ViMove2 is a wireless medical device that measures, records, and reports movements and muscle activity of the lower back / lumbar spine. The system also measures range of motion in the sagittal and coronal anatomical planes.
Device Story
ViMove2 is a wireless system for healthcare professionals to assess lumbar spine movement and muscle activity. It consists of four wireless sensors (two movement, two muscle activity) attached to the patient via disposable application pads. Sensors use accelerometers, gyroscopes, magnetometers, and surface EMG to capture data. Under clinician guidance, patients perform dynamic movements and static postures. Data streams via BTLE to a host device (PC, tablet, or smartphone). The ViMove software processes this data to generate a personalized assessment report, including comparisons to normative values. The system provides objective measurements to assist clinicians in evaluating patient movement and muscle function, facilitating clinical decision-making regarding physical therapy or rehabilitation.
Clinical Evidence
Bench testing only. Validation performed for lumbo-pelvic range of motion (ROM) and posture assessment, including flexion, extension, lateral flexion, lordotic angle, and pelvic tilt. Design verification conducted per 21 CFR 820.30, demonstrating compliance with ISO 14971, ISO 13485, IEC 60601-1, IEC 60601-1-2, ISO 15223, ISO 10993-5, ISO 10993-10, IEC 62366, and IEC 62304.
Technological Characteristics
Wireless system with 4 sensors (2 movement, 2 muscle activity). Movement sensors use accelerometer, gyroscope, and magnetometer; muscle sensors use surface EMG. Connectivity via BTLE. Power via rechargeable Lithium-Polymer batteries. Components include disposable application pads. Software-based analysis and reporting. Complies with IEC 60601-1, IEC 60601-1-2, ISO 10993-5, ISO 10993-10, and IEC 62304.
Indications for Use
Indicated for measuring, recording, and reporting lower back/lumbar spine movement and muscle activity, including range of motion in sagittal and coronal planes.
Regulatory Classification
Identification
A diagnostic electromyograph is a device intended for medical purposes, such as to monitor and display the bioelectric signals produced by muscles, to stimulate peripheral nerves, and to monitor and display the electrical activity produced by nerves, for the diagnosis and prognosis of neuromuscular disease.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, composed of three human profiles facing right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 14, 2017
Dorsavi Ltd % Bosmat Friedman Regulatory Consultant Pushmed LLC 1208-12 Rockford Rd Toronto, LAJ 7Y9 CA
Re: K163150 Trade/Device Name: ViMove2 Regulation Number: 21 CFR 890.1375 Regulation Name: Diagnostic Electromyograph Regulatory Class: Class II Product Code: IKN, HCC, KQX Dated: June 16, 2017 Received: June 16, 2017
Dear Bosmat Friedman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K163150
Device Name ViMove2
Indications for Use (Describe)
ViMove2 is a wireless medical device that measures, records, and reports movements and muscle activity of the lower back / lumbar spine. The system also measures range of motion in the sagittal and coronal anatomical planes.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary (as required by section 807.92(c))
## VIMOVE2
# 510(K) NUMBER
## K163150
#### I. SUBMITTER
dorsaVi Ltd Level 1, 120 Jolimont Road East Melbourne 3002, Australia Establishment Registration No. 3010872373 Owner Operator Number: 10046462
Contact Person: Dan Ronchi Phone: +61-(0)3-9652-2191 Fax: +61-(0)3-8610-1024 Email: dr@dorsavi.com
Date Prepared: Oct 15, 2016
#### II. DEVICE
Trade Name:
Name of Device:
ViMove2
ViMove2
BioFeedback System
Common or Usual Name:
Classification:
Diagnostic Electromyograph Product Code: IKN Regulation No: 890.1375 Requlatory Class: II Classification Panel: Physical Medicine
Electromyograph and Goniometer with
Biofeedback Device Product Code: HCC Regulation No: 882.5050 Requlatory Class: II Classification Panel: Neurology
Goniometer Product Code: KQX Regulation No: 888.1500 Regulatory Class: I Classification Panel: Orthopedic
{4}------------------------------------------------
## III. PREDICATE DEVICE
Predicate Device(s):
ViMove manufactured by dorsaVi Ltd (K142494) (This predicate has not been subject to any product recalls).
#### IV. DEVICE DESCRIPTION
ViMove2 is a wireless electronic device used by healthcare professionals to accurately measure, record and analyze movement and muscle activity of the lower back. The device obiectively measures, records and analyzes angular movement and muscle activity.
Under the direction of the healthcare professional, the software quides the patient through a series of dynamic movements and static postures in standing and sitting positions. ViMove2 then measures the range of motion and muscle activity, streams the data live via BTLE to a host device, e.g. PC, tablet or smart phone, and generates a personalized assessment report including comparisons to normative values.
ViMove2 is comprised of the following key components:
- · 4 Wireless Sensors (2 movement and 2 muscle activity)
- · Disposable Application Pads for attaching sensors to the patient.
- · ViMove software package
#### V. INDICATIONS FOR USE
ViMove2 is a wireless medical device that measures, records, and reports movements and muscle activity of the lower back / lumbar spine. The system also measures range of motion in the sagittal and coronal anatomical planes.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The modified ViMove presented in this 510(k) is substantially equivalent to the predicate in terms of intended use, fundamental scientific technology, operating principles and mechanism of action. The main differences between the devices are a reduction in size of the movement (MDM) and muscle activity (MDE) sensors; changes to battery, electronic components, circuitry and DAPs to accommodate the smaller sensors, the addition of Bluetooth compatibility (BTLE), changes to the user interface and format of the assessment report. The RFD Biofeedback device utilized in the predicate device are no longer required for the modified ViMove2 device.
