COMBI 8 MAX

K070888 · Cec Electronica S.R.L. · IPF · Jul 3, 2008 · Physical Medicine

Device Facts

Record IDK070888
Device NameCOMBI 8 MAX
ApplicantCec Electronica S.R.L.
Product CodeIPF · Physical Medicine
Decision DateJul 3, 2008
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 890.5850
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Depending on device model, and the specific waveform applied - COMBI pulse generators are indicated for the following uses: COMBI 4/8 DIGITAL, COMBI TREND 4/8 . Relaxation of muscle spasms . Prevention or retardation of disuse atrophy . . Increasing local blood circulation ● . Muscle re-education . . Immediate post-surgical stimulation of calf muscles to prevent venous . thrombosis . . Maintaining or increasing range of motion. . COMBI 4/8 MAX . Relaxation of muscle spasms . Prevention or retardation of disuse atrophy . . Increasing local blood circulation ● . Muscle re-education ● . Immediate post-surgical stimulation of calf muscles to prevent venous . thrombosis . Maintaining or increasing range of motion. . For TENS waveform: . . Relief of chronic, intractable pain . Adjunctive treatment of post surgical or post traumatic acute pain .

Device Story

Multi-wave pulse generator line (COMBI Digital, Trend, Max) delivering electrical pulses via skin-contact adhesive electrodes; 4 or 8 independent channels. Produces various waveforms: Russian currents, symmetric/asymmetric TENS (including burst modes), bipolar interferential, rectangular/triangular impulses, and direct current. Used in clinical settings by healthcare providers to manage muscle conditions and pain. Output parameters controlled by device settings; provider monitors patient response to adjust stimulation intensity/mode. Benefits include improved muscle function, circulation, and pain management.

Clinical Evidence

Bench testing only. Device performance and safety validated against FDA guidance for Powered Muscle Stimulators and IEC 60601-2-10 standards. No clinical data presented.

Technological Characteristics

Multi-wave pulse generator; 4-8 independent channels. Waveforms: Russian (1:1, 2:1, 4:1), TENS (symmetric/asymmetric), bipolar interferential, rectangular/triangular impulses, direct current. Connectivity: Standalone. Safety standard: IEC 60601-2-10. Software validation performed per FDA guidance for software in medical devices.

Indications for Use

Indicated for patients requiring muscle stimulation for spasm relaxation, prevention of disuse atrophy, increased local blood circulation, muscle re-education, post-surgical venous thrombosis prevention, and range of motion maintenance. TENS waveform indicated for relief of chronic, intractable pain and adjunctive treatment of post-surgical or post-traumatic acute pain.

