P200019 · Ventana Medical Systems · QNH · Apr 22, 2021 · Hematology
Device Facts
Record ID
P200019
Device Name
VENTANA MMR RxDx Panel
Applicant
Ventana Medical Systems
Product Code
QNH · Hematology
Decision Date
Apr 22, 2021
Decision
APPR
Regulation
21 CFR 864.1860
Device Class
Class 3
Indications for Use
VENTANA MMR RxDx Panel is a qualitative immunohistochemistry test intended for use in the assessment of mismatch repair (MMR) proteins (MLH1, PMS2, MSH2 and MSH6) in formalin-fixed, paraffin-embedded (FFPE) endometrial carcinoma tissue by light microscopy. The OptiView DAB IHC Detection Kit is used for MLH1, MSH2 and MSH6, and the OptiView DAB IHC Detection Kit with the OptiView Amplification Kit is used for PMS2 on a VENTANA BenchMark ULTRA instrument. VENTANA MMR RxDx Panel includes VENTANA anti-MLH1 (M1) Mouse Monoclonal Primary Antibody, VENTANA anti-PMS2 (A16-4) Mouse Monoclonal Primary Antibody, VENTANA anti-MSH2 (G219-1129) Mouse Monoclonal Primary Antibody, and VENTANA anti-MSH6 (SP93) Rabbit Monoclonal Primary Antibody. VENTANA MMR RxDx Panel is indicated as an aid in identifying patients eligible for treatment with JEMPERLI (dostarlimab-gxly) as listed in Table 1 in accordance with the approved therapeutic product labeling. Table 1. VENTANA MMR RxDx Panel companion diagnostic indications. Indication for use: Endometrial Carcinoma (EC); Therapy: JEMPERLI (dostarlimab-gxly); MMR Status: deficient MMR (dMMR). Test results of this panel should be interpreted by a qualified pathologist in conjunction with histological examination, relevant clinical information, and proper controls. This product is intended for in vitro diagnostic (IVD) use.
Device Story
The VENTANA MMR RxDx Panel is an immunohistochemistry (IHC) assay used to detect MMR protein expression in FFPE endometrial carcinoma tissue. The device utilizes four mouse/rabbit monoclonal primary antibodies (anti-MLH1, anti-PMS2, anti-MSH2, anti-MSH6) on the BenchMark ULTRA automated staining platform. The system performs deparaffinization, antigen retrieval, and sequential application of antibodies, followed by OptiView DAB detection (with signal amplification for PMS2). A qualified pathologist interprets the presence or absence of nuclear staining via light microscopy. The output is a binary status (Intact or Loss) for each protein, determining overall MMR status (proficient or deficient). This status identifies patients with dMMR endometrial carcinoma eligible for JEMPERLI (dostarlimab-gxly) therapy. The device aids clinical decision-making by selecting patients likely to benefit from immunotherapy, potentially improving response rates in recurrent/advanced disease.
Clinical Evidence
Clinical performance was established using the GARNET trial (4010-01-001), a multicenter, open-label study evaluating dostarlimab in recurrent/advanced endometrial cancer. The study included 167 patients tested with the VENTANA MMR RxDx Panel. Effectiveness was demonstrated by comparing ORR and DOR in dMMR patients (n=71) identified by the device. ORR was 42.3% (95% CI: 30.6, 54.6) with a median DOR not reached. Concordance between the device and the clinical trial assay (CTA) showed PPA of 92.1% and NPA of 98.9% in the ITD population. Analytical validation included precision, robustness, and stability studies.
Technological Characteristics
IHC assay using four monoclonal antibodies (MLH1, PMS2, MSH2, MSH6) on FFPE tissue. Automated staining on VENTANA BenchMark ULTRA platform. Detection via OptiView DAB IHC Detection Kit (indirect, biotin-free, HRP-based); PMS2 requires OptiView Amplification Kit. Requires 4μm thick FFPE sections on positively charged slides. Standardized protocols for antigen retrieval (CC1) and incubation. Software-controlled (VSS version 12.5.4 or earlier).
Indications for Use
Indicated for assessment of MMR proteins (MLH1, PMS2, MSH2, MSH6) in FFPE endometrial carcinoma tissue to identify patients eligible for JEMPERLI (dostarlimab-gxly) therapy. No contraindications.
Regulatory Classification
Identification
Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.
Special Controls
(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
Related Devices
P210001 — VENTANA MMR RxDx Panel · Ventana Medical Systems, Inc. (Roche Tissue Diagnostics) · Aug 17, 2021
P250004 — MMR IHC Panel pharmDx (Dako Omnis) · Agilent Technologies, Inc. · Aug 15, 2025
K213348 — BOND MMR Antibody Panel · Leica Biosystems Newcastle, Ltd. · Feb 21, 2023
Submission Summary (Full Text)
{0}
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
# I. GENERAL INFORMATION
Device Generic Name: Immunohistochemistry test, DNA mismatch repair (MMR) Protein assay
Device Trade Name: VENTANA MMR RxDx Panel
Device Procode: QNH
Applicant's Name and Address: Ventana Medical Systems, Inc.
1910 East Innovation Park Drive
Tucson, AZ 85755
Date of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P200019
Date of FDA Notice of Approval: April 22, 2021
# II. INDICATIONS FOR USE
VENTANA MMR RxDx Panel is a qualitative immunohistochemistry test intended for use in the assessment of mismatch repair (MMR) proteins (MLH1, PMS2, MSH2 and MSH6) in formalin-fixed, paraffin-embedded (FFPE) endometrial carcinoma tissue by light microscopy. The OptiView DAB IHC Detection Kit is used for MLH1, MSH2 and MSH6, and the OptiView DAB IHC Detection Kit with the OptiView Amplification Kit is used for PMS2 on a VENTANA BenchMark ULTRA instrument.
VENTANA MMR RxDx Panel includes VENTANA anti-MLH1 (M1) Mouse Monoclonal Primary Antibody, VENTANA anti-PMS2 (A16-4) Mouse Monoclonal Primary Antibody, VENTANA anti-MSH2 (G219-1129) Mouse Monoclonal Primary Antibody, and VENTANA anti-MSH6 (SP93) Rabbit Monoclonal Primary Antibody.
VENTANA MMR RxDx Panel is indicated as an aid in identifying patients eligible for treatment with JEMPERLI (dostarlimab-gxly) as listed in Table 1 in accordance with the approved therapeutic product labeling.
Table 1. VENTANA MMR RxDx Panel companion diagnostic indications.
| Indication for use | Therapy | MMR Status |
| --- | --- | --- |
| Endometrial Carcinoma (EC) | JEMPERLI (dostarlimab-gxly) | deficient MMR (dMMR) |
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{1}
Test results of this panel should be interpreted by a qualified pathologist in conjunction with histological examination, relevant clinical information, and proper controls.
This product is intended for in vitro diagnostic (IVD) use.
## III. CONTRAINDICATIONS
There are no known contraindications associated with these tests.
## IV. WARNINGS AND PRECAUTIONS
Warnings and precautions can be found in the anti-MLH1 (M1) Mouse Monoclonal Primary Antibody, anti-PMS2 (A16-4) Mouse Monoclonal Primary Antibody, VENTANA anti-MSH2 (G219-1129) Mouse Monoclonal Primary Antibody, and VENTANA anti-MSH6 (SP93) Rabbit Monoclonal Primary Antibody product labeling.
## V. DEVICE DESCRIPTION
### A. Device Kit Components
The VENTANA MMR RxDx Panel is comprised of four primary antibodies used to detect the MMR proteins MLH1, PMS2, MSH2 and MSH6 in endometrial tissue specimens. The primary antibodies are used in combination with individually optimized detection reagents and in conjunction with ancillary reagents in order to complete specimen testing. The VENTANA MMR RxDx Panel is optimized to run on the VENTANA BenchMark Ultra platform with OptiView DAB detection kit or in the case of PSM2 antibody the OptiView DAB detection Kit with the OptiView Amplification Kit. The presence or absence of target proteins is determined by visual examination of the specimen slide under light microscope by a qualified pathologist.
The VENTANA MMR RxDx Panel antibodies are packaged as individual products in single ready to use reagent dispensers. The VENTANA MMR RxDx Panel uses four separate endometrial tissue slides that are stained on the BenchMark ULTRA instrument. The primary antibody reagents are listed below.
1. **VENTANA anti-MLH1 (M1) Mouse Monoclonal Primary Antibody**
The anti-MLH1 (M1) antibody is a mouse monoclonal antibody produced against a full-length recombinant MLH1 protein with a glutathione-S-transferase tag. The antibody binds to the MLH1 protein in FFPE tissue sections, where it can be localized using the OptiView DAB IHC Detection Kit. Antibody concentration is ~1 µg/mL.
2. **VENTANA anti-PMS2 (A16-4) Mouse Monoclonal Primary Antibody**
The VENTANA anti-PMS2 (A16-4) antibody is a mouse monoclonal antibody raised against a recombinant PMS2 protein. The antibody binds to PMS2 protein
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{2}
in FFPE tissue sections, where it can be localized using the OptiView DAB IHC Detection Kit and OptiView Amplification Kit. Antibody concentration is ~1 µg/mL.
## 3. VENTANA anti-MSH2 (G219-1129) Mouse Monoclonal Primary Antibody
The anti-MSH2 (G219-1129) antibody is a mouse monoclonal antibody produced against a recombinant human MSH2 protein. The antibody binds to the MSH2 protein in FFPE tissue sections, where it can be localized using the OptiView DAB IHC Detection Kit. Antibody concentration is ~20 µg/mL.
## 4. VENTANA anti-MSH6 (SP93) Rabbit Monoclonal Primary Antibody
The VENTANA anti-MSH6 (SP93) antibody is a rabbit monoclonal antibody raised against a recombinant MSH6 protein. It binds to MSH6 protein in FFPE tissue sections, where it can be localized using the OptiView DAB IHC Detection Kit. Antibody concentration is ~1 µg/mL.
Detection and ancillary reagents required but not provided with VENTANA MMR RxDx Panel are listed below:
- OptiView DAB IHC detection Kit containing the following components
- OptiView peroxidase Inhibitor
- OptiView HQ universal Linker
- OptiView HRP Multimer
- OptiView H₂O₂
- OptiView DAB
- OptiView Copper
- OptiView Amplification kit
- OptiView Amplification (0.003% HQ conjugated tyramide complex)
- OptiView H₂O₂
- OptiView Multimer
- Hematoxylin II
- Bluing Reagent
- Reaction Buffer (10x)
- EZ Prep Reagent (10x)
- ULTRA Cell Conditioning (CC1) (Pre-dilute)
- ULTRA Liquid Cover Slip (LCS) (Pre-dilute)
- Negative reagent control mouse monoclonal antibody
- Negative reagent control rabbit monoclonal antibody
## B. Device Instrument and Software
The VENTANA MMR RxDx Panel test is fully automated. The VENTANA MMR RxDx panel antibodies are for use on the BenchMark ULTRA instrument using Ventana System Software (VSS) software version 12.5.4 or earlier.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{3}
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 4
C. Specimen Preparation
Routinely processed, formalin-fixed, paraffin-embedded (FFPE) tissues are suitable for use with the MMR primary antibodies when used with VENTANA OptiView DAB detection kit (with OptiView Amplification Kit for PMS2) and BenchMark ULTRA instruments. Tissue should be fixed in 10% neutral buffered formalin (NBF) for at least 6 hours and for a maximum of 72 hours at room temperature (15-25 °C). Fixation times of less than 6 hours may result in a loss of staining for the MMR antibodies. The amount of NBF used should be 15 to 20 times the volume of tissue. Fixatives such as alcohol-formalin-acetic acid (AFA), PREFER fixative, or other alcohol-containing fixatives are not recommended for use with this assay.
