ARROW FWD POWER WHEELCHAIR

K991168 · Invacare Corp. · ITI · Jun 25, 1999 · Physical Medicine

Device Facts

Record IDK991168
Device NameARROW FWD POWER WHEELCHAIR
ApplicantInvacare Corp.
Product CodeITI · Physical Medicine
Decision DateJun 25, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of the Invacare Arrow Front Wheel Drive Power Wheelchair is to provide mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair.

Device Story

The Invacare Arrow is a battery-powered, motor-driven, rigid (non-folding) front-wheel-drive power wheelchair. It utilizes drive motors connected directly to front wheels and pivoting rear casters to enhance maneuverability and provide a natural turning feel. The device is operated by the user via a joystick-based motor controller that engages system motion and steering. It is intended for use by individuals with mobility limitations who are capable of operating the controls. The device provides independent mobility, allowing users to navigate environments where they are seated. It functions as a standard powered mobility aid.

Clinical Evidence

Bench testing only. The device was tested for electromagnetic compatibility in accordance with ISO EMC Draft Standard 7176-14 (April 3, 1995). The device met all required performance criteria.

Technological Characteristics

Battery-powered, motor-driven, front-wheel-drive power wheelchair. Rigid, non-folding frame. Features joystick-operated motor controllers, motor/gearbox drive mechanisms, and pivoting casters. Powered by DC batteries.

Indications for Use

Indicated for persons limited to a sitting position who possess the capability to operate a powered wheelchair.

Regulatory Classification

Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KGG 1168 #### 510(k) SUMMARY ## Invacare Corporation's Model Arrow Front Wheel Drive Power Wheelchair ﻣﺜﻞ ﻣﻦ ﻣﻨﻬﺎ ﺍﻟﻤﺘﻮﻗﻒ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍ # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared. Invacare Corporation One Invacare Way PO Box 4028 Elyria, Ohio 44036 Phone: (440) 329-6000 Facsimile: (440) 365-4558 Contact Person: Edward A. Kroll Director, TQM and Regulatory Affairs Date Prepared: April 5, 1999 ## Name of Device and Name/Address of Sponsor Models Arrow Front Wheel Drive Power Wheelchair Invacare Corporation One Invacare Way Elyria, Ohio 44036-2028 Phone: (440) 329-6000 Facsimile: (440) 365-4558 ## Common or Usual Name Power Wheelchair ## Classification Name Wheelchair, Powered #### Predicate Devices Invacare Corporations' Action Storm Ranger II and Action Storm Power 9000 Front Wheel Drive Power Wheelchairs (K982064 , October 5, 1998) and Permobil's Chairman Front Wheel Drive Power Wheelchair (K960951, April 4, 1997). {1}------------------------------------------------ ## Intended Use The intended use of the Invacare Arrow Front Wheel Drive Power Wheelchair is to provide mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair. ## Technological Characteristics and Substantial Equivalence ## A. Device Description The Invacare Model Arrow Front Wheel Drive (FWD) Power Wheelchair is a battery powered, motor driven device with the intended function and use of providing mobility to those persons limited to a sitting position that have the capability of operating a power wheelchair. It is a rigid or " non- folding" type power wheelchair, with front wheel drive capability. Front wheel drive means that the drive motors are connected directly to the front wheels of the wheelchair, as opposed to a rear wheel drive chair which connects the drive motors to the rear wheels. Mounting the drive wheels on the front of the wheelchair, results in a chair that is easier to maneuver than the traditional rear wheel drive wheelchair. Because the center line of the pivoting casters is now behind the user, the user experiences a more natural feel when turning the wheelchair. ## B. Substantial Equivalence The Invacare Model Arrow FWD Power Wheelchair is substantially equivalent to other devices of comparable type that are currently being legally marketed within the United States. The comparable devices include those manufactured by Invacare as well as its' competitors. Products which are substantially equivalent to these devices are Invacare Corporations' Action Storm Ranger II and Action Storm Power 9000 Front Wheel Drive Power Wheelchairs (K982064 , October 5, 1998) and Permobil's Chairman Front Wheel Drive Power Wheelchair (K960951, April 4, 1997). Each of these products are battery powered, motor driven, front wheel drive powered wheelchairs with the same intended function and use which is to provide mobility to persons limited to a sitting position, that have the capability of operating a powered Similarities include large wheels with attached motor/gearbox drive wheelchair. mechanisms, smaller pivoting casters for turning, and joystick operated motor controllers to engage system motion and steer the wheelchair. They are all constructed from the same basic materials, have the same basic operational principles and all use DC batteries as their source of power. {2}------------------------------------------------ ## PERFORMANCE DATA As required by FDA's July 26, 1995, draft publication entitled "Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorized Three- Wheeled Vehicles", the Invacare Arrow Front Wheel Drive Power Wheelchair was tested in accordance with ISO EMC Draft Standard 7176-14 (Titled "Draft ISO EMC Group Proposal: Electromagnetic Compatibility Addition" And Dated April 3, 1995) for powered wheelchairs and motorized scooters. In all instances, the Invacare Arrow Front Wheel Drive Power Wheelchair met the required performance criteria and functioned as intended. i {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol consists of three abstract human profiles arranged in a flowing, wave-like pattern. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 25 1999 Mr. Edward A. Kroll Director, TQM and Regulatory Affairs Invacare Corporation One Invacare Way P.O. Box 4028 Elyria, Ohio 44036-2125 Re: K991168 Arrow Front Wheel Drive (FWD) Power Wheelchair Requlatory Class: II Product Code: ITI Dated: April 5, 1999 Received: April 7, 1999 Dear Mr. Kroll: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {4}------------------------------------------------ Page 2 - Mr. Edward A. Kroll This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M. Witten, Ph.D., M.D. a M. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 510(k) Number (if known): TBD K991168 Device Name: Invacare Model Arrow Front Wheel Drive (FWD) Power Wheelchair Indications For Use: Its intended use is to provide mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) 510(k) Number _ Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Division Sign-Off) (Optional Format 1-2-96) Division of General Restorative Devices K991168
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...