THE MARGRON HIP REPLACEMENT SYSTEM

K992815 · Portland Orthopaedics Pty. , Ltd. · MEH · Feb 7, 2000 · Orthopedic

Device Facts

Record IDK992815
Device NameTHE MARGRON HIP REPLACEMENT SYSTEM
ApplicantPortland Orthopaedics Pty. , Ltd.
Product CodeMEH · Orthopedic
Decision DateFeb 7, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Margron™ Hip Replacement System is intended for primary or revision restruction, without bone cement, of the femoral portion of a severely disabled and/or very painful hip ioint. where radiographic evidence of sufficient sound bone is present. The patient should be skeletally mature. The patient's condition should be due to one or more of the following: - Osteoarthritis . - Rheumatoid arthritis ● - Ankylosing spondylitis . - Psoriatic arthritis ● - Old osteomvelitis with a long infection free period and a normal WBC, C-. Reactive protein and ESR - Non union of femoral neck fracture or avascular necrosis of the femoral head ● - Post-traumatic fracture/dislocation of the hip . - Revision of an unsuccessful arthrodesis with either poor positioning or pain in ● the hip. or where low back pain or knee pain is becoming disabling - Revision of an unsuccessful uncemented or cemented hip replacement stem, ● providing sufficient bone stock is present - Revision of a previous unsuccessful femoral osteotomy, Girdlestone resection, . cup arthroplasty or hemi arthroplasty.

Device Story

Modular femoral hip replacement system; designed for uncemented fixation in high-quality bone. Components include tapering stem cone with external threads and longitudinal derotation columns; adjustable neck component for anteversion selection. Surgical procedure involves precision milling and tapping of femur; stem insertion via 'screw home' rotation. Unique instrumentation provided for insertion and extraction. Used by orthopedic surgeons in clinical settings. Provides stable femoral fixation; restores joint function; reduces pain in patients with severe hip disability. Materials include Chrome Cobalt (ASTM F-799-96) with Calcium Phosphate coating on upper stem half.

Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence supported by fatigue, torque, and corrosion testing.

Technological Characteristics

Modular femoral prosthesis; Chrome Cobalt (ASTM F-799-96) construction; Calcium Phosphate coating on upper stem. Features tapering stem cone with external threads and longitudinal derotation columns. Mechanical fixation via 'screw home' rotation. No software or electronic components.

