K160537 · Blue Belt Technologies, Inc. · OLO · May 19, 2016 · Neurology
Device Facts
Record ID
K160537
Device Name
Navio
Applicant
Blue Belt Technologies, Inc.
Product Code
OLO · Neurology
Decision Date
May 19, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Navio system is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
Device Story
Computer-assisted orthopedic surgical navigation and burring system; uses passive infrared tracking camera to establish bone surface models and plan implant locations based on intraoperative landmarks. System controls surgical bur (Anspach eMax2 Plus) motion via software-defined spatial boundaries; operates in Exposure Control mode (bur retracts into guard) or Speed Control mode (bur speed regulated/stops at target surface). Used in orthopedic surgical suites by trained surgeons. Provides orientation/reference information; limits bone removal to shape condyles, tibial plateau, or patellofemoral joint. For TKA, system guides placement of femoral and tibial cutting guides; surgeon completes bone preparation using standard surgical saws. Benefits include precise, software-defined bone preparation for implant placement.
Clinical Evidence
No human clinical testing was conducted. Evidence consists of non-clinical bench testing, including software database reviews, drawing inspections, and clinical simulation (usability testing) using sawbones and cadavers. Testing confirmed that surgeons, physician assistants, and technical support personnel could successfully use the system to place total knee implants per manufacturer specifications.
Technological Characteristics
Computer-assisted navigation system using passive infrared tracking. Components include a standard off-the-shelf surgical drill motor and bur (Anspach eMax2 Plus). Operates via image-free/non-CT intraoperative data collection to create bone models. Software-defined spatial boundaries control bur motion (retraction or speed control). System is networked for data logging; logs are anonymized and exportable to CD. Sterilization requirements follow standard surgical instrumentation protocols.
Indications for Use
Indicated for surgical knee procedures, including unicondylar knee replacement (UKR), patellofemoral arthroplasty (PFA), and total knee arthroplasty (TKA), where stereotactic surgery is appropriate and reference to rigid anatomical bony structures can be determined. Indicated for use with cemented implants only.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 19, 2016
Blue Belt Technologies, Incorporation Mr. Richard Confer Vice President of Regulatory Affairs 2905 Northwest Boulevard, Suite 40 Plymouth, Minnesota 55441
Re: K160537 Trade/Device Name: Navio® Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: Class II Product Code: OLO, HSX, HRY, KRR, NPJ, JWH Dated: February 24, 2016 Received: February 26, 2016
Dear Mr. Confer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K160537 XIXIXXXXXXX
Device Name Navio
Indications for Use (Describe)
The Navio system is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
The Navio system is indicated for use in surgical knee procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include unicondylar knee replacement (UKR), patellofemoral arthroplasty (PFA), and total knee arthroplasty (TKA).
The Navio system is indicated for use with cemented implants only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The overall design is clean and modern.
2905 Northwest Boulevard, Suite 40, Plymouth, Minnesota 55441 USA Tel 763.452.4950 Fax 763.452.4675 www.bluebelttech.com
# 510(k) Summary
| 510(k) Owner | Blue Belt Technologies, Inc.<br>2905 Northwest Blvd Ste. 40<br>Plymouth, MN 55441 USA<br>Tel: (763) 452-4950<br>Fax: (763) 452-4675 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Richard G. Confer<br>Vice President of Regulatory Affairs<br>Tel: (412) 683-3844 x 4106<br>Email: Rick.Confer@smith-nephew.com |
| Date of Submission | February 24, 2016 |
| Classification Reference | 21 CFR 882.4560 |
| Product Code | OLO |
| Supported Codes | HSX, HRY, KRR, NPJ, JWH |
| Common/Usual Name | Orthopedic Sterotaxic Instrument |
| Trade/Proprietary Name | Navio® |
| Predicate Device(s) | Navio® (K152574) |
| Reason for Submission | Expanded indications for use to include total knee<br>arthroplasty (TKA), which adds JWH as a supported<br>product code |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters. The logo is clean and professional, with a focus on the company name.