A summary of the differences between the ViMove predicate and ViMove2 is provided in Table 1 below. A more detailed comparison is provided in Attachment G.
{5}------------------------------------------------
| | Predicate ViMove | ViMove2 |
|---------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K142494 | K163150 |
| Manufacturer | dorsaVi Ltd | dorsaVi Ltd |
| Product codes | IKN<br>HCC<br>KQX | IKN<br>HCC<br>KQX |
| Indications for Use | ViMove is a wireless<br>medical device that<br>measures, records, and<br>reports movements and<br>muscle activity of the lower<br>back / lumbar spine. The<br>system also measures<br>range of motion in the<br>sagittal and coronal<br>anatomical planes. | ViMove is a wireless<br>medical device that<br>measures, records, and<br>reports movements and<br>muscle activity of the lower<br>back / lumbar spine. The<br>system also measures<br>range of motion in the<br>sagittal and coronal<br>anatomical planes. |
| Non-invasive medical device | Yes | Yes |
| System Components | 4 Wireless Sensors (2 movement MDMs and 2 muscle activity MDEs) Disposable Application Pads (DAPs) for holding the wireless sensors (DAP-M for movement and DAP-E for muscle sensors) Recording and Feedback Device (RFD) Recharging Cradle with AC Adaptor USB ViMove software | 4 Wireless Sensors (2 movement MDMs and 2 muscle activity MDEs) Disposable Application Pads (DAPs) for holding the wireless sensors (DAP-M for movement and DAP-E for muscle sensors) AC Adaptor USB ViMove software Recharging Cradle with AC Adaptor |
| Performance data submitted to<br>support range of motion<br>measurements | Flexion<br>Extension<br>Lateral Flexion<br>Pelvic Tilt<br>Single Leg Standing<br>Sitting (normal, upright and<br>slouched) | Flexion<br>Extension<br>Lateral Flexion<br>Pelvic Tilt<br>Single Leg Standing<br>Sitting (normal, upright and<br>slouched) |
| Assessment report includes | Yes | Yes |
| normative values for ROM<br>Provides real time biofeedback to<br>the user | Yes (via RFD device) | Yes (directly to host<br>device) |
| Movement Sensor Type | Accelerometer<br>Gyroscope<br>Magnetometer | Accelerometer<br>Gyroscope<br>Magnetometer |
| Wireless Type | ANT | BTLE |
| Muscle Activity sensor | Surface EMG | Surface EMG |
| Battery Type | Lithium-Polymer | Lithium-Polymer |
| Includes DAPs for holding the<br>wireless sensors | Yes | Yes |
| Charging | Recharging Cradle or AC<br>Adaptor (5VDC) | AC Adaptor, USB Port<br>(5VDC) or Recharging Cradle |
| Includes dedicated software | Yes | Yes |
| Rechargeable batteries in the<br>sensors | Yes | Yes |
| Real-time objective measurement<br>of range of motion | Yes | Yes |
| Report generated for Healthcare<br>Professional | Yes | Yes |
| Includes Bluetooth compatibility<br>(BTLE) | No | Yes |
Summary Comparison of the Modified ViMove and Predicate Table 1.
{6}------------------------------------------------
#### dorsaVi Ltd. ViMove2 Special 510(k)
#### VII. PERFORMANCE DATA
In support of the modifications presented in this 510(k), testing was performed to validate ViMove2 for assessment of lumbo-pelvic range of motion (ROM) and posture, including flexion, extension, lateral flexion, normal standing lordotic angle, pelvic tilt (sitting and standing) and various sitting postures (normal, upright and slouched).
Additional design verification was performed on the ViMove2 in accordance with 21 CFR 820.30, including testing to demonstrate compliance testing to the following standards:
| ISO 14971 | Medical Devices - Application of Risk Management to Medical<br>Devices |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 13485 | Medical devices - Quality management systems - Requirements for<br>regulatory purposes |
| IEC 60601-1 | Medical electrical equipment, Part 1: General requirements for basic<br>safety and essential performance |
| IEC 60601-1-2 | Medical electrical equipment - Part 1-2: General requirements for<br>basic safety and essential performance - Collateral standard:<br>Electromagnetic compatibility - Requirements and tests |
| ISO 15223 | Medical devices - Symbols to be used with medical device labels,<br>labeling, and information to be supplied - Part 1: General<br>requirements. |
| ISO 10993-5 | Biological evaluation of medical devices - Part 5: Tests for in vitro<br>cytotoxicity. |
| ISO 10993-10 | Biological evaluation of medical devices - Part 10: Tests for irritation<br>and skin sensitization |
| IEC 62366 | Medical devices - Application of usability engineering to medical<br>devices. |
| IEC 62304 | Medical device software - Software life cycle processes. |
{7}------------------------------------------------
## VIII. CONCLUSIONS
Based on the information provided in this submission, the modified ViMove2 is substantially equivalent to the predicate ViMove device and does not raise any questions relating to safety and/or effectiveness.
The modifications proposed in the ViMove2 do not affect the intended use, fundamental scientific technology, principles of operation or clinical application when compared to the predicate device.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.