P210001 · Ventana Medical Systems, Inc. (Roche Tissue Diagnostics) · QNH · Aug 17, 2021 · Hematology
Device Facts
Record ID
P210001
Device Name
VENTANA MMR RxDx Panel
Applicant
Ventana Medical Systems, Inc. (Roche Tissue Diagnostics)
Product Code
QNH · Hematology
Decision Date
Aug 17, 2021
Decision
APPR
Regulation
21 CFR 864.1860
Device Class
Class 3
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
P210001 · Aug 17, 2021
VENTANA MMR RxDx Panel
Ventana Medical Systems, Inc. (Roche Tissue Diagnostics)
Archived FFPE tumor specimens from GARNET clinical trial participants
Retrospective retesting of archived clinical trial samples (Concordance Population) to evaluate the positive and negative percent agreement between the original enrolling clinical trial assays (CTA) and the VENTANA MMR RxDx Panel (CDx).
GARNET Clinical Bridging Study; Retrospective concordance study using archived clinical trial samples; Follow-up/Duration: Not applicable; Study Period: Samples collected prior to March 1, 2020
Patients with recurrent or advanced dMMR solid tumors (including EC and non-EC) from the GARNET trial; Sample Size: 156 (73 EC, 83 non-EC); Number of Sites: Not applicable (central laboratory retesting)
Clinical Trial Assay (CTA) (local or central IHC testing)
Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) between CDx and CTA
Indications for Use
VENTANA MMR RxDx Panel is a qualitative immunohistochemistry test intended for use in the assessment of mismatch repair (MMR) proteins (MLH1, PMS2, MSH2 and MSH6) in formalin-fixed, paraffin-embedded (FFPE) tissue specimens by light microscopy. The OptiView DAB IHC Detection Kit is used for MLH1, MSH2 and MSH6, and the OptiView DAB IHC Detection Kit with the OptiView Amplification Kit is used for PMS2 on a VENTANA BenchMark ULTRA instrument. VENTANA MMR RxDx Panel includes VENTANA anti-MLH1 (M1) Mouse Monoclonal Primary Antibody, VENTANA anti-PMS2 (A16-4) Mouse Monoclonal Primary Antibody, VENTANA anti-MSH2 (G219-1129) Mouse Monoclonal Primary Antibody, and VENTANA anti-MSH6 (SP93) Rabbit Monoclonal Primary Antibody. VENTANA MMR RxDx Panel is indicated as an aid in identifying patients eligible for treatment with the therapy listed in Table 1 for the indication and MMR status in accordance with the approved therapeutic product labeling. Table 1. VENTANA MMR RxDx Panel companion diagnostic indications Indication for use: Solid Tumors; Therapy: JEMPERLI (dostarlimab-gxly); Status: deficient MMR (dMMR). Results of the VENTANA MMR RxDx Panel should be interpreted by a qualified pathologist in conjunction with histological examination, relevant clinical information, and proper controls. This product is intended for in vitro diagnostic (IVD) use.
Device Story
The VENTANA MMR RxDx Panel is an immunohistochemistry (IHC) test used to detect MMR proteins (MLH1, PMS2, MSH2, MSH6) in FFPE solid tumor tissue. The device uses four mouse/rabbit monoclonal primary antibodies on the BenchMark ULTRA instrument. The process involves deparaffinization, heat-induced antigen retrieval, and sequential application of antibodies, followed by an indirect, biotin-free detection system (OptiView DAB) that produces a brown precipitate at antigen sites. PMS2 requires additional signal amplification. A qualified pathologist examines the stained slides via light microscopy to assess nuclear staining. The scoring algorithm is binary: 'Intact' (proficient) if all four proteins are expressed, or 'Loss' (deficient) if at least one protein is absent. This status aids clinicians in identifying patients eligible for JEMPERLI therapy. The device benefits patients by enabling targeted immunotherapy for those with dMMR solid tumors, potentially improving response rates compared to standard therapies.
Clinical Evidence
Clinical performance was established via the GARNET trial (NCT02715284), a multicenter, open-label study. The efficacy population included 209 patients with dMMR solid tumors. The ORR was 41.6% (95% CI: 34.9, 48.6). A clinical bridging study compared the VENTANA MMR RxDx Panel to a Clinical Trial Assay (CTA) in a concordance population (n=156). PPA for dMMR solid tumors was 87.8% (95% CI: 81.8, 92.1) and NPA was 98.7% (95% CI: 92.9, 99.8). Analytical validation included precision studies (within-run, between-day, between-site, between-reader) showing high concordance (OPA >99%).
Technological Characteristics
The device consists of four monoclonal primary antibodies (MLH1, PMS2, MSH2, MSH6) optimized for the BenchMark ULTRA automated staining platform. It utilizes the OptiView DAB IHC Detection Kit (indirect, biotin-free) and OptiView Amplification Kit (for PMS2). Specimens are 4μm FFPE tissue sections. The system is fully automated, performing deparaffinization, antigen retrieval, staining, and counterstaining. Connectivity is standalone via the BenchMark ULTRA instrument.
Indications for Use
Indicated for adult patients with recurrent or advanced solid tumors to identify those eligible for treatment with JEMPERLI (dostarlimab-gxly) based on deficient mismatch repair (dMMR) status.
Regulatory Classification
Identification
Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.
Special Controls
(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
{1}
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
# I. GENERAL INFORMATION
Device Generic Name: Immunohistochemistry test, DNA mismatch repair (MMR) Protein assay
Device Trade Name: VENTANA MMR RxDx Panel
Device Procode: QNH
Applicant's Name and Address: Ventana Medical Systems, Inc. (Roche Tissue Diagnostics) 1910 East Innovation Park Drive Tucson, AZ 85755
Date of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P210001
Date of FDA Notice of Approval: August 17, 2021
# II. INDICATIONS FOR USE
VENTANA MMR RxDx Panel is a qualitative immunohistochemistry test intended for use in the assessment of mismatch repair (MMR) proteins (MLH1, PMS2, MSH2 and MSH6) in formalin-fixed, paraffin-embedded (FFPE) tissue specimens by light microscopy. The OptiView DAB IHC Detection Kit is used for MLH1, MSH2 and MSH6, and the OptiView DAB IHC Detection Kit with the OptiView Amplification Kit is used for PMS2 on a VENTANA BenchMark ULTRA instrument.
VENTANA MMR RxDx Panel includes VENTANA anti-MLH1 (M1) Mouse Monoclonal Primary Antibody, VENTANA anti-PMS2 (A16-4) Mouse Monoclonal Primary Antibody, VENTANA anti-MSH2 (G219-1129) Mouse Monoclonal Primary Antibody, and VENTANA anti-MSH6 (SP93) Rabbit Monoclonal Primary Antibody.
VENTANA MMR RxDx Panel is indicated as an aid in identifying patients eligible for treatment with the therapy listed in Table 1 for the indication and MMR status in accordance with the approved therapeutic product labeling.
Table 1. VENTANA MMR RxDx Panel companion diagnostic indications
| Indication for use | Therapy | Status |
| --- | --- | --- |
| Solid Tumors | JEMPERLI (dostarlimab-gxly) | deficient MMR (dMMR) |
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{2}
Results of the VENTANA MMR RxDx Panel should be interpreted by a qualified pathologist in conjunction with histological examination, relevant clinical information, and proper controls.
This product is intended for in vitro diagnostic (IVD) use.
## III. CONTRAINDICATIONS
There are no known contraindications associated with these tests.
## IV. WARNINGS AND PRECAUTIONS
Warnings and precautions can be found in the anti-MLH1 (M1) Mouse Monoclonal Primary Antibody, anti-PMS2 (A16-4) Mouse Monoclonal Primary Antibody, VENTANA anti-MSH2 (G219-1129) Mouse Monoclonal Primary Antibody, and VENTANA anti-MSH6 (SP93) Rabbit Monoclonal Primary Antibody product labeling.
## V. DEVICE DESCRIPTION
### A. Device Kit Components
The VENTANA MMR RxDx Panel is comprised of four primary antibodies used to detect the MMR proteins MLH1, PMS2, MSH2 and MSH6 in solid tumor tissue specimens. The primary antibodies are used in combination with individually optimized detection reagents and in conjunction with ancillary reagents in order to complete specimen testing. The VENTANA MMR RxDx Panel is optimized to run on the VENTANA BenchMark Ultra platform with OptiView DAB detection kit or in the case of PSM2 antibody the OptiView DAB detection Kit with the OptiView Amplification Kit. The presence or absence of target proteins is determined by visual examination of the specimen slide under light microscope by a qualified pathologist.
The VENTANA MMR RxDx Panel antibodies are packaged as individual products in single ready to use reagent dispensers. The VENTANA MMR RxDx Panel uses four separate intended use tissue slides that are stained on the BenchMark ULTRA instrument. The primary antibody reagents are listed below.
### 1. VENTANA anti-MLH1 (M1) Mouse Monoclonal Primary Antibody
The anti-MLH1 (M1) antibody is a mouse monoclonal antibody produced against a full-length recombinant MLH1 protein with a glutathione-S-transferase tag. The antibody binds to the MLH1 protein in FFPE tissue sections, where it can be localized using the OptiView DAB immunohistochemistry (IHC) Detection Kit. Antibody concentration is ~1 µg/mL.
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{3}
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 3
2. **VENTANA anti-PMS2 (A16-4) Mouse Monoclonal Primary Antibody**
The VENTANA anti-PMS2 (A16-4) antibody is a mouse monoclonal antibody raised against a recombinant PMS2 protein. The antibody binds to PMS2 protein in FFPE tissue sections, where it can be localized using the OptiView DAB IHC Detection Kit and OptiView Amplification Kit. Antibody concentration is ~1 µg/mL.
3. **VENTANA anti-MSH2 (G219-1129) Mouse Monoclonal Primary Antibody**
The anti-MSH2 (G219-1129) antibody is a mouse monoclonal antibody produced against a recombinant human MSH2 protein. The antibody binds to the MSH2 protein in FFPE tissue sections, where it can be localized using the OptiView DAB IHC Detection Kit. Antibody concentration is ~20 µg/mL.
4. **VENTANA anti-MSH6 (SP93) Rabbit Monoclonal Primary Antibody**
The VENTANA anti-MSH6 (SP93) antibody is a rabbit monoclonal antibody raised against a recombinant MSH6 protein. It binds to MSH6 protein in FFPE tissue sections, where it can be localized using the OptiView DAB IHC Detection Kit. Antibody concentration is ~1 µg/mL.
