COMMAND 2 MICOELECTRIC SURGICAL SYSTEM

K961970 · Stryker Corp. · HWE · Jul 26, 1996 · General, Plastic Surgery

Device Facts

Record IDK961970
Device NameCOMMAND 2 MICOELECTRIC SURGICAL SYSTEM
ApplicantStryker Corp.
Product CodeHWE · General, Plastic Surgery
Decision DateJul 26, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4820
Device ClassClass 1
AttributesTherapeutic

Intended Use

The Command2 MicroElectric System is intended for use in the cutting, drilling, reaming, decorticating, and smoothing of bone and other bone related tissue in a variety of surgical procedures including: ENT, dental, orthopedic, maxiofacial, spinal, and plastics. It is also usable in the placement or cutting of screws, wires, pins, and other fixation devices.

Device Story

System consists of electrically powered surgical instruments, control console, and power source; used for cutting, drilling, reaming, decorticating, and smoothing bone/tissue; facilitates placement/cutting of fixation hardware (screws, wires, pins). Operated by surgeons in clinical/OR settings. Provides mechanical power to surgical attachments; enables precise control over surgical contact location and extent. Benefits include surgical precision in diverse specialties.

Clinical Evidence

Bench testing only; no clinical data presented. Equivalence established through design comparison, material consistency, and adherence to recognized safety standards (CSA 22.2 No. 601.1-M90, UL 2601-1, IEC 601-1 series).

Technological Characteristics

Electrically powered surgical system; includes console, handpieces, and accessories. Materials equivalent to pre-1976 predicate devices. Standards: CSA 22.2 No. 601.1-M90, UL 2601-1, IEC 601-1, IEC 601-1-2, CISPR 11, IEC 1000-4-2/3/4/5. Sterilization provided for cutting accessories.

Indications for Use

Indicated for patients requiring bone or bone-related tissue surgery, including ENT, dental, orthopedic, maxillofacial, spinal, and plastic procedures, or placement/cutting of fixation devices like screws, wires, and pins.

Regulatory Classification

Identification

Surgical instrument motors and accessories are AC-powered, battery-powered, or air-powered devices intended for use during surgical procedures to provide power to operate various accessories or attachments to cut hard tissue or bone and soft tissue. Accessories or attachments may include a bur, chisel (osteotome), dermabrasion brush, dermatome, drill bit, hammerhead, pin driver, and saw blade.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K961970 stryker INSTRUMENTS 4100 East Milham Avenue Kalamazoo, MI 49001-6197 (616) 323-700 (800) 253-3210 JUL 26 1996 ## Device Name: Classification Names: Surgical Instrument Motors and Accessories/Attachments: 21 CFR 878.4820, Class I and Class II Common/Usual Name: Surgical instrument Motors and Accessories/Attachments Proprietary Name: Stryker Command2 MicroElectric Surgical System ## Device Sponsor: Stryker Corporation Instruments Division 4100 E. Milham Avenue Kalamazoo, MI 49001 Registration No: 1811755 Stryker Puerto Rico Las Palmas Industrial Park Highway 3, 130.2 Arroyo, Puerto Rico 00615 Registration No.: 2648666 ## Regulatory Classification: Class II ## Summary of Safety and Effectiveness: The Command2 MicroElectric System is intended for use in the cutting, drilling, reaming, decorticating, and smoothing of bone and other bone related tissue in a variety of surgical procedures including: ENT, dental, orthopedic, maxiofacial, spinal, and plastics. It is also usable in the placement or cutting of screws, wires, pins, and other fixation devices. The Stryker Command2 MicroElectric System, consisting of electrically powered surgical instruments, controls, power source, and other system accessories, is equivalent in intended use, safety, and effectiveness to powered instruments, controls, and associated accessories that were marketed by Stryker prior to the 1976 Medical Devices Amendment. The cutting accessories available to be used with the Command2 MicroElectric System will be provided sterile. The materials of construction, cutting action, and cutting surface configuration are equivalent to devices distributed by Stryker prior to the 1976 Medical Devices Amendment. As proven in surgical procedures, these designs offer the required precision required for controlling the location and extent of the surgical contact. {1} The additional power source/console feature described in this submission is equivalent to existing power sources distributed by MicroAire. Intended use, function, and safety risks are all substantially equivalent. Safety warnings, malfunction troubleshooting, and operating duty cycles are listed in the Command2 MicroElectric System Operator's Manual. In addition, the Stryker Command2 MicroElectric System is designed to meet the following Standards: CSA 22.2 No. 601.1-M90, UL 2601-1 First Edition, IEC 601-1 with amendments, IEC 601-1-2, CISPR 11, IEC 1000-4-2, IEC 1000-4-3, IEC 1000-4-4, IEC 1000-4-5. The Stryker Command2 MicroElectric System does not raise any new safety and efficacy concerns when compared to similar devices already being marketed. Therefore, the Stryker Command2 MicroElectric System is substantially equivalent to these existing devices. Diane Davis Regulatory Affairs Representative Stryker Instruments
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...