The intended function and use is to provide mobility to persons limited to a sitting position that have the capability of operating a powered wheelchair.
Device Story
Battery-powered, motorized mobility vehicles (Nutron and Pronto series) modified with MKV-NX series electronic controllers. Controller acts as microcomputer-based motion control device; activates and regulates wheelchair movement. User interface allows adjustment and programming of performance characteristics (e.g., speed). Two versions: MKV-NX (full programming) and MKV-NX LP (limited to forward, turning, and reverse speeds). Programmed via external MIV remote programmer; software prevents wheelchair operation during programming. Used by patients in various settings to provide mobility and transportation. Output is controlled mechanical motion of the wheelchair, directly affecting patient mobility and independence.
Clinical Evidence
Bench testing only. Device performance validated against ANSI/RESNA WC/14 (1991) and ISO 7176: 1993(E) standards for power and control systems of electric wheelchairs.
Technological Characteristics
Electronic, microcomputer-based motion control device. Battery-powered. Two versions: MKV-NX (full programming) and MKV-NX LP (limited programming). Connectivity via MIV remote programmer. Complies with ANSI/RESNA WC/14 and ISO 7176 standards.
Indications for Use
Indicated for physically challenged persons limited to a sitting position who have the capability of operating a powered wheelchair.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
K253849 — Electric wheelchair (JJW-7002, JJW-7003) · ZheJiang J&J Mobility Co., Ltd. · Mar 20, 2026
Submission Summary (Full Text)
{0}------------------------------------------------
# KOZI680
10Fd
### 510(k) SUMMARY
Nutron and Pronto Series Power Wheelchairs (M50, M51, M71, M91) with MKV-NX series of wheelchair controllers
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared.
Invacare Corporation One Invacare Way PO Box 4028 Elyria, Ohio 44036 Phone: (440) 329-6000 Facsimile: (440) 365-4558
ﺗﻢ ﺗﻘ
> Contact Person: Rae Ann Farrow
Manager, Regulatory Compliance
Date Prepared: May 21, 2002
#### Name of Device and Name/Address of Sponsor
Nutron and Pronto Series Power Wheelchairs (M50, M51, M71, M91) with MKV-NX series of wheelchair controllers
Invacare Corporation One Invacare Way Elyria, Ohio 44036-2028 Phone: (440) 329-6000 Facsimile: (440) 365-4558
### Common or Usual Name
Power Wheelchair
## Classification Name
Wheelchair, Powered
#### Predicate Devices
The Invacare powered wheelchairs with the MK-NX controller are substantially equivalent to devices that are currently being legally marketed within the United States, including Invacare Xterra GT power wheelchair (K012909, 10/24/2001) and the Invacare Pronto M71 power wheelchair (K012927).
#### Intended Use
The intended function and use is to provide mobility to persons limited to a sitting position that have the capability of operating a powered wheelchair.
{1}------------------------------------------------
K 021680
2 OF d
## Technological Characteristics and Substantial Equivalence
#### A. Device Description
The purpose of this pre-market notification is to request marketing clearance from FDA to change the controller of certain power wheelchairs models, manufactured by Invacare Corporation, which have already been granted marketing clearance by FDA under separate submissions
The controller change applies to the Invacare Pronto Powered Wheelchair Series (K012927, 9/19/2001), and the Nutron Powered Wheelchair Series (K900565, 2/15/1990). Each of these devices are battery powered, motorized mobility vehicles. Their intended function and use, which has not changed, is to provide mobility and transportation to physically challenged persons that may be restricted to a seated position.
The controller for the Nutron and Pronto Series of Wheelchairs is being modified from the Invacare MKIV-RII controller to the Invacare MKV-NX controller. The Invacare MKV-NX controller, which is the subject of this premarket notification, is an electronic, microcomputer based. motion control device designed for use with Invacare powered wheelchairs. The intended function of the controller is to activate and control powered wheelchair motions. Additionally, it provides a method of adjusting, selecting and programming certain powered wheelchair performance characteristics to better suit the specific control needs of the user.
Two versions of the controller will be available. These are the MKV-NX and the MKY-NX LP. The MKV NX version has full programming capability, while the MKV-NX LP version has limited programming capability. Programming capability for the LP version is limited to forward speed, turning speed, and reverse speed only. Otherwise the controllers are identical.
The Invacare MKV-NX and MKV-NX LP controllers are programmed using the Invacare MIV remote programmer, previously cleared with the Model MCC-MKIV Micro Computer Control (K940972, 6/2/1994). The controller software and electronics will not allow wheelchair operation during programming.
#### B. Substantial Equivalence
The Invacare powered wheelchairs with the MK-NX controller are substantially equivalent to devices that are currently being legally marketed within the United States, including Invacare Xterra GT power wheelchair (K012909, 10/24/2001) and the Invacare Pronto M71 power wheelchair (K012927).
### PERFORMANCE DATA
The Invacare Elevating Seat meets the applicable requirements specified in the Rehabilitation Engineering Society of North America (RESNA) Standard ANSI/RESNA WC/14 (1991) and ISO Standard ISO 7176: 1993(E), "ISO Standard, Wheelchairs -Requirements and Test Methods for the Power and Control Systems of Electric Wheelchairs."
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Rae Ann Farrow Manager, Regulatory Compliance Invacare Corporation One Invacare Way PO Box 4028 Elyria, OH 44036-2125
Re: K021680
Trade Name: Invacare Powered Wheelchairs, Models M50, M51, M71 and M91 Regulatory Number: 890.3860 Regulatory Name: Powered wheelchair Regulatory Class: II Product Code: ITI Dated: May 21, 2002 Received: May 22, 2002
JUN 1 8 2002
Dear Ms. Farrow:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Rae Ann Farrow
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 310(k) Number (if known): TBD
Device Name: Nutron and Pronto Series Power Wheelchairs (M50; M71, M91) with MKV-NX series wheelchair controllers
## Indications For Use:
The intended function and use is to provide mobility to persons limited to a sitting position that have capability of operating a powered wheelchair.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
**Prescription Use** OR
(Per 21 CFR 801.109) Mark n Milken
Over-The-Counter Use
(Optional Format 1-2-96)
(Division Sign-Off)
Division of General, Restorative
and Neurological Devices
510(k) Number
K021680
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.