GEN_2 MECHANICAL WHEELCHAIR

K113713 · Free Wheelchair Mission · IOR · May 7, 2012 · Physical Medicine

Device Facts

Record IDK113713
Device NameGEN_2 MECHANICAL WHEELCHAIR
ApplicantFree Wheelchair Mission
Product CodeIOR · Physical Medicine
Decision DateMay 7, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3850
Device ClassClass 1

Indications for Use

The GEN 2 Mechanical Wheelchair is intended to provide mobility to persons restricted to a seated position.

Device Story

The GEN 2 Mechanical Wheelchair is a manually operated, all-purpose mobility device designed for use in developing countries. It features a powder-coated steel frame, adjustable seating, large 8-inch castor wheels, 26-inch spoke wheels with pneumatic tires, and an extra-thick cushion. The device is designed to provide a personalized fit for users, with adjustable footrests and back upholstery. It is intended for use by individuals restricted to a seated position to improve their mobility and quality of life. The device is operated by the user or a caregiver. It is a non-powered, mechanical device that does not require complex clinical workflows or electronic processing.

Clinical Evidence

Bench testing only. The device underwent mechanical performance testing according to international standards, including ISO 7176-1 (static stability), ISO 7176-3 (effectiveness of brakes), ISO 7176-5 (dimensions/mass), ISO 7176-7 (seating dimensions), ISO 7176-8 (static, impact, and fatigue strength), ISO 7176-15 (documentation), and ISO 7176-16 (resistance to ignition). No clinical data was required or provided.

Technological Characteristics

Powder-coated steel frame (ASTM A 53/A 53M-2005, JIS G 3444-2004); manual operation; 26-inch spoke wheels with pneumatic tires; 8-inch polypropylene castor wheels; polypropylene footplates; adjustable back upholstery. Standards applied: ISO 7176-1, 3, 5, 7, 8, 15, 16. Weight limit: 220 lbs. No electronic components or software.

Indications for Use

Indicated for persons restricted to a seated position requiring mobility assistance.