Sections should be cut 4μm thick and mounted on positively charged glass slides. Slides should be stained promptly, as antigenicity of cut tissue sections may diminish over time.
D. Test Controls
Run controls should be included in each staining run to establish the validity of the test results. The following controls should be run with the assay:
Pre-qualified endometrial cancer (EC) tissue with an MMR status of intact may be used as a positive system-level control for MMR antibodies to detect the intact protein.
Alternatively, pre-qualified normal endometrial tissue fixed and processed in the same manner as the patient tissue can also be used as a positive system-level control. Normal endometrium will stain positive for all antibodies in the MMR IHC Panel. Since the MSH6, PMS2, MSH2 and MLH1 proteins are expressed in all tissues, a normal negative tissue control does not exist for these biomarkers. For a negative system level control, EC tissue with loss of an MMR protein can be used as an appropriate tissue control for mismatch repair protein deficiency status. However, lymphocytes, fibroblast and epithelial cells should exhibit staining and serve as positive internal control cells in EC samples with MMR protein deficiency (dMMR).
1. Internal Positive Controls
Normal tissue elements (e.g. lymphocytes, fibroblasts, or normal epithelium) in the immediate vicinity of the tumor may serve as internal positive controls. Unequivocal nuclear staining in these cells validates the staining run. If the internal positive controls fail to demonstrate appropriate staining, results with the test specimen should be considered invalid.
2. Negative Reagent Control
Negative reagent control should be used to stain an adjacent section of the patient specimen tissue on a separate slide from the VENTANA MMR antibody stained slides. A negative reagent control mouse monoclonal antibody (for MLH1, PMS2, and MSH2) and a negative reagent control rabbit monoclonal antibody (for MSH6) is recommended for use in place of the primary antibodies to evaluate nonspecific staining.
{4}
Additional information about positive and negative controls are available in the product labeling.
## E. Principles of Procedure
The VENTANA MMR RxDx Panel is an immunohistochemistry test system used to stain FFPE EC specimens to detect expression of the MMR proteins (MLH1, PMS2, MSH2 and MSH6). The 4 antibodies of the VENTANA MMR RxDx Panel have individualized staining protocols that are created using available staining parameters provided in staining procedures in the VSS software that drives the BenchMark ULTRA automated staining platform. The panel test is run individually on 4 separate tissue sections and the test process involves sequential application of specific primary antibodies against the panel protein, followed by detection reagents and chromogen deposition for visualization of the target protein expression.
The VENTANA MMR RxDx Panel is automated for use on the BenchMark ULTRA automated slide stainer from deparaffinization through counterstaining. Patient FFPE tissue specimens are cut 4μm thick and mounted on positively charged glass slides. These slides are loaded into the Benchmark ULTRA instrument. This system first removes the paraffin wax from the tissue, and then subjects the tissue to heated antigen retrieval (cell conditioning). Antigen retrieval is the process by which the ability of antibodies to bind to the epitopes is restored to formalin-fixed tissues.
Endogenous peroxidases that could potentially react with the horseradish peroxidase conjugates (HRP) are blocked with OptiView Inhibitor (3% H2O2). After the endogenous peroxidase block, the Antibody is dispensed during the antibody incubation step and allowed to bind to its antigen. The slides are then incubated with the reagents in the OptiView DAB IHC Detection Kit and OptiView Amplification Kit to achieve preferred staining of target cells. The OptiView DAB IHC Detection Kit is an indirect, biotin-free system for detecting mouse IgG, mouse IgM, and rabbit primary antibodies which produces a visible dark brown precipitate (3,3'-Diaminobenzidine) via an HRP enzymatic reaction at the antigen site. The PMS2 test uses the OptiView amplification in addition to the OptiView DAB detection system for signal amplification. The OptiView Amplification Kit includes an HQ hapten conjugate (OptiView Amplifier), corresponding substrate (OptiView Amplification H2O2), and mouse anti-HQ monoclonal antibody containing HRP (OptiView Amplification Multimer). Tissues are then counterstained blue using Hematoxylin II and Bluing Reagent to create brown/blue contrast to aid the pathologist when reviewing the slides using bright field microscopy.
Table 1. VENTANA MMR RxDx Panel Staining Protocol
| Protocol Parameter | MLH1 | PMS2 | MSH2 | MSH6 |
| --- | --- | --- | --- | --- |
| Deparaffinization | Selected | Selected | Selected | Selected |
| Cell Conditioning | Cell Conditioning 1, 64 minutes, 100°C | Cell Conditioning 1, 92 minutes, 100°C | Cell Conditioning 1, 40 minutes, 100°C | Cell Conditioning 1, 48 minutes, 100°C |
| Pre-primary antibody peroxidase | Selected | Selected | Selected | Selected |
| Antibody Incubation or Negative Reagent Control | 24 minutes, 36°C | 32 minutes, 36°C | 12 minutes, 36°C | 16 minutes, 36°C |
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{5}
| Protocol Parameter | MLH1 | PMS2 | MSH2 | MSH6 |
| --- | --- | --- | --- | --- |
| OptiView HQ Linker | 8 minutes (default) | 8 minutes (default) | 8 minutes (default) | 8 minutes (default) |
| OptiView HRP Multimer | 8 minutes (default) | 8 minutes (default) | 8 minutes (default) | 8 minutes (default) |
| OptiView Amplification | Not Selected | Selected | Not Selected | Not Selected |
| Amplifier and Amplification H2O2 | Not Selected | 4 minutes | Not Selected | Not Selected |
| Amplification Multimer | Not Selected | 4 minutes | Not Selected | Not Selected |
| Hematoxylin II | 4 minutes | 4 minutes | 4 minutes | 4 minutes |
| Bluing Reagent | 4 minutes | 4 minutes | 4 minutes | 4 minutes |
F. Staining Interpretation
Stained slides are interpreted by a qualified pathologist and the MMR status (intact or loss) for each of the MMR proteins (MLH1, PMS2, MSH2 and MSH6) is assigned based on the presence or absence of specific nuclear staining in the tumor. An MMR status of "Intact" is assigned to cases with unequivocal nuclear staining in viable tumor cells in the presence of acceptable internal positive controls (nuclear staining in lymphocytes, fibroblasts or normal epithelium in the vicinity of the tumor). An MMR status of "Loss" is assigned to cases with unequivocal loss of nuclear staining or focal weak equivocal nuclear staining in the viable tumor cells in the presence of internal positive controls.
If unequivocal nuclear stain is absent in internal positive controls and/or background staining interferes with interpretation, the assay should be considered unacceptable and repeated. Punctate nuclear staining of tumor cells should be considered negative.
MMR Intact (or proficient) Status: Detection of all four proteins (MLH1, PMS2, MSH2 and MSH6) in the tumor.
MMR Loss (or deficient) Status: Loss of expression of at least one protein (MLH1, PMS2, MSH2 and MSH6) in the tumor.
Interpretation of Challenging Cases: The VENTANA MMR RxDx Panel scoring algorithm for each of the four MMR antibodies is binary: either the protein is expressed (a clinical status of Intact) or it is not expressed (a clinical status of Loss). Once each of the four antibodies is interpreted, the case may be assigned an overall status of proficient (all four antibodies are intact) or deficient (at least one antibody is a loss). As a result, there are no true "borderline" cases. While the vast majority of cases stained with VENTANA MMR RxDx Panel are clearly proficient or deficient in their staining results, a few cases have been observed that present a challenge in interpretation. These cases may be challenging due to the following issues: Non-specific background staining, focal staining, punctate staining, speckling, and tissue or staining artifacts resulting from sample processing and microtomy processes.
Interpretation for MMR proteins status is detailed in Table 2 below.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{6}
Table 2. Protein Expression in Endometrial Carcinoma (EC) Tissue Stained with the VENTANA MMR RxDx Panel
| Clinical Status | Description |
| --- | --- |
| Intact MMR Protein Expression | Unequivocal nuclear staining in viable tumor cells, in the presence of acceptable internal positive controls (e.g. nuclear staining in lymphocytes, fibroblasts, or normal epithelium in the vicinity of the tumor) |
| Loss of MMR Protein Expression | Unequivocal loss of nuclear staining or focal weak equivocal nuclear staining in the viable tumor cells in the presence of acceptable internal positive controls. Punctate nuclear staining will be considered negative. |
VI. ALTERNATIVE PRACTICES AND PROCEDURES
Other devices utilizing immunohistochemistry, next generation sequencing, or fluorescent in situ hybridization for detection of MMR deficiency are commercially available.
VII. MARKETING HISTORY
VENTANA MMR RxDx Panel is currently marketed globally in several countries, including in the United States as a Class II device for the identification of Lynch syndrome (DEN170030). The device in the US and ex-US products contain the same reagents. The device has not been withdrawn to date from the market in any country for reasons relating to safety and effectiveness of the device.
VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
As with any IVD test, the potential risks are associated with an incorrect test result or incorrect interpretation of results, rather than with the device itself. Failure of the device to perform as expected or failure to correctly interpret results may lead to improper patient management decisions.
IX. SUMMARY OF NONCLINICAL STUDIES
Nonclinical studies were performed using the VENTANA MMR RxDx Panel to support the analytical performance of the device for the EC indication. These studies were performed using EC specimens (as well as a variety of other tumor tissue for certain studies). Since the prevalence of EC with dMMR status is low, some studies were supplemented with proficient and deficient cases as needed, from a variety of additional tumor types. Studies were conducted to characterize the assay, demonstrate the impact of pre-analytical variables on assay performance, verify precision and robustness of the assay, and establish assay stability.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{7}
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 8
# A. Laboratory Studies
## 1. Analytical Sensitivity
Prevalence of the absence of staining (loss) in individual MMR antibodies were assessed on 227 EC specimens. Prevalence of MMR loss cases (MMR Panel status), as well as VENTANA MMR RxDx Panel failure rates (Not Evaluable in the below Table) within EC tissue samples were also assessed. The results of this assessment for the overall MMR Panel status is shown in the last row of Table 3 below.
