K161862 · Zimmer Medizinsysteme GmbH · IMJ · Nov 14, 2016 · Physical Medicine
Device Facts
Record ID
K161862
Device Name
ThermoPro
Applicant
Zimmer Medizinsysteme GmbH
Product Code
IMJ · Physical Medicine
Decision Date
Nov 14, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5290
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Indications for use in applying therapeutic deep heat in body tissues for the treatment of selected medical conditions such as disorders of the musculoskeletal system, muscle spasm, joint stiffness, contractures, and chronic inflammatory or infective conditions such as tenosynovitis, bursitis, synovitis and chronic inflammatory pelvic diseases. Generally accepted indications for use: Pain Relief; Contractures; Reduce Muscle Spasm; Localized increase Blood Flow; Chronic Inflammatory Conditions; Bursitis; Decrease Joint Stiffness; Tenosynovitis; Synovitis; Chronic Inflammatory Pelvic Disease
Device Story
ThermoPro is a shortwave diathermy device delivering electromagnetic energy at 27.12 MHz to induce deep tissue heating via eddy currents. The device consists of a metal enclosure with a touchscreen interface and a monode electrode applicator. A sine wave synthesizer and RF power amplifier generate energy in constant or pulsed modes. Operated by clinicians in a medical setting, the user selects therapy parameters (mode, pulse frequency, duty cycle, power, duration) via the touchscreen. A coupling indicator on the display assists the user in positioning the applicator relative to the patient. The device provides pain relief, reduces muscle spasms, increases local blood flow, and treats chronic inflammatory conditions. Clinical benefit is derived from the conversion of electrical energy into heat directly within the tissue.
Clinical Evidence
Bench testing only. The device was evaluated for compliance with ANSI/AAMI/ISO 10993-1 (biocompatibility), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-1-6 (usability), IEC 60601-2-3 (shortwave therapy performance), IEC 62366-1 (usability engineering), ISO 62304 (software lifecycle), and ISO 14971 (risk management). All components and systems met required specifications.
Technological Characteristics
Shortwave diathermy; 27.12 MHz frequency; 100W output power (200W pulsed); metal enclosure; monode electrode applicator; Type BF applied part; Class I protection; touchscreen interface; constant wave and pulsed modes; 100-240 VAC power source. Complies with IEC 60601-2-3, IEC 60601-1, and ISO 62304.
Indications for Use
Indicated for patients requiring therapeutic deep heat for musculoskeletal disorders, muscle spasms, joint stiffness, contractures, and chronic inflammatory or infective conditions including tenosynovitis, bursitis, synovitis, and chronic inflammatory pelvic diseases.
Regulatory Classification
Identification
A nonthermal shortwave therapy is a prescription device that applies to the body pulsed electromagnetic energy in the RF bands of 13.56 MHz or 27.12 MHz and that is intended for adjunctive use in the palliative treatment of postoperative pain and edema of soft tissue by means other than the generation of deep heat within body tissues as described in paragraph (a) of this section.
Special Controls
*Classification: Class II (special controls).* The device is classified as class II. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible.
(ii) Appropriate analysis/testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(iii) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Non-clinical performance testing must characterize the output waveform of the device and demonstrate that the device meets appropriate output performance specifications. The output characteristics and the methods used to determine these characteristics, including the following, must be determined:
(A) Peak output power;
(B) Pulse width;
(C) Pulse frequency;
(D) Duty cycle;
(E) Characteristics of other types of modulation that may be used;
(F) Average measured output powered into the RF antenna/applicator;
(G) Specific absorption rates in saline gel test load or other appropriate model;
(H) Characterization of the electrical and magnetic fields in saline gel test load or other appropriate model for each RF antenna and prescribed RF antenna orientation/position; and
(I) Characterization of the deposited energy density in saline gel test load or other appropriate model.
(iv) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device in its intended use.
(v) Labeling must include the following:
(A) Output characteristics of the device;
(B) Recommended treatment regimes, including duration of use; and
(C) A detailed summary of the clinical testing pertinent to the use of the device and a summary of the adverse events and complications.
(vi) Nonthermal shortwave therapy devices marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.