Detection and ancillary reagents required but not provided with VENTANA MMR RxDx Panel are listed below:
- OptiView DAB IHC detection Kit has the following components
- OptiView peroxidase Inhibitor
- OptiView HQ universal Linker
- OptiView HRP Multimer
- OptiView H₂O₂
- OptiView DAB
- OptiView Copper
- OptiView Amplification kit
- OptiView Amplification (0.003% HQ conjugated tyramide complex)
- OptiView H₂O₂
- OptiView Multimer
- Hematoxylin II
- Bluing Reagent
- Reaction Buffer (10x)
- EZ Prep Reagent (10x)
- ULTRA Cell Conditioning (CC1) (Pre-dilute)
- ULTRA Liquid Cover Slip (LCS) (Pre-dilute)
- Negative reagent control mouse monoclonal antibody
- Negative reagent control rabbit monoclonal antibody
{4}
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 4
## B. Device Instrument and Software
The VENTANA MMR RxDx Panel test is fully automated. The VENTANA MMR RxDx panel antibodies are for use on the BenchMark ULTRA instrument using Ventana System Software (VSS) software version 12.5.4 or earlier.
## C. Specimen Preparation
Routinely processed, formalin-fixed, paraffin-embedded (FFPE) tissues are suitable for use with the MMR primary antibodies when used with VENTANA OptiView DAB detection kit (with OptiView Amplification Kit for PMS2) and BenchMark ULTRA instruments. Tissue should be fixed in 10% neutral buffered formalin (NBF) for at least 6 hours and for a maximum of 72 hours at room temperature (15-25 °C). Fixation times of less than 6 hours may result in a loss of staining for the MMR antibodies. The amount of NBF used should be 15 to 20 times the volume of tissue. Fixatives such as alcohol-formalin-acetic acid (AFA), PREFER fixative, or other alcohol-containing fixatives are not recommended for use with this assay.
Sections should be cut 4μm thick and mounted on positively charged glass slides. Slides should be stained promptly, as antigenicity of cut tissue sections may diminish over time.
## D. Test Controls
Run controls should be included in each staining run to establish the validity of the test results. The following controls should be run with the assay:
Pre-qualified solid tumor tissue including endometrial cancer (EC) tissue with an MMR status of intact or tonsil may be used as a positive system-level control for MMR antibodies to detect the intact protein. Alternatively, pre-qualified normal endometrial tissue fixed and processed in the same manner as the patient tissue can also be used as a positive system-level control. Normal endometrium will stain positive for all antibodies in the MMR IHC Panel. Since the MSH6, PMS2, MSH2 and MLH1 proteins are expressed in all tissues, a normal negative tissue control does not exist for these biomarkers. For a negative system level control, EC tissue with loss of an MMR protein can be used as an appropriate tissue control for mismatch repair protein deficiency status. However, lymphocytes, fibroblast and epithelial cells should exhibit staining and serve as positive internal control cells in EC samples with MMR protein deficiency (dMMR).
### 1. Internal Positive Controls
Normal tissue elements (e.g., lymphocytes, fibroblasts, or normal epithelium) in the immediate vicinity of the tumor may serve as internal positive controls. Unequivocal nuclear staining in these cells validates the staining run. If the internal positive controls fail to demonstrate appropriate staining, results with the test specimen should be considered invalid.
### 2. Negative Reagent Control
Negative reagent control should be used to stain an adjacent section of the patient specimen tissue on a separate slide from the VENTANA MMR antibody stained
{5}
slides. A negative reagent control mouse monoclonal antibody (for MLH1, PMS2, and MSH2) and a negative reagent control rabbit monoclonal antibody (for MSH6) is recommended for use in place of the primary antibodies to evaluate nonspecific staining.
Additional information about positive and negative controls are available in the product labeling.
## E. Principles of Procedure
The VENTANA MMR RxDx Panel is an immunohistochemistry test system used to stain FFPE solid tumor specimens to detect expression of the MMR proteins (MLH1, PMS2, MSH2 and MSH6). The 4 antibodies of the VENTANA MMR RxDx Panel have individualized staining protocols that are created using available staining parameters provided in staining procedures in the VSS software that drives the BenchMark ULTRA automated staining platform. The panel test is run individually on 4 separate tissue sections and the test process involves sequential application of specific primary antibodies against the panel protein, followed by detection reagents and chromogen deposition for visualization of the target protein expression.
The VENTANA MMR RxDx Panel is automated for use on the BenchMark ULTRA automated slide stainer from deparaffinization through counterstaining. Patient FFPE tissue specimens are cut 4μm thick and mounted on positively charged glass slides. These slides are loaded into the Benchmark ULTRA instrument. This system first removes the paraffin wax from the tissue, and then subjects the tissue to heated antigen retrieval (cell conditioning). Antigen retrieval is the process by which the ability of antibodies to bind to the epitopes is restored to formalin-fixed tissues.
Endogenous peroxidases that could potentially react with the horseradish peroxidase conjugates (HRP) are blocked with OptiView Inhibitor (3% H2O2). After the endogenous peroxidase block, the Antibody is dispensed during the antibody incubation step and allowed to bind to its antigen. The slides are then incubated with the reagents in the OptiView DAB IHC Detection Kit and OptiView Amplification Kit to achieve preferred staining of target cells. The OptiView DAB IHC Detection Kit is an indirect, biotin-free system for detecting mouse IgG, mouse IgM, and rabbit primary antibodies which produces a visible dark brown precipitate (3,3'-Diaminobenzidine) via an HRP enzymatic reaction at the antigen site. The PMS2 test uses the OptiView amplification in addition to the OptiView DAB detection system for signal amplification. The OptiView Amplification Kit includes an HQ hapten conjugate (OptiView Amplifier), corresponding substrate (OptiView Amplification H2O2), and mouse anti-HQ monoclonal antibody containing HRP (OptiView Amplification Multimer). Tissues are then counterstained blue using Hematoxylin II and Bluing Reagent to create brown/blue contrast to aid the pathologist when reviewing the slides using bright field microscopy.
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 5
{6}
Table 1. VENTANA MMR RxDx Panel Staining Protocol
| Protocol Parameter | MLH1 | PMS2 | MSH2 | MSH6 |
| --- | --- | --- | --- | --- |
| Deparaffinization | Selected | Selected | Selected | Selected |
| Cell Conditioning | Cell Conditioning 1, 64 minutes, 100°C | Cell Conditioning 1, 92 minutes, 100°C | Cell Conditioning 1, 40 minutes, 100°C | Cell Conditioning 1, 48 minutes, 100°C |
| Pre-primary antibody peroxidase | Selected | Selected | Selected | Selected |
| Antibody Incubation or Negative Reagent Control | 24 minutes, 36°C | 32 minutes, 36°C | 12 minutes, 36°C | 16 minutes, 36°C |
| OptiView HQ Linker | 8 minutes (default) | 8 minutes (default) | 8 minutes (default) | 8 minutes (default) |
| OptiView HRP Multimer | 8 minutes (default) | 8 minutes (default) | 8 minutes (default) | 8 minutes (default) |
| OptiView Amplification | Not Selected | Selected | Not Selected | Not Selected |
| Amplifier and Amplification H2O2 | Not Selected | 4 minutes | Not Selected | Not Selected |
| Amplification Multimer | Not Selected | 4 minutes | Not Selected | Not Selected |
| Hematoxylin II | 4 minutes | 4 minutes | 4 minutes | 4 minutes |
| Bluing Reagent | 4 minutes | 4 minutes | 4 minutes | 4 minutes |
F. Staining Interpretation
Stained slides are interpreted by a qualified pathologist and the MMR status (intact or loss) for each of the MMR proteins (MLH1, PMS2, MSH2 and MSH6) is assigned based on the presence or absence of specific nuclear staining in the tumor. An MMR status of "Intact" is assigned to cases with unequivocal nuclear staining in viable tumor cells in the presence of acceptable internal positive controls (nuclear staining in lymphocytes, fibroblasts or normal epithelium in the vicinity of the tumor). An MMR status of "Loss" is assigned to cases with unequivocal loss of nuclear staining or focal weak equivocal nuclear staining in the viable tumor cells in the presence of internal positive controls.
If unequivocal nuclear stain is absent in internal positive controls and/or background staining interferes with interpretation, the assay should be considered unacceptable and repeated. Punctate nuclear staining of tumor cells should be considered negative.
MMR Intact (or proficient) Status: Detection of all four proteins (MLH1, PMS2, MSH2 and MSH6) in the tumor.
MMR Loss (or deficient) Status: Loss of expression of at least one protein (MLH1, PMS2, MSH2 and MSH6) in the tumor.
Interpretation of Challenging Cases: The VENTANA MMR RxDx Panel scoring algorithm for each of the four MMR antibodies is binary: either the protein is expressed (a clinical status of Intact) or it is not expressed (a clinical status of Loss). Once each of the four antibodies is interpreted, the case may be assigned an overall status of proficient (all four antibodies are intact) or deficient (at least one antibody is a loss). As a result, there are no true "borderline" cases. While the vast majority of cases stained with VENTANA MMR RxDx Panel are clearly proficient or deficient in their staining results, a few cases have been observed that present a challenge in interpretation. These cases
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{7}
may be challenging due to the following issues: Non-specific background staining, focal staining, punctate staining, speckling, and tissue or staining artifacts resulting from sample processing and microtomy processes.
Interpretation for MMR proteins status is detailed in Table 2 below.
Table 2. Interpretation of MMR Protein Expression in Solid Tumor Tissue Stained with the VENTANA MMR RxDx Panel
| Clinical Status | Description |
| --- | --- |
| Intact MMR Protein Expression | Unequivocal nuclear staining in viable tumor cells, in the presence of acceptable internal positive controls (e.g. nuclear staining in lymphocytes, fibroblasts, or normal epithelium in the vicinity of the tumor) |
| Loss of MMR Protein Expression | Unequivocal loss of nuclear staining or focal weak equivocal nuclear staining in the viable tumor cells in the presence of acceptable internal positive controls. Punctate nuclear staining will be considered negative. |
VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are currently no FDA approved companion diagnostic devices to assess MMR status in solid tumors for treatment with JEMPERLI. This device is also approved for the EC indication. See Summary of Safety and Effectiveness Data for P200019.
VII. MARKETING HISTORY
VENTANA MMR RxDx Panel is currently marketed globally in several countries, including in the United States as a Class II device for the identification of Lynch syndrome (DEN170030) and as a Class III device for identifying patients with dMMR EC for treatment with JEMPERLI. The device in the US and ex-US products contain the same reagents. The device has not been withdrawn to date from the market in any country for reasons relating to safety and effectiveness of the device.
The VENTANA MMR RxDx Panel to assess MMR status in solid tumors for treatment with JEMPERLI is not marketed in any foreign country.
VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
As with any IVD test, the potential risks are associated with an incorrect test result or incorrect interpretation of results, rather than with the device itself. Failure of the device to perform as expected or failure to correctly interpret results may lead to improper patient management decisions.
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{8}
IX. SUMMARY OF NON-CLINICAL STUDIES
Non-clinical studies were performed using the VENTANA MMR RxDx Panel to support the analytical performance of the device for the solid tumor indication. These studies were performed using a variety of solid tumor tissue. Studies were conducted to characterize the assay, demonstrate the impact of pre-analytical variables on assay performance, verify precision and robustness of the assay, and establish assay stability.
