K250904 · Wandercraft SAS · PHL · Oct 24, 2025 · Physical Medicine
Device Facts
Record ID
K250904
Device Name
Atalante X
Applicant
Wandercraft SAS
Product Code
PHL · Physical Medicine
Decision Date
Oct 24, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3480
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Atalante X is intended to perform ambulatory functions and mobility exercises, hands-free, in rehabilitation institutions under the supervision of a trained operator for the following populations: Individuals with hemiplegia due to cerebrovascular accident (CVA) Individuals with spinal cord injuries at levels C4 to L5 (SCI) Individuals with multiple sclerosis (MS) The operator must complete a training program prior to use of the device. Atalante X is intended to be used on adolescents of 18 years and older, and adults able to tolerate a stand-up position. The device is not intended for sports or stair climbing.
Device Story
Atalante X is a self-balancing, fully powered hip-knee-ankle lower body exoskeleton; 12 actuated degrees of freedom. Device uses dynamic-walking control to enable natural gait and power efficiency. Used in rehabilitation institutions; operated by trained clinicians. Input: patient physical state (stand-up tolerance). Output: assisted ambulation, mobility exercises (sit, stand, walk, turn, weight shift, squat). Provides visual/auditory feedback to operator/patient. Must be used with a safety system. Benefits: enables hands-free mobility for patients with SCI, CVA, or MS; facilitates rehabilitation exercises.
Clinical Evidence
Evidence includes four previous prospective studies (K232077), a retrospective study (n=33, SCI C4-T4), a prospective study (n=14, MS), and an interim report (n=4, MS). Real-world data from 978 patients (212 CVA, 136 SCI, 99 MS) supports safety. No serious adverse events reported; non-serious events included fatigue, dizziness, muscle/shoulder pain, and skin issues. Device demonstrated ambulatory function at speeds comparable to predicate.
Indicated for adolescents (18+) and adults with hemiplegia due to CVA, spinal cord injuries (C4-L5), or multiple sclerosis, who can tolerate a stand-up position. Must be used in rehabilitation institutions under supervision of a trained operator. Not for sports or stair climbing.
Regulatory Classification
Identification
A powered lower extremity exoskeleton is a prescription device that is composed of an external, powered, motorized orthosis that is placed over a person's paralyzed or weakened limbs for medical purposes.
Special Controls
In combination with the general controls of the FD&C Act, the Powered Exoskeleton is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Elements of the device materials that may contact the patient must be demonstrated to be biocompatible.
(2) Appropriate analysis/testing must validate electromagnetic compatibility/interference (EMC/EMI), electrical safety, thermal safety, mechanical safety, battery performance and safety, and wireless performance, if applicable.
(3) Appropriate software verification, validation, and hazard analysis must be performed.
(4) Design characteristics must ensure geometry and materials composition are consistent with intended use.
(5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must include:
(i) Mechanical bench testing (including durability testing) to demonstrate that the device will withstand forces, conditions, and environments encountered during use;
(ii) Simulated use testing (
*i.e.,* cyclic loading testing) to demonstrate performance of device commands and safeguard under worst case conditions and after durability testing;(iii) Verification and validation of manual override controls are necessary, if present;
(iv) The accuracy of device features and safeguards; and
(v) Device functionality in terms of flame retardant materials, liquid/particle ingress prevention, sensor and actuator performance, and motor performance.
(6) Clinical testing must demonstrate a reasonable assurance of safe and effective use and capture any adverse events observed during clinical use when used under the proposed conditions of use, which must include considerations for:
(i) Level of supervision necessary, and
(ii) Environment of use (
*e.g.,* indoors and/or outdoors) including obstacles and terrain representative of the intended use environment.(7) A training program must be included with sufficient educational elements so that upon completion of training program, the clinician, user, and companion can:
(i) Identify the safe environments for device use,
(ii) Use all safety features of device, and
(iii) Operate the device in simulated or actual use environments representative of indicated environments and use.
(8) Labeling for the Physician and User must include the following:
(i) Appropriate instructions, warning, cautions, limitations, and information related to the necessary safeguards of the device, including warning against activities and environments that may put the user at greater risk.
(ii) Specific instructions and the clinical training needed for the safe use of the device, which includes:
(A) Instructions on assembling the device in all available configurations;
(B) Instructions on fitting the patient;
(C) Instructions and explanations of all available programs and how to program the device;
(D) Instructions and explanation of all controls, input, and outputs;
(E) Instructions on all available modes or states of the device;
(F) Instructions on all safety features of the device; and
(G) Instructions for properly maintaining the device.