Predicate Devices
INTELECT SWD 100, MODEL 1600; SENIOR SOLUTIONS SHORTWAVE DIATHERMY, MODEL 1601 (K083433)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 14, 2016
Zimmer MedizinSysteme GmbH % Scott Blood Principal Consultant Quality & Regulatory Consulting 22 Nichols Street #2 Salem, Massachusetts 01970
Re: K161862
Trade/Device Name: ThermoPro Regulation Number: 21 CFR 890.5290 Regulation Name: Shortwave Diathermy Regulatory Class: Class II Product Code: IMJ Dated: October 5, 2016 Received: October 11, 2016
Dear Scott Blood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Michael J. Hoffmann -A
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K161862
Device Name ThermoPro
Indications for Use (Describe)
Indications for use in applying therapeutic deep heat in body tissues for the treatment of selected medical conditions such as disorders of the musculoskeletal system, muscle spasm, joint stiffness, contractures, and chronic inflammatory or infective conditions such as tenosynovitis, bursitis, synovitis and chronic inflammatory pelvic diseases.
Generally accepted indications for use:
o Pain Relief
- o Contractures
- o Reduce Muscle Spasm
- o Localized increase Blood Flow
- o Chronic Inflammatory Conditions
- o Bursitis
- o Decrease Joint Stiffness
- o Tenosynovitis
- o Synovitis
- o Chronic Inflammatory Pelvic Disease
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "z" and "i". Below the word "zimmer" is the phrase "MedizinSysteme" in a smaller, sans-serif font. The logo is in grayscale.
## 510(k) Summary
- 1. Basic Information-Submitter:
| 510(k) Owner: | Zimmer MedizinSysteme GmbH<br>Junkersstrasse 9<br>D-89231 Neu-Ulm<br>Germany |
|-------------------------------------|------------------------------------------------------------------------------|
| Establishment Registration: 8010720 | |
Official Contact: Mr. Armin Petraschka Project Manager Phone: +49-731-9761-140 Fax: +49-731-9761-4475 E-mail: a.petraschka@zimmer.de
Date Summary Prepared: 30 June 2016
- 2. Device Name:
Trade Name: ThermoPro Common Name: Shortwave Diathermy Classification Name: Diathermy, Shortwave, for use in applying therapeutic deep heat Regulation Number: 21 CFR 890.5290 Product Code: IMJ Classification: Class II
- 3. Predicate Devices:
INTELECT SWD 100, MODEL 1600; SENIOR SOLUTIONS SHORTWAVE DIATHERMY, MODEL 1601 - K083433
- 4. Device Description:
ThermoPro is a shortwave diathermy device for use in applying therapeutic deep heat for selected medical conditions by applying electromagnetic energy in the radio frequency band of 27.12 MHz. The ThermoPro is a device with a metal enclosure that includes a display touchscreen. The device also features an applicator that is a standard monode electrode. The applicator is connected to the device housing, where an electronic power module drives electromagnetic energy into applicator. This
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "z". Below the word "zimmer" is the text "MedizinSysteme" in a smaller, lighter font. The logo is in grayscale.
electromagnetic energy is generated using a sine wave synthesizer in combination with a RF power amplifier. The energy can be applied either in constant wave mode or in pulsed mode.
- 5. Indications for Use Statement:
Indications for use in applying therapeutic deep heat in body tissues for the treatment of selected medical conditions such as disorders of the musculoskeletal system, muscle spasm, joint stiffness, contractures, and chronic inflammatory or infective conditions such as tenosynovitis, bursitis, synovitis and chronic inflammatory pelvic diseases.
Generally accepted indications for use:
- Pain Relief ●
- . Contractures
- Reduce Muscle Spasm ●
- Localized increase Blood Flow ●
- Chronic Inflammatory Conditions ●
- . Bursitis
- . Decrease Joint Stiffness
- Tenosynovitis ●
- Synovitis ●
- Chronic Inflammatory Pelvic Disease ●
- 6. Technological Characteristics:
Shortwave diathermic therapy relies on the conversion of electric power into heat energy that takes place directly inside the tissue, as opposed to a typical heat treatment method in which heat (e.g. heat packs) is supplied from the outside of the tissue. A highfrequency field generates eddy currents in the treated tissues that lead to molecular excitations and convert the electrical energy into heat.
The energy can be applied either in constant wave mode or in pulsed mode. To run a therapy, the user selects the mode, pulse frequency and duty cycle (if applicable), power level and duration of therapy .The coupling is dependent on the distance between the applicator and the patient and can easily be controlled by supervising the coupling indicator on the display.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "z". Below the word "zimmer" is the text "MedizinSysteme" in a smaller, lighter font.