Regulatory Classification

Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the "Free Wheelchair Mission". The logo consists of a drawing of two wheels, one larger than the other, with the words "Free Wheelchair Mission" written below. Underneath the name is the text "TRANSFORMING LIVES THROUGH THE GIFT OF MOBILITY". KII3713 PAGE 1 OF 4 MAY - 7 2012 # 510(K) SUMMARY (21 CFR 807.92) ## GEN 2 MECHANICAL WHEELCHAIR Free Wheelchair Mission 510(k) Owner: 15279 Alton Parkway, Suite 300 Irvine, CA 92618 Tel: 949-273-0858 Fax: 949-273-8471 Contact Person: Sharon Rockwell Tel: 714-695-9269 E-mail: srockwell@writeme.com December, 2011 Date Prepared: GEN 2 Mechanical Wheelchair Trade Name: Mechanical Wheelchair Common Name: Classification Name: Mechanical wheelchair per 21 CFR 890.3850, IOR Ki Mobility Catalyst Wheelchair, K062660 Predicate Devices: The GEN 2 mechanical wheelchair is a highly adaptable all-Device Description: purpose wheelchair created for use in developing countries, while allowing for a healthy, personalized fit. The wheelchair features adjustable seating, large castor wheels, an extra-thick cushion and adjustable footrests. The GEN_2 is a manually operated device intended to provide Intended Use: mobility for persons restricted to a sitting position. > The Indications for Use are identical to those of the predicate devices. {1}------------------------------------------------ ### Technological Characteristics: Technological Characteristics: The GEN_2 mechanical wheelchair is substantially equivalent in design, materials, and intended use to the Ki Mobility Catalyst mechanical wheelchair as described in the table below: | Feature | Catalyst | GEN 2 | |-----------------------|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------| | Intended Use | As a means of mobility for<br>persons restricted to a sitting<br>position. | To provide mobility to persons<br>restricted to a seated position. | | Frame material | 6061-T6 Aluminum | Powder coated steel frame that<br>meets ASTM A 53/A 53M-2005<br>and JIS G 3444-2004 standards | | Frame widths | 14-20" | 13.5-19.5" | | Overall widths | 20.5-26.5" | 29" | | Seat depths | 14-20" | 11.5-17" | | Back heights | 8.5-19" | 11.8-19" | | Weight limit | 220 lbs | 220 lbs | | Chair weight | 23 lbs (without footrests) | 36 lbs (with footrests) | | Armrests | Flip back height adjustable desk<br>and full length arm pads.<br>Tubular swing-away. | Fixed in place to serve as<br>armrests but not to restrict<br>transfers. | | Front end type | Swing-away, non-swing-away | Swing-away | | Back type | Standard | Standard | | Footrest hangers | 70°, elevating leg rest | NA | | Footplates | Composite, foam, angle<br>adjustable | Polypropylene, 280 mm<br>adjustable range with angles<br>adjustable to 0, ± 7°, ± 14°, and ±<br>21°. | | Extension tubes | Extra short, short, med, long | NA | | Back upholstery | Low, med, tall, adjustable | Adjustable to 4 different heights | | Axle plates | Standard, curved, amputee,<br>offset | NA | | Wheel sizes | 22, 24, 26 | 26" | | Wheel types | Spoke, composite mag, Octopus<br>(performance spoke) | Spoke | | Tire types | Pneumatic, full profile<br>polyurethane, iron cap<br>(puncture resistant), high<br>pressure | Pneumatic | | Handrims | Aluminum, plastic coated<br>projections | Powder coated steel | | Feature | Catalyst | GEN 2 | | Caster sizes | 4, 5, 6, 8" | 8" | | Caster types | Poly, pneumatic, pneumatic<br>with airless insert | Polypropylene | | Forks sizes | 4, 5, 6, 7" frog legs | 5" | | Wheel locks | Push to lock, pull to lock, low<br>profile | Pull to lock | | Anti tips tubes | Yes | No | | Standards applied | Static Stability and Fatigue<br>Strength, ANSI/RESNA std<br>sections 1, 5, 7, 8, 16 and 93 | ISO 7176-1, 3, 5, 7, 8, 15, 16 | | Tubing wall thickness | 0.070" on frame except front<br>vertical tube which is 0.1". X-<br>tube is 0.070" | 0.079" throughout | | Tube properties | 6061-T4 prior to welding.<br>After welding, the component is<br>treated up to T-6. | ASTM A 53/A 53M-2005<br>JIS G 3444-2004 | {2}------------------------------------------------ The indications for use for the GEN 2 and the predicate device are identical. The size ranges in terms of frame width, seat depth, back depths, chair weight and weight limit overlap those of the predicate device. Most other features are very similar. The fundamental technology, welded tube frame, fire retardant upholstery, and wheel locks are the same. The information provided supports a substantial equivalence decision. Any differences in features to the predicate do not affect the performance of the device and do not raise new types of safety or effectiveness questions. - Non-Clinical Non-clinical testing has been performed to mechanical wheelchair Performance Data: international standards including: 1) ISO 7176-1:1999 Determination of static stability 2) ISO 7176-3:2006 Determination of effectiveness of breaks 3) ISO 7176-5:2008 Determination of dimensions, mass and manoeuvring space 4) ISO 7176-7:1998 Measurement of seating and wheel dimensions 5) ISO 7176-8:1998 Requirements and test methods for static, impact and fatigue 6) ISO 7176-15:1996 Requirements for information disclosure, documentation and labeling 7) ISO 7176-16: 1997 Resistance to ignition of upholstered parts --Requirements and test methods {3}------------------------------------------------ The ISO tests were conducted with methods that met the following standards: 1) ISO 7176-11:1992 Test dummies 2) ISO 7176-13: 1989 Determination of coefficient of friction of test surfaces The non-clinical test results demonstrate the GEN 2 Mechanical Conclusions: Wheelchair does not raise any issues regarding safety and effectiveness. The testing supports a determination of substantial equivalence to mechanical wheelchairs previously cleared by FDA. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like figure with three wing-like shapes extending upwards and to the right. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Free Wheelchair Mission % Ms. Sharon Rockwell Consultant 5582 Chalon Road Yorba Linda, California 92886 MAY - 7 2012 、Re: K113713 Trade/Device Name: Free Wheelchair Mission GEN 2 Mechanical Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical wheelchair Regulatory Class: Class I Product Code: IOR Dated: March 26, 2012 Received: March 27, 2012 Dear Ms. Rockwell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ #### Page 2 - Ms. Sharon Rockwell If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Eunel Keith Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Free Wheelchair Mission GEN_2 Mechanical Wheelchair Indications for Use: The GEN 2 Mechanical Wheelchair is intended to provide mobility to persons restricted to a seated position. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ X (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) signature Division Sign-Off Page ( of ) (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K113713
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...