ELECTRIC WHEELCHAIR (QDWP-A01,QDWP-B02)

K251248 · Jinhua Qidian Vehicle Co., Ltd. · ITI · Jun 20, 2025 · Physical Medicine

Device Facts

Record IDK251248
Device NameELECTRIC WHEELCHAIR (QDWP-A01,QDWP-B02)
ApplicantJinhua Qidian Vehicle Co., Ltd.
Product CodeITI · Physical Medicine
Decision DateJun 20, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3860
Device ClassClass 2

Indications for Use

It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.

Device Story

Electric wheelchair (models QDWP-A01, QDWP-B02) provides mobility for disabled or elderly users. Device consists of frame (aluminum or steel), wheels, seat, armrests, battery (lithium or lead-acid), motor, and electronic controller. User operates device via joystick to control speed and direction; rear wheels are driven; front wheels facilitate rotation and maneuvering. Features include foldable design and flip-back armrests. Automatic electromagnetic brake system engages for safety. Used in indoor and outdoor environments. Output is physical transport of the user. Benefits include increased independence for individuals with limited mobility.

Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through testing against ISO 7176 series (stability, braking, energy consumption, dimensions, speed, strength, climatic, obstacle climbing, power/control systems, EMC, and battery/charger safety) and ISO 16840-10 (flame retardancy). Biocompatibility confirmed per ISO 10993-5, -10, and -23.

Technological Characteristics

Powered wheelchair with aluminum (QDWP-A01) or steel (QDWP-B02) frame. Energy source: 24V lithium-ion or lead-acid batteries. Drive system: rear-wheel drive with brushless or brushed DC motors. Control: joystick-operated electronic controller. Braking: automatic electromagnetic system. Connectivity: none. Sterilization: N/A. Compliance: ISO 7176 series, ISO 10993, IEC 60601-1-2.

Indications for Use

Indicated for disabled or elderly persons limited to a seated position requiring indoor and outdoor mobility assistance.

