K062354 · Encore Medical L.P. · IPF · Jan 26, 2007 · Physical Medicine
Device Facts
Record ID
K062354
Device Name
VECTRA GENISYS VMS FR
Applicant
Encore Medical L.P.
Product Code
IPF · Physical Medicine
Decision Date
Jan 26, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For VMS-(Pulsed Mode, Burst Mode or FR Mode), Russian, Monophasic Hi-Volt (NMES) & Interferential and Premodulated (IFS) Relaxation of Muscle Spasms Prevention or retardation of disuse atrophy Increasing local blood circulation Muscle re-education Maintaining or increasing range of motion Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis Additionally for Microcurrent, Interferential, Premodulated (IFS), VMS-(Pulsed Mode, Burst Mode or FR Mode), Asymmetrical Biphasic (TENS), and Symmetrical Biphasic (TENS) Symptomatic relief or management of chronic, intractable pain Post-traumatic acute pain Post-surgical acute pain For FES Stimulation of the muscles in the leg and ankle of partially paralyzed patients to provide flexion of the foot and thus improve the patient's gait For DC Continuous Mode Relaxation of muscle spasm For EMG To determination the activation timing of muscles for: a) retraining of muscle activation b) coordination of muscle activation An indication of the force produced by muscle for control and maintenance of muscle contractions Relaxation muscle training Muscle re-education For EMG triggered Stim Stroke rehab by muscle re-education Relaxation of muscle spasms Prevention or retardation of disuse atrophy Increase local blood circulation Muscle re-education Maintaining or increasing range of motion For Ultrasound Application of therapeutic deep heat for the treatment of selected sub-chronic and chronic medical conditions such as: 1. Relief of pain, muscle spasms and joint contractures 2. Relief of pain, muscle spasms and joint contractures that may be associated with: a) Adhesive capsulitis b) Bursitis with slight calcification c) Myositis d) Soft tissue injuries e) Shortened tendons due to past injuries and scar tissues 3. Relief of sub-chronic and chronic pain and joint contractures resulting from: a) Capsular tightness b) Capsular scarring
Device Story
Vectra Genisys is a modular clinical electrotherapy system combining muscle stimulation, ultrasound, and sEMG biofeedback. Input modalities include surface EMG signals and user-selected therapy parameters. The device transforms these inputs via integrated clinical protocols and on-board indications to guide waveform selection, parameter settings, and electrode placement. Outputs include therapeutic electrical stimulation (various waveforms), ultrasound energy (1 or 3.3 MHz), and visual biofeedback. Used in clinical settings by healthcare providers to manage pain, muscle spasms, and facilitate rehabilitation (e.g., stroke recovery, muscle re-education). Clinicians use the 5-inch TFT LCD display to monitor patient data and adjust treatment. The system benefits patients by providing integrated, multi-modal therapy options in a single compact unit, improving clinical workflow and treatment precision.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and compliance with consensus standards (bench testing).
Indicated for patients requiring pain management, muscle spasm relaxation, muscle re-education, increased local circulation, range of motion maintenance, or prevention of disuse atrophy. Includes stroke patients for rehabilitation, patients with chronic/intractable/post-surgical/post-traumatic pain, and partially paralyzed patients requiring FES for gait improvement.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Summary - 510(k) # K062354
Chattanooga Group, SUBMITTER: A Division of Encore Medical, L.P. 4717 Adams Road Hixson, TN 37343
JAN 2 6 2007
#### EXTABLISHMENT 1022819 REGISTRATION:
CONTACT PERSON: Michael Treas Manager of Regulatory Affairs Phone: (423) 870-2281 Fax: (423) 870-7404
DATE PREPARED: October 20, 2006
#### DEVICE TRADE NAME: Vectra Genisys
CLASSIFICATION: Class II
PRODUCT CODES: IPF, IMG, GZJ, HCC, GZI, LIH
21 CFR 890.5850- Stimulator, Muscle, Powered REGULATION NUMBERS
21 CFR 890.5860- Ultrasound and muscle stimulator AND COMMON NAMES:
21 CFR 882.5890- Transcutaneous electrical nerve stimulator for pain relief
21 CFR 882.5050- Biofeedback device
21 CFR 882.5810- External functional neuromuscular stimulator
21 CFR 876.5320- Non-implantable electrical continence device
89- Physical Medicine, 84- Neurology & 78- Gastroenterology PANELS:
PREDICATE: Omnistim FX2 - K945509 Omnistim FX2 Pro - K945508
Page 1 of 3
{1}------------------------------------------------
# Summary - 510(k) # K062354
Description: Vectra Genisys electrotherapy product lines offer clinicians a modular design of muscle stimulation, ultrasound, and biofeedback modalities in one combination device. These clinical product lines are designed to give the most treatment options in one compact and integrated package. The award winning design offers a 5 inch TFT LCD vibrant color display screen and hand held accessories.
Clinicians have a variety of choices to best suit the needs of the individual practice. Below is an overview of the system choices. The electrotherapy mode offers one of the largest selections of multiple waveforms cleared to market by the FDA. The numeric pain scales can be recorded with the patient data management system. The therapy system cart provides six concealed storage bins to conveniently house clinical essentials.
The electrotherapy module offers multiple waveforms; Interferrential, Premodulated, Asymmetrical Biphasic, Microcurrent, VMS-(Pulsed Mode, Burst Mode, and FR Mode), Russian, High Voltage Pulsed Current, Symmetrical Biphasic, Direct Current.
