K062660 · Ki Mobility, LLC · IOR · Oct 24, 2006 · Physical Medicine
Device Facts
Record ID
K062660
Device Name
CATALYSY
Applicant
Ki Mobility, LLC
Product Code
IOR · Physical Medicine
Decision Date
Oct 24, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Attributes
3rd-Party Reviewed
Indications for Use
The Catalyst wheelchair is a manually operated device intended to be used as a means of mobility for persons restricted to a sitting position.
Device Story
Catalyst is a folding, ultra-lightweight manual wheelchair constructed from 6061-T6 aluminum tubing. It features a cross-brace frame design that, when opened, locks into a rigid box-like assembly using 4 hooks. The device includes adjustable components such as axle plates, backtubes, armrests, and swing-away footrests. It is operated by the user or a caregiver to provide mobility for individuals restricted to a sitting position. The device is intended for use on solid surfaces. Clinical benefit is derived from providing independent or assisted mobility to patients with physical limitations. The device is a mechanical, non-powered mobility aid.
Clinical Evidence
Bench testing only. The device was tested for compliance with ANSI/RESNA WC Volume 1 1998 standards, specifically sections 1 (static stability), 5 (dimensions/mass), 7 (seating/wheel dimensions), 8 (static/impact/fatigue strength), 16 (flammability), and 93 (maximum overall dimension).
Indicated for persons restricted to a sitting position requiring a manually operated mobility device.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{1}------------------------------------------------
K06260
Image /page/0/Picture/1 description: The image shows the logo for KI, a company that provides furniture and architectural wall systems. The logo consists of the letters "Ki" in a bold, sans-serif font. The letters are black and are surrounded by an oval shape. The oval shape is also black and is slightly tilted to the right.
4925 Coye Drive Suite D Stevens Point, WI 54481 715-254-0991 715-254-0996 (fax) dhmunsey@kimobility.com
0C1 2 4 2006
October 13, 2006
### 510(k) Summary of Safety and Effectiveness
#### Submitter
Ki Mobility 4925 Coye Drive Suite D Stevens Point, WI 54481 USA
Telephone: (715) 254-0991 Fax: (715) 254-0996
Contact: Douglas H. Munsey
510(k) Number: K062660
Trade Name: Catalyst
Common Name: Folding Ultra-Lightweight Manual Wheelchair
Classification Name: Mechanical wheelchair
Classification: Class I
### Predicate Device
The predicate device is the Quickie 2 wheelchair. K850536
### Device Description
The Catalyst is a traditional folding cross-brace wheelchair. It is made of lightweight aluminum tubing similar to what is used on existing products. The frame utilizes two cross tube sets and, when opened, nestles inside of the frame onto 4 hooks to create a box like rigid assembly. Upon the outside of this framework, and to the rear, are assembled 2 machined aluminum axle plates. Stainless steel receivers are mounted within the axle plates. Wheels of varying size and type are connected to the receivers via stainless steel axles.
{2}------------------------------------------------
- Page 2
On the front end of the frame are assembled 2 aluminum housings. Caster forks are mounted to these housings via stainless steel axles. A variety of caster wheels and tires are then connected to the forks.
Upon the top of the frame is attached a seat sling. Into the rear of the frame assembly 2 backtubes are inserted to the desired height. Backrest upholstery is affixed to the backtubes.
Armrests receivers bolt onto the rear of each side frame. Individual Swing Away tubular arms are then inserted into the receivers. If flipback armrests are chosen a receiver is mounted onto the from of each side frame to catch the forward most section of the arm and the rear mounting is pivoting to allow the arm to rotate backward. Height Adjustable arms fit into a receiver that mounts onto the outside of the frame.
Wheel locks mount onto the upper side frame and are adjusted so that the brake arm engages with the tire and when in the locked position prevents to the wheel from rotating.