Table 3. Prevalence of individual MMR antibodies
| Antibody | Assessed | Evaluable | Not Evaluable | Loss Cases | Prevalence % (95% CI) | Failure Rate % (95% CI) |
| --- | --- | --- | --- | --- | --- | --- |
| MLH1 | 227 | 227 | 0 | 35 | 15.4 (11.3, 20.7) | 0.0 (0.0, 1.7) |
| PMS2 | 227 | 211 | 16 | 37 | 17.5 (13.0, 23.2) | 7.0 (4.4, 11.1) |
| MSH2 | 227 | 220 | 7 | 6 | 2.7 (1.3, 5.8) | 3.1 (1.5, 6.2) |
| MSH6 | 227 | 226 | 1 | 17 | 7.5 (4.7, 11.7) | 0.4 (0.1, 2.5) |
| Panel | 227 | 208 | 19 | 50 | 24.0 (18.7, 30.3) | 8.4 (5.4, 12.7) |
## 2. Analytical Specificity
### a. Western Blot
Western blots analyses were conducted to demonstrate that the antibodies specifically detect the proteins of predicted molecular weight for each of the 4 VENTANA MMR RxDx Panel antibodies using cell lines with known MMR loss or intact status. Cell lines used in the study were matched pair of human cell line HD PAR595 that expressed wild type MMR proteins or were engineered with frame shift knockouts for PMS2 and MLH1 and complete knockout for MSH6 and MSH2. Western Blots confirmed presence of reactive bands at expected molecular weighs for each of the 5 panel markers.
IHC tests using the same cell lines formalin-fixed, paraffin-embedded conducted to assess nonspecific binding in the context of use. The results of the IHC with engineered cell lines was consistent with expected reactivity.
The combined results from western blots and cell line IHC demonstrated specific antibody reactivity for each of the 4 markers included in the VENTANA MMR RxDx Panel.
### b. Peptide Inhibition
The objective of this study was to confirm that the VENTANA MMR RxDx Panel antibodies bind their epitopes within the corresponding protein biomarker. The presence of excess peptide containing the specific epitope will inhibit the antibody from binding to the specific epitope within the corresponding protein biomarker. This information contributes to the characterization of the primary antibody, confirming the antibody binding site and demonstrating the staining specificity of the antibody for the biomarker
{8}
in an immunohistochemical assay. The tissue was stained in duplicate with one lot of primary antibody co-incubated with and without varying concentrations of its corresponding peptide using OptiView detection kit chemistry (and OptiView Amplification where applicable) on the BenchMark ULTRA automated slide staining platform. Samples were scored according to stain intensity.
MLH1 peptide inhibition data: In this peptide inhibition study, complete inhibition of VENTANA anti-MLH1 (M1) antibody binding was observed in the presence of 5×10-5 to 5×10-8 M concentrations of the MLH1 peptide, and partial restoration of staining was observed at 5×10-9 M peptide. No inhibition was observed in the presence of the non-specific, unrelated PMS2 cross-react peptide at 5×10-7 M. All controls and background levels were acceptable.
PMS2 peptide inhibition data: Complete inhibition of VENTANA anti-PMS2 (A16-4) antibody binding was observed in the presence of 5×10-5 to 5×10-6 M concentrations of the PMS2-specific 19-mer peptide, partial restoration of staining was observed at 5×10-7 M peptide, and full restoration of staining at 5×10-8 M peptide. No inhibition was observed in the presence of unrelated peptide at 5×10-7 M. All controls and background levels were acceptable.
MSH2 peptide inhibition data: Similarly, complete inhibition of VENTANA anti-MSH2 (G219-1129) antibody binding was observed in the presence of 5×10-5 to 5×10-8 M concentrations of the MSH2 peptide, and partial restoration of staining was observed at 5×10-9 M peptide. No inhibition was observed in the presence of the non-specific, unrelated PMS2 cross-react peptide at 5×10-7 M. All controls and background levels were acceptable.
MSH6 peptide inhibition data: In this assay the 19mer MSH6 peptide successfully inhibited the VENTANA anti-MSH6 (SP93) antibody by showing complete loss of specific staining at high peptide concentrations (5×10-6, 5×10-7, and 5×10-8 M). Partial recovery of staining was observed when the peptide concentration decreased to 5×10-9 M. These results demonstrate that the epitope for the antibody is contained within the 19mer MSH6 peptide. All controls and background levels were acceptable.
c. Immunoreactivity in Human Tissues
The purpose of this study was to assess the analytical specificity of VENTANA MMR RxDx Panel on non-neoplastic and neoplastic tissue samples performed in accordance with the FDA's "Guidance for Submission of Immunohistochemistry Applications to the FDA" recommendation, including non-specific staining, background, cross-reactivity and biomarker prevalence. No unexpected staining was observed with VENTANA MMR RxDx Panel on the normal and neoplastic tissues. As expected, since mismatch repair is present in all actively proliferating cells, most normal and neoplastic tissues demonstrated positive staining (and acceptable levels of nonspecific staining that did not interfere with sample interpretation). The results are presented in Tables 4 and 5 below.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 9
{9}
Table 4. Specificity of VENTANA MMR RxDx Panel in FFPE normal tissues
| Tissue | MLH1 # Positive / Total Cases | PMS2 # Positive / Total Cases | MSH2 # Positive / Total Cases | MSH6 # Positive / Total Cases |
| --- | --- | --- | --- | --- |
| Adrenal gland | 3/3 | 3/3 | 3/3 | 3/3 |
| Bladder | 3/3 | 3/3 | 3/3 | 3/3 |
| Bone marrow | 3/3 | 3/3 | 3/3 | 3/3 |
| Ovary | 5/5 | 4/4 | 5/5 | 5/5 |
| Breast | 3/3 | 3/3 | 3/3 | 3/3 |
| Cerebellum | 3/3 | 3/3 | 3/3 | 3/3 |
| Cerebrum | 3/3 | 3/3 | 3/3 | 3/3 |
| Cervix | 3/3 | 3/3 | 3/3 | 3/3 |
| Colon | 3/3 | 3/3 | 3/3 | 3/3 |
| Endometrium | 3/3 | 3/3 | 3/3 | 2/3 |
| Esophagus | 3/3 | 3/3 | 3/3 | 3/3 |
| Heart | 3/3 | 2/3 | 1/3 | 3/3 |
| Hypophysis (Pituitary) | 3/3 | 3/3 | 3/3 | 3/3 |
| Intestine | 3/3 | 3/3 | 3/3 | 3/3 |
| Kidney | 3/3 | 3/3 | 3/3 | 3/3 |
| Liver | 3/3 | 3/3 | 3/3 | 3/3 |
| Lung | 4/4 | 3/3 | 3/3 | 4/4 |
| Lymph node | 3/3 | 3/3 | 3/3 | 3/3 |
| Mesothelium | 4/4 | 2/3 | 3/3 | 3/3 |
| Pancreas | 3/3 | 3/3 | 3/3 | 3/3 |
| Parathyroid gland | 5/5 | 3/3 | 3/3 | 3/3 |
| Peripheral nerve | 5/5 | 4/4 | 5/5 | 5/5 |
| Prostrate | 3/3 | 3/3 | 3/3 | 3/3 |
| Skeletal muscle | 3/3 | 2/3 | 3/3 | 3/3 |
| Skin | 3/3 | 3/3 | 3/3 | 3/3 |
| Spleen | 3/3 | 3/3 | 3/3 | 3/3 |
| Stomach | 3/3 | 3/3 | 3/3 | 3/3 |
| Testis | 3/3 | 3/3 | 3/3 | 3/3 |
| Thymus | 3/3 | 3/3 | 3/3 | 3/3 |
| Thyroid | 4/4 | 4/4 | 3/3 | 4/4 |
| Tongue/Salivary gland | 3/3 | 3/3 | 2/3 | 3/3 |
| Tonsil | 3/3 | 3/3 | 3/3 | 3/3 |
Note: Mismatch repair proteins such as MLH1, PMS2, MSH6, and MSH2 are present in all actively proliferating cells. For all tissues, positive/negative staining was determined for tissue specific elements in the presence of positive staining in normal control cells (lymphocytes, fibroblasts, and epithelial cells).
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{10}
Table 5. Specificity of VENTANA MMR RxDx Panel in FFPE neoplastic tissues
| Tissue | MLH1 #
Positive / Total Cases | PMS2 #
Positive / Total Cases | MSH2 #
Positive / Total Cases | MSH6 #
Positive / Total Cases |
| --- | --- | --- | --- | --- |
| Glioblastoma (Cerebrum) | 1/1 | 1/1 | 1/1 | 1/1 |
| Meningioma (Cerebrum) | 1/1 | 0/0 | 1/1 | 1/1 |
| Ependymoma (Cerebrum) | 1/1 | 1/1 | 1/1 | 1/1 |
| Oligodendroglioma (Cerebrum) | 1/1 | 1/1 | 1/1 | 1/1 |
| Serous adenocarcinoma (Ovary) | 1/1 | 1/1 | 1/1 | 1/1 |
| Adenocarcinoma (Ovary) | 1/1 | 1/1 | 1/1 | 1/1 |
| Pancreatic neuroendocrine neoplasm (Pancreas) | 1/1 | 1/1 | 1/1 | 1/1 |
| Adenocarcinoma (Pancreas) | 0/0 | 0/0 | 1/1 | 0/0 |
| Seminoma (Testis) | 1/1 | 2/2 | 2/2 | 2/2 |
| Medullary carcinoma (Thyroid) | 1/1 | 1/1 | 1/1 | 1/1 |
| Papillary carcinoma (Thyroid) | 1/1 | 1/1 | 1/1 | 1/1 |
| Ductal carcinoma in situ (Breast) | 1/1 | 1/1 | 1/1 | 1/1 |
| Microinvasive ductal carcinoma (Breast) | 1/1 | 1/1 | 1/1 | 1/1 |
| Invasive ductal carcinoma (Breast) | 1/1 | 1/1 | 1/1 | 1/1 |
| B-cell lymphoma, NOS (spleen) | 0/0 | 0/0 | 1/1 | 1/1 |
| Small cell carcinoma (Lung) | 1/1 | 1/1 | 1/1 | 1/1 |
| Squamous cell carcinoma (Lung) | 1/1 | 1/1 | 1/1 | 1/1 |
| Neuroendocrine carcinoma (Esophagus) | 1/1 | 1/1 | 1/1 | 1/1 |
| Adenocarcinoma (Lung) | 0/0 | 0/0 | 1/1 | 1/1 |
| Adenocarcinoma (Esophagus) | 1/1 | 0/0 | 1/1 | 1/1 |
| Signet ring carcinoma (Stomach) | 1/1 | 1/1 | 1/1 | 1/1 |
| Adenocarcinoma (Small intestine) | 1/1 | 1/1 | 1/1 | 1/1 |
| Stromal sarcoma (Small intestine) | 1/1 | 1/1 | 1/1 | 1/1 |
| Gastrointestinal stromal tumor (GIST) (Small intestine) | 1/1 | 0/0 | 1/1 | 0/0 |
| Adenocarcinoma (Colon) | 1/1 | 1/1 | 1/1 | 1/1 |
| Adenocarcinoma (Rectum) | 1/1 | 1/1 | 1/1 | 1/1 |
| Gastrointestinal stromal tumor (GIST) (Rectum) | 1/1 | 1/1 | 1/1 | 1/1 |
| Hepatoblastoma (Liver) | 1/1 | 1/1 | 1/1 | 1/1 |
| Hepatocellular carcinoma (Liver) | 0/0 | 0/0 | 0/0 | 1/1 |
| Clear cell carcinoma (Kidney) | 1/1 | 1/1 | 1/1 | 1/1 |
| Adenocarcinoma (Prostate) | 2/2 | 1/1 | 2/2 | 2/2 |
| Leiomyoma (Uterus) | 1/1 | 0/0 | 1/1 | 0/0 |
| Adenocarcinoma (Uterus) | 1/1 | 0/0 | 1/1 | 1/1 |
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{11}
| Tissue | MLH1 #
Positive / Total Cases | PMS2 #
Positive / Total Cases | MSH2 #
Positive / Total Cases | MSH6 #
Positive / Total Cases |
| --- | --- | --- | --- | --- |
| Squamous cell carcinoma (Cervix) | 2/2 | 1/1 | 2/2 | 2/2 |
| Clear cell carcinoma (Endometrium) | 0/0 | 0/0 | 0/0 | 1/1 |
| Embryonal rhabdomyosarcoma (Striated muscle) | 1/1 | 1/1 | 1/1 | 1/1 |
| Squamous cell carcinoma (Skin) | 1/1 | 1/1 | 1/1 | 1/1 |
| Neurofibroma (Lumbar) | 1/1 | 0/0 | 0/0 | 0/0 |
| Neuroblastoma (Retroperitoneum) | 1/1 | 1/1 | 1/1 | 1/1 |
| Mesothelioma (Peritoneum) | 1/1 | 1/1 | 1/1 | 1/1 |
| B-cell lymphoma; NOS (Lymph node) | 3/3 | 2/2 | 2/2 | 2/2 |
| Hodgkin’s lymphoma (Lymph node) | 1/1 | 1/1 | 1/1 | 1/1 |
| Leiomyosarcoma (Bladder) | 1/1 | 1/1 | 1/1 | 1/1 |
| Osteosarcoma | 1/1 | 1/1 | 1/1 | 1/1 |
| Spindle cell rhabdomyosarcoma (Peritoneum) | 1/1 | 0/0 | 0/0 | 1/1 |
| Leiomyosarcoma (Smooth muscle) | 1/1 | 1/1 | 1/1 | 1/1 |
Note: Mismatch repair proteins are present in all actively proliferating cells. For all tissues, positive/negative staining was determined for tumor cells in the presence of positive staining in normal control cells (lymphocytes, fibroblasts, and epithelial cells).