Regulatory Classification

Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION June 20, 2025 Jinhua Qidian Vehicle Co. Ltd % Ariel Xiang Consultant Shanghai Sungo Management Consulting Co. Ltd. 14th Floor, 1500# Century Avenue Shanghai, Shanghai 200122 China Re: K251248 Trade/Device Name: Electric Wheelchair (qdwp-a01,qdwp-b02) Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: April 23, 2025 Received: April 23, 2025 Dear Ariel Xiang: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251248 - Ariel Xiang Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K251248 - Ariel Xiang Page 3 Sincerely, Julia E. Slocomb 2025.06.20 -5 11:44:26 -04'00' for, Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K251248 Device Name ELECTRIC WHEELCHAIR (QDWP-A01,QDWP-B02) Indications for Use (Describe) It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} JINHUA QIDIAN VEHICLE CO.LTD #8.NO308. WeiyeRoad, Jiangtang, Wucheng, Jinhua, Zhejiang, China # 510(K) Summary **Document Prepared Date:** 2025/4/21 ## A. Applicant Name: JINHUA QIDIAN VEHICLE CO.LTD Address: #8.NO308. WeiyeRoad, Jiangtang, Wucheng, Jinhua, Zhejiang, China Contact Person: Qunbao Yang Email: 568438678@qq.com Submission Correspondent: Primary contact: Ms. Ariel Xiang Title: Senior Consultant Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: shouqiu.xiang@sungoglobal.com Secondary contact: Mr. Raymond Luo Title: Technical Director Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: zxfda@sungoglobal.com ## B. Device Trade Name: Electric wheelchair Common Name: Powered wheelchair Model(s): QDWP-A01, QDWP-B02 Regulatory Information Classification Name: Powered wheelchair Classification: Class II. Product code: ITI Regulation Number: 890.3860 Review Panel: Physical Medicine ## C. Predicate device 510Knumber: K113463 power wheelchair, PL00I SUZHOU KD Medical Appliance Co. Ltd. {5} Reference Device K211647 Electrically Powered Wheelchair, HP358EA Kunshan Hi-Fortune Health Products Co., Ltd D. Indications for use of the device It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. E. Device Description: This product consists of frame, wheels, seat, armrest, lithium battery, motor and controller with a lightweight and compact design. The wheelchair can easily fold and unfold for transportation or storage. The armrest can be flipped backward, which is convenient for the elderly to move. Users can drive the wheelchair by themselves through the control device. The QDWP-A01 uses lithium batteries as its power source, The QDWP-B02 uses lead acid batteries as its power source. The controller controls the drive left/right motor to realize the wheelchair forward, backward and turn functions. The frame of QDWP-A01 is aluminum, the frame of QDWP-B02 is steel. The front wheels are driven wheels suitable for rotation, acceleration, retrograde and other actions of the wheelchair. The front wheels movement will be achieved by thrust generated from the rear wheels. The rear wheels are driving wheels to control the speed and direction. When in use, the operator drives the motor of the rear wheel by operating the joystick to achieve the rear wheels movement. The motor and brake system are fixed on the rear wheels. The max loading of the device is 120KG. Only for one person sit. F. Comparison with predicate Device Table 1 General Comparison | Elements of Comparison | Subject Device (K251248) | Predicate Device (K113463) | Reference Device (K211647) | Remark | | --- | --- | --- | --- | --- | | Manufacturer | JINHUA QIDIAN VEHICLE CO.LTD | SUZHOU KD Medical Appliance Co. Ltd. | Kunshan Hi-Fortune Health Products Co., Ltd | -- | | Common or | Electric wheelchair | Power Wheelcha | Electrically Powered | -- | {6} | Usual name | | | ir | Wheelchair | | | --- | --- | --- | --- | --- | --- | | Model(s) | QDWP-A01 | QDWP-B02 | PL00I | HP358EA | -- | | Classification | Class II | | Class II | Class II | Same | | Classification regulation | 21 CFR890.3860 | | 21 CFR890.3860 | 21 CFR890.3860 | Same | | Product code | ITI | | ITI | ITI | Same | | Indications for use | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | S.E. | | Intended user | disabled or elderly person limited to a seated position. | | disabled or elderly person limited to a seated position. | disabled or elderly person limited to a seated position. | S.E. | {7} | | | position. | | | | --- | --- | --- | --- | --- | | Use condition | indoor and outdoor use | indoor and outdoor use | indoor and outdoor use | S.E | | Number of wheels | 4,including two front wheels and two rear Wheels | 4,including two front wheels and two rear Wheels | 6, including two front wheels and two rear wheels, two anti-tip wheels | S.E | | Function of wheels | Front wheels:driven wheels suitable for rotation, acceleration, retrograde Rear wheels: driving wheels to control the speed and direction | Front wheels:driven wheels suitable for rotation, acceleration, retrograde Rear wheels: driving wheels to control the speed and direction | Front wheels:driven wheels suitable for rotation, acceleration, retrograde Rear wheels: driving wheels to control