Indications for Use

Indicated for skeletally mature patients requiring primary or revision femoral hip reconstruction without bone cement due to osteoarthritis, rheumatoid arthritis, ankylosing spondylitis, psoriatic arthritis, old osteomyelitis, femoral neck fracture non-union, avascular necrosis, post-traumatic hip dislocation, or revision of prior failed hip procedures (arthrodesis, uncemented/cemented stems, osteotomy, Girdlestone resection, cup/hemi-arthroplasty). Requires sufficient sound bone stock.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ FEB r .. 2000 K 99+8/5 1083 # ummary of Safety and Effectiveness Submitter Name and Address: Contact Person: Date Summary Prepared: Proprietary Name: Common Name: Classification Name: Product Codes: Portland Orthopaedics Pty. Ltd. Suite 3 Level 5 St. George Private Hospital 1 South Street Kogarah NSW 2217 Australia Dr. Ronald Sekel 011 (61) 2 9588 2121 March 12, 1999 Margron™ Hip Replacement System Modular Femoral Hip Replacement System Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis. 21 CFR 888.3353 LPY/NEH 87 JDO - Device, Fixation, Proximal Femoral, Implant Predicate Devices: The Margron Hip is similar in design, materials and intended use to: | Manufacturer | Metagen | Biomet | Joint<br>Medical/<br>J&J | Wright<br>Medical<br>Technology | S&N<br>Richards | | |--------------|--------------------------------------------------|---------------------|-------------------------------|--------------------------------------------|-----------------------|----------------------------------------------| | System Name | Modular<br>Femoral<br>Hip<br>System<br>(K980020) | Impact<br>(K942027) | Mallory-<br>Head<br>(K921181) | S-ROM<br>(K851422,<br>K912713,<br>K913231) | Infinity<br>(K942115) | Modular<br>Hip System<br>(RMHS)<br>(K912593) | {1}------------------------------------------------ K992815 2083 ## Device Description: The Margron™ hip replacement stem has been designed to be used in an uncemented mode in bone which is of good quality physiologically and which will heal quickly. The Margron™ Femoral Stem is a modular prosthesis consisting of: - A tapering stem cone with two different speed external threads and longitudinal . derotation columns. - A neck component which allows the optimum angle of anteversion to be selected . after stem insertion. - . Precision milling and tapping is used to prepare the femur, and stem insertion is by 'screw home' rotation. - A unique set of instruments have been designed to insert and extract the . prosthesis. ## Intended Use/Indications: The Margron™ Hip Replacement System is intended for primary or revision restruction, without bone cement, of the femoral portion of a severely disabled and/or very painful hip joint, where radiographic evidence of sufficient sound bone is present. The patient should be skeletally mature. The patient's condition should be due to one or more of the following: - Osteoarthritis ● - . Rheumatoid arthritis - Ankylosing spondylitis - Psoriatic arthritis - Old osteomyelitis with a long infection free period and a normal WBC, C-. Reactive protein and ESR - Non union of femoral neck fracture or avascular necrosis of the femoral head . - Post-traumatic fracture/dislocation of the hip - Revision of an unsuccessful arthrodesis with either poor positioning or pain in . the hip, or where low back pain or knee pain is becoming disabling - . Revision of an unsuccessful uncemented or cemented hip replacement stem, providing sufficient bone stock is present - Revision of a previous unsuccessful femoral osteotomy, Girdlestone resection, . cup arthroplasty or hemi arthroplasty. {2}------------------------------------------------ KSS 2815 3083 ## Materials Materials for components of the Margron™ Hip Replacement System are implant grade materials. Material Composition of the Components: Stem Chrome Cobalt (Cr Co) . Calcium Phosphate Coating on upper half of stem . Neck Chrome Cobalt (Cr Co) . Chrome Cobalt (CrCo) per ASTM F-799-96 # Mechanical Characteristics The Margron Hip System has been evaluated by use of the fatigue, torque and corrosion tests. It has been shown to be substantially equivalent to the predicate devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the words "Public Health Service" in bold, black font. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line. The text appears to be extracted from a document or sign. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 7 2000 FEB > Dr. Ronald Sekel Director Portland Orthopedics Pty. Ltd. Suite 3 Level 5 St. George Private Hospital 1 South Street Kogarah NSW 2217 Australia Re: K992815 > Trade Name: Margron Hip Replacement System Regulatory Class: II Product Code: MEH & LPH Dated: November 27, 1999 Received: November 30, 1999 Dear Dr. Sekel: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ ## Page 2 - Dr. Ronald Sekel This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely vours. Nunell Sayer Sames E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Page 1 of 1 K992815 510(k) Number (if known): New Device Name: # Margron Hip Replacement System Indication For Use: The Margron™ Hip Replacement System is intended for primary or revision restruction, without bone cement, of the femoral portion of a severely disabled and/or very painful hip ioint. where radiographic evidence of sufficient sound bone is present. The patient should be skeletally mature. The patient's condition should be due to one or more of the following: - Osteoarthritis . - Rheumatoid arthritis ● - Ankylosing spondylitis . - Psoriatic arthritis ● - Old osteomvelitis with a long infection free period and a normal WBC, C-. Reactive protein and ESR - Non union of femoral neck fracture or avascular necrosis of the femoral head ● - Post-traumatic fracture/dislocation of the hip . - Revision of an unsuccessful arthrodesis with either poor positioning or pain in ● the hip. or where low back pain or knee pain is becoming disabling - Revision of an unsuccessful uncemented or cemented hip replacement stem, ● providing sufficient bone stock is present - Revision of a previous unsuccessful femoral osteotomy, Girdlestone resection, . cup arthroplasty or hemi arthroplasty. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sion-Off) Div Restorative Devices 510(k) Number `(9) 77877 **Prescription Use** (Per 21 CFR 801.109)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...