IMPULSE-ADJUSTING INSTRUMENT, CBP ADJUSTING INSTRUMENT, NEUROMECHANICAL ADJUSTING INSTRUMENT, MODELS 2003

K023462 · Neuromechanical Innovations, LLC · LXM · Jan 13, 2003 · PM

Device Facts

Record IDK023462
Device NameIMPULSE-ADJUSTING INSTRUMENT, CBP ADJUSTING INSTRUMENT, NEUROMECHANICAL ADJUSTING INSTRUMENT, MODELS 2003
ApplicantNeuromechanical Innovations, LLC
Product CodeLXM · PM
Decision DateJan 13, 2003
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

Impulse™ (aka Neuromechanical™, aka CBP®) Adjusting Instrument is for adjustment, mobilization, or manipulation of the musculoskeletal joints of the spine and/or extremities by a licensed health care professional. The device is for external use only.

Device Story

Hand-held electromechanical chiropractic adjusting instrument; used by licensed healthcare professionals for spinal/extremity joint manipulation. Device consists of high-impact plastic housing, internal solenoid, circuit board, and interchangeable stainless steel styluses with neoprene tips. Operator places stylus on target area; applies ~15 N preload to compress internal spring; manually triggers device. Solenoid activation delivers single thrust (50 N, 150 N, or 265 N) in <5 ms. Safety mechanism prevents activation unless preload spring is maximally compressed. Output provides controlled mechanical force to joints; assists clinician in performing chiropractic adjustments. Benefits include consistent, controlled force delivery compared to manual techniques.

Clinical Evidence

Bench testing only. Comparative study performed using a 2200 N quartz force ring (PCB model 201A03) to measure force transmission of the subject device against Harrison Hand Held and Activator II instruments. 90 total trials (30 per device) across three force settings. Results confirmed subject device produces peak forces (55.8 N, 150.5 N, 254.4 N) comparable to predicate devices.

Technological Characteristics

Electromechanical chiropractic adjusting instrument. Materials: high-impact plastic housing, stainless steel styluses, neoprene rubber tips. Energy source: electrical power (solenoid). Dimensions: 15 cm shaft, 10 cm handle. Features: force adjustment switch (50/150/265 N), preload control spring safety mechanism. Connectivity: none. Sterilization: not specified.