Swing Away footrests with footplates and heel loops attach at the front of the chair. There is a plug at the top of the footrest hanger that inserts into an opening at the top front of the wheelchair frame. The footrest hanger is turned 90 degrees away from the frame. Hold the latch, which is midway down the hanger, tightly against the front of the frame and rotate the footrest back towards center. The latch will engage once the footrest hanger is centered.
| Feature | Description | Materials |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Catalyst wheelchair is a manually operated device intended to be used as a means of mobility for persons restricted to a sitting position. | |
| Primary Frame Materials | 6061-T6 Aluminum | 6061-T6 Aluminum |
| Folding Method | Collapsible Cross-brace | 6061-T6 Aluminum |
| Frame Widths | 14" to 20" | |
| Overall Width | 20.5" to 26.5" | |
| Seat Depths | 14" to 20" | |
| Back Heights | 8.5" to 19" | |
| Weight Limit | 220 lbs | |
| Seat Height | 15" to 21" | |
| Chair Weight | 23 lbs (w/o footrests) | |
| Warranty | Lifetime on frame | |
| Armrests | 1. Flip Back Height Adjustable -- Desk & Full Length Arm pads<br>2. Flip Back Desk & Full Length Arm pads<br>3. Height Adjustable- Desk & Full Length Arm pads<br>4. Tubular Swing-Away | 1,2,3 - 6061-T6 Aluminum tubing, ERW 1020 Steel upper, plastic side guard, 303 stainless steel pin, Plastic armrest pad<br>4. - 6061-T6 Aluminum |
| Front End Type | Swing-Away, Non-Swing Away | Welded 6061-T6 Aluminum |
| Back Type | Straight with push handles, 8 degree bend with push handles, no push handles | 6061-T6 Aluminum |
| Footrests Hangers | 70°, Elevating legrest | Welded 6061-T6 Aluminum |
| Footplates | Composite, Foam, Angle Adjustable | Composite - Nylon 66 with 15% short glass.<br>Foam - 6061-T6 Aluminum with PU foam cover<br>Angle Adj - 6061-T6 Aluminum plate |
| Extension Tubes | Ex. Short, Short, Medium, Long | 6061-T6 Aluminum |
| Back Heights | 8.5" to 19" in 3 adjustable sections | |
{3}------------------------------------------------
## ● Page 3
| Back Upholstery | Sewn Nylon Parapak 420D w/ ¼" polyurethane foam batting | Sewn Nylon Parapak 420D w/ ¼" polyurethane foam batting |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Seat Upholstery | Sewn Dacron Polyester Sailcloth - No filling | Sewn Dacron Polyester Sailcloth - No filling |
| Cushion | 2' foam cushion | Manufactured and distributed by Hudson Industries. Model - Pressure<br>Eez 2 General Seat Cushion |
| Axle Plates | Standard, Curved, Amputee, Offset | Machined 6061-T6 Aluminum |
| Axle | Quick Release or Fixed | Machined 303 Stainless steel |
| Wheel Sizes | 22,24,26 | |
| Wheel Types | Spoke, Composite Mag, Octopus (performance spoke) | Spoke & Octopus- Machined 6061-T6 Aluminum rims, Stainless steel spokes, Machined 6061 - T6 Aluminum hub<br>Mag - Nylon 66 with 15% short Glass |
| Tire Types | Pneumatic, Pneumatic w/ airless insert, Full profile Polyurethane, Low Profile Polyurethane, Iron Cap (puncture resistant), High Pressure | Synthetic rubber and Urethane |
| Handrims | Aluminum, Plastic Coated, Projections | 6061-T6 Aluminum, anodized or covered with PVC coating |
| Caster Sizes | 4", 5", 6", 7", 8" | |
| Caster Types | Poly, Pneumatic, Pneumatic w/ airless insert | Nylon 66 with 15-30% short glass<br>Synthetic rubber and Urethane tires |
| Forks Sizes | 4", 5", 6", 7" | Machined 6061-T6 Aluminum |
| Fork Stem Sizes | Std, +3/4", +1 ½" | Machined 303 Stainless steel |
| Wheel Locks | Push to Lock, Pull To Lock, Scissor Lock | Machined 6061-T6 Aluminum |
| Anti Tips Tubes | Locate off of the axle to insure safety at all seat heights | Cast aluminum housing with 6061-T6 tubes and plastic wheels |
| Where used | Solid surface | |
| Target Population | Persons restricted to a sitting position. | |
| Standards | Static Stability and Fatigue Strength: ANSI/RESNA<br>Wheelchair Std. Vol. 1<br>Sections 1, 5, 7, 8, 16 & 93 | |
| Tubing thicknesses | .070" on frame except front vertical tube, which is .1". X-tube is .070. | |
| Tube properties | 6061-T4 prior to welding. After welding the | |
# Indications for Use
The Catalyst wheelchair is a manually operated device intended to be used as a means of mobility for persons restricted to a sitting position.