The dual frequency ultrasound module offers Pulsed and Continuous Duty Cycles (10%, 20%, 50%, and 100%), Low BNR (5:1), Four different size ultrasound applicators, 1cm², 2cm², and 10 cm².
The sEMG biofeedback module provides two channels of surface EMG. Feedback can be stored onto the sEMG Data Card. The sEMG features a clinician chosen trigger point that activates therapeutic stimulation. The sEMG feature is often used to treat stroke patients and for muscle re-education.
The online-guided assistance through Clinical Protocols and On-Board Indications to help guide therapy selections: electrotherapy waveform rationale, parameter selections, electrode placement images, ultrasound applicator recommendations.
The combination electrotherapy is used for the management of pain and muscle spasm. All functions of 1 or 3.3 MHz Ultrasound can be combined with Interferrential, Premodulated, Asymmetrical Biphasic, VMS-(Pulsed Mode, Burst Mode, ), and High Voltage Pulsed Current.
{2}------------------------------------------------
# Summary - 510(k) # K062354
# Declarations of Conformity
The Vectra Genisys devices are in compliance with the following FDA recognized consensus standards:
UL 60601-1: 2003, Standards for Medical Equipment Part 1: General Requirements for Safety, l \$1 Edition
IEC 60601-1-2: 2001, Medical Electrical Equipment Part 1 - 2: General requirements for Safety - Collateral Standard, Electromagnetic Compatibility - Requirements and Tests, 2nd Edition
# Truthful and Accurate Statement
A statement attesting to the truthfulness and accuracy of the information was included in the submission.
### Further Information
In the event that additional information is required, please contact:
Michael Treas Manager of Regulatory Affairs Chattanooga Group A Division of Encore Medical, L.P. 4717 Adams Road Hixson, TN 37343-0489 U.S.A. Telephone: (423) 870-2281 Fax: (423) 870-7404 E-mail: Michael Treas@ChattGroup.com
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three curved lines representing its body and wings. The eagle faces to the right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
FEB 2 2 2007
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Encore Medical, L.P. Chattanooga Group c/o Mr. Michael Treas 4717 Adams Road Hixson, Tennessee 37343
Re: K062354
Trade Name: Vectra GENiSYS Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: II Product Code: IPF, GZJ, HCC, IMG, GZI Dated: January 26, 2007 Received: January 26, 2007
Dear Mr. Treas:
This letter corrects our substantially equivalent letter of January 26, 2007. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Michael Treas
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours,
for
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure .
{5}------------------------------------------------
| 510(k) Number | K062354 |
|---------------|----------------|
| Device Name: | Vectra Genisys |
### Indications for Use: (Page 1 of 2)
# For VMS-(Pulsed Mode, Burst Mode or FR Mode), Russian, Monophasic Hi-Volt (NMES) & Interferential and Premodulated (IFS)
Relaxation of Muscle Spasms
Prevention or retardation of disuse atrophy
Increasing local blood circulation
Muscle re-education
Maintaining or increasing range of motion
Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis
# Additionally for Microcurrent, Interferential, Premodulated (IFS), VMS-(Pulsed Mode, Burst Mode or FR Mode), Asymmetrical Biphasic (TENS), and Symmetrical Biphasic (TENS)
Symptomatic relief or management of chronic, intractable pain
Post-traumatic acute pain
Post-surgical acute pain
#### For FES
Stimulation of the muscles in the leg and ankle of partially paralyzed patients to provide flexion of the foot and thus improve the patient's gait
### For DC Continuous Mode
Relaxation of muscle spasm
# For EMG
To determination the activation timing of muscles for:
- a) retraining of muscle activation
b) coordination of muscle activation
An indication of the force produced by muscle for control and maintenance of muscle contractions
Relaxation muscle training
Muscle re-education
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-)
Division of General. Restorative,
and Neurological Devices
Prescription Us510(k) Number K062354
(Per 21 CFR 801.109) Over-The-Counter Use
(Optional Forma
(Optional Format 1-2-96)
{6}------------------------------------------------
| 510(k) Number | K062354 |
|---------------|----------------|
| Device Name: | Vectra Genisys |
# Indications for Use: (Page 2 of 2)
# For EMG triggered Stim
| Stroke rehab by muscle re-education |
|---------------------------------------------|
| Relaxation of muscle spasms |
| Prevention or retardation of disuse atrophy |
| Increase local blood circulation |
| Muscle re-education |
| Maintaining or increasing range of motion |
# For Ultrasound
| Application of therapeutic deep heat for the treatment of selected sub-chronic and | | |
|-----------------------------------------------------------------------------------------|--|--|
| chronic medical conditions such as: | | |
| 1. Relief of pain, muscle spasms and joint contractures | | |
| 2. Relief of pain, muscle spasms and joint contractures that may be associated<br>with: | | |
| a) Adhesive capsulitis | | |
| b) Bursitis with slight calcification | | |
| c) Myositis | | |
| d) Soft tissue injuries | | |
| e) Shortened tendons due to past injuries and scar tissues | | |
| 3. Relief of sub-chronic and chronic pain and joint contractures resulting from: | | |
| a) Capsular tightness | | |
| b) Capsular scarring | | |
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use J (Por 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Mark A. Milliman
(Division Sign. 47) (Division of General. Restorative, and Neurological Devices
**510(k) Number** K 06
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.