Indications for Use

Indicated for adjustment, mobilization, or manipulation of musculoskeletal joints of the spine and/or extremities in patients requiring chiropractic care. Intended for use by licensed healthcare professionals only. External use only.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JAN 1 3 2003 ## 510(k) Summary Impulse " Adjusting Instrument; also known as (aka) Neuromechanical Adjusting Instrument"; aka CBP® Adjusting Instrument Neuromechanical Innovations, LLC Manufacturer: 11011 S. 48th St., Suite 205 Phoenix, AZ 85044 - Christopher J. Colloca, D.C. Contact: Phone: 480-785-8442 Fax: 480-893-2412 Email: cicolloca(@neuromechanical.com Jeff Green Contract Manufacturer: FA Green 9742 352nd Ave., S.E. Snoqualmie, WA 98065 Contract Manufacturer Contact: Trade Name(s): Impulse™ Adjusting Instrument aka (also known as) Neuromechanical Adjusting Instrument™ aka (also known as) CBP® Adjusting Instrument Manipulator, Plunger-like Joint K003185 - Full Spectrum Activator III K973506 ~ Activator II Adjusting Instrument K930431 - Arthrostim K010851 - Harrison Hand Held Adjusting Instrument K001476 - Torque Specific Cervical Adjusting Instrument Telephone: (425) 888-0007 Facsimile: (425) 888-0675 Email: sensors@centurytel.net Chiropractic Adjusting Instrument Common or Usual Name: Classification Name: Predicate Devices for Substantial Equivalence: Applied for herein. Establishment Registration #: Classification: Unclassified Physical Medicine Panel: R. A. Fisher, Natural Selection and Variation Performance Standards: None known established {1}------------------------------------------------ ### Device Description & Specifications: (aka Neuromechanical™, aka CBP®) Adjusting Instrument is a hand-held Impulse "" electromechanical chiropractic adjusting instrument intended to be used for chiropractic adjustment or manipulation/mobilization of the joints of the spine and extremities. The device is only intended for use from a health care professional licensed by the law of the state that he or she practices. The instrument's shaft measures approximately 15 cm in length and 3.5 cm in width. The instrument's handle is rigidly attached to the shaft and measures approximately 10 cm in length. The external housing of the device consists of high impact plastic that surrounds its main internal components. The major external components of the device consist of the instrument's shaft, handle, force adjustment switch, trigger, preload control nose, stylus and power cord. Three types of interchangeable stainless steel stylus' attach to the preload control nose which make contact with the patient by means of a neoprene rubber end. The major internal components of the device consist of a circuit board, power supply, internal thrust element, solenoid, internal Both external and internal views of the device are shown in Figure 1. Stylus spring. specifications are shown in Figure 2, illustrating the three kinds of stylus', a single stylus, and two double contact stylus' that allows for dual contact to each side of the spine simultaneously. These styluses have been designed for appropriate anatomical contact with the spine. The instrument is manually triggered by the operator and contains a safety mechanism consisting of a triggering mechanism that does not allow for activation of the device unless the preload control spring is maximally compressed. In this manner, the clinician (operator) is able to test the patient's tolerance to contact prior to the instrument being able to be activated. The device is equipped with a force adjustment switch which allows for control of force produced by the instrument, high, medium, and low, representing transmitted peak forces of 265 N, 150 N, and 50 N respectively. Peak forces for all three settings are delivered in less than 5 ms. The device is operated in the following manner. After locating the target area to be treated the rubber end of the stylus makes contact with the treatment location. An approximate 15 N of preload force compresses the spring allowing for activation of the instrument's trieger mechanism. The operator is then able to manually trigger the device to deliver a single thrust ranging from 50 - 265 N depending on the force setting used. ### Substantial Equivalence Comparison: Impulse™ (aka Neuromechanical™, aka CBP®) Adjusting Instrument is substantially equivalent to other FDA registered predicate hand-held chiropractic adjusting instruments. Specifically, Impulse™ (aka Neuromechanical™, aka CBP®) Adjusting Instrument has the same intended use, similar technological characteristics, and similar force-time profiles as predicate devices. The current device, perhaps, is most similar in structure and function to the Harrison Hand Held Adjusting Instrument (K010851). Both hand-held adjusting instruments generate the force to protrude the stylus a controlled distance by means of electrical power that transiently charges a solenoid. The current device, however, has several noteworthy technological characteristics that are similar to predicate devices. Therefore, Table 1 has been prepared to provide a comparison of technological characteristics of the the current device and predicate devices. {2}------------------------------------------------ | Feature | Impulse* | Harrison | Arthrostim | Activator<br>II | Activator<br>III | |-------------------------------------------------------------|----------|----------|------------|-----------------|------------------| | Intended use is chiropractic<br>adjustment of the spine? | Yes | Yes | Yes | Yes | Yes | | Hand-held adjusting device? | Yes | Yes | Yes | Yes | Yes | | Manually activated thrust? | Yes | Yes | Yes | Yes | Yes | | Single