Terminology used in the analysis:
An MMR status (intact or loss) was assigned to cases for each of the 4 individual MMR protein status (MLH1, PMS2, MSH2 and MSH6). Panel-level status (proficient or deficient) was assigned to cases where proficient status implies detection of all four proteins (MLH1, PMS2, MSH2 and MSH6) in the tumor, and a deficient status implies loss of expression of at least one protein (MLH1, PMS2, MSH2 or MSH6) in the tumor.
A. Laboratory Studies
1. Analytical Sensitivity Study
See Summary of Safety and Effectiveness Data for P200019.
An additional analytical sensitivity study for solid tumors will be performed in the post-market setting.
2. Analytical Specificity
a. Western Blot
See Summary of Safety and Effectiveness Data for P200019.
b. Peptide Inhibition
See Summary of Safety and Effectiveness Data for P200019.
c. Immunoreactivity in Human Tissues
See Summary of Safety and Effectiveness Data for P200019.
3. Robustness
a. Tissue Thickness
See Summary of Safety and Effectiveness Data for P200019.
b. Protocol Limitations and Failure Modes
See Summary of Safety and Effectiveness Data for P200019.
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 8
{9}
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 9
# 4. Precision
The purpose of this study was to evaluate precision of the VENTANA MMR RxDx Panel in solid tumors. Assay precision was evaluated for each of the 4 panel antibodies individually using identical study design.
The sample distribution was as follows:
MLH1: 33 solid tumors (21 intact and 18 loss): urinary (6 cases: 4 intact and 2 loss), reproductive (9 cases: 6 intact and 3 loss), gastrointestinal (6 cases: 1 intact and 5 loss), hepato-pancreatobiliary (3 cases: 2 intact and 1 loss) soft tissue (1 case: 1 intact), and thoracic (2 cases: 1 intact and 1 loss), EC (6 cases: 3 intact and 3 loss).
PMS2: 32 solid tumors (17 intact and 15 loss): urinary (3 cases: 1 intact and 2 loss), reproductive (9 cases: 4 intact and 5 loss), gastrointestinal (7 cases: 2 intact and 5 loss), endocrine (2 cases: 2 intact), hepato-pancreatobiliary (1 case: 1 intact), soft tissue (2 cases: 2 intact), and thoracic (2 cases: 2 intact), EC (6 cases: 3 intact and 3 loss).
MSH2: 33 solid tumors (18 intact and 15 loss): urinary (6 cases: 4 intact and 2 loss), reproductive (12 cases: 4 intact and 8 loss), gastrointestinal (4 cases: 2 intact and 2 loss), hepato-pancreatobiliary (1 case: 1 intact), soft tissue (2 cases: 2 intact), and thoracic (2 cases: 2 intact), EC (6 cases: 3 intact and 3 loss).
MSH6: 34 solid tumors (18 intact and 16 loss): urinary (4 cases: 1 intact and 3 loss), reproductive (11 cases: 5 intact and 6 loss), gastrointestinal (2 cases: 2 intact), endocrine (3 cases: 1 intact and 2 loss), hepato-pancreatobiliary (2 cases: 2 intact) soft tissue (4 cases: 2 intact and 2 loss), and thoracic (2 cases: 2 intact), EC (6 cases: 3 intact and 3 loss).
Total number of observations for each test condition is shown in Tables 4-8 below.
The following parameters were tested:
- Three lots of antibody (between-antibody lots)
- Three lots of OptiView DAB IHC Detection Kits (between-detection kits) and OptiView Amplification (where applicable)
- Three ULTRA instruments (between-instruments)
- Across 3 days (between-day)
- Across all intermediate precision conditions (within-run).
Each case was assigned one mode (optimal prediction) based on the samples aggregated per test condition for:
- Between-antibody lots
- Between-detection kit lots
- Between-instruments
{10}
- Between-day
Each case was compared within its duplicate samples per test run for the within-run precision study.
For each case there were 4 antibody components of the MMR RxDx Panel and a negative reagent control (NRC) antibody. For the combined intermediate precision studies, the number of non-evaluable slides and the reasons for non-evaluable observations were as follows:
MLH1: 3 slides (no staining in any cells).
PMS2: 9 slides (insufficient internal controls staining for evaluation).
MSH2: 1 slide (no internal control cells staining).
MSH6: 0
Analyses included evaluation of overall percent agreement (OPA), positive percent agreement (PPA) for Loss cases, and negative percent agreement (NPA) for Intact cases between antibody lot, between detection kit lot, between day, between instrument and within run.
Table 4. Precision, Between-antibody lots
| | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 100.0
(162/162) | (97.7, 100.0) | 100.0
(156/156) | (97.6, 100.0) | 100.0
(162/162) | (97.7, 100.0) | 100
(168/168) | (97.8, 100.0) |
| Intact | 100.0
(96/96) | (96.2, 100.0) | 100.0
(84/84) | (95.6, 100) | 100.0
(90/90) | (95.9, 100.0) | 100 (90/90) | (95.9, 100.0) |
| Loss | 100.0
(66/66) | (94.5, 100.0) | 100.0
(72/72) | (94.9, 100) | 100.0
(72/72) | (94.9, 100.0) | 100 (78/78) | (95.3, 100.0) |
Table 5. Precision, Between-detection kit lots
| | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 100.0
(160/160) | (97.7, 100.0) | 100.0
(147/147) | (97.5, 100.0) | 100.0
(162/162) | (97.7, 100.0) | 100
(168/168) | (97.8, 100.0) |
| Intact | 100.0
(95/95) | (96.1, 100.0) | 100.0
(82/82) | (95.5, 100.0) | 100.0
(90/90) | (95.9, 100.0) | 100 (90/90) | (95.9, 100.0) |
| Loss | 100.0
(65/65) | (94.4, 100.0) | 100.0
(65/65) | (94.4, 100.0) | 100.0
(72/72) | (94.9, 100.0) | 100 (78/78) | (95.3, 100.0) |
Table 6. Precision, Between-instruments
| | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 100.0
(160/160) | (97.7, 100.0) | 100.0
(156/156) | (97.6, 100.0) | 100.0
(162/162) | (97.7, 100.0) | 100
(168/168) | (97.8, 100.0) |
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{11}
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Table 7. Precision, Between-day
| | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 100.0 (160/160) | (97.7, 100.0) | 100.0 (156/156) | (97.6, 100.0) | 100.0 (162/162) | (97.7, 100.0) | 100 (168/168) | (97.8, 100.0) |
| Intact | 100.0 (96/96) | (96.2, 100.0) | 100.0 (84/84) | (95.6, 100) | 100.0 (90/90) | (95.9, 100.0) | 100 (90/90) | (95.9, 100.0) |
| Loss | 100.0 (64/64) | (94.3, 100.0) | 100.0 (72/72) | (94.9, 100) | 100.0 (72/72) | (94.9, 100.0) | 100 (78/78) | (95.3, 100.0) |
Table 8. Precision, Within-Run
| | MLH1 | | PMS2 | | MSH2 | | MSH6 | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI | %(n/N) | 95% CI |
| Overall | 98.8 (240/243) | (97.3, 100.0) | 100.0 (228/228) | (98.3, 100.0) | 100.0 (243/243) | (98.4, 100.0) | 100.0 (252/252) | (98.5, 100.0) |
| Intact | 99.3 (143/144) | (97.8, 100.0) | 100.0 (125/125) | (97.0, 100.0) | 100.0 (135/135) | (97.2, 100.0) | 100.0 (135/135) | (97.2, 100.0) |
| Loss | 98.0 (97/99) | (94.9, 100.0) | 100.0 (103/103) | (96.4, 100.0) | 100.0 (108/108) | (96.6, 100.0) | 100.0 (117/117) | (96.8, 100.0) |
Between-Day Precision for VENTANA MMR RxDx Panel
In addition to the between-day precision described above, a supplemental between-day precision to assess the test performance of VENTANA MMR RxDx Panel on solid tumor samples over 5 non-consecutive days was also performed. Agreement results are presented in Table 9 below.
The between-day precision study over 5 non-consecutive days consisted of 24 FFPE solid tumor cases distributed as follows:
MLH1: 12 intact cases [4 endometrium, 1 bladder (a challenging case), and 7 gastric (including 2 challenging cases)] and 12 loss cases (3 endometrium, 1 bladder, 5 gastric (including 1 challenging case), 1 ureter, 1 kidney (a challenging case) and 1 melanoma (a challenging case)].
Page 11
{12}
PMS2: 12 intact cases (4 gastric, 3 endometrium, 1 ovary, 3 bladder, and 1 pancreas) and 12 loss cases (5 gastric, 3 endometrium, 3 bladder, and 1 pancreas). The agreement results are presented below.
MSH2: 12 intact cases [3 endometrium, 1 bladder, 5 gastric (including 1 challenging case), 1 kidney, 1 ureter (a challenging case), and 1 melanoma] and 12 loss cases [3 endometrium, 1 bladder, 5 gastric (including 1 challenging case), 1 ureter, 1 kidney (a challenging case), and 1 melanoma (a challenging case)].
MSH6: 12 intact cases [4 endometrium (including 1 challenging case), 1 ovary, 1 bladder, 3 gastric (including 1 challenging case), 1 kidney, 1 ureter, and 1 melanoma] and 12 loss cases [4 endometrium, 1 ovary, 1 bladder, 3 gastric (including 1 challenging case), 1 kidney, 1 ureter, and 1 melanoma (a challenging case)].
For this supplemental between-day precision study for solid tumor cases, there was one non-evaluable slide:
MSH2: 1 slide (no internal control cells staining).
Table 9. Precision, Between-Day for VENTANA MMR RxDx Panel – Solid Tumor Cases
| Overall Agreement Between-Days | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Antibody | Slide-Level Marker Status | Marker Mode Status | | | Measure | %(n/N) | 95% CI |
| | | Loss | Intact | Total | | | |
| MLH1 | Loss | 120 | 0 | 120 | PPA | 100.0 (120/120) | (96.9, 100.0) |
| | Intact | 0 | 120 | 120 | NPA | 100.0 (120/120) | (96.9, 100.0) |
| | Total | 120 | 120 | 240 | OPA | 100.0 (240/240) | (98.4, 100.0) |
| PMS2 | Loss | 120 | 0 | 120 | PPA | 100.0 (120/120) | (96.9, 100.0) |
| | Intact | 0 | 120 | 120 | NPA | 100.0 (120/120) | (96.9, 100.0) |
| | Total | 120 | 120 | 240 | OPA | 100.0 (240/240) | (98.4, 100.0) |
| MSH2 | Loss | 119 | 0 | 119 | PPA | 100.0 (119/119) | (96.9, 100.0) |
| | Intact | 0 | 120 | 120 | NPA | 100.0 (120/120) | (96.9, 100.0) |
| | Total | 119 | 120 | 239 | OPA | 100.0 (239/239) | (98.4, 100.0) |
| MSH6 | Loss | 120 | 0 | 120 | PPA | 100.0 (120/120) | (96.9, 100.0) |
| | Intact | 0 | 120 | 120 | NPA | 100.0 (120/120) | (96.9, 100.0) |
| | Total | 120 | 120 | 240 | OPA | 100.0 (240/240) | (98.4, 100.0) |
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{13}
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 13
# 5. Reader Precision
Between-Reader and Within-Reader precision was assessed by evaluating concordance of VENTANA MMR RxDx Panel status between 3 readers and within individual readers using 162 (100 proficient and 62 deficient) cases from a variety of solid tumors as follows:
Urinary system-15 cases (9 proficient and 6 deficient), reproductive system-48 cases, gastrointestinal system-56 cases (28 proficient and 28 deficient, endocrine system-7 cases (7 proficient), hepato-pancreatobiliary system-13 cases (9 proficient and 4 deficient), soft tissue/skin-9 cases (7 proficient and 2 deficient), thoracic-9 cases (7 proficient and 2 deficient) and head and neck cancer-5 cases (5 proficient).