## 3. Robustness
### a. Tissue Thickness
Staining performance of the VENTANA MMR RxDx Panel was assessed on FFPE solid tumor tissue sections including EC tissues at the standard tissue thickness of 4 µm. These results show that a tissue thickness of 4 µm is suitable for consistent VENTANA MMR RxDx Panel antibody staining performance, however, no other tissue thicknesses have been validated.
### b. Protocol Limitations and Failure Modes
The purpose of this study was to identify protocol conditions that might lead to a potential false positive, false negative, or unacceptable result and prevent these conditions from affecting the end user. Tumor Stain Intensity, Internal Control Stain Intensity, Background Staining, and Antibody Status were used to evaluate the impact of varied protocol conditions on solid tumor tissue. The conditions tested were; primary antibody incubation time and concentration, cell conditioning duration, counterstain times, critical dispense failure modes, and HRP Multimer and HQ Linker incubation times. Since the prevalence of EC with dMMR status is low, the protocol limitations and failure modes studies were supplemented with a variety of additional tumor types.
The sample distribution was as follows:
Four EC samples (2 intact and 2 loss) for each biomarker of the MMR panel.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{12}
MLHI: 9 (5 intact and 4 loss); ovary (2 cases), stomach (1 case), bladder (2 cases).
PMS2: 10 (5 intact and 5 loss); gastric (2 case), ovary (2 case) and bladder (2 case), and intact cases included a challenging bladder case.
MSH2: 10 (5 intact and 5 loss); ovary (2 cases), stomach (2 cases), bladder (1 case), ureter (1 case).
MSH6: 9 (5 intact and 4 loss); ovary (1 case), ureter (1 case), gastric (2 cases), renal pelvis (1 case).
Results from failure modes tested for each antibody of the VENTANA MMR RxDx Panel are summarized in Tables 6-9. Standard staining protocol conditions for the VENTANA MMR RxDx Panel antibodies are provided in Table 1 (refer to sub-section E (Principles of Procedure) under Section V (Device Description) above.
Table 6. MLH1 Protocol Limitations and Failure Modes
| Protocol Limitations and Failure Modes for the VENTANA anti-MLH1 (M1) Primary Antibody | |
| --- | --- |
| Protocol Step | Results |
| Cell Conditioning (CC1) | Standard Antigen Retrieval: 64 min.
Antigen retrieval duration (40 min CC1 and 80 min CC1) resulted in no change of MLH1 status. No protocol limitation required. |
| Primary Antibody | Standard anti-MLH1 Antibody Incubation: 24 min.
Primary antibody Incubation (4 min and 32 min) resulted in no change of MLH1 status. No protocol limitation required.
Ab Titer: 1/4x of standard resulted in greater than 1.0 pt. difference in the stain intensity as compared to the reference.
This has been identified as a potential failure mode. |
| OptiView HQ Linker | No reagent on slide (no dispense) resulted in unevaluable MLH1 status.
This has been identified as a potential failure mode. |
| OptiView Multimer | No reagent on slide (no dispense) resulted in unevaluable MLH1 status.
This has been identified as a potential failure mode. |
| Counterstain: Hematoxylin II | Standard Counterstain: 4 min
Counterstain duration (32 min) resulted in no change of MLH1 status. No protocol limitation required. |
| Post Counterstain: Bluing Reagent | Standard Post Counterstain: 4 min
Post Counterstain duration (32 min) resulted in no change of MLH1 status. No protocol limitation required. |
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{13}
Table 7. PMS2 Protocol Limitations and Failure Modes
| Protocol Limitations and Failure Modes for the VENTANA anti-PMS2 (A16-4) Primary Antibody | |
| --- | --- |
| Protocol Step | Results |
| Cell Conditioning (CC1) | Standard Antigen Retrieval: 92 min. Antigen retrieval duration (56 min CC1 and 104 min CC1) resulted in no change of PMS2 status. No protocol limitation required. |
| Primary Antibody | Standard PMS2 Antibody Incubation: 32 min. 1) Ab incubation duration (8 min) resulted in unevaluable PMS2 status. This has been identified as a potential protocol limitation. 2) Ab Titer: 0.25x and 0.5x the optimal titer resulted in greater than 1.0 pt. difference in the stain intensity as compared to the reference but there was no change in status. 1.5x and 2x the optimal titer resulted in unacceptable high background. This has been identified as a potential failure mode. |
| OptiView HQ Linker | No reagent on slide (no dispense) resulted in unevaluable PMS2 status. This has been identified as a potential failure mode. |
| OptiView HRP Multimer | No reagent on slide (no dispense) resulted in unevaluable PMS2 status. This has been identified as a potential failure mode. |
| OptiView Amplification | No reagent on slide (no dispense) resulted in either unevaluable PMS2 status or greater than 1.0 pt. difference in the stain intensity as compared to the reference without change in status. This has been identified as a potential failure mode. |
| Counterstain: Hematoxylin II | Standard Counterstain: 4 min. Counterstain duration (32 min) resulted in no change of PMS2 status. No protocol limitation required. |
| Post Counterstain: Bluing Reagent | Standard Post Counterstain: 4 min. Post Counterstain duration (32 min) resulted in no change of PMS2 status. No protocol limitation required. |
Table 8. MSH2 Protocol Limitations and Failure Modes
| Protocol Limitations and Failure Modes for the VENTANA anti-MSH2 (G219-1129) Primary Antibody | |
| --- | --- |
| Protocol Step | Results |
| Cell Conditioning (CC1) | Standard Antigen Retrieval: 40 min. Antigen retrieval duration (24 min CC1 and 72 min CC1) resulted in unevaluable MSH2 status. These conditions have been identified as potential protocol limitations. |
| Primary Antibody | Standard Primary Antibody Incubation: 12 min. Ab incubation duration (4 min) resulted in unevaluable MSH2 status. This condition has been identified as a potential protocol limitation. Ab Titer: 1/4x, 1.5x, 2x of standard resulted in unevaluable MSH2 status. This condition has been identified as a potential failure mode. |
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{14}
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
| Protocol Limitations and Failure Modes for the VENTANA anti-MSH2 (G219-1129) Primary Antibody | |
| --- | --- |
| OptiView HQ Linker | No reagent on slide (no dispense) resulted in unevaluable MSH2 status. This condition has been identified as a potential failure mode. |
| OptiView Multimer | No reagent on slide (no dispense) resulted in unevaluable MSH2 status. This condition has been identified as a potential failure mode. |
| Counterstain: Hematoxylin II | **Standard** Counterstain: 4 min.
Counterstain duration (32 min) resulted in no change of MSH2 status. No protocol limitation required. |
| Post Counterstain: Bluing Reagent | **Standard** Post Counterstain: 4 min.
Post Counterstain duration (32 min) resulted in no change of MSH2 status. No protocol limitation required. |
Table 9. MSH6 Protocol Limitations and Failure Modes
| Protocol Limitations and Failure Modes for the VENTANA anti-MSH6 (SP93) Rabbit Monoclonal Primary Antibody | |
| --- | --- |
| Protocol Step | Results |
| Cell Conditioning (CC1) | **Standard Antigen Retrieval: 48 min.**
Antigen retrieval duration (32 min and 96 min) resulted in no change of MSH6 status. No protocol limitation required. |
| Primary Antibody | Standard Primary Antibody Incubation: 16 min.
Antibody Incubation duration (32 min) resulted in no change of status. No protocol limitation required.
Antibody Incubation duration (4 min) resulted in a greater than 1.0 pt. stain intensity for one case (VT0000237595) when compared to the reference slide but there was no change in MSH6 status. This has been identified as a potential protocol limitation.
Ab Titer of 2x that of the optimal titer resulted in unacceptable background and non-evaluable internal controls for one case when compared to the reference slide and there was also a change in MSH6 status. This has been identified as a potential failure mode. |
| OptiView HQ Linker | No reagent on slide (no dispense) resulted in change of MSH6 status. This has been identified as a potential failure mode. |
| OptiView Multimer | No reagent on slide (no dispense) resulted in change of MSH6 status. This has been identified as a potential failure mode. |
| Counterstain: Hematoxylin II | **Standard** Counterstain: 4 min.
Counterstain duration of 32 min resulted in no change of MSH6 status. No protocol limitation required. |
| Post Counterstain: Bluing Reagent | **Standard** Post Counterstain: 4 min.
Post Counterstain duration of 32 min resulted in no change of MSH6 status. No protocol limitation required. |
Page 15
{15}
Results from Failure Modes tested for each Antibody of the VENTANA MMR RxDx Panel (primary antibody titer and failed reagent dispenses for critical components) did result in failures and the internal controls were able to accurately identify the failures. A higher than optimal concentration of antibody could lead to a false status change (loss to intact). Varied protocol conditions (primary antibody incubation time, cell conditioning duration, counterstain time and HRP Multimer and HQ Linker incubation times) do not adversely affect the ability to properly interpret and determine loss or intact status.
## 4. Precision
The purpose of this study was to evaluate precision of the VENTANA MMR RxDx Panel. For each biomarker of the MMR panel, the intermediate precision studies used 6 cases of EC with a balanced loss/intact status distribution. Since the prevalence of EC with dMMR status is low, the intermediate precision studies were supplemented with a variety of additional tumor types.