(iii) Information on the patient population for which the device has been demonstrated to have a reasonable assurance of safety and effectiveness.
(iv) Pertinent non-clinical testing information (
*e.g.,* EMC, battery longevity).(v) A detailed summary of the clinical testing including:
(A) Adverse events encountered under use conditions,
(B) Summary of study outcomes and endpoints, and
(C) Information pertinent to use of the device including the conditions under which the device was studied (
*e.g.,* level of supervision or assistance, and environment of use (*e.g.,* indoors and/or outdoors) including obstacles and terrain).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 24, 2025
Wandercraft SAS
Mélanie Combes
Head of Regulatory and Quality Affairs
88 Rue de Rivoli
Paris, Ile-de-France, 75004
France
Re: K250904
Trade/Device Name: Atalante X
Regulation Number: 21 CFR 890.3480
Regulation Name: Powered lower extremity exoskeleton
Regulatory Class: Class II
Product Code: PHL
Dated: September 22, 2025
Received: September 22, 2025
Dear Mélanie Combes:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250904 - Mélanie Combes
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K250904 - Mélanie Combes
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Atalante X
Page 11 of 54
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K250904 | ? |
| Please provide the device trade name(s). | | ? |
| Atalante X | | |
| Please provide your Indications for Use below. | | ? |
| Atalante X is intended to perform ambulatory functions and mobility exercises, hands-free, in rehabilitation institutions under the supervision of a trained operator for the following populations:
* Individuals with hemiplegia due to cerebrovascular accident (CVA)
* Individuals with spinal cord injuries at levels C4 to L5 (SCI)
* Individuals with multiple sclerosis (MS)
The operator must complete a training program prior to use of the device.
Atalante X is intended to be used on adolescents of 18 years and older, and adults able to tolerate a stand-up position.
The device is not intended for sports or stair climbing. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
Page 1 of 6
510(k) Summary
K250904
1. Submission Sponsor
Wandercraft SAS
88 rue de Rivoli
75004 Paris, France
Contact: Matthieu Masselin
Title: Chief Executive Officer
2. Submission Correspondent
Wandercraft SAS
88 rue de Rivoli
75004 Paris, France
Email: reglementaire@wandercraft.health
Contact: Mélanie Combes
Title: Head of Regulatory and Quality Affairs
3. Date Prepared
March 25th, 2025
4. Device Identification
Trade/Proprietary Name: Atalante X
Common/Usual Name: Powered Exoskeleton
Classification Name: Powered lower extremity exoskeleton
Regulation Number: 890.3480
Product Code: PHL
Class: II
Classification Panel: Neurological and Physical Medicine Devices (OHTS)
{5}
Page 2 of 6
# 5. Legally Marketed Predicate and Reference Devices
| | Predicate Device | Reference Device |
| --- | --- | --- |
| Device name | Atalante X | EksoNR |
| 510(K) number | K232077 | K220988 |
| Regulation Number | 21 CFR 890.3480 | 21 CFR 890.3480 |
| Regulation Name | Powered Exoskeleton | Powered Exoskeleton |
| Product code | PHL | PHL |
| Review Panel | Neurology | Neurology |
| Manufacturer | Wandercraft SAS | Ekso Bionics Inc. |
# 6. Indication for Use Statement
Atalante X is intended to perform ambulatory functions and mobility exercises, hands-free, in rehabilitation institutions under the supervision of a trained operator for the following populations:
- Individuals with hemiplegia due to cerebrovascular accident (CVA)
- Individuals with spinal cord injuries at levels C4 to L5 (SCI)
- Individuals with multiple sclerosis (MS)
The operator must complete a training program prior to use of the device.
Atalante X is intended to be used on adolescents of 18 years and older and adults, able to tolerate a stand-up position.
The device is not intended for sports or stair climbing.
# 7. Device Description
Atalante X is a completely self-balancing walking system for people with mobility disabilities. It is a fully powered hip-knee-ankle lower body exoskeleton with 12 actuated degrees of freedom. Atalante X is self-balancing and includes dynamic-walking control. Dynamic-walking allows the Atalante X to consume significantly less power and have a more natural gait.
# 8. Substantial Equivalence Discussion
The following table compares Atalante X to the predicate and reference devices with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate.
