PHL · Powered Exoskeleton

Physical Medicine · 21 CFR 890.3480 · Class 2

Overview

Product CodePHL
Device NamePowered Exoskeleton
Regulation21 CFR 890.3480
Device ClassClass 2
Review PanelPhysical Medicine

Identification

A powered lower extremity exoskeleton is a prescription device that is composed of an external, powered, motorized orthosis that is placed over a person's paralyzed or weakened limbs for medical purposes.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the Powered Exoskeleton is subject to the following special controls:

*Classification.* Class II (special controls). The special controls for this device are:(1) Elements of the device materials that may contact the patient must be demonstrated to be biocompatible. (2) Appropriate analysis/testing must validate electromagnetic compatibility/interference (EMC/EMI), electrical safety, thermal safety, mechanical safety, battery performance and safety, and wireless performance, if applicable. (3) Appropriate software verification, validation, and hazard analysis must be performed. (4) Design characteristics must ensure geometry and materials composition are consistent with intended use. (5) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must include: (i) Mechanical bench testing (including durability testing) to demonstrate that the device will withstand forces, conditions, and environments encountered during use; (ii) Simulated use testing ( *i.e.,* cyclic loading testing) to demonstrate performance of device commands and safeguard under worst case conditions and after durability testing;(iii) Verification and validation of manual override controls are necessary, if present; (iv) The accuracy of device features and safeguards; and (v) Device functionality in terms of flame retardant materials, liquid/particle ingress prevention, sensor and actuator performance, and motor performance. (6) Clinical testing must demonstrate a reasonable assurance of safe and effective use and capture any adverse events observed during clinical use when used under the proposed conditions of use, which must include considerations for: (i) Level of supervision necessary, and (ii) Environment of use ( *e.g.,* indoors and/or outdoors) including obstacles and terrain representative of the intended use environment.(7) A training program must be included with sufficient educational elements so that upon completion of training program, the clinician, user, and companion can: (i) Identify the safe environments for device use, (ii) Use all safety features of device, and (iii) Operate the device in simulated or actual use environments representative of indicated environments and use. (8) Labeling for the Physician and User must include the following: (i) Appropriate instructions, warning, cautions, limitations, and information related to the necessary safeguards of the device, including warning against activities and environments that may put the user at greater risk. (ii) Specific instructions and the clinical training needed for the safe use of the device, which includes: (A) Instructions on assembling the device in all available configurations; (B) Instructions on fitting the patient; (C) Instructions and explanations of all available programs and how to program the device; (D) Instructions and explanation of all controls, input, and outputs; (E) Instructions on all available modes or states of the device; (F) Instructions on all safety features of the device; and (G) Instructions for properly maintaining the device. (iii) Information on the patient population for which the device has been demonstrated to have a reasonable assurance of safety and effectiveness. (iv) Pertinent non-clinical testing information ( *e.g.,* EMC, battery longevity).(v) A detailed summary of the clinical testing including: (A) Adverse events encountered under use conditions, (B) Summary of study outcomes and endpoints, and (C) Information pertinent to use of the device including the conditions under which the device was studied ( *e.g.,* level of supervision or assistance, and environment of use (*e.g.,* indoors and/or outdoors) including obstacles and terrain).

Recent Cleared Devices (20 of 24)

Showing 20 most recent of 24 cleared devices.

RecordDevice NameApplicantDecision DateDecision
K250904Atalante XWandercraft SASOct 24, 2025SESE
K241822ReWalk® 7 Personal Exoskeleton (50-20-0005)Rewalk Robotics Ltd. Dba LifewardMar 12, 2025SESE
K233695Medical HAL Lower Limb Type (HAL-ML)Cyberdyne, Inc.May 7, 2024SESE
K232077Atalante XWandercraft SASDec 13, 2023SESE
K221696ReWalk P6.0Re Walk Robotics , Ltd.Mar 2, 2023SESE
K221859AtalanteWandercraft SASDec 29, 2022SESE
K220988EksoNREkso Bionics, Inc.Jun 9, 2022SESE
K213452GEMS-HSamsung Electronics Co., Ltd.Apr 21, 2022SESE
K201473ExoAtlet-IIExoatlet Asia Co. , Ltd.Jul 16, 2021SESE
K201559HAL for Medical Use(Lower Limb type)Cyberdyne, Inc.Oct 2, 2020SESE
K201539Keeogo Dermoskeleton SystemB-Temia, Inc.Sep 9, 2020SESE
K200574EksoNREkso Bionics, Inc.Jun 19, 2020SESE
K200032ReWalk P6.0Re Walk Robotics , Ltd.May 26, 2020SESE
K190337ReWalk RestoreRe Walk Robotics , Ltd.Jun 3, 2019SESE
K183152PhoenixUs Bionics, Inc. (Dba Suitx)Apr 17, 2019SESE
K181294Honda Walking Assist DeviceHonda Motor Company, Ltd.Dec 14, 2018SESE
K173530Indego(R)Parker-Hannifin CorporationJan 31, 2018SESE
K171909HAL for Medical Use (Lower Limb Type)Cyberdyne, Inc.Dec 17, 2017SESE
K171334IndegoParker Hannifin CorporationSep 8, 2017SESE
K160987ReWalk(TM)Rewalk Robotics, Inc.Jul 22, 2016SESE

Top Applicants

Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...