The sample distribution was as follows:
Six EC samples (3 intact and 3 loss) for each biomarker of the MMR panel.
MLH1: 27 (15 intact and 12 loss): urinary (6 cases), reproductive (9 cases), gastrointestinal (6 cases), hepato-pancreatobiliary (3 cases) soft tissue (1 case), and thoracic (2 cases).
PMS2: 26 (14 intact and 12 loss): urinary (3 cases), reproductive (9 cases), gastrointestinal (7 cases), endocrine (2), hepato-pancreatobiliary (1 case), soft tissue (2 cases), and thoracic (2 cases).
MSH2: 27 (15 loss and 12 intact loss): urinary (6 cases), reproductive (12 cases), gastrointestinal (4 cases), hepato-pancreatobiliary (1 case), soft tissue (2 cases), and thoracic (2 cases).
MSH6: 28 (15 intact and 13 loss): urinary (4 cases), reproductive (11 cases), gastrointestinal (2 cases), endocrine (3 cases), hepato-pancreatobiliary (2 cases) soft tissue (4 cases), and thoracic (2 cases).
Total number of observations for each test condition is depicted in Tables 10-14 below.
The following parameters were tested:
- Three lots of antibody (between-antibody lots)
- Three lots of OptiView DAB IHC Detection Kits (between-detection kits) and OptiView Amplification (where applicable)
- Three ULTRA instruments (between-instruments)
- Across 3 days (between-day)
- Across all intermediate precision conditions (within-run).
Each case was assigned one mode based on the samples aggregated per test condition for:
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 16
{16}
- Between-antibody lots
- Between-detection kit lots
- Between-instruments
- Between-day
Each case was compared within its duplicate samples per test run for:
- Within-run
Analyses included evaluation of overall percent agreement (OPA), positive percent agreement (PPA) for Loss cases, and negative percent agreement (NPA) for Intact cases between antibody lot, between detection kit lot, between day, between instrument and within run. Table a is analysis for EC tissue. However, since the prevalence of each of the biomarkers of the dMMR panel is low for EC, the study was supplemented with samples from other tissue types. Therefore, Table b for all test conditions below is analyses of other tissues including EC.
Tables 10 (a and b). Precision, Between-antibody lots
| Table 10a. Between-Antibody Lots - EC Tissue | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Mode | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 100.0
(36/36) | (90.4, 100.0) | 100.0
(36/36) | (90.4, 100.0) | 100.0
(36/36) | (90.4, 100.0) | 100.0
(36/36) | (90.4, 100.0) |
| Intact | 100.0
(18/18) | (82.4, 100.0) | 100.0
(18/18) | (82.4, 100.0) | 100.0
(18/18) | (82.4, 100.0) | 100.0
(18/18) | (82.4, 100.0) |
| Loss | 100.0
(18/18) | (82.4, 100.0) | 100.0
(18/18) | (82.4, 100.0) | 100.0
(18/18) | (82.4, 100.0) | 100.0
(18/18) | (82.4, 100.0) |
| Table 10b. Between-Antibody Lots - Variety of Tumor Tissues Including EC | | | | | | | | |
| Mode | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 100.0
(162/162) | (97.7, 100.0) | 100.0
(156/156) | (97.6, 100.0) | 100.0
(162/162) | (97.7, 100.0) | 100
(168/168) | (97.8, 100.0) |
| Intact | 100.0
(96/96) | (96.2, 100.0) | 100.0
(84/84) | (95.6, 100) | 100.0
(90/90) | (95.9, 100.0) | 100 (90/90) | (95.9, 100.0) |
| Loss | 100.0
(66/66) | (94.5, 100.0) | 100.0
(72/72) | (94.9, 100) | 100.0
(72/72) | (94.9, 100.0) | 100 (78/78) | (95.3, 100.0) |
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{17}
Tables 11 (a and b). Precision, Between-detection kit lots
| Table 11a. Between-Detection Kit Lots - EC Tissue | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Mode | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 100.0(36/36) | (90.4, 100.0) | 100.0(36/36) | (90.4, 100.0) | 100.0(36/36) | (90.4, 100.0) | 100.0(36/36) | (90.4, 100.0) |
| Intact | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) |
| Loss | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) |
| Table 11b. Between-Detection Kit Lots - Variety of Tumor Tissues Including EC | | | | | | | | |
| Mode | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 98.8(160/162) | (96.9, 100.0) | 94.2(147/156) | (89.7, 98.1) | 100.0(162/162) | (97.7, 100.0) | 100(168/168) | (97.8, 100.0) |
| Intact | 99.0 (95/96) | (96.7, 100.0) | 97.6 (82/84) | (92.9, 100) | 100.0(90/90) | (95.9, 100.0) | 100 (90/90) | (95.9, 100.0) |
| Loss | 98.5 (65/66) | (95.2, 100.0) | 94.2 (65/72) | (83.3, 97.2) | 100.0(72/72) | (94.9, 100.0) | 100 (78/78) | (95.3, 100.0) |
Tables 12 (a and b). Precision, Between-instruments
| Table 12a. Between-Instruments - EC Tissue | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Mode | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 100.0(36/36) | (90.4, 100.0) | 100.0(36/36) | (90.4, 100.0) | 100.0(36/36) | (90.4, 100.0) | 100.0(36/36) | (90.4, 100.0) |
| Intact | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) |
| Loss | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) |
| Table 12b. Between-Instruments - Variety of Tumor Tissues Including EC | | | | | | | | |
| Mode | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 98.8(160/162) | (96.9, 100.0) | 100.0(156/156) | (97.6, 100.0) | 100.0(162/162) | (97.7, 100.0) | 100(168/168) | (97.8, 100.0) |
| Intact | 100.0(96/96) | (96.2, 100.0) | 100.0(84/84) | (95.6, 100) | 100.0(90/90) | (95.9, 100.0) | 100 (90/90) | (95.9, 100.0) |
| Loss | 97.0 (64/66) | (92.4, 100.0) | 100.0(72/72) | (94.9, 100) | 100.0(72/72) | (94.9, 100.0) | 100 (78/78) | (95.3, 100.0) |
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{18}
Tables 13 (a and b). Precision, Between-days
| Table 13a. Between-Days - EC Tissue | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Mode | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 100.0(36/36) | (90.4, 100.0) | 100.0(36/36) | (90.4, 100.0) | 100.0(36/36) | (90.4, 100.0) | 100.0(36/36) | (90.4, 100.0) |
| Intact | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) |
| Loss | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) | 100.0(18/18) | (82.4, 100.0) |
| Table 13b. Between-Days - Variety of Tumor Tissues Including EC | | | | | | | | |
| Mode | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 99.4(161/162) | (98.1, 100.0) | 100.0(156/156) | (97.6, 100.0) | 100.0(162/162) | (97.7, 100.0) | 100(168/168) | (97.8, 100.0) |
| Intact | 100.0(96/96) | (96.2, 100.0) | 100.0(84/84) | (95.6, 100) | 100.0(90/90) | (95.9, 100.0) | 100(90/90) | (95.9, 100.0) |
| Loss | 98.5(65/66) | (95.2, 100.0) | 100.0(72/72) | (94.9, 100) | 100.0(72/72) | (94.9, 100.0) | 100(78/78) | (95.3, 100.0) |
Table 14 (a and b). Precision, Within Run
| Table 14a. Within-Run - EC Tissue | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Mode | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 100.0(54/54) | (93.4, 100.0) | 100.0(54/54) | (93.4, 100.0) | 100.0(54/54) | (93.4, 100.0) | 100.0(54/54) | (93.4, 100.0) |
| Intact | 100.0(27/27) | (87.5, 100.0) | 100.0(27/27) | (87.5, 100.0) | 100.0(27/27) | (87.5, 100.0) | 100.0(27/27) | (87.5, 100.0) |
| Loss | 100.0(27/27) | (87.5, 100.0) | 100.0(27/27) | (87.5, 100.0) | 100.0(27/27) | (87.5, 100.0) | 100.0(27/27) | (87.5, 100.0) |
| Table 14b. Within-Run - Variety of Tumor Tissues Including EC | | | | | | | | |
| Mode | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 98.8(240/243) | (97.3, 100.0) | 100.0(228/228) | (98.3, 100.0) | 100.0(243/243) | (98.4, 100.0) | 100.0(252/252) | (98.5, 100.0) |
| Intact | 99.3(143/144) | (97.8, 100.0) | 100.0(125/125) | (97.0, 100.0) | 100.0(135/135) | (97.2, 100.0) | 100.0(135/135) | (97.2, 100.0) |
| Loss | 98.0(97/99) | (94.9, 100.0) | 100.0(103/103) | (96.4, 100.0) | 100.0(108/108) | (96.6, 100.0) | 100.0(117/117) | (96.8, 100.0) |
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{19}
Between Day Precision for VENTANA MMR RxDx Panel
The precision of the VENTANA MMR RxDx Panel on EC samples when stained across multiple days (5 non-consecutive) was also evaluated. For biomarkers of the MMR panel, the between-day precision study used 6 cases of EC for PMS2 and MSH2, and 8 cases of EC for MLH1 and MSH6 with a balanced loss/intact status distribution for each biomarker. OPA, PPA, and NPA were assessed, and the agreement results are presented in Table 15 below.
Table 15. Precision, Between Day for VENTANA MMR RxDx Panel
| Overall Agreement Between Days | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Antibody | Slide-Level Marker Status | Marker Mode Status | | | Measure | %(n/N) | 95% CI |
| | | Loss | Intact | Total | | | |
| MLH1 | Loss | 40 | 0 | 40 | PPA | 100.0 (40/40) | (91.2, 100.0) |
| | Intact | 0 | 40 | 40 | NPA | 100.0 (40/40) | (91.2, 100.0) |
| | Total | 40 | 40 | 80 | OPA | 100.0 (80/80) | (95.4, 100.0) |
| PMS2 | Loss | 30 | 0 | 30 | PPA | 100.0 (30/30) | (88.6, 100.0) |
| | Intact | 0 | 30 | 30 | NPA | 100.0 (30/30) | (88.6, 100.0) |
| | Total | 30 | 30 | 60 | OPA | 100.0 (60/60) | (94.0, 100.0) |
| MSH2 | Loss | 30 | 0 | 30 | PPA | 100.0 (30/30) | (88.6, 100.0) |
| | Intact | 0 | 30 | 30 | NPA | 100.0 (30/30) | (88.6, 100.0) |
| | Total | 30 | 30 | 60 | OPA | 100.0 (60/60) | (94.0, 100.0) |
| MSH6 | Loss | 40 | 0 | 40 | PPA | 100.0 (40/40) | (91.2, 100.0) |
| | Intact | 0 | 40 | 40 | NPA | 100.0 (40/40) | (91.2, 100.0) |
| | Total | 40 | 40 | 80 | OPA | 100.0 (80/80) | (95.4, 100.0) |
Staining with the VENTANA MMR RxDx Panel antibodies on EC tissue met study requirements, which stated that the primary antibody shall produce concordant staining results in at least 90% of samples.
