K142388 · New Era Orthopedics, LLC · JWH · Dec 5, 2014 · Orthopedic
Device Facts
Record ID
K142388
Device Name
Neo Total Knee System
Applicant
New Era Orthopedics, LLC
Product Code
JWH · Orthopedic
Decision Date
Dec 5, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NEO Total Knee System components are indicated for use in skeletally mature individuals, with severe knee pain and disability, undergoing surgery for total knee replacement due to: . Osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, moderate deformities and femoral condyle osteonecrosis. . Failed osteotomies, failed partial knee replacement, or failed total knee replacement whose age, weight and activity level are compatible with an adequate long-term result. Post-traumatic loss of joint configuration, particularly when there is . patellofemoral erosion, dysfunction or prior patellectomy. Moderate valgus, varus, or flexion deformities. . The cruciate retaining (CR) femur with CR tibial insert is for use when the posterior cruciate ligament (PCL) is intact. The CR femur with UC tibial inserts is for use when the PCL is sacrificed. The posteriorly stabilized (PS) femur with PS tibial insert is for use when the PCL is sacrificed. The NEO Total Knee System components are indicated for use only with cement and are single use devices.
Device Story
NEO Total Knee System is a cemented, single-use total knee replacement prosthesis. Components include CoCr femoral components (CR and PS designs), titanium alloy tibial trays, and UHMWPE tibial inserts (CR, UC, PS) and patellar components. Device restores joint function in patients with severe knee pain/disability. Surgeons implant the device in an OR setting. System supports up to 150° active clinical flexion. Output is a mechanical joint replacement; clinical decision-making relies on surgeon assessment of patient anatomy and PCL status to select appropriate component configuration (CR, UC, or PS). Benefits include pain relief and improved joint mobility.
Clinical Evidence
Bench testing only. Evaluations included tibial tray fatigue strength, insert locking mechanism strength, tibial post strength, femorotibial and patellofemoral range of motion, constraint, contact areas, and contact stress.
Indicated for skeletally mature individuals with severe knee pain and disability requiring total knee replacement due to osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, moderate deformities, femoral condyle osteonecrosis, or failed prior knee surgeries (osteotomies, partial or total replacements). Suitable for patients with post-traumatic joint loss, patellofemoral erosion/dysfunction, or moderate valgus/varus/flexion deformities. Contraindicated for patients whose age, weight, or activity level are incompatible with long-term results.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
NEO Total Knee System – Line Extension (UC inserts) (K131368)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized human figure with three faces in profile, representing the department's focus on health and well-being. The logo is simple, clean, and easily recognizable.
December 5, 2014
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
New Era Orthonedics, LLC % Mr. Terry Powell, Sr. Project Manager M. Squared Associates, Inc. 575 Eight Avenue Suite 1212 New York, New York 10018
Re: K142388
Trade/Device Name: NEO Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: October 30, 2014 Received: November 3, 2014
Dear Mr. Powell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
Page 2 - Mr. Terry Powell, Sr.
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use Statement
510(k) Number: K142388
Device Name: NEO Total Knee System
#### Indications for Use:
The NEO Total Knee System components are indicated for use in skeletally mature individuals, with severe knee pain and disability, undergoing surgery for total knee replacement due to:
- . Osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, moderate deformities and femoral condyle osteonecrosis.
- . Failed osteotomies, failed partial knee replacement, or failed total knee replacement whose age, weight and activity level are compatible with an adequate long-term result.
- Post-traumatic loss of joint configuration, particularly when there is . patellofemoral erosion, dysfunction or prior patellectomy.
- Moderate valgus, varus, or flexion deformities. .
The cruciate retaining (CR) femur with CR tibial insert is for use when the posterior cruciate ligament (PCL) is intact. The CR femur with UC tibial inserts is for use when the PCL is sacrificed. The posteriorly stabilized (PS) femur with PS tibial insert is for use when the PCL is sacrificed.