Specimens were blinded and randomized prior to evaluation of each of the 4 individual antibody status (intact or loss) and panel-level status (proficient or deficient) for each case using the VENTANA MMR RxDx Panel scoring algorithm (Staining Interpretation Section F of the Device Description Section V, and Table 2 above). Readers scored all specimens twice, with a minimum of two weeks wash-out period between reads. The agreement rates between the readers and within-reader are summarized in Table 10.
For the reader precision studies, the number of non-evaluable slides and the reasons for non-evaluable observations were as follows:
Case 1: Unevaluable for PMS2 (no internal control cells staining).
Case 2: Unevaluable for all 4 antibodies (high DAB staining on NRC). Due to unacceptable NRCs, each antibody-stained slide was not evaluated.
Case 3: same as above.
Case 4: Unevaluable for all 4 antibodies (internal control very minimally present).
Table 10. Within-Reader and Between-Reader Precision of the VENTANA MMR RxDx Panel on solid tumor tissues, by MMR Status (Proficient/ Deficient)
| Precision | Clinical Status | Agreement | | | |
| --- | --- | --- | --- | --- | --- |
| | | Type | n/N | % | 95% CI |
| Within-Reader | Deficient | APA | (364/366) | 99.5 | (98.6,100.0) |
| | Proficient | ANA | (598/600) | 99.7 | (99.2,100.0) |
| | Total | OPA | (481/483) | 99.6 | (99.0,100.0) |
| Between-Reader | Deficient | APA | (364/366) | 99.5 | (98.3,100.0) |
| | Proficient | ANA | (596/598) | 99.7 | (99.0,100.0) |
| | Total | OPA | (480/482) | 99.6 | (98.8,100.0) |
{14}
Note: Average Positive Agreement (ANA), Average Negative Agreement (ANA), Overall Percent Agreement (OPA).
2-sided 95% confidence interval (CI) was calculated using the percentile bootstrap method from 2000 bootstrap samples.
Table 11. Within-Reader and Between-Reader Precision of the VENTANA MMR RxDx Panel on solid tumor tissues, by marker (Intact/ Loss)
| Marker | Parameter | Clinical Status | Agreement | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | Type | n/N | % | 95% CI |
| MLH1 | Within-Reader | Loss | APA | 288/290 | 99.3 | (98.2,100.0) |
| | | Intact | ANA | 676/678 | 99.7 | (99.3,100.0) |
| | | Total | OPA | 482/484 | 99.6 | (99.0,100.0) |
| | Between-Reader | Loss | APA | 286/290 | 98.6 | (96.4,100.0) |
| | | Intact | ANA | 674/678 | 99.4 | (98.5,100.0) |
| | | Total | OPA | 480/484 | 99.2 | (97.9,100.0) |
| PMS2 | Within-Reader | Loss | APA | 292/294 | 99.3 | (98.3,100.0) |
| | | Intact | ANA | 670/672 | 99.7 | (99.3,100.0) |
| | | Total | OPA | 481/483 | 99.6 | (99.0,100.0) |
| | Between-Reader | Loss | APA | 292/294 | 99.3 | (97.8,100.0) |
| | | Intact | ANA | 668/670 | 99.7 | (99.1,100.0) |
| | | Total | OPA | 480/482 | 99.6 | (98.8,100.0) |
| MSH2 | Within-Reader | Loss | APA | 48/50 | 96.0 | (90.2,100.0) |
| | | Intact | ANA | 916/918 | 99.8 | (99.0,100.0) |
| | | Total | OPA | 482/484 | 99.6 | (99.0,100.0) |
| | Between-Reader | Loss | APA | 44/50 | 88.0 | (69.0,100.0) |
| | | Intact | ANA | 912/918 | 99.3 | (98.5,100.0) |
| | | Total | OPA | 478/484 | 98.8 | (97.1,100.0) |
| MSH6 | Within-Reader | Loss | APA | 72/73 | 98.6 | (95.5,100.0) |
| | | Intact | ANA | 894/895 | 99.9 | (99.7,100.0) |
| | | Total | OPA | 483/484 | 99.8 | (99.4,100.0) |
| | Between-Reader | Loss | APA | 66/72 | 91.7 | (81.4,100.0) |
| | | Intact | ANA | 890/896 | 99.3 | (98.4,100.0) |
| | | Total | OPA | 478/484 | 98.8 | (97.1,100.0) |
Note: Average Positive Agreement (ANA), Average Negative Agreement (ANA), Overall Percent Agreement (OPA).
2-sided 95% confidence interval (CI) was calculated using the percentile bootstrap method from 2000 bootstrap samples.
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{15}
Reader precision study met all the requirements and acceptance criteria for Within-Reader and Between-Reader precision studies.
# 6. External Reproducibility
The reproducibility of VENTANA MMR RxDx Panel was evaluated in 3 independent Inter-Laboratory Reproducibility (ILR) studies (diverse solid tumors, EC only and CRC only) conducted using the same study design. In each study, a set of de-identified FFPE tumor specimens was stained on a BenchMark ULTRA instrument at each of 3 external laboratories on each of 3 non-consecutive days (spanning at least 20 days in total). Each staining day at each site produced a 5-slide panel [4 biomarker antibody-stained slides and 1 slide stained with mouse NRC using the PMS2 staining protocol] that was independently evaluated for the status of each marker (Intact or Loss) and for MMR status (Deficient or Proficient) by 2 pathologists at the site.
The solid tumor ILR study used 60 diverse solid tumor cases (30 dMMR cases and 30 pMMR cases), of which 6 were considered challenging; EC only and CRC only ILR studies used 30 EC cases and 30 CRC cases respectively. The EC and CRC studies each used 15 dMMR cases and 15 pMMR cases, and each included 4 cases that were considered challenging.
The specimen distribution is shown the Table 12 below.
Table 12. Specimen Distribution, External Reproducibility Study
| Study | Total Number of Cases in the Study | Organ System | Tissue Type | MMR Panel Clinical Status | | Total number of Cases |
| --- | --- | --- | --- | --- | --- | --- |
| | | | | Number of Proficient Cases | Number of Deficient Cases | |
| Solid tumor | 60 | Gastrointestinal | Gastric/Stomach | 6 | 9 | 15 |
| | | | Esophagus | 0 | 2 | 2 |
| | | | Colon | 0 | 5 | 5 |
| | | Urinary | Bladder | 5 | 1 | 6 |
| | | | Renal Pelvis | 1 | 0 | 1 |
| | | Reproductive | Endometrium | 0 | 5 | 5 |
| | | | Cervix | 1 | 0 | 1 |
| | | | Ovary | 5 | 5 | 10 |
| | | Hepato-pancreatobiliary | Pancreas | 3 | 1 | 4 |
| | | | Liver | 0 | 1 | 1 |
| | | | Lung | 3 | 1 | 4 |
| | | Endocrine | Pancreas (Neuroendocrine) | 1 | 0 | 1 |
| | | | Thyroid | 1 | 0 | 1 |
| | | Soft Tissue/Skin | Soft tissue | 1 | 0 | 1 |
| | | | Skin | 3 | 0 | 3 |
| EC | 30 | Reproductive | EC | 15 | 15 | 30 |
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{16}
| | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| CRC | 30 | Gastrointestinal | CRC | 15 | 15 | 30 |
Note: The solid tumor study cohort represented a diverse range of histologies and organ systems, including 5 CRC cases and 5 EC cases.
One sample was common between the solid-tumor and EC ILR studies, but is being counted as separate for analysis purposes.
Each set of 5 stained slides per case per staining day (4 biomarker-stained slides and 1 mouse NRC stained with PMS2 staining protocol) was combined with a case-matched hematoxylin and eosin (H&E)-stained slide and provided as a 6-slide panel to 2 trained, blinded pathologists (readers) at the site. Each pathologist independently interpreted each case panel for the marker status (Intact or Loss) for each of the 4 VENTANA MMR RxDx Panel biomarkers (MLH1, PMS2, MSH2 and MSH6) and for the overall MMR panel status of the case (dMMR or pMMR). As noted above in section V.F. (Staining Interpretation), a case was considered pMMR if all 4 biomarkers were Intact and was considered dMMR if any one of the four biomarkers had a Loss status.
PPA and NPA rates for MMR status (dMMR or pMMR) across all evaluable observations, using the case-level reader modes for MMR status as the reference were assessed (see below for a description of non-evaluable observations and reasons for non-evaluable). Study results are presented for panel-level status (dMMR or pMMR), and for marker-level status (for 4 VENTANA MMR RxDx Panel biomarkers MLH1, PMS2, MSH2 and MSH6) for Solid Tumor, EC and CRC indications in Tables 13-20 below.
For the external reproducibility studies, the number of non-evaluable slides and the reasons for non-evaluable observations were as follows:
Case 1: Solid tumor ILR study (site C; day 1, reader 2): Unevaluable for PMS2 (Internal control not acceptable.
Case 2: CRC ILR study (site A; day 1, readers 1 and 2): Unevaluable for MSH2 (No tissue on slide). The backup slide was used, and repeat staining was evaluable.
Case 3: CRC ILR study (site C; day 2, reader 1): Unevaluable for PMS2 (Internal control not acceptable.
Table 13. External Reproducibility, VENTANA MMR RxDx Panel, Solid Tumors**
| Analysis | Overall | | Between-Site | | Between-Reader | |
| --- | --- | --- | --- | --- | --- | --- |
| | % (n/N) | 95% CI* | % (n/N) | 95% CI* | % (n/N) | 95% CI* |
| PPA | 99.6 (537/539) | (99.1, 100.0) | 99.6 (537/539) | (99.1, 100.0) | 99.6 (537/539) | (99.1, 100.0) |
| NPA | 99.3 (536/540) | (98.3, 100.0) | 99.3 (536/540) | (98.3, 100.0) | 99.3 (536/540) | (98.3, 100.0) |
| OPA | 99.4 (1073/1079) | (99.0, 99.9) | 99.4 (1073/1079) | (99.0, 99.9) | 99.4 (1073/1079) | (99.0, 99.9) |
* Two-sided 95% CIs for point estimates (PEs) of 100% were calculated using the Wilson score method; for other PEs, they were calculated using the percentile bootstrap method with 2000 replicates.