5. **Reader Precision**
Between-Reader and Within-Reader precision was assessed by evaluating concordance of VENTANA MMR RxDx Panel status between 3 readers and within individual readers using 134 (72 proficient and 62 deficient) cases from a variety of tumor types including 34 (17 proficient and 17 deficient) EC cases.
The sample distribution was as follows:
Thirty-four EC samples (17 proficient and 17 deficient) for each biomarker of the MMR panel.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{20}
134 (72 proficient and 62 deficient): urinary (14 cases), reproductive (46 cases), gastrointestinal (33 cases), endocrine (7 cases), hepato-pancreatobiliary (11 cases), soft tissue/skin (9 cases), thoracic (9 cases) and other (head and neck- 5 cases)
Specimens were blinded and randomized prior to evaluation of each of the 4 individual antibody status (intact or loss) and panel-level status (proficient of deficient) for each case using the VENTANA MMR RxDx Panel scoring algorithm (Staining Interpretation Section F of the Device Description Section V, and Table 2 above). Readers scored all specimens twice, with a minimum of two weeks between reads. The agreement rates between the readers and within-reader are summarized in Table 16 for EC tissues and Table 17 for variety of tumor tissues including EC.
Table 16. Within-Reader and Between-Reader Precision of the VENTANA MMR RxDx Panel on EC tissues as measured by MMR (Proficient/ Deficient)
| Precision | Clinical Status | Agreement | | | |
| --- | --- | --- | --- | --- | --- |
| | | Type | n/N | % | 95% CI |
| Within-Reader | Deficient | APA | 100/101 | 99.0 | (97.0,100.0) |
| | Proficient | ANA | 102/103 | 99.0 | (97.1,100.0) |
| | Total | OPA | 101/102 | 99.0 | (97.1,100.0) |
| Between-Reader | Deficient | APA | 98/100 | 98.0 | (93.8,100.0) |
| | Proficient | ANA | 102/104 | 98.1 | (94.4,100.0) |
| | Total | OPA | 100/102 | 98.0 | (94.1,100.0) |
Note: Average Positive Agreement (ANA), Average Negative Agreement (ANA), Overall Percent Agreement (OPA).
2-sided 95% confidence interval (CI) was calculated using the percentile bootstrap method from 2000 bootstrap samples.
Table 17. Within-Reader and Between-Reader Precision of the VENTANA MMR RxDx Panel on a variety of tumor tissues including EC as measured by MMR Status (Proficient/ Deficient)
| Precision | Clinical Status | Agreement | | | |
| --- | --- | --- | --- | --- | --- |
| | | Type | n/N | % | 95% CI |
| Within-Reader | Deficient | APA | (362/365) | 99.2 | (98.2,100.0) |
| | Proficient | ANA | (432/435) | 99.3 | (98.5,100.0) |
| | Total | OPA | (397/400) | 99.3 | (98.4,100.0) |
| Between-Reader | Deficient | APA | (356/362) | 98.3 | (96.6,100.0) |
| | Proficient | ANA | (428/434) | 98.6 | (97.2,100.0) |
| | Total | OPA | (392/398) | 98.5 | (96.9,100.0) |
Note: Average Positive Agreement (ANA), Average Negative Agreement (ANA), Overall Percent Agreement (OPA).
2-sided 95% confidence interval (CI) was calculated using the percentile bootstrap method from 2000 bootstrap samples.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{21}
Table 18. Within-Reader and Between-Reader Precision of the VENTANA MMR RxDx Panel on EC tissue as measured for each marker (Intact/ Loss)
| Marker | Parameter | Clinical Status | Agreement | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | Type | n/N | % | 95% CI |
| MLH1 | Within-Reader | Loss | APA | 78/78 | 100.0 | (95.3, 100.0) |
| | | Intact | ANA | 126/126 | 100.0 | (97.0, 100.0) |
| | | Total | OPA | 102/102 | 100.0 | (96.4, 100.0) |
| | Between-Reader | Loss | APA | 78/78 | 100.0 | (95.3, 100.0) |
| | | Intact | ANA | 126/126 | 100.0 | (97.0, 100.0) |
| | | Total | OPA | 102/102 | 100.0 | (96.4, 100.0) |
| PMS2 | Within-Reader | Loss | APA | 78/78 | 100.0 | (95.3, 100.0) |
| | | Intact | ANA | 126/126 | 100.0 | (97.0, 100.0) |
| | | Total | OPA | 102/102 | 100.0 | (96.4, 100.0) |
| | Between-Reader | Loss | APA | 78/78 | 100.0 | (95.3, 100.0) |
| | | Intact | ANA | 126/126 | 100.0 | (97.0, 100.0) |
| | | Total | OPA | 102/102 | 100.0 | (96.4, 100.0) |
| MSH2 | Within-Reader | Loss | APA | 16/17 | 94.1 | (80.0, 100.0) |
| | | Intact | ANA | 186/187 | 99.5 | (98.2, 100.0) |
| | | Total | OPA | 101/102) | 99.0 | (97.1, 100.0) |
| | Between-Reader | Loss | APA | 14/16 | 87.5 | (50.0, 100.0) |
| | | Intact | ANA | 186/188 | 98.9 | (96.6, 100.0) |
| | | Total | OPA | 100/102 | 98.0 | (94.1, 100.0) |
| MSH6 | Within-Reader | Loss | APA | 22/23 | 95.7 | (84.6, 100.0) |
| | | Intact | ANA | 180/181 | 99.4 | (98.2, 100.0) |
| | | Total | OPA | 101/102 | 99.0 | (97.1, 100.0) |
| | Between-Reader | Loss | APA | 20/22 | 90.9 | (63.6, 100.0) |
| | | Intact | ANA | 180/182 | 98.9 | (96.4, 100.0) |
| | | Total | OPA | 100/102 | 98.0 | (94.1, 100.0) |
Note: Average Positive Agreement (ANA), Average Negative Agreement (ANA), Overall Percent Agreement (OPA).
2-sided 95% confidence interval (CI) was calculated using the percentile bootstrap method from 2000 bootstrap samples.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{22}
Table 19. Within-Reader and Between-Reader Precision of the VENTANA MMR RxDx Panel on a variety of tumor tissues including EC as measured for each marker (Intact/ Loss)
| Marker | Parameter | Clinical Status | Agreement | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | Type | n/N | % | 95% CI |
| MLH1 | Within-Reader | Loss | APA | 290/291 | 99.7 | (98.9,100.0) |
| | | Intact | ANA | 510/511 | 99.8 | (99.4,100.0) |
| | | Total | OPA | 400/401 | 99.8 | (99.3,100.0) |
| | Between-Reader | Loss | APA | 286/290 | 98.6 | (96.4,100.0) |
| | | Intact | ANA | 506/510 | 99.2 | (98.0,100.0) |
| | | Total | OPA | 396/400 | 99.0 | (97.5,100.0) |
| PMS2 | Within-Reader | Loss | APA | 290/293 | 99.0 | (97.7,100.0) |
| | | Intact | ANA | 504/507 | 99.4 | (98.7,100.0) |
| | | Total | OPA | 397/400 | 99.3 | (98.3,100.0) |
| | Between-Reader | Loss | APA | 284/290 | 97.9 | (95.3,100.0) |
| | | Intact | ANA | 500/506 | 98.8 | (97.4,100.0) |
| | | Total | OPA | 392/398 | 98.5 | (96.5,100.0) |
| MSH2 | Within-Reader | Loss | APA | 44/47 | 93.6 | (86.7,100.0) |
| | | Intact | ANA | 752/755 | 99.6 | (99.1,100.0) |
| | | Total | OPA | 398/401 | 99.3 | (98.4,100.0) |
| | Between-Reader | Loss | APA | 40/46 | 87.0 | (69.2,100.0) |
| | | Intact | ANA | 748/754 | 99.2 | (98.1,100.0) |
| | | Total | OPA | 394/400 | 98.5 | (96.5,100.0) |
| MSH6 | Within-Reader | Loss | APA | 70/72 | 97.2 | (93.3,100.0) |
| | | Intact | ANA | 728/730 | 99.7 | (99.3,100.0) |
| | | Total | OPA | 399/401 | 99.5 | (98.8,100.0) |
| | Between-Reader | Loss | APA | 62/70 | 88.6 | (76.9,97.7) |
| | | Intact | ANA | 722/730 | 98.9 | (97.7,99.7) |
| | | Total | OPA | 392/400 | 98.0 | (96.0,99.5) |
Note: Average Positive Agreement (ANA), Average Negative Agreement (ANA), Overall Percent Agreement (OPA).
2-sided 95% confidence interval (CI) was calculated using the percentile bootstrap method from 2000 bootstrap samples.
Reader concordance of VENTANA MMR RxDx Panel staining on a variety of tumor cases met all the requirements and acceptance criteria for Within-Reader and Between-Reader precision studies.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{23}
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 24
# 6. External Reproducibility
The reproducibility of VENTANA MMR RxDx Panel, as used on the BenchMark ULTRA instrument with OptiView DAB detection and OptiView amplification, in determining MMR status in EC was evaluated.
Thirty archival, de-identified, FFPE EC specimens were used in the study. Fifteen of the cases were deficient for overall MMR status (dMMR) and 15 cases were proficient for overall MMR status (pMMR). Four cases considered to be challenging to interpret (refer to sub-section F (Staining Interpretation) under section V (Device Description) were included.
Multiple slides containing serial sections of the specimens were distributed to each of 3 external clinical laboratories (study sites) for staining with VENTANA MMR RxDx Panel on BenchMark ULTRA automated staining instruments. On each of 3 non-consecutive staining days at each site, site study staff stained 5 sections of each of the 30 cases, using the 4 antibody components of MMR RxDx Panel and a negative control antibody. The first and last staining days at a site were at least 20 days apart. All sites used the same investigational product lots.
Each set of 5 stained slides per case per staining day was combined with a case-matched hematoxylin and eosin (H&E)-stained slide and provided as a 6-slide panel to 2 trained, blinded pathologists (readers) at the site. Each pathologist independently interpreted each case panel for the status (Intact or Loss) of each of the 4 VENTANA MMR RxDx Panel biomarkers (MLH1, PMS2, MSH2 and MSH6) and for the overall MMR status of the case (dMMR or pMMR). A case was considered pMMR if all 4 biomarkers were Intact and was considered dMMR if any one of the four biomarkers had a Loss status.
PPA and NPA rates for MMR status (dMMR or pMMR) across all evaluable observations, using the case-level reader modes for MMR status as the reference were assessed. Study results are presented for panel-level status in Table 20, and for marker-level status in Tables 21 below.
Table 20. External Reproducibility Study for VENTANA MMR RxDx Panel
| | Overall | | 1. Between-Site | | Between-Reader | |
| --- | --- | --- | --- | --- | --- | --- |
| Analysis | % (n/N) | 95% CI* | % (n/N) | 95% CI* | % (n/N) | 95% CI* |
| PPA | 98.1 (263/268) | (95.5, 100.0) | 98.1 (263/268) | (95.5, 100.0) | 99.2 (263/265) | (98.1, 100.0) |
| NPA | 100.0 (269/269) | (98.6, 100.0) | 100.0 (269/269) | (98.6, 100.0) | 100.0 (272/272) | (98.6, 100.0) |
| OPA | 99.1 (532/537) | (97.8, 100.0) | 99.1 (532/537) | (97.8, 100.0) | 99.6 (535/537) | (99.1, 100.0) |
* Two-sided 95% CIs for point estimates (PEs) of 100% were calculated using the Wilson score method; for other PEs, they were calculated using the percentile bootstrap method with 2000 replicates.