## Intended Use
The Navio system is intended to assist the surgeon in providing software-defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
## Indications for Use
The Navio system is indicated for use in surgical knee procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be determined. These procedures include unicondylar knee replacement (UKR), patellofemoral arthroplasty (PFA), and total knee arthroplasty (TKA).
The Navio system is indicated for use with cemented implants only.
The Intended Use Statement is consistent with that of Navio®, the predicate device, cleared via K121936, K140596, K143668, and K152574. This expanded Indications for Use Statement includes the addition of total knee arthroplasty and support for product code JWH, per regulation number 888.3560.
## Device Description
The Navio system is a computer-assisted orthopedic surgical navigation and surgical burring system. The system uses established technologies of navigation, via a passive infrared tracking camera, to aid the surgeon in both establishing a bone surface model for the target surgery and in planning the surgical implant location, based on intraoperatively-defined bone landmarks and known geometry of the surgical implant. The Navio system then aids the surgeon in executing the surgical plan by using a standard off-the-shelf surgical drill motor and bur (Anspach eMax2 Plus System, cleared via K080802), which has been adapted using a tracking system.
The surgical bur is inserted into a handpiece, which allows the bur to move within the handpiece. The Navio system software controls the position of the surgical bur relative to the end of a guard attached to the handpiece and prohibits the bur from cutting bone as it approaches the planned target surface. As the planned surface is reached, the tip of the bur is fully retracted within the guard. This is referred to as Exposure Control mode.
An alternate mode of operation is the Speed Control mode. In this mode, the speed of the bur is controlled and the bur stops as the planned target surface is reached. In this mode of operation, the bur does not retract into the guard. This mode of operation is useful in shaping surfaces of the condyle as well as placing post holes and fixation features.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The overall design is clean and modern.
The Navio computer system maintains a log of the patient data and procedure data. Each entry is date and time stamped. Data log entries include date and time stamp for each data line entry, patient and procedure ID, implant ID, step in process, and error messages received by the user during the procedure. This data can be archived to a CD upon demand at the end of the procedure and is anonymized.
The following diagram shows the primary workflow steps in each application: UKR, PFA, and TKA. The three procedures are mutually independent and cannot be planned or completed in parallel.
Note: The workflow for UKR and PFA procedures has not changed from the predicate device, cleared via K152574.
ReshapingMobility™
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The logo is clean and modern, with a focus on the company name.
2905 Northwest Boulevard, Suite 40, Plymouth, Minnesota 55441 USA Tel 763.452.4950 Fax 763.452.4675 www.bluebelttech.com
### Figure 007-1a. Comparison of Workflows (Part 1 of 2)
Image /page/6/Figure/3 description: The image shows three flowcharts that describe the operative procedures for Navio-UKR, Navio-PFA, and Navio-TKA. Each flowchart lists the steps involved in the procedure, starting with hardware connection and ending with evaluating knee ROM. The Navio-UKR procedure includes steps such as malleoli point collection, femur kinematic axis, and gap planning, while the Navio-PFA procedure includes hip center calculation and prosthesis placement. The Navio-TKA procedure includes surgery options and a selection between femur first or tibia first.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The overall design is clean and modern.
2905 Northwest Boulevard, Suite 40, Plymouth, Minnesota 55441 USA Tel 763.452.4950 Fax 763.452.4675 www.bluebelttech.com
Image /page/7/Figure/2 description: The image shows two flowcharts comparing the steps for Navio-TKA for femur first and tibia first. The femur first flowchart includes the following steps: Joint Laxity Extension/Flexion, Place Femur Implant, Place Tibia Implant, Gap Planning, Place Femur Cut Guide, Place Tibia Cut Guide, Checkpoint Verification, Femur Bone Removal, Checkpoint Verification, Tibia Bone Removal, and Evaluate Knee ROM. The tibia first flowchart includes the following steps: Place Tibia Implant, Place Tibia Cut Guide, Checkpoint Verification, Tibia Bone Removal, Joint Laxity Extension/Flexion, Place Femur Implant, Gap Planning, Place Femur Cut Guide, Checkpoint Verification, Femur Bone Removal, and Evaluate Knee ROM.