the speed and direction Anti-tip wheels: preventing the wheelchair from tipping turning over when driving on the slope. Non-adjustable | S.E | | Movement control | By Joystick control | By Joystick control | By Joystick control | S.E | {8} | method | | | | | | --- | --- | --- | --- | --- | | Driving system | Direct drive on the rear wheels | Direct drive on the rear wheels | Direct drive on the rear wheels | S.E | | Brake system | Automatic electromagnetic brake system | Automatic electromagnetic brake system | Automatic electromagnetic brake system | S.E | | Braking distance | 1.2m | 1.5 m | 0.8m | **Analysis** : Minor difference on braking distance will not cause different performance. Shorter distance for braking will be more safety. | | Maximum degree | 10° | 9° | 10° | S.E | | Armrest | PP | PU | PU | **Analysis** : | {9} | | | | | | Different material used for parts in contact with user, which such differences will not impact the safety and effectiveness of the subject device as biocompatibility tests are carried out according to ISO 10993 series. | | --- | --- | --- | --- | --- | --- | | Main frame material | Aluminium alloy | Steel | Aluminum alloy | Aluminum alloy | QDWP-A01 is | {10} ```html | | | | | | Same as Predicate Device; QDWP-B02 is different. **Analysis** : Different material used for frame, that such difference will not impact the safety and effective ness of the subject device as the performance tests are conducted according to ISO 7176 | | --- | --- | --- | --- | --- | --- | {11} | | | | | | series. | | --- | --- | --- | --- | --- | --- | | Back cushion | Nylon | | PU foam covered by nylon fabric cloth | Terylene | S.E Same as Predicate Device. | | Seat cushion | Nylon | | PU foam covered by nylon fabric cloth | Terylene | | | Overall length | 1120mm | 970mm | 880mm | 990mm | Analysis : Minor difference on wheel chair dimension will not cause different performance. All safety and performance have been validated with the maximum rated weight dummy. | | Overall width | 600mm | 660mm | 570mm | 620mm | | | Folded Dimension (length*width*height) | 710*600*450mm | 680*460*770mm | 750*570*400mm | 620mmX550mmX850mm | | | | | | | | | | | | | | | | {12} | Front wheel size/type | 200mm×5 0mm PU Solid tire (8 inch) | 200mm×45mm (8 inch) PU Solid tire | 6" x 2"/PU Solid tire | 8" PU tire | S.E Same as Reference Device. | | --- | --- | --- | --- | --- | --- | | Rear wheel size/type | 10inch×2. 0inch Rubber Filler | 305mm×43mm (12 inch) PU Solid tire | 8"x 2.4"/ PU Solid tire | 12" Pneumatic tires | DWP-B02 is same as reference device. Analysis : Minor difference on dimension of wheels will not cause different performance. | | Max speed forward | 1.8 m/s (6.5km/h) | | Up to 6 km/h (1.6 m/s), adjustable | 1.4 m/s | Analysis : Minor difference on Max speed forward of | {13} | | | | | | wheels will not cause different performance. | | --- | --- | --- | --- | --- | --- | | Max loading weight | 120kg | | 114kg (251lbs) | 120kg | S.E | | Maximum distance of travel on the fully charged battery | 20.3km | 10.0km | 20 km | 10km | S.E | | Battery | Lithium-ion Battery 24V, 10Ah,2PCS | Lead Acid Battery, 12V, 12Ah,2PCS | li-ion battery pack; rechargeable, 24 VDC 20Ah | li-ion battery; rechargeable, 25.2VDC, 10.4Ah | Analysis: QDWP-A01 is same, QDWP-B02 is different. The performance of batteries and charger of device meet the Requirements of ISO | | Battery charger | Off-board charger, Input: 100-240V~1.8 A, 50/60Hz Output: 24V, 3.0A | Off-board charger, Input:100/240V,5 0/60Hz/55W Output:24V-29.6V,1.8A | Off-board charger Input: 110-240V, 50-60Hz, Output: 24 Vdc, 2A; | Off-board charger, Input :100-240VAC, 50-60Hz, 1.2-0.5A DC output: 24Vdc2A | | {14} | | | | | | 7176-25. | | --- | --- | --- | --- | --- | --- | | Motor | Brushless Motor, 24VDC, 250W | Brushed motor, 24VDC, 250W | Brushless DC motor; 24VDC; 180W; 2pcs | DC Brushed Motor, 24Vdc 250W;2pcs | QDWP-B02 is same as reference device. The type of QDWP-A01 is predicate device. **Analysis**: Minor difference on motor power. Both meet the require of ISO 7176-14 standard. It will not cause new safety and effective ness concerns. | | Electronic controller | 24V 40A | 24V35A | Brushless dual-drive rocker controller | 24V 35A | QDWP-B02 is same as reference device. **Analysis** | {15} | | | | | | Both of the control systems are evaluated according to standard ISO 7176-14:2008 and software validation requirement and there are no new safety and effective new concerns due to the difference. | | --- | --- | --- | --- | --- | --- | | Turning Radius | 900mm | 950mm | 800 mm | 500mm | Analysis: | {16} | | | | | | Minor difference on turning radius will not cause new safety and effective ness concerns. | | --- | --- | --- | --- | --- | --- | | Maximum obstacle climbing | 25mm | | 30 mm | 30mm | Analysis: Minor difference on obstacle climbing will not cause new safety and effective ness concerns. | Table 2 safety comparison {17} | Item | Proposed Device | Predicate Device | Results | | --- | --- | --- | --- | | Biocompatibility | All user directly contacting materials are compliance