{24}
Table 21. External Reproducibility Study for VENTANA MMR RxDx Panel on EC tissue as measured for each marker
| | | Agreement with Mode [b] | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Marker | PPA | | NPA | | OPA | |
| | | % (n/N)[a] | 95% CI [c] | % (n/N) | 95% CI | % (n/N) | 95% CI |
| Agreement of Reader Biomarker Status with the Case-Level Reader Modal Status | MLH1 | 98.8 (160/162) | (97.2, 100.0) | 100.0 (377/377) | (99.0, 100.0) | 99.6 (537/539) | (99.1, 100.0) |
| | PMS2 | 98.1 (159/162) | (96.3, 100.0) | 100.0 (377/377) | (99.0, 100.0) | 99.4 (536/539) | (98.9, 100.0) |
| | MSH2 | 96.2 (102/106) | (90.0, 100.0) | 99.5 (429/431) | (98.9, 100.0) | 98.9 (531/537) | (97.6, 99.8) |
| | MSH6 | 88.9 (96/108) | (78.9, 98.1) | 99.3 (428/431) | (98.6, 100.0) | 97.2 (524/539) | (95.2, 99.3) |
| Within Site Agreement of Biomarker Status with the Case-Level Mode | MLH1 | 98.8 (160/162) | (97.2, 100.0) | 100.0 (377/377) | (99.0, 100.0) | 99.6 (537/539) | (99.1, 100.0) |
| | PMS2 | 98.1 (159/162) | (96.3, 100.0) | 100.0 (377/377) | (99.0, 100.0) | 99.4 (536/539) | (98.9, 100.0) |
| | MSH2 | 96.2 (102/106) | (90.0, 100.0) | 99.5 (429/431) | (98.9, 100.0) | 98.9 (531/537) | (97.6, 99.8) |
| | MSH6 | 94.1 (96/102) | (88.2, 98.6) | 99.3 (434/437) | (98.6, 100.0) | 98.3 (530/539) | (97.0, 99.3) |
| Within-Reader Agreement of Biomarker Status with the Case-Level Mode | MLH1 | 98.8 (160/162) | (97.2, 100.0) | 100.0 (377/377) | (99.0, 100.0) | 99.6 (537/539) | (99.1, 100.0) |
| | PMS2 | 98.1 (159/162) | (96.3, 100.0) | 100.0 (377/377) | (99.0, 100.0) | 99.4 (536/539) | (98.9, 100.0) |
| | MSH2 | 99.0 (102/103) | (96.9, 100.0) | 99.5 (432/434) | (98.9, 100.0) | 99.4 (534/537) | (98.9, 100.0) |
| | MSH6 | 97.0 (96/99) | (94.4, 100.0) | 99.3 (437/440) | (98.7, 100.0) | 98.9 (533/539) | (98.1, 99.6) |
NPA = negative percent agreement; PPA = positive percent agreement OPA = overall percent agreement
[a] Counts indicate numbers of reader observations, not numbers of unique cases.
[b] For the purpose of agreement rate calculations, a biomarker status of Loss was considered Positive and a biomarker status of Intact was considered Negative.
[c] Two-sided $95\%$ CIs were calculated using the percentile bootstrap method with 2000 replicates, except that those for point estimates of $100\%$ were calculated using the Wilson score method.
Results of the External Reproducibility Study for VENTANA MMR RxDx Panel showed that one of the biomarkers, MSH6, presented relatively low PPA with point estimate $88.9\%$ and lower bound for its $95\%$ CI as $78.9\%$ . It was observed that the discordant observations for MSH6 (relative to a Loss mode) were largely concentrated at a single Site C (11 of 15 discordant MSH6 observations) and mostly associated with a single Reader C1 (7 of the 11 discordant Site C observations). Further, the decreased PPA for MSH6 relative to that for the other MMR biomarkers had little effect on MMR status reproducibility, as all study cases with a Loss (positive) MSH6 status also had a Loss MSH2 status, and the PPA for MSH2 status was high $(\mathrm{PPA} = 96.2\%; 95\% \mathrm{CI}: 90.0, 100.0)$ . Based on these observations, the relatively low PPA of MSH6 may not be assay related or a major issue for the final assessment of the VENTANA MMR RxDx Panel assay results.
# 7. Stability Studies
# a. Cut Slide Stability
The purpose of the cut slide stability study was to determine the stability of VENTANA MMR RxDx Panel antibody epitopes in FFPE EC tissues sections. Two EC tissues with
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{25}
intact MMR status were sectioned and stored at two different conditions for the duration of the study (5±3°C and 30±5°C). Slides were stained with each of the VENTANA MMR RxDx Panel antibodies at different time points and compared to the Time 0 reference as follows: Day 0 (reference), Day 15, Day 30 and Day 45.
Based on the study results, the cut slide stability is 45 days at both the 5±3°C and 30±5°C storage conditions.
## b. Real Time Stability
The purpose of this study was to evaluate the stability (shelf-life, in-use and shipping) of VENTANA MMR antibodies (MLH1, PMS2, MSH2 and MSH6). Stability testing evaluated staining performance on FFPE tumor tissue. Three production lots of each MMR panel antibody were subjected to the ship stress conditions and tested at specified intervals until 26 months or failure, whichever occurred first. The conditions tested were as follows:
- Ship Stress
- Intended Storage (2°C -8°C)
- Hot Ship Stress (30°C±5°C 192 hours)
- Hot Ship Stress (15°C±5°C 192 hours)
- Cold Ship Stress Freeze/Thaw (-20°C±5°C 192 hours)
- On-Board Stability
- Open Vial Stability
FFPE tumor tissues were tested in triplicate at multiple time points. Based on the data that is provided the assigned reagent stability is as follows:
1. MLH1- 24 months
2. PMS2 - 12 months
3. MSH2 - 24 months
4. MSH6 - 24 months
## B. Animal Studies
None
## C. Additional Studies
### 1. Fixative Type and Time
This study evaluated the effect of fixative type, fixation time, and delay to fixation on each of the VENTANA MMR RxDx Panel antibodies using a BenchMark ULTRA instrument with OptiView DAB IHC detection (with OptiView Amplification where applicable). Tonsil tissues were fixed for 1 hour (hr), 6 hr, 12 hr, 24 hr, 48 hr, or 72 hr in each of six different fixatives: 10% neutral buffered formalin [NBF], zinc formalin, 95% alcohol, alcohol-formalin-acetic acid [AFA], Z-5, or PREFER. Additionally, tonsil tissue was initially harvested and then stored for 0 min, 30 min, 1 hr, 2 hr, 6 hr, or 24 hr prior to fixation in 10% NBF.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{26}
Based on the results of this study, fixation of tissue in 10% NBF, zinc formalin, or Z-5 for 6 to 72 hours produced acceptable staining results with the VENTANA MMR antibody assays. Tissue fixation in 95% alcohol, AFA, or PREFER resulted in weak and variable staining. These fixatives (95% alcohol, AFA, PREFER) are not recommended for use with the VENTANA MMR RxDx Panel antibody assays.
## 2. Tissue Heterogeneity
This study investigated the prevalence of case heterogeneity in EC tissue blocks when stained with VENTANA MMR RxDx Panel Assay on the BenchMark ULTRA instrument.
For the sample set of 44 tissues (22 pairs) evaluated, 10 of 44 EC tissues demonstrated dMMR status. All 22 pairs exhibited equivalent marker and panel level MMR status. Therefore, case heterogeneity was not observed in any of patient pairs for overall panel-level nor marker-level MMR status. Although the trend in this study indicates case heterogeneity is unlikely to be observed in EC, the small sample size of the available dMMR tissues provide a limited assessment of case heterogeneity for MMR panel status.
## 3. Primary versus Metastatic
This study assessed the concordance of MMR panel status between FFPE matched EC primary and metastatic tumors when stained with the MMR RxDx Panel assay on the BenchMark ULTRA instrument.
For the sample set of 20 tissue cases (10 pairs) evaluated, 4 cases exhibited an MMR status of deficient (dMMR), and 16 cases exhibited an MMR status of proficient (pMMR). All 10 pairs exhibited equivalent marker and panel level MMR status. Therefore, concordance between each matched primary vs. metastatic pair was observed with no change in individual status. Due to the difficulty of procuring matching patient samples from commercial vendors and the overall low prevalence of MMR biomarkers, the sample distribution for MMR status in this study was unequal, with only 4 of 20 cases representing a dMMR status. Although the trend in this study indicates discordance is unlikely to be observed between EC primary and their matched metastatic tumor, the small sample size of the available dMMR tissues provide a limited assessment of primary vs. metastatic concordance for MMR panel status. An OPA of 100% and the corresponding 95% Wilson Score confidence intervals of 72.2-100.0% were observed.
## X. SUMMARY OF CLINICAL STUDIES
The clinical performance of VENTANA MMR RxDx Panel as a CDx device for the EC indication for dostarlimab was based on the clinical trial 4010-01-001 (GARNET) – Subpart 2B, Cohort A1.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{27}
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 28
A. Study Design
GARNET is a multicenter, open-label, study with Subpart 2B, designed to evaluate the antitumor activity of dostarlimab in patients with recurrent or advanced mismatch repair deficient (dMMR)/microsatellite instability-high (MSI-H) cancers, including dMMR/MSI-H EC. The Subpart 2B expansion cohorts A1 and A2 are specific for dMMR (loss or biomarker-positive) and pMMR (intact or biomarker-negative) EC, respectively.
1. Clinical Inclusion and Exclusion Criteria (Cohorts A1 and A2)
Key Trial Inclusion Criteria
1. Histologically or cytologically proven recurrent or advanced EC. All EC histologies are allowed except endometrial sarcoma (including carcinosarcoma).
2. Patients who have progressed on or after platinum doublet therapy
3. Patients have received no more than 2 lines of anti-cancer therapy for recurrent or advanced (≥Stage IIIB) disease. Prior treatment with hormone therapies is acceptable and does not count towards the number of anti-cancer therapies noted in the criterion above for this cohort.
4. Patients must submit 2 scans demonstrating increase in tumor measurement that meet criteria for PD on or after the latest systemic anticancer therapy based on RECIST 1.1 to Central Radiology prior to the first dose of dostarlimab.