| Feature | Quickie 2 | Catalyst |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Quickie wheelchair is a manually<br>operated device intended to be used as a<br>means of mobility for persons restrict to a<br>sitting position. | The Catalyst wheelchair is a manually<br>operated device intended to be used as a<br>means of mobility for persons restrict to a<br>sitting position. |
| Primary Materials | 6061-T6 Aluminum | 6061-T6 Aluminum |
| Folding Method | Collapsible Cross-brace | Collapsible Cross-brace |
| Frame Widths | 11" to 22" | 14" to 20" |
| Overall Width | 20.5" to 28.5" | 20.5" to 26.5" |
| Seat Depths | 10" to 20" | 14" to 20" |
| Back Heights | 8.5" to 19" | 8.5" to 19" |
| Weight Limit | 250 lbs (350lbs. Heavy Duty ) | 220 lbs |
| Seat Height | 16.75" to 22.75" | 15" to 21" |
| Chair Weight | 27 lbs (w/o footrests) | 23 lbs (w/o footrests) |
| Warranty | Lifetime on frame | Lifetime on frame |
| Armrests | 1. Flip Back Height Adjustable - Desk & Full Length Arm pads<br>2. Flip Back<br>3. Height Adjustable- Desk & Full Length Arm pads<br>4. Tubular Swing-Away<br>5. Adjustable Locking Flip-up<br>6. Length Adjustable Locking Flip-up | 5. Flip Back Height Adjustable - Desk & Full Length Arm pads<br>6. Flip Back Desk & Full Length Arm pads<br>7. Height Adjustable- Desk & Full Length Arm pads<br>8. Tubular Swing-Away |
| Front End Type | Swing-Away, Non-Swing Away | Swing-Away, Non-Swing Away |
| Back Type | Std, Angle Adjustable, Depth Adjustable | Std |
| Footrests Hangers | 60°, 70°, 70°V, 90°, Articulating Legrest, Elevating legrest, 90° Elevating, 1-Piece | 70°, Elevating legrest |
| Footplates | Composite, Foam, Aluminum, Angle Adjustable, Platform Flip-up, Locking Angle Adjustable | Composite, Foam, Angle Adjustable |
| Extension Tubes | Ex Short, Short, Medium, Long | Ex Short, Short, Medium, Long |
| Back Upholstery | Low, Medium, Tall, Adjustable, Reinforced | Low, Medium, Tall, Adjustable |
| Axle Plates | Standard, Curved, Amputee, Offset | Standard, Curved, Amputee, Offset |
| Wheel Sizes | 20,22,24,26 | 22,24,26 |
| Wheel Types | Spoke, Composite Mag, Spinergy (performance spoke), One Arm Drive | Spoke, Composite Mag, Octopus (performance spoke) |
| Tire Types | Pneumatic, Pneumatic w/ airless insert, Full profile Polyurethane, Low Profile Polyurethane, Kevlar (puncture resistant), High Pressure | Pneumatic, Pneumatic w/ airless insert, Full profile Polyurethane, Low Profile Polyurethane, Iron Cap (puncture resistant), High Pressure |
| Handrims | Aluminum, Plastic Coated, Projections | Aluminum, Plastic Coated, Projections |
| Caster Sizes | 4", 5", 6", 8", 8x2 | 4", 5", 6", 8" |
| Caster Types | Poly, Semi-Pneumatic, Soft Roll, Pneumatic, Pneumatic w/ airless insert | Poly, Pneumatic, Pneumatic w/ airless insert |
| Forks Sizes | 3", 4", 5 ¼", 6", 7", Frog Legs | 4", 5", 6", 7", Frog Legs |
| Fork Stem Sizes | Std, +3/4", +1 ½" | Std, +3/4", +1 ½" |
| Caster Options | Multi-Position Fork, Caster Pin Locks, Quick Release Caster Stems | NA |
| Wheel Locks | Push to Lock, Pull To Lock, Scissor Lock, Grade Aid | Push to Lock, Pull To Lock, Low Profile |
| Anti Tips Tubes | Yes | Yes |
| Power Adapter | Quickie Extender | NA |
| Where used | Solid surface. | Solid surface |
| Target Population | Persons restricted to a sitting position. | Persons restricted to a sitting position. |
| Standards | Unknown | Static Stability and Fatigue Strength ANSI/RESNA Wheelchair Std. Vol. 1 Sections 1, 5, 7, 8, 16 & 93 |