with ISO10993-5 ,ISO10993-10 ,ISO10993-23 requirements. | All user directly contacting materials are compliance with ISO10993-5 and ISO10993-10 requirements. | S.E. | | EMC | ISO7176-21 and IEC 60601-1-2 | ISO7176-21 and IEC 60601-1-2 | S.E. | | Performance | ISO7176 series | ISO7176 series | S.E. | | Label and labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | S.E. | ## G. Substantial Equivalence Discussion The proposed device and predicate device are complying to the same ISO standards, ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 16840-10, ISO 7176-21, ISO7176-22, ISO 7176-25, IEC 60601-1-2 and FDA guidance Submission for Power Wheelchair. The indications for use, design, structure, composition, principle of operation and energy source of proposed device and predicate device are identical. The dimensions and performance parameters are similar. The main differences are motor and controller. The design principles of the controller and Driving system are the same, and both meet the requirements of the ISO 7176-14. Software validation is carried out on both control systems. Brake system and speed control are designed in the same way as well, and both meet the requirements of the ISO 7176-3. Turning radius, Maximum obstacle climbing and Maximum safe operational incline are slightly different while such differences will not impact the safety and effectiveness of the subject device or raise new safety and effectiveness concerns as well as both meet the requirements of the ISO 7176-2 and ISO 7176-10. The materials in contact with human tissue are identical for both proposed device and predicate device. The biocompatibility of the proposed device and predicate device meet the requirements of the ISO 10993-5:2009 & ISO 10993-10:2021 & ISO 10993-23: 2021. {18} The flame-retardant test of the seat cushion/ backrest of both subject device and predicate device is carried out according to the ISO 16840-10 test. Therefore, both devices are assured to be under the same safety level. In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the device substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness. ## H. Product Performance ### a) Product Material Safety To ensure the safety of the wheelchair material, the bio-compatibility testing was done to the products. The testing was conducted according to the following standards: - ISO 10993-5 Biological evaluations of medical devices -- Part 5: Tests for In Vitro cytotoxicity - ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for skin sensitization - ISO 10993-23 Biological evaluation of medical devices - Part 23: Tests for irritation From the test result, we can find the material are safety and can meet the requirements. ### b) Performance of the products The following performance data were provided to verify that the subject device met all design specifications and provided support of the substantial equivalence determination. Risk Analysis developed in accordance with ISO 14971: 2019. #### Software validation - ISO 7176-1:2014 Wheelchairs - Part 1: Determination of static stability - ISO 7176-2:2017 Wheelchairs - Part 2: Determination of dynamic stability of electric wheelchairs - ISO 7176-3:2012 Wheelchairs - Part 3: Determination of effectiveness of brakes - ISO 7176-4:2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range - ISO 7176-5:2008 Wheelchairs - Part 5: Determination of dimensions, mass and maneuvering space - ISO 7176-6:2018 Wheelchairs - Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs - ISO 7176-7:1998 Wheelchairs - Part 7: Measurement of seating and wheel dimensions - ISO 7176-8:2014 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strength - ISO 7176-9:2009 Wheelchairs - Part 9: Climatic tests for electric wheelchairs {19} ISO 7176-10:2008 Wheelchairs - Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs ISO 7176-11:2012 Wheelchairs -- Part 11: Test dummies ISO 7176-13:1989 Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces. ISO 7176-14:2022 Wheelchairs -- Part 14: Power and control systems for electrically powered wheelchairs and scooters -- Requirements and test methods ISO 7176-15:1996 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling. ISO 16840-10:2021 Wheelchair seating - Part 10: Resistance to ignition of postural support devices - Requirements and test method ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters ISO 7176-22: 2014 Wheelchairs-Part 22: Set-up proceduresd ISO 7176-25: 2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs. Electromagnetic Compatibility Testing in accordance with IEC 60601-1-2:2020 8.11. I. Conclusion The proposed device and predicate device have the same technological characteristics, features, specifications, materials, mode of operation, and intended use of the device. To ensure all the key characteristic of the products can meet the requirements, the testing was done and all the results indicate the positive conclusion. Based on the analysis above, we confirm these two products are substantially equivalent.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...