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{17}
** Does not include EC ILR study and CRC ILR study tumors
Table 14. External Reproducibility Study for VENTANA MMR RxDx Panel on Solid Tumor* tissue, by each marker
| | | Agreement with Mode [b] | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Marker | PPA | | NPA | | OPA | |
| | | % (n/N)[c] | 95% CI [c] | % (n/N) | 95% CI | % (n/N) | 95% CI |
| Agreement of Reader Biomarker Status with the Case-Level Reader Modal Status | MLH1 | 99.7 (359/363) | (99.1, 100.0) | 99.4 (716/717) | (98.7, 100.0) | 99.5 (1075/1080) | (99.1, 99.9) |
| | PMS2 | 98.6 (354/357) | (97.3, 99.7) | 99.6 (717/722) | (99.1, 100.0) | 99.3 (1071/1079) | (98.7, 99.7) |
| | MSH2 | 98.3 (177/185) | (96.7, 100.0) | 99.1 (892/895) | (98.0, 99.9) | 99.0 (1069/1080) | (98.0, 99.7) |
| | MSH6 | 97.0 (192/195) | (94.4, 99.1) | 99.7 (879/885) | (99.3, 100.0) | 99.2 (1071/1080) | (98.6, 99.6) |
| Within Site Agreement of Biomarker Status with the Case-Level Mode | MLH1 | 99.7 (359/363) | (99.1, 100.0) | 99.4 (716/717) | (98.7, 100.0) | 99.5 (1075/1080) | (99.1, 99.9) |
| | PMS2 | 98.6 (354/357) | (97.3, 99.7) | 99.6 (717/722) | (99.1, 100.0) | 99.3 (1071/1079) | (98.7, 99.7) |
| | MSH2 | 98.3 (177/185) | (96.7, 100.0) | 99.1 (892/895) | (98.0, 99.9) | 99.0 (1069/1080) | (98.0, 99.7) |
| | MSH6 | 97.0 (192/195) | (94.4, 99.1) | 99.7 (879/885) | (99.3, 100.0) | 99.2 (1071/1080) | (98.6, 99.6) |
| Within-Reader Agreement of Biomarker Status with the Case-Level Mode | MLH1 | 99.7 (359/363) | (99.1, 100.0) | 99.4 (716/717) | (98.7, 100.0) | 99.5 (1075/1080) | (99.1, 99.9) |
| | PMS2 | 98.6 (354/357) | (97.3, 99.7) | 99.6 (717/722) | (99.1, 100.0) | 99.3 (1071/1079) | (98.7, 99.7) |
| | MSH2 | 98.4 (180/185) | (96.7, 100.0) | 99.4 (892/895) | (98.9, 99.9) | 99.3 (1072/1080) | (98.7, 99.7) |
| | MSH6 | 97.0 (192/195) | (94.4, 99.1) | 99.7 (879/885) | (99.3, 100.0) | 99.2 (1071/1080) | (98.6, 99.6) |
NPA = negative percent agreement; PPA = positive percent agreement OPA = overall percent agreement
[a] Counts indicate numbers of reader observations, not numbers of unique cases.
[b] For the purpose of agreement rate calculations, a biomarker status of Loss was considered Positive and a biomarker status of Intact was considered Negative.
[c] Two-sided $95\%$ CIs were calculated using the percentile bootstrap method with 2000 replicates, except that those for point estimates of $100\%$ were calculated using the Wilson score method.
*Does not include EC ILR study and CRC ILR study tumors
Table 15. External Reproducibility, VENTANA MMR RxDx Panel, Endometrial cancer
| | Overall | | Between-Site | | Between-Reader | |
| --- | --- | --- | --- | --- | --- | --- |
| Analysis | % (n/N) | 95% CI* | % (n/N) | 95% CI* | % (n/N) | 95% CI* |
| PPA | 98.1 (263/268) | (95.5, 100.0) | 98.1 (263/268) | (95.5, 100.0) | 99.2 (263/265) | (98.1, 100.0) |
| NPA | 100.0 (269/269) | (98.6, 100.0) | 100.0 (269/269) | (98.6, 100.0) | 100.0 (272/272) | (98.6, 100.0) |
| OPA | 99.1 (532/537) | (97.8, 100.0) | 99.1 (532/537) | (97.8, 100.0) | 99.6 (535/537) | (99.1, 100.0) |
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{18}
* Two-sided 95% CIs for point estimates (PEs) of 100% were calculated using the Wilson score method; for other PEs, they were calculated using the percentile bootstrap method with 2000 replicates.
Table 16. External Reproducibility Study for VENTANA MMR RxDx Panel on EC tissue, by each marker
| | | Agreement with Mode [b] | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Marker | PPA | | NPA | | OPA | |
| | | % (n/N)[c] | 95% CI [c] | % (n/N) | 95% CI | % (n/N) | 95% CI |
| Agreement of Reader Biomarker Status with the Case-Level Reader Modal Status | MLH1 | 98.8(160/162) | (97.2,100.0) | 100.0(377/377) | (99.0,100.0) | 99.6(537/539) | (99.1,100.0) |
| | PMS2 | 98.1(159/162) | (96.3,100.0) | 100.0(377/377) | (99.0,100.0) | 99.4(536/539) | (98.9,100.0) |
| | MSH2 | 96.2(102/106) | (90.0,100.0) | 99.5(429/431) | (98.9,100.0) | 98.9(531/537) | (97.6,99.8) |
| | MSH6 | 88.9(96/108) | (78.9,98.1) | 99.3(428/431) | (98.6,100.0) | 97.2(524/539) | (95.2,99.3) |
| Within Site Agreement of Biomarker Status with the Case-Level Mode | MLH1 | 98.8(160/162) | (97.2,100.0) | 100.0(377/377) | (99.0,100.0) | 99.6(537/539) | (99.1,100.0) |
| | PMS2 | 98.1(159/162) | (96.3,100.0) | 100.0(377/377) | (99.0,100.0) | 99.4(536/539) | (98.9,100.0) |
| | MSH2 | 96.2(102/106) | (90.0,100.0) | 99.5(429/431) | (98.9,100.0) | 98.9(531/537) | (97.6,99.8) |
| | MSH6 | 94.1(96/102) | (88.2,98.6) | 99.3(434/437) | (98.6,100.0) | 98.3(530/539) | (97.0,99.3) |
| Within-Reader Agreement of Biomarker Status with the Case-Level Mode | MLH1 | 98.8(160/162) | (97.2,100.0) | 100.0(377/377) | (99.0,100.0) | 99.6(537/539) | (99.1,100.0) |
| | PMS2 | 98.1(159/162) | (96.3,100.0) | 100.0(377/377) | (99.0,100.0) | 99.4(536/539) | (98.9,100.0) |
| | MSH2 | 99.0(102/103) | (96.9,100.0) | 99.5(432/434) | (98.9,100.0) | 99.4(534/537) | (98.9,100.0) |
| | MSH6 | 97.0 (96/99) | (94.4,100.0) | 99.3(437/440) | (98.7,100.0) | 98.9(533/539) | (98.1,99.6) |
NPA = negative percent agreement; PPA = positive percent agreement OPA = overall percent agreement
[a] Counts indicate numbers of reader observations, not numbers of unique cases.
[b] For the purpose of agreement rate calculations, a biomarker status of Loss was considered Positive and a biomarker status of Intact was considered Negative.
[c] Two-sided $95\%$ CIs were calculated using the percentile bootstrap method with 2000 replicates, except that those for point estimates of $100\%$ were calculated using the Wilson score method.
Table 17. External Reproducibility, VENTANA MMR RxDx Panel, Colorectal Cancer
| Analysis | Overall | | Site-Stratified | | Reader-Stratified | |
| --- | --- | --- | --- | --- | --- | --- |
| | % (n/N) | 95% CI* | % (n/N) | 95% CI* | % (n/N) | 95% CI* |
| PPA | 99.3 (267/269) | (98.1, 100.0) | 99.3 (267/269) | (98.1, 100.0) | 99.3 (267/269) | (98.1, 100.0) |
| NPA | 100.0 (270/270) | (98.6, 100.0) | 100.0 (270/270) | (98.6, 100.0) | 100.0 (270/270) | (98.6, 100.0) |
| OPA | 99.6 (537/539) | (99.1, 100.0) | 99.6 (537/539) | (99.1, 100.0) | 99.6 (537/539) | (99.1, 100.0) |
* Two-sided 95% CIs for point estimates (PEs) of 100% were calculated using the Wilson score method; for other PEs, they were calculated using the percentile bootstrap method with 2000 replicates.
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{19}
Table 18. External Reproducibility Study for VENTANA MMR RxDx Panel on CRC tissue as measured for each marker
| | | Agreement with Mode [b] | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Marker | PPA | | NPA | | OPA | |
| | | % (n/N)[c] | 95% CI [c] | % (n/N) | 95% CI | % (n/N) | 95% CI |
| Agreement of Reader Biomarker Status with the Case-Level Reader Modal Status | MLH1 | 98.6 (142/144) | (96.7, 100.0) | 100.0 (396/396) | (99.0, 100.0) | 99.6 (538/540) | (99.1, 100.0) |
| | PMS2 | 98.6 (142/144) | (96.7, 100.0) | 100.0 (395/395) | (99.0, 100.0) | 99.6 (537/539) | (99.1, 100.0) |
| | MSH2 | 92.1 (116/126) | (78.9, 100.0) | 99.3 (411/414) | (98.2, 100.0) | 97.6 (527/540) | (94.2, 99.8) |
| | MSH6 | 91.7 (132/144) | (79.9, 99.1) | 100.0 (396/396) | (99.0, 100.0) | 97.8 (528/540) | (94.3, 99.8) |
| Within-Site or Within-reader Agreement of Biomarker Status with the Case-Level Mode | MLH1 | 98.6 (142/144) | (96.7, 100.0) | 100.0 (396/396) | (99.0, 100.0) | 99.6 (538/540) | (99.1, 100.0) |
| | PMS2 | 98.6 (142/144) | (96.7, 100.0) | 100.0 (395/395) | (99.0, 100.0) | 99.6 (537/539) | (99.1, 100.0) |
| | MSH2 | 96.7 (116/120) | (91.3, 100.0) | 99.3 (417/420) | (98.3, 100.0) | 98.7 (533/540) | (97.4, 99.8) |
| | MSH6 | 95.7 (132/138) | (90.7, 99.1) | 100.0 (402/402) | (99.1, 100.0) | 98.9 (534/540) | (97.6, 99.8) |
NPA = negative percent agreement; PPA = positive percent agreement OPA = overall percent agreement
[a] Counts indicate numbers of reader observations, not numbers of unique cases.
[b] For the purpose of agreement rate calculations, a biomarker status of Loss was considered Positive and a biomarker status of Intact was considered Negative.
[c] Two-sided $95\%$ CIs were calculated using the percentile bootstrap method with 2000 replicates, except that those for point estimates of $100\%$ were calculated using the Wilson score method.
