeutic pulling forces on the patient's body.
Chattanooga Group Triton/Tru-Trac/Tx/Triton Dts Traction (K053223)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Pivotal Health Solutions. The logo consists of a stylized human figure on the left, followed by the word "PIVOTAL" in large, bold letters. Below "PIVOTAL" are the words "Health Solutions" in a smaller font. The logo is black and white.
### Section 5: 510(k) Summary
- 1. Submitted By:
Phillip Klein Pivotal Health Solutions 724 Oakwood Road Watertown SD, 57201 Phone: 605-753-0110 Fax: 605-882-8398
AUG 27 2010
Date Prepared: November 23, 2009
- 2. Trade Name: Decompression' of Choice (DOC) and AAAR 2K10 (referred to collectively as the "Devices" in this document) Common Name: Power Traction Equipment Classification Name: TTH, Power Traction Equipment, 21 CFR, Section 890-5900
#### 3. Predicate Device Information:
The Devices are substantially equivalent to the Saunders Group 3D Active Trac. K001712 and Chattanooga Group Triton/Tru-Trac/Tx/Triton Dts Traction K053223.
- Device Description: বঁ
The Devices are a non surgical traction or also called decompression' systems. It accurately controls tensions designed to relax Para spinal musculature and to allow distractive- traction forces to decompress inter vertebral spinal disk space. The Devices do this by creating a negative pressure in the disc due to the distractive forces and treatment protocols. Bulges or herniation are reduced back into the center or annular part of the disc. Fibroblasts then form a matrix over the protryded area to form a "seal" over the previously injured area.
The load cell constantly updates the user interface provided which controls a gear-motor/linear actuator to immediately and safely apply forces as determined by qualified healthcare personnel. The load cell feedback is also used to further verify and adjust traction forces allowing for variations in patient postures and outside influencing forces. This method of operation allows safe, continuous and smooth operation of the unit and proper comfortable and safe appliance of said forces to the patient.
An extensive manual of suggested treatment protocols and safety warnings is furnished to the end user.
- ഗ് Intended Use Statement:
The Decompression' of Choice (DOC) and AAAR 2K10 is a non invasive method of applying distractive forces to the spine through controlled tensions.
It is designed to apply decompression' forces to intervertebral discs.
The Decompression' of Choice (DOC) and AAAR 2K10 may be used for back pain, neck pain, hemiated discs, protruding discs, degenerative disc disease, sciatica and posterior facet syndromes.
1 Decompression is unloading due to distraction and positioning and/or as non-surgical in nature.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Pivotal Health Solutions. The logo consists of a stylized human figure on the left, followed by the word "PIVOTAL" in large, bold letters. Below "PIVOTAL" are the words "Health Solutions" in a smaller font.
#### 6. Technological Characteristics Comparison Summary
The Devices and predicate device K001712 are similar in shape and can be divided into similar smaller sections. They have a base, elevation, cervical, thoracic, pelvic and a controls section. The base supports the lifting mechanism and also provides a stable surface during table operation. The elevation section lowers the table surface so it is easier to get on and off of the table. The elevation also elevates so the operator is working in an ergonomically correct position. The cervical section applies distractive forces and is capable of P-A flexion for targeting. The thoracic section is stationary and provides a mounting place for the cervical and pelvic section applies distractive forces and is capable of P-A flexion and axial rotation for targeting. The controls sections move and monitor the table during treatment.
The Devices and predicate device K001712 use similar material and construction methods. They are constructed from structural steel that has been welded and bolted together. All of the steel has been coated with a protective coating to prevent rust. The cushions are constructed from plywood, foam and vinyl. The Plywood is the base of the cushion and mounts directly to the steel. The foam is secured to the plywood and provides a comfortable surface for the patient to lie on. Vinyl covers the foam and is secured to the plywood.
The Devices and predicate devices K001712 use different sources of power for the distractive forces. The Devices use a power supply and linear actuator to generate the force for distraction. The predicate device uses an air compressor and pneumatic cylinder. The force delivered to the patient is measured using a load cell, similar to a scale, and a pic interface for the devices. The predicate device uses an air pressure regulator and calculation to create the required force.
The Devices and predicate device K053223 perform similar traction pulls. Both the Devices and predicate K053223 have static, intermittent and cyclic ramp up, treatment and ramp down. They have similar hold period and rest period settings as well as ramp up and ramp down steps. The maximum traction of the predicate K053223 is 200 lbs. and the lumbar section of the devices is 200 lbs. The cervical section of the devices is limited to 50 lbs for safety of the patient.
7. Performance and Compliance Data
> To prove the Devices are effective and as safe as the predicate performance testing was done on the cervical and lumbar sections. A calibrated scale was placed between the cervical and lumbar sections to simulate a patient. The scale was used to measure the force being applied to the patient. Using manual distraction the table was ramped up to target poundage and the readings where noted.
The Devices are in compliance with the following safety standards.
- 1. IEC 60601-1. Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1. 1991-11. Amendment 2, 1995
- 2. IEC 60601-2-46, Medical Electrical Equipment; Part 2-46: Particular requirements for safety of Operating Tables, First Edition: 1998
- 3. IEC 60601-1-4:2000 Consol. Ed. 1.1, Medical electrical equipment Part 1-4: General requirements for safety - Collateral standard: Programmable electrical medical systems, edition 1.1
- 4. UL 60601-1, Medical Electrical Equipment; Part 1: General Requirements for Safety, 1ª Edition: 2003 with revisions dated April 26, 2006.
- 5. IEC 60601-1-2, (Second Edition, 2001), Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Pivotal Health Solutions. The logo consists of a stylized human figure on the left, followed by the word "PIVOTAL" in large, bold letters. Below "PIVOTAL" are the words "Health Solutions" in a smaller font. The logo is black and white.
### 8. Biocompatibility Information
Materials that come in direct contact with the patient are the table top and cervical capture. The materials of those devices are used on marketed device with no known adverse effects. Because the materials are commonly used on other products biocompatibility testing is not warranted.
..
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines representing its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66-G609 Silver Spring, MD 20993-0002
Pivotal Health Solutions % Intertek Testing Services NA, Inc. Mr. William J. Sammons 2307 E. Aurora Road Unit B7 Twinsburg, OH 44087
AUG 27 2010
Re: K101889
Trade/Device Name: Decompression of Choice and AAAR 2K10 Regulation Number: 21 CFR 890.5900 Regulation Name: Power traction equipment Regulatory Class: II Product Code: ITH Dated: August 11, 2010 Received: August 13, 2010
Dear Mr. Sammons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 - Mr. William J. Sammons
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Mink M. Mllanson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Pivotal Health Solutions. The logo consists of a stylized human figure on the left, with the word "PIVOTAL" in large, bold letters to the right. Below "PIVOTAL" are the words "Health Solutions" in a smaller font. The logo is black and white.
K101 889
# Indications for Use
K101889 510(k) Number:
The Decompression' of Choice (DOC) and AAAR 2K10 Device Name:
Indications For Use:
The Decompression' of Choice (DOC) and AAAR 2K10 is a non invasive method of applying distractive forces to the spine through controlled tensions.
It is designed to apply decompression is unloading due to distraction and positioning and/or as non-surgical in nature) forces to intervertebral discs.
The Decompression' of Choice (DOC) and AAAR 2K10 may be used for back pain, neck pain, hemiated discs, protruding discs, degenerative disc disease, sciatica and posterior facet syndromes.
1 Decompression is unloading due to distraction and positioning and/or as non-surgical in nature.
Prescription Use × (Part 21 CFR 801 Subpart D)
Over-The Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
---
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K101899
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.