{6}
Table 1 - Comparison of Characteristics
| FEATURE | Subject Device Atalante X | Predicate Device Atalante X | Reference Device EksoNR |
| --- | --- | --- | --- |
| 510(k) Number | K250904 | K232077 | K220988 |
| Manufacturer | Wandercraft SAS | Wandercraft SAS | Ekso Bionics Inc. |
| Product Code | PHL | PHL | PHL |
| Regulation | 890.3480 | 890.3480 | 890.3480 |
| Regulation Name | Powered Exoskeleton | Powered Exoskeleton | Powered Exoskeleton |
| Indications for Use | Atalante X is intended to perform ambulatory functions and mobility exercises, hands-free, in rehabilitation institutions under the supervision of a trained operator for the following populations: - Individuals with hemiplegia due to cerebrovascular accident (CVA). - Individuals with spinal cord injuries at levels C4 to L5 (SCI). - Individuals with multiple sclerosis (MS). The operator must complete a training program prior to use of the device. Atalante X is intended to be used on adolescents of 18 years and older, and adults able to tolerate a stand-up position. The device is not intended for sports or stair climbing. | Atalante X is intended to perform ambulatory functions and mobility exercises, hands-free, in rehabilitation institutions under the supervision of a trained operator for the following populations: - Individuals with hemiplegia due to cerebrovascular accident (CVA). - Individuals with spinal cord injuries at levels T5 to L5 (SCI). The operator must complete a training program prior to use of the device. Atalante X is intended to be used on adolescents of 18 years and older, and adults able to tolerate a stand-up position. The device is not intended for sports or stair climbing. | The EksoNR is intended to perform ambulatory functions in rehabilitation institutions under the supervision of a trained physical therapist for the following populations: - Individuals with multiple sclerosis (upper extremity motor function of at least 4/5 in at least one arm). - Individuals with acquired brain injury, including traumatic brain injury and stroke (upper extremity motor function of at least 4/5 in at least one arm). - Individuals with spinal cord injuries at levels T4 to L5 (upper extremity motor function of at least 4/5 in both arms). - Individuals with spinal cord injuries at levels of C7 to T3 (ASIA D with upper extremity motor function of at least 4/5 in both arms). The therapist must complete a training program prior to use of the device. The devices are not intended for sports or stair climbing. |
| Population | Adolescents of 18 years and older, and adults | Adolescents of 18 years and older, and adults | Adolescents of 18 years and older, and adults |
| Environment | Rehabilitation institutions | Rehabilitation Institutions | Rehabilitation Institutions |
{7}
| FEATURE | Subject Device
Atalante X | Predicate Device
Atalante X | Reference Device
EksoNR |
| --- | --- | --- | --- |
| Body Coverage | Worn over legs and upper body
with rigid torso | Worn over legs and upper body
with rigid torso | Worn over legs and upper body
with rigid torso |
| Size of
Components | Adjustable upper leg, lower leg.
Non-adjustable hip width | Adjustable upper leg, lower leg.
Non-adjustable hip width | Adjustable upper leg, lower leg,
and hip width |
| Control Unit | Control unit integrated into the
torso | Control unit integrated into the
torso | Control unit integrated into the
torso |
| Mobility Aid | None | None | Walker, Crutches, Cane |
| User Mobility | Sit, stand, walk, turn, exercise
(weight shift, squat),
repositioning | Sit, stand, walk, turn, exercise
(weight shift, squat),
repositioning | Sit, stand, walk, and turn |
| Walking Speed | ~2km/hr | ~2km/hr | ~2 km/hr |
| Type of Surface | Smooth | Smooth | Smooth, cement, carpet |
| Range of
Motion | Hip: 90° flexion,
5° extension;
17° abduction,
10° adduction;
10° medial rotation,
20° lateral rotation
Knee: 110° Flexion;
-5°Extension
Ankle: 0° dorsiflexion,
9° plantar flexion | Hip: 90° flexion,
5° extension;
17° abduction,
10° adduction;
10° medial rotation,
20° lateral rotation
Knee: 110° Flexion;
-5°Extension
Ankle: 0° dorsiflexion,
9° plantar flexion,
18° pronation and supination | Hip: 135° flexion to
20° extension
Knee: 130° flexion to 0°
extension
Ankles: 10° flexion to 10°
extension |
| Rechargeable
Battery | Rechargeable Lithium-ion
battery
46.8 V, 6,4Ah
Usage duration: 2h of
continuous usage per charge | Rechargeable Lithium-ion
battery
46.8 V, 6,4 Ah
Usage duration: 2h of
continuous usage per charge | Rechargeable Lithium-ion
battery.