### Figure 007-1b. Comparison of Workflows (Part 2 of 2)
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue circular design on the left, followed by the words "BLUE BELT" in blue, and the word "TECHNOLOGIES" in gray below. The circular design appears to be a stylized representation of a belt or a ring.
Currently supported implants include:
| Implant Model Name | Manufacturer | 510(k) number |
|-----------------------|---------------------|---------------|
| NEO Total Knee System | New Era Orthopedics | K142388 |
| Journey II CR | Smith and Nephew | K121443 |
### Table 007-1. Summary of Technological Similarities with Predicate
| Devices | Subject Device | Predicate Device |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Navio with Total Knee<br>Arthroplasty Application | Navio [K152574] |
| Intended use | Same as Predicate | The Navio system is intended to<br>assist the surgeon in providing<br>software-defined spatial<br>boundaries for orientation and<br>reference information to<br>anatomical structures during<br>orthopedic procedures. |
| Indications for Use | The Navio System is indicated<br>for use in surgical knee<br>procedures in which the use of<br>stereotactic surgery may be<br>appropriate, and where<br>reference to rigid anatomical<br>bony structures can be<br>determined.<br>These procedures include<br>unicondylar knee replacement,<br>patellofemoral arthroplasty, and<br>total knee arthroplasty. | The Navio System is indicated<br>for use in surgical knee<br>procedures in which the use of<br>stereotactic surgery may be<br>appropriate, and where<br>reference to rigid anatomical<br>bony structures can be<br>determined.<br>These procedures include<br>unicondylar knee replacement<br>and patellofemoral arthroplasty.<br><br>The Navio system is indicated<br>for use with cemented implants<br>only. |
| | The Navio system is indicated<br>for use with cemented implants<br>only. | |
| Supported Product Code(s) | HSX, HRY, KRR, NPJ, JWH | HSX, HRY, KRR, NPJ |
| Devices | Subject Device | Predicate Device |
| | Navio with Total Knee<br>Arthroplasty Application | Navio [K152574] |
| Environment of Use | Same as Predicate | Intended for use by trained<br>orthopedic surgeons in an<br>orthopedic surgical suite. |
| Technological Characteristics | The Navio system uses<br>established technologies to<br>prepare bone for attachment of<br>implant components, or in the<br>case of a total knee<br>arthroplasty, the bone surface is<br>prepared to receive the femoral<br>and tibial cutting guides. | The Navio system uses<br>established technologies to<br>prepare bone for attachment of<br>implant components. |
| | Navio uses intraoperative data<br>collection (image-free or non-CT<br>data generation) to create a<br>model of the patient's femur<br>and/or tibia, dependent on the<br>procedure being performed, and<br>allows the surgeon to prepare a<br>surgical plan. | Navio uses intraoperative data<br>collection (image-free or non-CT<br>data generation) to create a<br>model of the patient's femur<br>and/or tibia, dependent on the<br>procedure being performed, and<br>allows the surgeon to prepare a<br>surgical plan. |
| | The Navio system uses pre-<br>defined boundaries generated<br>during the planning process to<br>control the motion of the<br>surgical bur and limit the<br>amount of bone removed, in<br>order to shape the condyles,<br>tibial plateau, or patellofemoral<br>joint in preparation for<br>placement of the surgical<br>implant, or in the case of a total<br>knee arthroplasty, the bone<br>surface is prepared to receive<br>the femoral and tibial cutting<br>guides.<br>Bone surface for<br>receiving the implant is | The Navio system uses pre-<br>defined boundaries generated<br>during the planning process to<br>control the motion of the<br>surgical bur and limit the<br>amount of bone removed, in<br>order to shape the condyles,<br>tibial plateau, or patellofemoral<br>joint in preparation for<br>placement of the surgical<br>implant. |
| Devices | Subject Device | Predicate Device |
| | Navio with Total Knee<br>Arthroplasty Application | Navio [K152574] |
| | prepared using a standard<br>surgical saw. | |
| | Bur cutting is controlled either<br>by retracting the bur in a guard,<br>or by controlling the speed of<br>the bur as the target surface is<br>approached. | Bur cutting is controlled either<br>by retracting the bur in a guard,<br>or by controlling the speed of<br>the bur as the target surface is<br>approached. |
ReshapingMobility™
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, bold, blue letters. Below "BLUE BELT" is the word "TECHNOLOGIES" in smaller, gray letters. The logo is simple and modern, and the use of blue and gray gives it a professional look.