5. Presence of at least 1 measurable lesion on baseline scan will be confirmed by central radiology review.
6. Status of tumor MMR status as determined by immunohistochemistry (IHC) testing.
Key Trial Exclusion Criteria
1. Patient has received prior therapy with an anti-PD-1, anti-PD-L1, or anti-PD-L2 agent.
2. Patient has known uncontrolled central nervous system metastases and/or carcinomatous meningitis.
3. Patient has a known additional malignancy that progressed or required active treatment within last 2 years.
2. Follow-up Schedule
Up to 2 years, or until the subject meets protocol specific discontinuation criteria
3. Clinical Endpoints
Primary endpoints:
Objective response rate (ORR) and duration of response (DOR) by independent blinded central review using Response Evaluation Criteria in Solid Tumors (RECIST) v1.1
Secondary endpoints:
Progression-free survival (PFS) based on independent blinded central review using RECIST v1.1
{28}
Immune-related overall response rate (irORR) based on Investigators' assessment using irRECIST
## B. Accountability of PMA Cohort
Study participants with recurrent or advanced endometrial carcinoma in the GARNET study comprise the primary analysis population for this application. Of the 249 patients enrolled in the study, 175 subjects were in the BLA efficacy population and 74 were in the BLA non-efficacy population. 167 (118 in the BLA efficacy population and 49 in the BLA non-efficacy population) were able to be tested at the central laboratory to determine their MMR status, including 71 CDx+ and 96 CDx-, although 44 of the samples were associated with one or more deviations from the protocol-defined sample eligibility requirements for testing. Of 118 patients in BLA efficacy population retested by CDx, 53 was detected as CDx+ and 65 as CDx- (Complete cases population). After excluding the cases with protocol deviation, 42 cases were identified as CDx+ and 57 as CDx- (Complete cases IU population).
Table 22. Accountability of PMA Cohort
| | Number of Study Subjects |
| --- | --- |
| Enrolled in GARNET Trial
MMR Proficient (pMMR) by clinical trial assay (CTA)
MMR Deficient (dMMR) by CTA | 249
145 (BLA Cohort A2)
104 (BLA Cohort A1) |
| Tumor sample collected at baseline
Tissue available but excluded from VENTANA MMR RxDx
Panel (CDx) testing [i.e., from Intent-to-Diagnose (ITD) population]
Tested by CDx (ITD population)
Total ITD with evaluable CDx MMR assessment
pMMR
dMMR | 185
18*
167
167
96 (CDx Cohort A2)
71 (CDx Cohort A1) |
| In ITD but associated with diagnostic protocol deviation [i.e., excluded from Intended Use (IU) population]
Total ITD with evaluable CDx MMR assessment
pMMR
dMMR | 44**
123
72 (CDx Cohort A2–IU)
51 (CDx Cohort A1–IU) |
| Specimens with insufficient tissue | 4 HE not acceptable, 14 test results not complete |
| Sample Procedure
Biopsy
Resection
Unknown | 4
25
94 |
*Sample characteristics unsuitable for CDx testing (4 cases) or CDx testing not complete as of diagnostic data snapshot date (14 cases).
**Sample tested but excluded from IU population due to diagnostic protocol deviation(s) concerning sample eligibility for CDx testing. For 43 cases in the ITD population, the deviation was staining of a sample for which the fixative was unknown. Additional deviations concerning 5 cases with unknown fixatives also occurred; An additional case was stained with the CDx test outside of the cut slide stability date.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{29}
# C. Study Population Demographics and Baseline Parameters
The table below summarizes patient demographic information and MMR status for patients tested with VENTANA MMR RxDx Panel (the ITD population).
Table 23. Study Population Demographics and Baseline Parameters
| Characteristic | VENTANA MMR RxDx Panel Status | | Overall (N=167) |
| --- | --- | --- | --- |
| | dMMR (CDx Cohort A1) (N=71) | pMMR (Cohort A2) (N=96) | |
| Age (yr) | | | |
| Mean (SD) | 64.0 (8.31) | 64.3 (9.54) | 64.1 (9.01) |
| Median | 65.0 | 66.0 | 66.0 |
| Minimum, maximum | 39, 80 | 30, 80 | 30, 80 |
| Age Group (yr), n (%) | | | |
| <65 years | 35 (49.3) | 39 (40.6) | 74 (44.3) |
| ≥65 years to <75 years | 28 (39.4) | 48 (50.0) | 76 (45.5) |
| ≥75 years | 8 (11.3) | 9 (9.4) | 17 (10.2) |
| Ethnicity, n (%) | | | |
| Hispanic or Latino | 3 (4.2) | 3 (3.1) | 6 (3.6) |
| Not Hispanic or Latino | 56 (78.9) | 87 (90.6) | 143 (85.6) |
| Not reported | 11 (15.5) | 5 (5.2) | 16 (9.6) |
| Unknown | 1 (1.4) | 1 (1.0) | 2 (1.2) |
| Race, n (%) | | | |
| American Indian or Alaska Native | 3 (4.2) | 1 (1.0) | 4 (2.4) |
| Asian | 2 (2.8) | 4 (4.2) | 6 (3.6) |
| Black | 0 | 8 (8.3) | 8 (4.8) |
| White | 56 (78.9) | 74 (77.1) | 130 (77.8) |
| Other | 0 | 3 (3.1) | 3 (1.8) |
| Not reported | 10 (14.1) | 5 (5.2) | 15 (9.0) |
| Unknown | 0 | 1 (1.0) | 1 (0.6) |
| Country, n (%) | | | |
| Canada | 14 (19.7) | 15 (15.6) | 29 (17.4) |
| Denmark | 1 (1.4) | 4 (4.2) | 5 (3.0) |
| France | 9 (12.7) | 2 (2.1) | 11 (6.6) |
| Italy | 2 (2.8) | 0 | 2 (1.2) |
| Poland | 0 | 1 (1.0) | 1 (0.6) |
| Spain | 21 (29.6) | 19 (19.8) | 40 (24.0) |
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{30}
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 31
| Characteristic | VENTANA MMR RxDx Panel Status | | Overall (N=167) |
| --- | --- | --- | --- |
| | dMMR (CDx Cohort A1) (N=71) | pMMR (Cohort A2) (N=96) | |
| United Kingdom of Great Britain | 5 (7.0) | 7 (7.3) | 12 (7.2) |
| United States | 19 (26.8) | 48 (50.0) | 67 (40.1) |
| Clinical Trial Assay (CTA) MMR Status ^{[a]}, n (%) | | | |
| MMR-deficient (dMMR) | 70 (98.6) | 6 (6.3) | 76 (45.5) |
| MMR-proficient (pMMR) | 1 (1.4) | 90 (93.8) | 91 (54.5) |
| Stage at Initial Diagnosis, n (%) | | | |
| Stage I | 29 (40.8) | 24 (25.0) | 53 (31.7) |
| Stage II | 4 (5.6) | 6 (6.3) | 10 (6.0) |
| Stage III | 25 (35.2) | 33 (34.4) | 58 (34.7) |
| Stage IV | 12 (16.9) | 32 (33.3) | 44 (26.3) |
| Unknown | 1 (1.4) | 1 (1.0) | 2 (1.2) |
| Grade Endometrial Cancer at First Diagnosis, n (%) | | | |
| Grade 1 | 21 (29.6) | 7 (7.3) | 28 (16.8) |
| Grade 2 | 25 (35.2) | 14 (14.6) | 39 (23.4) |
| Grade 3 | 21 (29.6) | 56 (58.3) | 77 (46.1) |
| Not Assessable | 4 (5.6) | 19 (19.8) | 23 (13.8) |
| Baseline ECOG Performance, n (%) | | | |
| 0 | 27 (38.0) | 55 (57.3) | 82 (49.1) |
| 1 | 44 (62.0) | 41 (42.7) | 85 (50.9) |
| Histology, n (%) | | | |
| Adenocarcinoma | 1 (1.4) | 0 | 1 (0.6) |
| Adenosquamous | 1 (1.4) | 0 | 1 (0.6) |
| Carcinoma epidermoide | 0 | 1 (1.0) | 1 (0.6) |
| Carcinosarcoma | 0 | 1 (1.0) | 1 (0.6) |
| Clear cell carcinoma | 0 | 8 (8.3) | 8 (4.8) |
| Endometrial carcinoma Type II | 14 (19.7) | 15 (15.6) | 29 (17.4) |
| Endometrial neuroendocrine carcinoma | 0 | 1 (1.0) | 1 (0.6) |
| Endometrioid carcinoma (type unknown) | 1 (1.4) | 0 | 1 (0.6) |
| Endometrioid carcinoma Type I | 46 (64.8) | 24 (25.0) | 70 (41.9) |
| High-grade uterine carcinoma | 0 | 1 (1.0) | 1 (0.6) |
| Mixed carcinoma | 2 (2.8) | 5 (5.2) | 7 (4.2) |
| Moderately differentiated adenocarcinoma | 0 | 1 (1.0) | 1 (0.6) |
{31}
| Characteristic | VENTANA MMR RxDx Panel Status | | Overall (N=167) |
| --- | --- | --- | --- |
| | dMMR (CDx Cohort A1) (N=71) | pMMR (Cohort A2) (N=96) | |
| Papillary serous carcinoma | 0 | 1 (1.0) | 1 (0.6) |
| Serous carcinoma | 2 (2.8) | 31 (32.3) | 33 (19.8) |
| Squamous carcinoma | 0 | 3 (3.1) | 3 (1.8) |
| Undifferentiated carcinoma | 3 (4.2) | 4 (4.2) | 7 (4.2) |
| Unknown | 1 (1.4) | 0 | 1 (0.6) |
| Number of Prior Therapies, n (%) | | | |
| One prior therapy | 43 (60.6) | 54 (56.3) | 97 (58.1) |
| ≥Two prior therapies | 28 (39.4) | 42 (43.8) | 70 (41.9) |
| Prior Radiation, n (%) | | | |
| No prior radiation | 14 (19.7) | 34 (35.4) | 48 (28.7) |
| Prior radiation | 57 (80.3) | 62 (64.6) | 119 (71.3) |
| Prior Bevacizumab, n (%) | | | |
| No prior bevacizumab | 70 (98.6) | 90 (93.8) | 160 (95.8) |
| Prior bevacizumab | 1 (1.4) | 6 (6.3) | 7 (4.2) |
[a] MMR result from local or central testing using an IHC assay other than VENTANA MMR RxDx Panel.
## D. Safety and Effectiveness Results
### 1. Safety Results
Adverse events (AEs) in the GARNET trial were specific to the toxicity of the investigational agent. There were no device-specific adverse events in the diagnostic study. In this trial, observed AEs included events that were in line with those expected in subjects with recurrent or advanced endometrial cancer, as well as those consistent with reported safety profiles of monoclonal antibodies blocking the PD 1 interactions. Based on the treatment-emergent adverse events (TEAEs) in this trial, the product labeling for dostarlimab reflects that there were 5 (4.8%) TEAEs leading to permanent discontinuation of study drug. Please refer to Drugs@FDA for complete safety information on dostarlimab.
### 2. Effectiveness Results
The analysis of effectiveness was based on the BLA efficacy population (71 in cohort A1 (dMMR)). Table 24 shows the efficacy results.
PMA P200019: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{32}
Table 24: Efficacy Results in GARNET dMMR Endometrial Cancer Population
| Endpoints | JEMPERLI
N = 71 (dMMR) |
| --- | --- |
| Confirmed Overall Response Rate | |
| ORR | 42.3% |
| (95% CI) | (30.6, 54.6) |
| Complete response rate | 12.7% |
| Partial response rate | 29.6% |
| Duration of Response | |
| Median in months | Not reached |
| (range)[a] | (2.6, 22.4+) |
| Patients with duration ≥6 months | 93.3% |
Note: CI = Confidence interval, + = ongoing at last assessment.
[a] Median follow-up for DOR was 14.1 months, measured from time of first response.
##…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.