The MMR results for all cases in the 3 ILR studies (solid tumors, EC and CRC) were combined (120 cases in total) and analyzed for the same cases to assess the performance of VENTANA MMR RxDx Panel as a diagnostic device for determining MMR status in diverse solid tumor specimens. The combined analysis of MMR status across all readers, sites, and days included a total of 2155 observations. The summary of MMR status agreement rates across all evaluable observations vs the reader modes, using the reader modes as the reference, is shown in Table 19. The agreement rates for the pooled within-site and within-reader analyses vs the reader modes, using the within-site and within-reader modes as the references, are also shown in Table 20.
Table 19. Inter-Laboratory Reproducibility for overall agreement rates for VENTANA MMR RxDx Panel in variety of solid tumor tissues including EC
| Inter-Laboratory Reproducibility | Agreement | | | |
| --- | --- | --- | --- | --- |
| | Type | n/N | % | 95% CI |
| Overall | PPA | 1066/1076 | 99.1 | (98.3, 99.6) |
| | NPA | 1075/1079 | 99.6 | (99.2, 100.0) |
| | OPA | 2141/2155 | 99.4 | (98.9, 99.7) |
| Within-Site | PPA | 1066/1076 | 99.1 | (98.3, 99.6) |
| | NPA | 1075/1079 | 99.6 | (99.2, 100.0) |
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{20}
| Inter-Laboratory Reproducibility | Agreement | | | |
| --- | --- | --- | --- | --- |
| | Type | n/N | % | 95% CI |
| | OPA | 2141/2155 | 99.4 | (98.9, 99.7) |
| Within-Reader | PPA | 1066/1073 | 99.3 | (98.9, 99.7) |
| | NPA | 1078/1082 | 99.6 | (99.2, 100.0) |
| | OPA | 2144/2155 | 99.5 | (99.2, 99.8) |
Note: Positive Percent Agreement (PPA), Negative Percent Agreement (NPA), Overall Percent Agreement (OPA).
Note: Two-sided 95% CIs were calculated using the percentile bootstrap method with 2000 replicates.
Note: For the purposes of this study, a sample with a MMR Panel status of deficient was considered positive and a sample with a MMR Panel status of proficient was considered negative.
In addition, pairwise comparisons were made Between-site, Between-Reader and Between-Day for panel level MMR status. A summary of the results can be found in Table 20.
Table 20. Inter-Laboratory Reproducibility Pairwise Agreement Rates for the VENTANA MMR RxDx Panel in a variety of solid tumor tissues including EC
| Inter-Laboratory Reproducibility | Agreement | | | |
| --- | --- | --- | --- | --- |
| | Type | n/N | % | 95% CI |
| Between-Site | APA | 12628/12794 | 98.7 | (97.8, 99.4) |
| | ANA | 12840/13006 | 98.7 | (97.8, 99.4) |
| | OPA | 12734/12900 | 98.7 | (97.8, 99.4) |
| Between-Reader | APA | 1056/1068 | 98.9 | (98.1, 99.5) |
| | ANA | 1072/1084 | 98.9 | (98.1, 99.5) |
| | OPA | 1064/1076 | 98.9 | (98.1, 99.5) |
| Between-Day | APA | 2110/2132 | 99.0 | (98.3, 99.5) |
| | ANA | 2146/2168 | 99.0 | (98.3, 99.5) |
| | OPA | 2128/2150 | 99.0 | (98.3, 99.5) |
Note: Average Positive Agreement (APA), Average Negative Agreement (ANA), Overall Percent Agreement (OPA)
Note: Two-sided 95% confidence interval (CI) was calculated using the percentile bootstrap method from 2000 bootstrap samples.
Note: For the purposes of this study, a sample with an MMR Panel status of deficient was considered positive and a sample with a MMR Panel status proficient was considered negative.
Note: The same sites and readers were used for EC and CRC ILR studies. For the solid tumor ILR study, two sites were different from the EC and CRC ILR.
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{21}
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 21
7. **Stability Studies**
a. Cut Slide Stability
See Summary of Safety and Effectiveness Data for P200019.
b. Real Time Stability
See Summary of Safety and Effectiveness Data for P200019.
An additional analytical stability study for solid tumors will be performed in the post-market setting.
B. **Animal Studies**
None
C. **Additional Studies**
1. **Fixative Type and Time**
See Summary of Safety and Effectiveness Data for P200019.
2. **Tissue Heterogeneity**
This study evaluated the prevalence of case heterogeneity in various solid tumor tissue blocks from the same case when stained with VENTANA MMR RxDx Panel Assay on the BenchMark ULTRA instrument.
For the sample set of 300 tissues (150 pairs) evaluated, 10 of 44 EC tissues, 6 of 54 CRC tissues, and 2 of 202 of the solid-tumor tissues demonstrated dMMR status. All 150 pairs across all indications exhibited equivalent marker and panel level MMR status for an OPA of 100%. Therefore, case heterogeneity was not observed in any of patient pairs for overall panel level nor marker-level MMR status. Due to the difficulty of procuring matched patient samples from commercial vendors and the overall low prevalence of MMR biomarkers, the sample distribution for MMR status in this study was unequal, with only 18 of 300 cases representing a dMMR status. Although the trend in this study indicates, case heterogeneity is unlikely to be observed in EC, CRC, and solid-tumor tissues, the small sample size of the available dMMR tissues provide a limited assessment of case heterogeneity for MMR panel status.
3. **Primary versus Metastatic**
This study evaluated the concordance of MMR panel status between FFPE matched primary and metastatic solid tumors when stained with the MMR RxDx Panel assay on the BenchMark ULTRA instrument.
{22}
In this characterization study, matched primary vs. metastatic solid tumor tissue specimens were stained and compared to show either concordance or discordance in overall MMR panel status between each pair. For the sample set of 156 tissue cases (78 pairs) evaluated, 152 of these cases (76 pairs) were concordant. Out of these 152 cases, 4 cases (2 pairs) exhibited an MMR status of deficient (dMMR), 146 (73 pairs) cases exhibited an MMR status of proficient (pMMR), and 2 cases (1 pair) exhibited an MMR status of non-evaluable. Four cases (2 pairs) exhibited discordance between the matched primary and metastatic tumor tissue pairs.
## X. SUMMARY OF CLINICAL STUDIES
The clinical performance of VENTANA MMR RxDx Panel as a CDx device for the solid tumor indication for JEMPERLI was based on the GARNET clinical trial (NCT02715284) Subpart 2B. The primary efficacy population for the drug was based on Cohorts A1 and F.
## A. Study Design
GARNET is a multicenter, open-label, study with Subpart 2B, designed to evaluate the antitumor activity of JEMPERLI in patients with recurrent or advanced mismatch repair deficient (dMMR)/microsatellite instability-high (MSI-H) cancers, including dMMR/MSI-H EC (Cohort A1) and recurrent or advanced dMMR/MSI-H solid tumors (Cohort F). The clinical bridging study for the VENTANA RxDx Panel included retesting the biomarker positive (dMMR) patients (Cohorts A1 and F) and a subset of biomarker negative EC patients from cohort A2 (pMMR) (intact or biomarker-negative) EC, and CRC negative (pMMR) samples that were procured from outside the trial, to evaluate the positive and negative percent agreement between the enrolling tests and the final CDx. Refer to the clinical bridging study in the Effectiveness Results in Section D below for further details.
## 1. GARNET Clinical Inclusion and Exclusion Criteria
### Key Trial Inclusion Criteria
1. Patient has proven recurrent or advanced solid tumor and has disease progression after treatment with available anti-cancer therapies or is intolerant to treatment that meets the requirements for the part of the study, they will participate in per clinical trial protocol.
2. Patients who have progressed on or after platinum doublet therapy
3. Patients have received no more than 2 lines of anti-cancer therapy for recurrent or advanced (≥Stage IIIB) disease. Prior treatment with hormone therapies is acceptable and does not count towards the number of anti-cancer therapies noted in the criterion above for this cohort.
4. Patients must submit 2 scans demonstrating increase in tumor measurement that meet criteria for Progressive Disease (PD) on or after the latest systemic anticancer therapy based on RECIST 1.1 to Central Radiology prior to the first dose of JEMPERLI.
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{23}
5. Presence of at least 1 measurable lesion on baseline scan will be confirmed by central radiology review.
6. Status of tumor MMR status as determined by immunohistochemistry (IHC) testing.
7. Tumor MMR/MSI or POLE status: Patients can be screened based on local MMR/MSI testing results using IHC, polymerase chain reaction (PCR), or next generation sequencing (NGS) performed in a certified local laboratory, but patient eligibility is determined by MMR IHC results. For patients with available local MMR IHC results for the respective cohort(s), tumor samples have to be submitted to a central IHC laboratory and its quality has to be checked and cleared prior to Day 1 of chemotherapy treatment Cycle 1(C1D1). For patients without available local MMR IHC test results (patients with local PCR or NGS test results), tumor samples have to be submitted directly to central IHC laboratory and central IHC results have to confirm eligibility prior to proceeding with other screening procedures. After the central IHC test is completed, remaining tumor tissue may be sent to a central NGS laboratory for further testing.
8. Patients who are considered for the study based on POLE mutation must have the local results available showing tumor mutation(s) in the exonuclease domain of the POLE gene (amino acid residues 268-471) to begin screening. Patients must have the quality of submitted tumor samples checked and cleared by a central IHC laboratory prior to receiving study treatment.
## Key Trial Exclusion Criteria
1. Patient has received prior therapy with an anti-PD-1, anti-PD-L1, or anti-PD-L2 agent.
2. Patient has known uncontrolled central nervous system metastases and/or carcinomatous meningitis.
3. Patient has a known additional malignancy that progressed or required active treatment within last 2 years.
## 2. Follow-up Schedule
Up to 2 years, or until the subject meets protocol specific discontinuation criteria.
## 3. Clinical Endpoints
### Primary endpoints:
1. The primary efficacy endpoint is ORR, defined as the proportion of patients achieving complete response (CR) or partial response (PR) per RECIST v1.1 for dMMR/MSI-H (Cohorts A1 and F).
2. Duration of response (DOR) based on independent blinded central review using RECIST v1.1 for Cohorts A1 and F will be evaluated as a secondary endpoint.
### Secondary endpoints:
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{24}
1. Disease control rate will be assessed for Cohorts A1, A2, and F as a secondary endpoint and is defined as the proportion of patients achieving CR, PR, or stable disease (SD) as assessed per RECIST v1.1
2. Immune-related duration of response will be evaluated as a secondary endpoint for all cohorts in Part 2B and is defined as the time from first documentation of CR or PR by irRECIST until the time of first documentation of PD (subsequently confirmed) per irRECIST
3. Progression-free survival will be assessed as a secondary endpoint and is defined as the time from date of first dose to the earlier date of assessment of progression or death by any cause in the absence of progression based on: (1) the time of first documentation of PD per RECIST v1.1 (for cohorts A1, A2, and F only); and (2) the time of first documentation of PD (subsequently confirmed) per irRECIST for all cohorts in Part 2B
4. OS will be assessed as a secondary endpoint and is defined as the time from date of first dose of study treatment to the date of death by any cause.