48.1V, 30A peak current
Usage duration: 1h of
continuous usage per charge |
| Battery Charge
Time | 2h 30min | 2h30min | 1h |
| Training
Program | Yes | Yes | Yes |
| Certification
Program | Yes | Yes | Yes |
| User Feedback | Provides visual feedback on the
handheld controller and physio
interface, and auditory
feedback | Provides visual feedback on the
handheld controller and physio
interface, and auditory
feedback | Provides visual feedback on the
handheld controller and
auditory feedback |
| Fall Detection
and Mitigation | Must be used in combination
with safety system | Must be used in combination
with safety rail | None |
| Failsafe
Feature | (use of safety system) | (use of safety rail) | In the event of power failure
knees become locked and hips
free (similar to typical passive
leg braces) |
Page 4 of 6
{8}
| FEATURE | Subject Device
Atalante X | Predicate Device
Atalante X | Reference Device
EksoNR |
| --- | --- | --- | --- |
| Electrical Testing | IEC 60601-1 | ANSI AAMI ES60601-1 | IEC 60601-1 |
| EMC Testing | IEC 60601-1-2
AIM 7351731 | IEC 60601-1-2
AIM 7351731 | IEC 60601-1-2 |
## 9. Non-Clinical Performance Data
To demonstrate safety and effectiveness of Atalante X and to show substantial equivalence to the predicate and reference devices, Wandercraft SAS completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. Atalante X passed the testing in accordance with internal requirements, national standards, and international standards shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate:
- Cytotoxicity testing per ISO 10993-5: Passed
- Irritation and Sensitization testing per ISO 10993-10: Passed
- Electrical safety testing per IEC 60601-1: Passed
- Electromagnetic Interference (EMI) testing per IEC 60601-1-2: Passed
- Software verification and validation per IEC 62304 and FDA Guidance – the current version has been fully validated
- System (Device) Verification and Validation: Passed
- Electronics Sub-system Verification: Passed
- Mechanics Sub-system Verification: Passed
- Cycling Testing: Passed
- Thermal Testing: Passed
- Useful Life Testing – supports 5 years
- Transportation Testing per ASTM D4169 – demonstrates package integrity is maintained: Passed
## 10. Clinical Performance Data
The current device was evaluated in four previous prospective studies to demonstrate its safety and effectiveness for the indicated use (K232077). This submission builds upon those studies with additional clinical trials to support its safety and effectiveness for an expanded indication.
A retrospective study included 33 patients with spinal cord injuries (SCI) at levels C4 to T4 who used the Atalante X device as part of their rehabilitation program. Additionally, a prospective study involving 14 patients with multiple sclerosis (MS) was conducted. The device was shown to facilitate ambulatory function at speeds comparable to the predicate device, while allowing the upper limbs to remain free for other tasks. Furthermore, an interim report from a second prospective clinical trial involving four patients with MS provided supplementary safety and performance data.
Across these studies, non-serious adverse events included fatigue/dizziness, muscle pain, shoulder pain and minor skin issues. No serious adverse events were observed.
Page 5 of 6
{9}
These findings are further supported by real-world data from over 978 patients who completed at least three sessions with Atalante X, including 212 CVA patients, 136 SCI patients (7 at levels C2-C3, 59 at levels C4-T4, 67 at levels T5-L5 and 3 with unknown lesion levels) and 99 MS patients.
In conclusion, the clinical data demonstrate that the Atalante X can be used safely and effectively by patients with SCI at levels C4-T4 and by those with MS.
## 11. Training
Prior to the first use of Atalante X, to ensure safe and effective operation of the device, the operator must complete a certification training. To optimize the integration of the technology in the clinical setting, such training is then followed by frequent visits by Wandercraft Customer Care to further support the operator with the exoskeleton. The training program is offered for as many operators as agreed in the contract signed with the customer.
Training is composed by a theoretical part including presentation and description of the whole system and its mode of operations and a hands-on practice. This latter will be performed both in pairs – with an operator in the role of “operator” and another operator in the role of a “patient” – and in individual sessions. The double role allows the operators to understand the device from both the operator’s and patient’s perspectives.
An operator is considered certified, i.e., with proven competence to safely operate Atalante X alone with the patient, only after fulfilling all eligibility requirements and pass the certification evaluation.
## 12. Statement of Substantial Equivalence
Atalante X has the same intended use as the predicate and reference devices, Atalante X (K232077) and EksoNR (K220988), and the same or similar technological characteristics. Compared to the predicate device Atalante X (K232077), the new version differs in its indications for use. The new version includes expanded indications for individuals with high spinal cord injuries from C4 to L5 and for individuals with multiple sclerosis, as supported by clinical studies. These differences in indications and technological characteristics do not raise new or different questions of safety or effectiveness. Performance testing has demonstrated that the updated Atalante X is as safe and effective as the predicate device. Therefore, Atalante X is substantially equivalent to the predicate device.
Page 6 of 6
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.