The NEO Total Knee System components are indicated for use only with cement and are single use devices.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# NEO Total Knee System 510(k) Summary - K142388
| Device Proprietary Name: | NEO Total Knee System |
|----------------------------|--------------------------------------------------------------|
| Common Name: | Total Knee System |
| Classification regulation: | 888.3560 |
| Device Class: | Class II |
| Product Codes: | JWH |
| Submitter's Name: | New Era Orthopedics, LLC |
| Address: | 1214 Research Boulevard, Suite 1019<br>Hummelstown, PA 17036 |
| Contact Person: | Craig Corrance |
| Telephone Number: | 717 585 6785 |
| Fax Number: | 407 386 3348 |
Date Summary Prepared:
### Purpose of Submission
To obtain 510(k)s under the name of the New Era Orthopedics, LLC for the same NEO Total Knee System, with minor dimensional changes, as previously cleared in 510(k)s #K120313, K122500, and K131368.
### Device Description:
The NEO Total Knee System includes:
- Cruciate retaining (CR) Femoral Components and Posteriorly stabilized (PS) Femoral . Components in right and left configurations in sizes 1 to 10, manufactured from Cast CoCr.
- Tibial Trays manufactured from titanium alloy in sizes 1 – 10.
- . Tibial Inserts in cruciate retaining (CR), ultracongruent (UC), and posteriorly stabilized (PS) designs, manufactured from standard ultrahigh molecular weight polyethylene, in sizes A to E and in thicknesses 6 to 21 mm (total thickness with tibial tray is 9 to 24 mm).
- Patellar Components in diameters of 26 to 41 mm and in thicknesses of 8, 9 or 10 mm, manufactured from standard ultrahigh molecular weight polyethylene.
The NEO Total Knee System is designed with the potential to have an active clinical flexion of 150°.
### Intended Use:
The NEO Total Knee System components are indicated for use in skeletally mature individuals, with severe knee pain and disability, undergoing surgery for total knee
{4}------------------------------------------------
# NEO Total Knee System 510(k) Summary - K142388
replacement due to:
- . Osteoarthritis, rheumatoid arthritis, post-traumatic arthritis, moderate deformities and femoral condyle osteonecrosis.
- . Failed osteotomies, failed partial knee replacement, or failed total knee replacement whose age, weight and activity level are compatible with an adequate long-term result.
- . Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy.
- Moderate valgus, varus, or flexion deformities.
The cruciate retaining (CR) femoral component with CR tibial insert is for use when the posterior cruciate ligament (PCL) is intact. The CR femoral component with UC tibial insert is for use when the PCL is sacrificed. The posteriorly stabilized (PS) femoral component with PS tibial insert is for use when the PCL is sacrificed.
The NEO Total Knee System components are indicated for use only with cement and are single use devices.
#### Predicate Devices:
The predicate devices are:
| Trade/Proprietary Name | 510(K) # | Clearance Date |
|--------------------------------------------------------|----------|----------------|
| NEO CR Knee System | K120313 | 04-20-2012 |
| NEO PS Knee | K122500 | 05-03-2013 |
| NEO Total Knee System – Line Extension (UC<br>inserts) | K131368 | 10-03-2013 |
#### Technological Characteristics:
The NEO Total Knee System has the same indications for use, is manufactured from the same materials using the same methods, and has the same design features as the predicate knee systems, and uses traditional sterilization methods. The NEO Total Knee System described in the subject 510(k) is the same knee system, with minor dimensional modifications, previously cleared for marketing in 510(k) #K120313, K122500, and K131368.
#### Performance Testing:
The NEO Total Knee System has been evaluated for tibial tray fatigue strength, insert locking mechanism strength, tibial post strength (for PS design), femorotibial range of motion, femorotibial range of constraint, patellofemoral range of constraint, femorotibial contact areas/contact stress, and patellofemoral contact area and contact stress.
{5}------------------------------------------------
# NEO Total Knee System 510(k) Summary - K142388
### Substantial Equivalence Information:
The NEO Total Knee System is the same, except for minor dimensional changes, as the predicate NEO Total Knee System, and is therefore substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.