## B. Accountability of PMA Cohorts
### Accountability of EC Cohorts A1 and A2
Study participants with recurrent or advanced endometrial carcinoma in the GARNET study comprise one of the primary analysis population for this application. Out of the 126 patients in EC Cohort A1, 103 were in the efficacy population and 23 were in the non-efficacy safety population. Out of the 103, 57 were available for retesting using the CDx test (EC population). Additionally, 16 were available for retesting that were not derived from the efficacy population but instead from the Safety Population. This resulted in an overall available sample for retesting of 73 subjects (Concordance Population).
Table 21. Accountability of EC Cohorts A1 and A2 (EC)
| Analysis Population | Clinical Trial Assay [a] MMR Status | | |
| --- | --- | --- | --- |
| | dMMR (n) | pMMR (n) | Unknown (n) |
| Total Enrolled in Study [b] | 126 | 145 | 4 |
| Efficacy Population | 103 | 142 | 3 |
| Available Sample for Retesting from Efficacy Population Population | 57 | 74 | 0 |
| Concordance Population from Safety Population | 16 | N/A | N/A |
| Concordance Population | 73 | 75 | 0 |
[a] Any IHC MMR test performed at a local or central laboratory for the purpose of GARNET enrollment screening.
[b] All patients enrolled in GARNET Cohort A1 and A2 (EC) who received study treatment prior to the 01-March-2020 clinical data cutoff date (CCOD) for solid-tumor and had an evaluable clinical trial assay (CTA) testing result or were tested with VENTANA MMR RxDx Panel (the CDx assay).
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{25}
# Accountability of Solid Tumor Cohort F
Study participants with recurrent or advanced solid tumors in cohort F of the GARNET study comprise the primary analysis population for this application. Out of the 141 patients in non-EC Cohort F dMMR, 106 were included in the Efficacy population after exclusion due to 35 being in the Non-efficacy safety population. Out of the 106, 53 were available for retesting using the CDx test (EC population). A further 30 were available for retesting that were not derived from the Efficacy population but instead the Safety population. This resulted in an overall available sample for retesting of 83 subjects (Concordance Population).
Table 22. Accountability of Solid Tumor Cohort F (non-EC)
| Analysis Population | Clinical Trial Assay [a] MMR Status | | |
| --- | --- | --- | --- |
| | dMMR (n) | pMMR (n) [b] | Unknown (n) |
| Total Enrolled in Study [c] | 141 | 4 | 6 |
| Efficacy Population | 106 | 2 | 4 |
| Available Sample for Retesting from Efficacy Population Population[c] | 53 | 0 | 0 |
| Concordance Population from Safety Population | 30 | N/A | N/A |
| Concordance Population | 83 | 1 | 0 |
[a] Any IHC MMR test performed at a local or central laboratory for the purpose of GARNET enrollment screening.
[b] Includes 1 patient who received a pMMR result from the CTA but also an MSI-H result and 3 patients who received a pMMR result from the CTA but also a POLE-mut result.
[c] All patients enrolled in GARNET Cohort F (non-EC) who received study treatment prior to the 01-March-2020 clinical data cutoff date (CCOD) for the solid tumor BLA and had an evaluable clinical trial assay (CTA) testing result or were tested with VENTANA MMR RxDx Panel (the CDx assay).
# C. Study Population Demographics and Baseline Parameters
The table below summarizes patient characteristics by presence/absence of evaluable CDx results for dMMR solid-tumor patients tested with VENTANA MMR RxDx Panel (Efficacy population).
Table 23. Patient Characteristics by Presence/Absence of Evaluable CDx Results – Efficacy Population[a] – dMMR Solid Tumor[b]
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{26}
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 26
| Characteristic | Evaluable VENTANA MMR RxDx Panel Result Obtained? | | |
| --- | --- | --- | --- |
| | Yes (N=110) | No (N=99) | Overall (N=209) |
| Age (yr) | | | |
| n | 110 | 99 | 209 |
| Mean (SD) | 61.8 (11.03) | 62.2 (10.68) | 62.0 (10.84) |
| Median | 62.0 | 65.0 | 63.0 |
| Min, Max | 24, 85 | 25, 80 | 24, 85 |
| Missing | 0 | 0 | 0 |
| Age Group (yr) | | | |
| n | 110 | 99 | 209 |
| <65 years | 63 (57.3%) | 47 (47.5%) | 110 (52.6%) |
| >=65 years - <75 years | 36 (32.7%) | 43 (43.4%) | 79 (37.8%) |
| >=75 years | 11 (10.0%) | 9 (9.1%) | 20 (9.6%) |
| Sex | | | |
| n | 110 | 99 | 209 |
| Female | 86 (78.2%) | 75 (75.8%) | 161 (77.0%) |
| Male | 24 (21.8%) | 24 (24.2%) | 48 (23.0%) |
| Ethnicity | | | |
| n | 110 | 99 | 209 |
| Hispanic or Latino | 5 (4.5%) | 2 (2.0%) | 7 (3.3%) |
| Not Hispanic or Latino | 72 (65.5%) | 63 (63.6%) | 135 (64.6%) |
| Not Reported | 31 (28.2%) | 31 (31.3%) | 62 (29.7%) |
| Unknown | 2 (1.8%) | 3 (3.0%) | 5 (2.4%) |
| Race | | | |
| n | 110 | 99 | 209 |
| American Indian or Alaska Native | 2 (1.8%) | 1 (1.0%) | 3 (1.4%) |
| Asian | 4 (3.6%) | 2 (2.0%) | 6 (2.9%) |
| Black | 1 (0.9%) | 3 (3.0%) | 4 (1.9%) |
| White | 71 (64.5%) | 61 (61.6%) | 132 (63.2%) |
| Other | 0 | 1 (1.0%) | 1 (0.5%) |
| Not Reported | 30 (27.3%) | 31 (31.3%) | 61 (29.2%) |
| Unknown | 2 (1.8%) | 0 | 2 (1.0%) |
| Organ System | | | |
| n | 110 | 99 | 209 |
| Endocrine | 1 (0.9%) | 0 | 1 (0.5%) |
| Gastrointestinal | 47 (42.7%) | 43 (43.4%) | 90 (43.1%) |
| Hepatopancreatobiliary | 2 (1.8%) | 6 (6.1%) | 8 (3.8%) |
| Reproductive | 59 (53.6%) | 48 (48.5%) | 107 (51.2%) |
| Thoracic | 0 | 1 (1.0%) | 1 (0.5%) |
| Urinary | 1 (0.9%) | 0 | 1 (0.5%) |
| Unknown | 0 | 1 (1.0%) | 1 (0.5%) |
| Tumor Site | | | |
| n | 110 | 99 | 209 |
| Adrenal Cortical | 1 (0.9%) | 0 | 1 (0.5%) |
| Biliary Neoplasm | 1 (0.9%) | 0 | 1 (0.5%) |
{27}
| Characteristic | Evaluable VENTANA MMR RxDx Panel Result Obtained? | | |
| --- | --- | --- | --- |
| | Yes (N=110) | No (N=99) | Overall (N=209) |
| Breast Cancer | 0 | 1 (1.0%) | 1 (0.5%) |
| Carcinoma of Unknown Primary Origin | 0 | 1 (1.0%) | 1 (0.5%) |
| Colorectal Cancer | 32 (29.1%) | 37 (37.4%) | 69 (33.0%) |
| Endometrial Cancer | 57 (51.8%) | 46 (46.5%) | 103 (49.3%) |
| Esophageal Cancer | 0 | 1 (1.0%) | 1 (0.5%) |
| Gallbladder | 1 (0.9%) | 0 | 1 (0.5%) |
| Gastric Cancer | 5 (4.5%) | 3 (3.0%) | 8 (3.8%) |
| Genital Neoplasm Malignant Female | 0 | 1 (1.0%) | 1 (0.5%) |
| Liver Cancer | 0 | 2 (2.0%) | 2 (1.0%) |
| Ovarian Cancer | 2 (1.8%) | 0 | 2 (1.0%) |
| Pancreatic Carcinoma | 0 | 4 (4.0%) | 4 (1.9%) |
| Pleural | 0 | 1 (1.0%) | 1 (0.5%) |
| Renal Cell Carcinoma | 1 (0.9%) | 0 | 1 (0.5%) |
| Small Intestinal Cancer | 10 (9.1%) | 2 (2.0%) | 12 (5.7%) |
| Stage at Initial Diagnosis | | | |
| n | 110 | 99 | 209 |
| Stage I | 23 (20.9%) | 21 (21.2%) | 44 (21.1%) |
| Stage II | 9 (8.2%) | 8 (8.1%) | 17 (8.1%) |
| Stage III | 41 (37.3%) | 30 (30.3%) | 71 (34.0%) |
| Stage IV | 36 (32.7%) | 39 (39.4%) | 75 (35.9%) |
| Unknown | 1 (0.9%) | 1 (1.0%) | 2 (1.0%) |
| Cancer Stage (Most Recent) [d] | | | |
| n | 53 | 53 | 106 |
| Locoregional (Unresectable) | 1 (1.9%) | 4 (7.5%) | 5 (4.7%) |
| Metastatic | 52 (98.1%) | 49 (92.5%) | 101 (95.3%) |
| Missing | 57 | 46 | 103 |
| Baseline ECOG Performance | | | |
| n | 110 | 99 | 209 |
| 0 | 36 (32.7%) | 46 (46.5%) | 82 (39.2%) |
| 1 | 74 (67.3%) | 53 (53.5%) | 127 (60.8%) |
| Histology [c] | | | |
| n | 57 | 46 | 103 |
| Adenocarcinoma | 48 (84.2%) | 40 (87.0%) | 88 (85.4%) |
| Clear Cell Carcinoma | 1 (1.8%) | 0 | 1 (1.0%) |
| Mixed Carcinoma | 3 (5.3%) | 1 (2.2%) | 4 (3.9%) |
| Serous Carcinoma | 3 (5.3%) | 1 (2.2%) | 4 (3.9%) |
| Squamous Carcinoma | 0 | 1 (2.2%) | 1 (1.0%) |
| Undifferentiated Carcinoma | 1 (1.8%) | 3 (6.5%) | 4 (3.9%) |
| Unknown | 1 (1.8%) | 0 | 1 (1.0%) |
| Missing | 53 | 53 | 106 |
[a]VENTANA MMR RxDx Panel (CDx) results associated with exclusionary diagnostic (Dx) protocol deviations (PDs) [ie, IU CDx results] were treated as missing.
[b] All patients in the Efficacy Population who had a dMMR CTA MMR result.
[c] Variable(s) for EC patients only.
[d] Variable(s) for non-EC patients only.
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
{28}
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 28
## D. Safety and Effectiveness Results
### 1. Safety Results
Adverse events (AEs) in the GARNET trial were specific to the toxicity of the investigational agent. There were no device-specific adverse events in the diagnostic study. In this trial, observed AEs included events that were in line with those expected in subjects with recurrent or advanced solid tumors, as well as those consistent with reported safety profiles of monoclonal antibodies blocking the PD 1 interactions.
### 2. Effectiveness Results
The analysis of effectiveness was based on the efficacy population (209 in cohorts A1 and F (dMMR)).