Regulatory Classification

Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 11070888 JUL - 9 2008 ## SECTION 2: 510K SUMMARY Date of application: March 2007 Applicant's name: CEC Electronica S.R.L. Argentina Tel/Fax: +54351422492 Contact person: Benny Arazy Arazy Group Mizpe Aviv, Industrial Park 13 M. P. Misgav 20187, ISRAEL Tel: +972 (4) 994-0337 Fax: +972 (4) 994-4224 e-mail: benny.A@arazygroup.com Product Information Common name: Powered Muscle Stimulator Trade name: COMBI 8 MAX Product Codes: IPF, GZJ and LIH Regulation Numbers and Descriptions: 890.5850 - Powered muscle stimulator (IPF) 882,5890 - Transcutaneous electrical nerve stimulator for pain relief (GZJ) Interferential Current Therapy (LIH - Pre-Amendment, unclassified) Panel Code: 89 Classification Advisory Committee: Physical Medicine Establishment Registration Number: to be assigned #### Information on devices to which substantial equivalence is claimed: | 510K # | Trade or propriety name | Manufacturer | Class | |---------|-------------------------|----------------------------|-------| | K984114 | SYS STYM 294 | Mettler Electronics Corp. | II | | K062354 | Vectra Genisys | Chattanooga Group | II | | K960969 | Galva 5-T | Zimmer Elektromedizin Gmbh | II | Technological characteristics, for each output mode, were provided and compared to the predicate devices, in compliance with the guidelines included in FDA Guidance Document for Powered Muscle Stimulator 510(k)s. {1}------------------------------------------------ #### Device Definition The COMBI family by CEC, is a line of multi-wave pulse generators, intended for medical purposes. The COMBI generators are capable of producing different wave types, such as: three Russian currents (1:1, 2:1, 4:1), symmetric TENS, and symmetric in burst mode: asymmetric TENS, and asymmetric in burst mode; bipolar interferential; three types of impulses, rectangular, triangular, exp + direct current. The devices deliver the predetermined electrical pulses through skin contact adhesive clectrodes, connected to up to four or eight independent body channels, depending on the specific model. (The prefixes 4 and 8 designate the number of channels, respectively). Three different configurations of Multiwave Generator devices comprise the COMBI product line: - 1. COMBI Digital - 2. COMBI Trend - 3. COMBI Max The following table gives a summary of the waveforms featured by each one of the three device configurations: | | Russian 1:1 | Russian 1:1, 2:1 ,4:1 | TENS sym. & asym. | Tetrapolar Interferential | Bipolar Interferential | Rect. & Square Impulses | Direct Current | |---------------|-------------|-----------------------|-------------------|---------------------------|------------------------|-------------------------|----------------| | COMBI Digital | ✓ | | | | | | ✓ | | CMBI Max | | ✓ | ✓ | | | ✓ | ✓ | | COMBI Trend | | ✓ | | ✓ | ✓ | | | For more details regarding the device physical and performance sharacteristics please refer to Section 4 of the submission. {2}------------------------------------------------ ### Intended Use Depending on device model, and the specific waveform applied - COMBI pulse generators are indicated for the following uses: ## COMBI 4/8 DIGITAL, COMBI TREND 4/8 - . Relaxation of muscle spasms - Prevention or retardation of disuse atrophy . - Increasing local blood circulation ● - Muscle re-education . - Immediate post-surgical stimulation of calf muscles to prevent venous . - thrombosis . - Maintaining or increasing range of motion. . ### COMBI 4/8 MAX - Relaxation of muscle spasms . - Prevention or retardation of disuse atrophy . - Increasing local blood circulation ● - Muscle re-education ● - Immediate post-surgical stimulation of calf muscles to prevent venous . thrombosis - Maintaining or increasing range of motion. . #### For TENS waveform: - . Relief of chronic, intractable pain - Adjunctive treatment of post surgical or post traumatic acute pain . #### Declaration of Conformity The COMBI 8 MAX device is in compliance with the following FDA recognized standard: - IEC 60601-2-10 (1987-12) Medical electrical equipment Part 2: イ Particular requirements for the safety of nerve and muscle stimulators A full statement of Declaration of Conformity to recognized standards is included in the submission. {3}------------------------------------------------ #### Summarv The attached documents contain the information required in accordance with FDA regulations at 21 C.F.R. Part 807.87. As indicated above, throughout the submission preparation the Guidance document for Powered Muscle Stimulator 510(K)'s was used to address the risks associated with the device. In particular, the Technological Characteristics were written according the attachment II of the guidance, the Instructions for Use were written by taking into consideration all the precautions and contraindications that appear in the guidance. The device labeling was prepared according the attachment II- Labeling guidance. The standard indication for use for powered muscle stimulators was adopted from the guidance. The software validation was performed by using the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. The COMBI product line, subject of this submission, constitutes a safe, reliable and effective medical device meeting all the declared requirements of its intended use. The device presents no adverse health effect or safety risks to patients when used as intended. We trust that the information provided in this SE file will be sufficient to enable FDA to find the COMBI product line substantially equivalent to the predicate devices as: > The intended use is identical to that of the cleared predicate device. イ The fundamental scientific technology is identical to that of the cleared predicate device. > The modification relative to the cleared predicate device does not introduce any new hazards and does not affect the mitigations and CAPA. These modifications as compared with the predicate device are described in the Substantial Equivalence to Cleared Device. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is oriented to follow the curve of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 CEC Electronica S.R.L. % Mr. Benny Arazy Arazy Group Mizpe Aviv, Industrial Park 13 M.P. Migav 20187 Israel JUL - 3 2008 K070888 Re: Trade/Device Name: Combi 8 Max Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, GZJ, LIH Dated: June 19, 2008 Received: June 26, 2008 Dear Mr. Arazy: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2 - Mr. Benny Arazy This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark M. Mulhausen Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # SECTION 3: INDICATIONS FOR USE 510(k) Number: K070888 Device Name: COMBI 8 MAX Indication for Use: Depending on device model, and the specific waveform applied - COMBI pulse generators are indicated for the following uses: ### COMBI 4/8 DIGITAL, COMBI TREND 4/8 - Relaxation of muscle spasms - Prevention or retardation of disuse atrophy . - . Increasing local blood circulation - Muscle re-education . - . Immediate post-surgical stimulation of calf muscles to prevent venous - thrombosis . - Maintaining or increasing range of motion. . #### COMBI 4/8 MAX - . Relaxation of muscle spasms - Prevention or retardation of disuse atrophy . - . Increasing local blood circulation - . Muscle re-education - . Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis - Maintaining or increasing range of motion. . #### For TENS waveform: - Relief of chronic, intractable pain . - Adjunctive treatment of post surgical or post traumatic acute pain . (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | <div> <span>X</span> OR Over-(Division Site Use Off) </div> | |------------------|-------------------------------------------------------------------------| |------------------|-------------------------------------------------------------------------| Division of General, Restorative, (Per 21 CFR 801. and Neurological Devices | 510(k) Number | | |---------------|--| |---------------|--|
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...