M71 POWER WHEELCHAIR

K012927 · Invacare Corp. · ITI · Sep 19, 2001 · Physical Medicine

Device Facts

Record IDK012927
Device NameM71 POWER WHEELCHAIR
ApplicantInvacare Corp.
Product CodeITI · Physical Medicine
Decision DateSep 19, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3860
Device ClassClass 2

Intended Use

The intended use of the Invacare Model M71 Power Wheelchair is to provide mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair.

Device Story

The Invacare M71 is a rigid, non-folding, mid-wheel drive power wheelchair designed for mobility assistance. It features a steel tube frame, front-wheel walking beam suspension for obstacle negotiation, and rear anti-tipper casters. The device is powered by two 12V deep-cycle gel cell batteries driving two right-angle gear motors via a direct drive system. It can be disassembled into six pieces without tools for transport. The user operates the chair via an integrated controller. The device provides mobility to patients in various settings, enhancing independence by allowing navigation over small obstacles like curbs or ramps. It is maintenance-free regarding drive components, as it lacks belts or chains.

Clinical Evidence

Bench testing only. The device was tested in accordance with ISO EMC Draft Standard 7176-14 for electromagnetic compatibility. It met all required performance criteria.

Technological Characteristics

Rigid, non-folding steel frame (1 1/4" diameter x .07 wall tubing). Mid-wheel drive with front-wheel walking beam suspension. Powered by two 12V deep-cycle gel cell batteries. Direct drive system with two right-angle gear motors. On-board charger included. Disassembles into 6 pieces without tools. Color-coded joints.

Indications for Use

Indicated for individuals limited to a sitting position who possess the physical and cognitive capability to operate a powered wheelchair.

Regulatory Classification

Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K012927 ## SEP 1 9 2001 #### 510(k) SUMMARY ### Invacare Corporation's Model M71 Power Wheelchair ## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared. Invacare Corporation One Invacare Way PO Box 4028 Elyria, Ohio 44036 Phone: (440) 329-6000 Facsimile: (440) 365-4558 Contact Person: Rae Ann Farrow Manager, Regulatory Compliance Date Prepared: August 30, 2001 #### Name of Device and Name/Address of Sponsor Model M71 Power Wheelchair Invacare Corporation One Invacare Way Elyria, Ohio 44036-2028 Phone: (440) 329-6000 Facsimile: (440) 365-4558 #### Common or Usual Name Power Wheelchair #### Classification Name Wheelchair, Powered #### Predicate Devices Invacare Corporations' Action Arrow Front Wheel Drive Power Wheelchairs (K991168 June 25, 1999), Invacare Corporations' Ranger II M6 Power Wheelchair (K002157, and Permobil's Chairman Front Wheel Drive Power Wheelchair August 25, 2000), (K960951, April 4, 1997). {1}------------------------------------------------ #### Intended Use The intended use of the Invacare Model M71 Power Wheelchair is to provide mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair. #### Technological Characteristics and Substantial Equivalence #### A. Device Description The Invacare Model M71 power wheelchair is a battery powered, motor driven device with the intended function and use of providing mobility to those persons limited to a sitting position that have the capability of operating a power wheelchair. It is a rigid or "non- folding" type power wheelchair, with mid wheel drive capability. The wheelchair frame is of rigid or non-folding type design. The wheelchair frame is of rigid or non-folding type design. The frame consists predominately of round 1 1/4 " diameter x .07 wall, steel tubing. The frame is of welded construction. The chair can be disassembled into 6 pieces without the use of tools for transport in vehicles that are not equipped with wheel chair lifts and ramps. To easy re-assembly, the mating parts are color-coded or labeled at the joints. Painted shrouds are used to conceal the wheelchair frame, motors and controller. The wheelchair is designed with front wheel, walking beam type suspension. That is, the front swivel casters are mounted to two separate front extension frame members, and are thus, independent of each other. Each extension frame member pivots independently in a vertical plane about a horizontal axis. This allows the chair to better negotiate small obstacles such as curbs, or ramps, from a complete stop. The rear casters pivot as well, and also serve as wheelchair anti-tippers. The Invacare M71 wheelchair incorporates a direct, drive, which includes two separate right angle gear motors. Because it is a direct drive system, there are no belts, pulleys, chains, sprockets or the like, to periodically adjust or maintain. Both gearboxes are sealed and maintenance free. The Invacare M71 wheelchair is powered by two 12 volt, U-1 deep cycle gel cell batteries. The operating range; between battery re-charging is approximately 12 miles. Range will vary with battery condition, surface terrain and operator weight. The On Board Charger is standard. #### B. Substantial Equivalence The Invacare Model M71 Power Wheelchair is substantially equivalent to Invacare Corporations' Action Arrow K991168, June 25, 1999), Invacare Corporations' Ranger II M6 Power Wheelchair (K002157, August 25, 2000), and Permobil's Chairman Front Wheel Drive Power Wheelchair (K960951, April 4, 1997). {2}------------------------------------------------ #### PERFORMANCE DATA As required by FDA's July 26, 1995, draft publication entitled "Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorized Three- Wheeled Vehicles", the Invacare M71 Power Wheelchair was tested in accordance with ISO EMC Draft Standard 7176-14 (Titled "Draft ISO EMC Group Proposal: Electromagnetic Compatibility Addition" And Dated April 3, 1995) for powered wheelchairs and motorized scooters. In all instances, the Invacare M71 Power Wheelchair met the required performance criteria and functioned as intended. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing right, with flowing lines suggesting movement or connection. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 1 9 2001 Ms. Rae Ann Farrow Manager, Regulatory Compliance Invacare Corporation One Invacare Way Elyria, Ohio 44036-2028 Re: K012927 Trade/Device Name: M71 Power Wheelchair Regulation Number: 890.3860 Regulation Name: Wheelchair, Powered Regulatory Class: II Product Code: ITI Dated: August 30, 2001 Received: August 31, 2001 Dear Ms. Farrow: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ Page 2 - Ms. Rae Ann Farrow This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Susan Walk, us Image /page/4/Picture/5 description: The image contains a simple, hand-drawn symbol that resembles a cursive letter 'G' or a stylized crescent shape. The symbol is drawn with a thick line and has a slightly irregular, organic appearance. The drawing is presented against a plain white background, which provides a clear contrast and emphasizes the shape of the symbol. Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # i10(k) Number (if known): #BD KO12927 Device Name: Invacare M71 Power Wheelchair ## Indications For Use: the intended use of the Invacare M71 Power Wheelchair is to provide mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) SN (Division Sign-Off) Division of General, Restorative and Neurological DeviceR Over-The-Counter Use _ Prescription Use (Per 21 CFR 801.109) 510(k) Number_KO12927 (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...