Efficacy in dMMR solid tumors was performed in 209 patients with dMMR tumors (103 dMMR EC and 106 dMMR non-EC). The tumor response summary for the dMMR solid-tumor population overall is summarized in Table 24 below. The overall response rate (ORR) from the drug trail was 41.6%.
Table 24. Efficacy Results in GARNET dMMR Recurrent or Advanced Solid Tumors
| Endpoint | JEMPERLI
N = 209 |
| --- | --- |
| Confirmed overall response rate | |
| Overall response rate | 41.6% |
| (95% CI) | (34.9, 48.6) |
| Complete response rate | 9.1% |
| Partial response rate | 32.5% |
| Duration of response | |
| Median in months | 34.7 |
| (range)^{a} | 2.6, 35.8+ |
| Patients with duration ≥6 months | 95.4% |
### Clinical Bridging Study
Bridging study was performed to evaluate agreements of MMR status (PPA for dMMR status, and NPA for pMMR status) between VENTANA MMR RxDx Panel and the CTA (local or central testing using an IHC assay other than VENTANA MMR RxDx Panel) and to assess the estimated clinical efficacy (ORR, DOR) of the patients enrolled in the GARNET Study, had the CDx been the enrolling device.
{29}
The agreement of MMR status between CTA and CDx results was calculated in the Concordance (EC = 73 and non-EC = 83) population using the CTA results as reference. For the purpose of the analyses, a pMMR status was considered negative, and dMMR status was considered positive.
## Concordance between the CTA and CDx [VENTANA MMR RxDx Panel (Solid Tumors)]
PPA rates for the dMMR EC (n = 73), non-EC (n = 83), and solid tumor (n = 156) groups were 93.2% (95% CI: 84.9, 97.0), 83.1% (95% CI: 73.7, 89.7), and 87.8% (95% CI: 81.8, 92.1), respectively. NPA rates for the pMMR EC group (n = 75) was 98.7% (95% CI: 92.8, 99.8), and the NPA in the non-EC group (n = 1) was 100% (95% CI: 20.7, 100.0) and in the solid tumor group (n = 76) it was 98.7% (95% CI: 92.9, 99.8), respectively.
Table 25.: Solid Tumor MMR Status Concordance between the GARNET Clinical Trial Assay and VENTANA MMR RxDx Panel
| Group | CDx MMR Status | CTA Status | | | Agreement | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | dMMR | pMMR | Total | Measure | % (n/N) | 95% CI [a] |
| EC | dMMR | 68 | 1 | 69 | PPA | 93.2 (68/73) | 84.9, 97.0 |
| | pMMR | 5 | 74 | 79 | NPA | 98.7 (74/75) | 92.8, 99.8 |
| | Total | 73 | 75 | 148 | OPA | 95.9 (142/148) | 91.4, 98.1 |
| Non-EC | dMMR | 69 | 0 | 69 | PPA | 83.1 (69/83) | 73.7, 89.7 |
| | pMMR | 14 | 1 | 15 | NPA | 100 (1/1) | 20.7, 100 |
| | Total | 83 | 1 | 84 | OPA | 83.3 (70/84) | 73.9, 89.8 |
| dMMR Solid tumors | dMMR | 137 | 1 | 138 | PPA | 87.8 (137/156) | 81.8, 92.1 |
| | pMMR | 19 | 75 | 94 | NPA | 98.7(75/76) | 92.9, 99.8 |
| | Total | 156 | 76 | 232 | OPA | 91.4(212/232) | 87.1, 94.4 |
[a] Two-sided 95% CIs were calculated using the Wilson score method.
A sensitivity analysis was conducted to evaluate the robustness of the ORR estimates in dMMR patients in the GARNET clinical trial. Multiple imputation using logistic regression was conducted separately for EC, non-EC and solid tumor. In an examination of covariates between samples that were evaluable and non-evaluable with the VENTANA MMR RxDx Panel, distribution of the covariates supports missing at random assumption. In the bridging study, since a large proportion of dMMR in solid tumor includes EC and CRC, the NPA for solid tumors was calculated based on lower bound 95% CI of NPA from EC and CRC cohorts. The imputation-based ORR for solid tumor population was 31.8% with 95% CI (26.1%, 37.6%) with CTA+ prevalence of 9% under the worst-case scenario (no drug efficacy in CDx+/CTA- population) and 48.7% with 95% CI (41.6%, 55.7%) under the best-case scenario (full drug efficacy in CDx+/CTA-
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 29
{30}
population). The range of these results include the observed ORR for the GARNET clinical trial (41.6% [34.9%, 48.6%]).
3. **Subgroup Analyses**
There was no subgroup analysis performed.
4. **Pediatric Extrapolation**
In this premarket application, existing clinical data was not leveraged to support approval of a pediatric patient population.
E. **Financial Disclosure**
The Financial Disclosure by Clinical Investigators regulation (21 CFR 54) requires applicants who submit a marketing application to include certain information concerning the compensation to, and financial interests and arrangement of, any clinical investigator conducting clinical studies covered by the regulation. The pivotal clinical study included 1 principal investigator. None of the clinical investigators had disclosable financial interests/arrangements as defined in sections 54.2(a), (b), (c), and (f). The information provided does not raise any questions about the reliability of the data.
XI. **SUMMARY OF SUPPLEMENTAL CLINICAL INFORMATION**
Not applicable
XII. **PANEL MEETING RECOMMENDATION AND FDA'S POST-PANEL ACTION**
In accordance with the provisions of section 515(c)(3) of the act as amended by the Safe Medical Devices Act of 1990, this PMA was not referred to the Hematology and Pathology Devices Panel, an FDA advisory committee, for review and recommendation because the information in the PMA did not raise any new safety and effectiveness questions compared with information previously reviewed by this panel.
XIII. **CONCLUSIONS DRAWN FROM THE STUDIES**
A. **Effectiveness Conclusions**
Effectiveness of the VENTANA MMR RxDx Panel in determining MMR status (proficient or deficient) is based on the clinical performance and benefit to patients with recurrent or advanced solid tumors as assessed in the GARNET study which evaluated the safety and the efficacy of JEMPERLI in these patients who have no satisfactory alternative treatment options. The clinical benefit of the VENTANA MMR RxDx Panel was demonstrated for patients enrolled in GARNET clinical study, in which the MMR status was determined using CTA. Overall, the observed clinical benefit in the subset of patients tested with the VENTANA MMR RxDx Panel Assay was comparable to that observed in the BLA efficacy population. Additional sensitivity analyses combined with
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 30
{31}
multiple imputation approaches for missing CDx values consistently support the clinical benefit of the VENTANA MMR RxDx Panel.
The performance of the VENTANA MMR RxDx Assay was also supported by the analytical validation studies.
## B. Safety Conclusions
The risks of the device are based on nonclinical laboratory studies as well as data collected in a clinical study conducted to support PMA approval as described above.
The VENTANA MMR RxDx Panel is an in vitro diagnostic device, which involves tumor specimens collected from patients with recurrent or advanced solid tumors. The risks of the device are based on data collected in the clinical study conducted to support PMA approval as described above. Risks of the VENTANA MMR RxDx Panel are associated with failure of the device to perform as expected or failure to correctly interpret test results. As VENTANA MMR RxDx Panel is intended for use to identify patients for JEMPERLI therapy, if incorrect, or false, results are reported, then patients may not receive the proper treatment. Patients with false positive results may undergo treatment with JEMPERLI without much clinical benefit and may experience adverse reactions associated with JEMPERLI therapy. Patients with false negative results may not be considered for treatment with JEMPERLI, and therefore, may receive other treatment options. There is also a risk of delayed results, which may lead to a delay in treatment with JEMPERLI.
## C. Benefit-Risk Determination
The probable benefits of this device are based on the data collected in the GARNET clinical study. When the VENTANA MMR RxDx Panel is used in the designated intended use population, recurrent or advanced dMMR solid tumors, according to the approved instructions for use, and whose tumors demonstrate deficient MMR with this device, the benefits of the use of JEMPERLI are expected to be an overall response rate of approximately 31.8% (estimate based on worst-case scenario) to 48.7% (estimate based on best case scenario) with acceptable duration of response and with evidence of probable benefit.
The potential risk associated with the use of the device, mainly due to 1) false positive, false negative or failure to provide a result. The risk of use of the VENTANA MMR RxDx Panel in the designated intended use population, for the proposed indication for use include the risks of a false positive device result and subsequent use of JEMPERLI include a failure of the patient's tumor to respond to this therapy, and the experiencing of toxicity/adverse events. With respect to a false negative device result, such a patient could be deprived of a potentially beneficial solid tumor treatment and associated overall response rate and durability.
In this trial, observed AEs included events that were in line with those expected in subjects with recurrent or advanced solid tumors and endometrial cancer, as well as those
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 31
{32}
consistent with reported safety profiles of monoclonal antibodies blocking the PD 1 interactions.
In addition to a false positive device result, a failure to respond could also be due to the particular biology of the patient's tumor, or other idiosyncratic biological issues specific to the individual patient, even if the device result is a true positive, as not all true positives will respond to this therapy. With respect to a false negative device result, such a patient could be deprived of a potentially beneficial solid tumor or endometrial cancer treatment and associated overall response rate and durability. The risks are mitigated by the expected poor prognosis of recurrent or advanced solid tumors. The risks of false results are partially mitigated by the analytical validation results summarized above.
The overall clinical and analytical data support that for the VENTANA MMR RxDx Panel, and the indications noted in the intended use statement, the probable benefits outweigh the probable risks
Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
D. Overall Conclusions
The data in this application support the reasonable assurance of safety and effectiveness of this device when used in accordance with the indications for use. Study 4010-01-001 (GARNET) is a multi-center, open-label ongoing clinical study of JEMPERLI efficacy and safety in adult patients who have recurrent or advanced solid tumors with limited available treatment options. Data from this study support the use of the MMR RxDx Panel as an aid in selecting patients with MMR deficient solid tumors who are likely to benefit from JEMPERLI therapy. The response rate in patients with solid tumors that have deficient MMR expression is better than what would be expected of available therapy and represents an improvement that is reasonably likely to predict clinical benefit.
XIV. CDRH DECISION
CDRH issued an approval order on August 17, 2021.
The final conditions of approval cited in the approval order are described below.
1. Ventana Medical Systems, Inc. must provide additional data from an appropriately designed analytical sensitivity study for the VENTANA RxDx Panel for the solid tumor indication. For each marker (MLH1, PMS2, MSH2 and MSH6), ensure that the relevant solid tumors are adequately represented. The data from this study must be adequate to support analytical sensitivity of the 4 markers of the VENTANA MMR RxDx Panel in the intended use population.
PMA P210001: FDA SUMMARY OF SAFETY AND EFFECTIVENESS DATA
Page 32
{33}
2. Ventana Medical Systems, Inc. must provide additional data from well-designed stability studies (shelf-life, in-use and shipping) to support stability claims for the VENTANA MMR RxDx Panel for the solid tumor indication. These studies should use…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.