ULTRATHERM (908)
Device Facts
| Record ID | K964083 |
|---|---|
| Device Name | ULTRATHERM (908) |
| Applicant | Sie-Med, Inc. |
| Product Code | IMJ · Physical Medicine |
| Decision Date | Apr 3, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 890.5290 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Ultratherm 908i is a shortwave diathermy device intended to generate deep heat in body tissues for the treatment of medical conditions such as the relief of pain, muscle spasms and joint contractures.
Device Story
Shortwave diathermy device; generates deep heat in body tissues. Input: electrical power; Output: electromagnetic energy for therapeutic heating. Operated by clinicians in clinical settings. Replaces tube generator with compact transistor generator; reduces operating voltage and leakage. Features LED display and luminous band for output monitoring; touch-sensitive contacts for mode/intensity selection. Electrodes include non-flammable silicon rubber field electrodes and sheltered coil field electrodes. Enhances safety via reduced impulse peak (400W) and maximal effective output (200W). Assists in pain relief, muscle spasm reduction, and joint contracture management.
Clinical Evidence
No clinical data. Safety and effectiveness established via manufacturer's successful marketing history of similar device in Europe and Canada without recalls or significant adverse experience reports.
Technological Characteristics
Shortwave diathermy; transistor generator; maximal output 200W; impulse peak 400W. Components: non-flammable silicon rubber field electrodes, sheltered coil field electrodes, touch-sensitive contacts, LED control/display. Symmetrical and asymmetrical output sockets.
Indications for Use
Indicated for patients requiring deep tissue heating for the treatment of pain, muscle spasms, and joint contractures.
Regulatory Classification
Identification
A nonthermal shortwave therapy is a prescription device that applies to the body pulsed electromagnetic energy in the RF bands of 13.56 MHz or 27.12 MHz and that is intended for adjunctive use in the palliative treatment of postoperative pain and edema of soft tissue by means other than the generation of deep heat within body tissues as described in paragraph (a) of this section.
Special Controls
*Classification: Class II (special controls).* The device is classified as class II. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. (ii) Appropriate analysis/testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment. (iii) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Non-clinical performance testing must characterize the output waveform of the device and demonstrate that the device meets appropriate output performance specifications. The output characteristics and the methods used to determine these characteristics, including the following, must be determined: (A) Peak output power; (B) Pulse width; (C) Pulse frequency; (D) Duty cycle; (E) Characteristics of other types of modulation that may be used; (F) Average measured output powered into the RF antenna/applicator; (G) Specific absorption rates in saline gel test load or other appropriate model; (H) Characterization of the electrical and magnetic fields in saline gel test load or other appropriate model for each RF antenna and prescribed RF antenna orientation/position; and (I) Characterization of the deposited energy density in saline gel test load or other appropriate model. (iv) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device in its intended use. (v) Labeling must include the following: (A) Output characteristics of the device; (B) Recommended treatment regimes, including duration of use; and (C) A detailed summary of the clinical testing pertinent to the use of the device and a summary of the adverse events and complications. (vi) Nonthermal shortwave therapy devices marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.
Predicate Devices
- Ultratherm 808i (K862589)
Related Devices
- K162240 — ProMedTek Model C1400 Shortwave Diathermy Device · Promedtek, Inc. · Nov 17, 2016
- K161862 — ThermoPro · Zimmer Medizinsysteme GmbH · Nov 14, 2016
- K022846 — SW-180 SHORTWAVE THERAPY UNIT · Ito Co., Ltd. · Jan 28, 2003
- K042554 — AUTO THERM 390, MODEL ME 390 · Mettler Electronics Corp. · Oct 8, 2004
- K083433 — INTELECT SWD 100, MODEL 1600; SENIOR SOLUTIONS SHORTWAVE DIATHERMY, MODEL 1601 · Chattanooga Group · Mar 18, 2009