thrust per activation? | Yes | Yes | No | Yes | Yes | | Thrust force delivered from<br>spring energy? | No | No | No | Yes | Yes | | Thrust force delivered from<br>powered solenoid activation? | Yes | Yes | Yes | No | No | | Adjustable force transmission | Yes | Yes | Yes | Yes | Yes | | Preload control spring | Yes | No | No | No | Yes | | Rubber contact end on stylus | Yes | Yes | Yes | Yes | Yes | Table 1. Substantial equivalence comparison for the current device (*), the Harrison Hand Held Device (K010851), Arthrostim (K930431), Activator II (K973506), and Activator III (K003185). ## Force Transmission Comparison Among Chiropractic Adjusting Devices The force of the current device (Impulse" Adjusting Instrument) was compared to the Harrison Hand Held Adjusting Instrument (K010851), and the Activator® II Adjusting Instrument (K973506) in a laboratory setting. Ten thrusts were delivered into a load cell (2200 N quartz force ring, PCB model 201A03, PCB Piezotronics, Buffalo, NY') mounted to a table top for each of three force settings for each of the three devices. Collectively, ninety trials were performed, thirty trials with each device consisting of ten trials at each of the three force settings. Figure 3 provides a comparison of force transmission among the three devices for each of the three force settings. The Impulse "device was found to produce an average peak force of 55.8 N, 150.5 N, and 254.4 N respectively (pulse duration < 5 ms) for each of the three force settings that the device allows. As noted in Figure 3, these forces are very similar to the Activator® II, and the Harrison Hand-Held Adjusting Instrument. The results of this experiment clearly demonstrate that the peak forces transmitted with the current device are of similar magnitude (or lower) to predicate devices and therefore does not raise any new safety or efficacy issues. Figure 1. External view (left) and internal view (right) of the Impulse™ (aka Neuromechanical" aka CBP®) Adjusting Instrument. Image /page/2/Figure/8 description: The image shows two diagrams of an adjusting instrument, one displaying the external view and the other displaying the internal view. The external view diagram labels the handle, power cord, trigger, preload control nose, stylus, shaft, impulse, and switch. The internal view diagram labels the thrust element, solenoid, internal spring, preload control nose, trigger, circuit board, power cord lock, and opposite shell casing. {3}------------------------------------------------ 5 Image /page/3/Figure/1 description: The image shows a chiropractic adjustment tool with different stylus attachments. The tool itself is labeled with a length of 15 cm and a height of 10 cm. The stylus attachments include a single stylus, a lumbar dual stylus, and a cervical dual stylus, each with labeled dimensions in millimeters. Figure 2. Measurements of the adjusting instrument (left) and stylus attachments (right). The current device, Impulse" (aka Neuromechanical™ aka CBP®) Adjusting Figure 3. Instrument was compared to two other FDA registered hand-held chiropractic adjusting devices, the Harrison Hand-Held Adjusting Instrument and the Activator II Adjusting Instrument. Ten trials were performed at each of three force settings. Error bars represent the standard deviations. Image /page/3/Figure/4 description: This bar graph compares the force in Newtons of three different devices: Activator II, Harrison, and Impulse. The x-axis shows the setting, which ranges from 1 to 3. The y-axis shows the force in Newtons, which ranges from 0 to 350. At setting 3, the Activator II has a force of approximately 200 Newtons, Harrison has a force of approximately 300 Newtons, and Impulse has a force of approximately 250 Newtons. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # JAN 1 3 2003 Christopher J. Colloca, D.C. President Neuromechanical Innovations, LLC 11011 S 48th Street, Suite 205 Phoenix, Arizona 85044 Re: K023462 Trade Name: Impulse – Adjusting Instrument, aka (also known as) CBP Adjusting Instrument, and aka Neuromechanical Adjusting Instrument Regulatory Class: Unclassified Product Code: LXM Dated: June 3, 2002 Received: October 15, 2002 Dear Dr. Colloca: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (R.MA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2 - Christopher J. Colloca, D.C. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html. Sincerely yours, Sincerely yours, Mark A. Millherson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ### Indications for Use Statement | 510(k) Number: | K023462 | |-----------------|-----------------------------------------------------------------------------------------------------| | Applicant: | Neuromechanical Innovations, LLC<br>11011 S. 48th St., Suite 205<br>Phoenix, AZ 85044 | | Contact: | Christopher J. Colloca, D.C.<br>Telephone: 480-893-2400<br>Facsimile: 480-893-2412 | | Device Name(s): | Impulse™ Adjusting Instrument<br>Neuromechancial™ Adjusting Instrument<br>CBP® Adjusting Instrument | ### Indications For Use: (aka Neuromechanical™, aka CBP®) Adjusting Instrument is for adjustment, Impulse™ mobilization, or manipulation of the musculoskeletal joints of the spine and/or extremities by a licensed health care professional. The device is for external use only. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use _____________________________________________________________________________________________________________________________________________________________ OR (Per 21 CFR 801.109) Over-the-Counter-Use Division Sign-Off Division of General. Restorative and Neurological Devices 510(k) Number K023462
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...