The ThermoPro shares the same intended use and the same basic characteristics and features as the predicate device, the Intelect SWD100. The table below summarizes the comparison of those critical features of the ThermoPro device with the predicate device:
| Technological<br>Characteristics | SUBJECT DEVICE<br>Zimmer MedizinSysteme GmbH<br>Thermo Pro | PREDICATE DEVICE<br>Chattanooga Group<br>Intelect SWD 100<br>K083433 |
|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Dimensions (in.) | 33.6 h x 22.9 w x 18.0 d | 45.0 h x 16.5 w x 16.1 d |
| Weight (lbs) | 81.6 | 60.0 |
| Enclosure material | Metal | Metal |
| Applicator | Standard:<br>Monode electrode | Standard:<br>Monode electrode<br>Optional:<br>Flexible electrode,<br>Diplode electrode,<br>Capacitive electrodes |
| Frequency (MHz) | 27.12 | 27.12 |
| Power Source | 100 - 240 VAC, 50 / 60Hz | 100 - 240 VAC, 50 / 60Hz |
| Output power (W) | 100 | 100 |
| Pulsed Output Power (W) | 200<br>Limited to 64W peak mean<br>power | 200<br>Limited to 64W peak mean<br>power |
| Applied Part Type<br>(applicator) | BF | BF |
| Protection Class | I | I |
| Touch screen Interface | Yes | Yes |
| External Memory | No | No |
| Modifiable waveform<br>parameters / Customizable<br>treatment parameters | Yes | Yes |
| Pulse Rate | CW mode<br>10 - 1000 Hz<br>Adjustable from:<br>10 Hz – 200 Hz in 10 Hz<br>increments<br>200 Hz - 800 Hz in 100 Hz<br>increments<br>1000 Hz | CW mode<br>10 - 800 Hz<br>Adjustable in 10 Hz increments |
| Duty Cycle (%) | 10 - 90% | 0.02 - 32%<br>preset - not adjustable |
| Pulse Width | 250 $\mu$ s – 90ms<br>in 250 $\mu$ s increments | 20 – 400 $\mu$ s<br>in 20 $\mu$ s increments |
| Technological<br>Characteristics | SUBJECT DEVICE<br>Zimmer MedizinSysteme GmbH<br>ThermoPro | PREDICATE DEVICE<br>Chattanooga Group<br>Intelect SWD 100<br>K083433 |
| | preset - not adjustable | |
| Accuracy (%) | $+/- 20%$ | $+/- 20%$ |
| Maximum Treatment Time<br>(mins) | 30 | 30 |
| Favorites / named user<br>defined programs | Yes | Yes |
| Calculated and displayed<br>energy | Yes | Yes |
| Displayed treatment time | Yes | Yes |
Premarket Notification, ThermoPro Proprietary and Confidential Information of Zimmer MedizinSysteme GmbH
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the letters "zi". Below the word "zimmer" is the text "MedizinSysteme" in a smaller, sans-serif font.
Any differences in their technological characteristics are explained to demonstrate in this submission that these differences do not raise any new questions of safety and effectiveness.
- 7. Performance data:
The ThermoPro device has been tested against and complies with the following voluntary standards:
| ANSI/AAMI/ISO 10993-1 | Biological evaluation of medical devices –<br>Part 1:Evaluation and testing |
|-----------------------|--------------------------------------------------------------------------------------|
| IEC 60601-1 | Medical Electrical Equipment, Part 1 |
| IEC 60601-1-2 | Medical Electrical Equipment, Part 1-2,<br>Electromagnetic Compatibility |
| IEC 60601-1-6 | Medical Electrical Equipment, Part 1-6,<br>Usability |
| IEC 60601-2-3 | Medical Electrical Equipment, Part 2-3,<br>Shortwave Therapy Equipment |
| IEC 62366-1 | Medical Devices – Part 1, Application of<br>Usability Engineering to Medical Devices |
| ISO 62304 | Software Development Life Cycle |
| ISO 14971 | Medical devices -- Application of risk<br>management to medical devices |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Zimmer MedizinSysteme. The logo consists of the word "zimmer" in a bold, sans-serif font, with a horizontal line above the "z". Below the word "zimmer" is the text "MedizinSysteme" in a smaller, lighter font. The logo is in grayscale.
Testing has been performed and all components, subassemblies and/or full devices and systems have met the required specifications for the completed tests.
- 8. 510(k) Summary:
Zimmer MedizinSysteme GmbH has demonstrated that the ThermoPro device is substantially equivalent to the predicate device listed above.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.