| Tubing<br>thicknesses | .083" on frame except front vertical tube, which is .125". X-tube is variable in thickness from .083" to .125" as it is a custom extrusion | .070" on frame except front vertical tube which is .1". X-tube is .070. |
| Tube properties | 6061 - T6 prior to welding. Hardness | 6061-T4 prior to welding. After welding |
Technological Comparison to the Predicate Device
{4}------------------------------------------------
{5}------------------------------------------------
| . A minimal and comments of the many of the commendent of the commendent of the contribution of the country of the count | reduced to between T() and T2 after | Company of the county of the control county of possible in possible in possible aspective<br>the component is heat treated up to 7-6. | |
|--------------------------------------------------------------------------------------------------------------------------|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|--|
| | 11'<br>welcing. | | |
It is Ki Mobility's conclusion that the Catalyst manual wheelchair is substantially equivalent to the Quickie 2 manual wheelchair currently marketed by Sunrise Medical.
## Summary of Performance Testing
The Catalyst has been tested and found to comply with the ANSI-RESNA WC Volume 1 1998 standards sections 1, 5, 7, 8, 16 & 93.
## Conclusions
As stated above, Ki Mobility's conclusion is that the Catalyst wheelchair is safe, effective, complies with the appropriate medical device standards, and is substantially equivalent to the Quickie 2 wheelchair.
# This 510(k) Summary of Safety and Effectiveness may be copied and submitted to interested parties as required by 21CFR807.92.
| STANDARD | TITLE and COMMENTS |
|---------------------------------------------|-------------------------------------------------------------|
| ANSI-RESNA WC - Vol 1<br>-1998 Section 1 | Determination of static stability |
| ANSI-RESNA WC - Vol 1<br>-1998 Section 8 | Static, impact, and fatigue strengths |
| ANSI-RESNA WC- Vol 1<br>- 1998 Section 16 | Determination of Flammability |
| ANSI-RESNA WC -<br>Volume 1-1998 Section 5 | Determination of Overall Dimensions, Mass and Turning Space |
| ANSI-RESNA WC -<br>Volume 1-1998 Section 7 | Measurement of seating and wheel dimensions |
| ANSI-RESNA WC -<br>Volume 1-1998 Section 93 | Maximum Overall Dimension |
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The text is in all capital letters and is evenly spaced around the circle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ki Mobility LLC % Mr. Douglas H. Munsey 4925 Coye Drive, Suite D Stevens Point, Wisconsin 54481
OCT 2 4 2006
Re: K062660
Trade/Device Name: Catalyst Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: October 6. 2006 Received: October 10, 2006
Dear Mr. Munsey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2 – Mr. Douglas H. Munsey
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Kavita Buelow
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
#### INDICATIONS FOR USE
## Applicant:
Ki Mobility 4925 Coye Drive Suite D Stevens Point, WI 54481 USA
Telephone: (715) 343-1280 Fax: (715) 343-1280 510(k) Number: K062660
Device Name: Catalyst
#### Indications for Use:
The Catalyst wheelchair is a manually operated device intended to be used as a means of mobility for persons restricted to a sitting position.
Prescription Use (Per 21 CFR 801.116)
Over-The-Counter OR X
(Please do not write below this line -- continue on another page if needed)
bashase-Puchand
(Division Sign-Off) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division of General, Restorative, and Neurological Devices
**510(k) Number** K062460
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.