## Non-Clinical Testing (Bench)
Design verification tests were performed on the Blue Belt Technologies, Inc. Navio system to support total knee arthroplasty, as a result of the risk analysis and product requirements. Testing included software database reviews, bench testing, labeling inspection, drawing inspections, and a clinical simulation (usability testing). Simulateduse testing included simulated knee (sawbones) and cadaver laboratory testing. Users included surgeons, physician's assistants, and technical support personnel who were able to successfully use the Navio system to place total knee implant systems per Blue Belt Technologies' and implant manufacturer's specifications after being adequately trained.
## Clinical Testing
No human clinical testing was conducted to determine safety and effectiveness of the Navio system.
## Discussion of Similarities and Differences
The predicate, Navio, uses established technologies to prepare bone for attachment of implant components. The predicate uses intraoperative data collection (image-free or non-CT data generation) to create a model of the patient's femur and/or tibia and allow the surgeon to prepare a surgical plan. The predicate uses pre-defined boundaries generated during the planning process to control the motion of the surgical bur and limit the amount of bone removed in order to shape the condyles, tibial plateau, or patellofemoral joint in preparation for placement of the surgical implant.
This methodology is equivalent to what is used by the Navio, the subject of this 510(k), with the exception of the Navio's expanded indication to prepare the knee for a total
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Blue Belt Technologies. The logo consists of a blue swirl design on the left, followed by the words "BLUE BELT" in large, blue, sans-serif font. Below "BLUE BELT" is the word "TECHNOLOGIES" in a smaller, gray, sans-serif font. The logo is clean and modern, with a focus on the company name.
knee arthroplasty. During a total knee arthroplasty procedure, the Navio controls the cutting action of the bur during the placement of the cutting guides. The surgeon then uses the Navio-placed cutting guides to finish the preparation of the bone surface using standard surgical saws and the manual instrumentation technique established by each implant's manufacturer to prepare the remainder of the bone surface to receive the total knee implant components.
The Navio surgical system uses pre-defined boundaries generated during the planning process to control the motion of the surgical bur and limit the amount of bone removed in order to shape the condyles, tibial plateau, and/or the patellofemoral joint in preparation for placement of the surgical implant, or in the case of a total knee arthroplasty, the bone surface is prepared to receive the femoral and tibial cutting guides. This submission supports the expanded indication for use of the Navio system to place total knee implants using the same techniques used in the predicate Navio system.
The UKR, PFA, and TKA applications are mutually independent and cannot be planned or completed in parallel.
### Conclusions
The Navio system described in this submission has the same intended use and the same technological characteristics as the Navio system, cleared per K121936, K140596, K143668, and K152574, with the additional indication of total knee arthroplasty. Nonclinical testing was completed to verify that the use of the Navio system to assist with placement of total knee implants does not raise any new issues of safety or effectiveness. The information presented in this 510(k) premarket notification demonstrates that the Navio, when used to place total knee implants, is as safe and effective as the currently cleared Navio system.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.