MODIFICATION TO: THE MARGRON HIP REPLACEMENT SYSTEM
K032641 · Portland Orthopaedics Pty. , Ltd. · MEH · Jan 20, 2004 · Orthopedic
Device Facts
| Record ID | K032641 |
| Device Name | MODIFICATION TO: THE MARGRON HIP REPLACEMENT SYSTEM |
| Applicant | Portland Orthopaedics Pty. , Ltd. |
| Product Code | MEH · Orthopedic |
| Decision Date | Jan 20, 2004 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 888.3353 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The MARGRON™ Hip Replacement System is intended for primary or revision restruction, without bone cement, of the femoral portion of a severely disabled and/or very painful hip joint, where radiographic evidence of sufficient sound bone is present. The patient should be skeletally mature. The patient's condition should be due to one or more of the following: Osteoarthritis; Rheumatoid arthritis; Tumor conditions involving the upper third of the femur or the acetablym; Ankylosing spondylitis; Psoriatic arthritis; Old osteomyelitis - with a long infection-free period and a normal WBC. ESR and C-reactive protein; Non union of femoral neck fracture or avascular necrosis of the femoral head; Post-traumatic fracture/dislocation of the hip; Revision of an unsuccessful arthrodesis with either poor positioning or pain in the hip, or where low back pain or knee pain is becoming disabling; Revision of an unsuccessful cemented or un-cemented hip replacement stem, providing sufficient bone stock is present; Revision of a previous unsuccessful femoral osteotomy, Girdlestone resection, cup arthroplasty or hemi arthroplasty.
Device Story
Modular femoral hip replacement system; uncemented fixation. Components include tapering stem cone with dual-speed external threads and longitudinal derotation columns; adjustable neck component for anteversion selection. Surgical procedure involves precision milling and tapping of femur; stem insertion via 'screw home' rotation. Unique instrumentation provided for insertion and extraction. Used by orthopedic surgeons in clinical/OR settings. Provides structural replacement of femoral portion of hip joint; restores joint function; alleviates pain in patients with severe hip disability.
Clinical Evidence
Bench testing only. Physical testing included fatigue, torque, and corrosion testing. Biocompatibility testing performed. No clinical data provided.
Technological Characteristics
Modular femoral prosthesis; uncemented fixation. Features tapering stem cone with dual-speed external threads and longitudinal derotation columns. Includes adjustable neck component for anteversion. Sterile, single-patient-use. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring primary or revision femoral hip replacement without bone cement due to osteoarthritis, rheumatoid arthritis, tumor conditions, ankylosing spondylitis, psoriatic arthritis, old osteomyelitis, non-union of femoral neck fracture, avascular necrosis, post-traumatic fracture/dislocation, or revision of prior unsuccessful hip procedures (arthrodesis, cemented/uncemented stems, osteotomy, Girdlestone resection, cup/hemi arthroplasty).
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
- MARGRON™ Hip Replacement System (K992158)
Related Devices
- K992815 — THE MARGRON HIP REPLACEMENT SYSTEM · Portland Orthopaedics Pty. , Ltd. · Feb 7, 2000
- K072417 — SMITH & NEPHEW MIS HIP STEM · Smith & Nephew, Inc. · Jan 10, 2008
- K092406 — ABG II MODULAR HIP STEM · Howmedica Osteonics Corp. · Nov 4, 2009
- K201471 — M-Vizion Femoral Revision System Extension · Medacta International S.A. · Jan 11, 2021
- K030733 — MAYO CONSERVATIVE HIP PROSTHESIS, MODEL 8026 SERIES · Zimmer, Inc. · May 1, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
032641
#### Summary of Safety and Effectiveness 10.
| Manufacturer: | Portland Orthopaedics Pty. Ltd.<br>Unit 3, 44 McCauley Street<br>Matraville, NSW, 2036<br>Australia |
|------------------------|-----------------------------------------------------------------------------------------------------|
| Contact: | David Sekel<br>CEO, Portland Orthopaedics |
| Telephone Number: | 011 (61) 2 9666 8444 |
| Fax Number: | 011 (61) 2 9666 8544 |
| Date Summary Prepared: | August 15, 2003 |
| Product Trade Name: | MARGRON™ Hip Replacement System |
| Common Name: | Modular Femoral Hip Replacement System |
| Classification Name: | Hip joint/ceramic/polymer semi-constrained<br>Cemented or nonporous uncemented prosthesis |
| Product Codes: | 87 MEH & LPH - Device, Fixation, Proximal<br>Femoral, Implant |
| Predicate Device: | MARGRON™ Hip Replacement System<br>(K992158) |
Page 1 of 3
MARGRON™ Hip Replacement System Special 510(k) Notification
.
:
{1}------------------------------------------------
## KO32641
#### Device Description:
The MARGRON 100 hip replacement stem has been designed to be used in an uncemented mode in bone which is of good quality physiologically and which will heal quickly.
The MARGRON TM Femoral Stem is a modular prosthesis consisting of
- A tapering stem cone with two different speed external threads and longitudinal . derotation columns.
- A neck component which allows the optimum angle of anteversion to be selected ● after stem insertion.
- . Precision milling and tapping is used to prepare the femur and stem insertion is by "screw home" rotation.
- . A unique set of instruments have been designed to insert and extract the prosthesis.
#### Intended Use/Indications:
The MARGRONTM Hip Replacement System is intended for primary or revision restruction, without bone cement, of the femoral portion of a severely disabled and/or very painful hip joint, where radiographic evidence of sufficient sound bone is present.
The patient should be skeletally mature. The patient's condition should be due to one or more of the following:
- Osteoarthritis ●
- . Rheumatoid arthritis
- . Tumor conditions involving the upper third of the femur or the acetablym
- Ankylosing spondylitis .
- Psoriatic arthritis .
- � Old osteomyelitis - with a long infection-free period and a normal WBC. ESR and C-reactive protein
- . Non union of femoral neck fracture or avascular necrosis of the femoral head
- . Post-traumatic fracture/dislocation of the hip
- . Revision of an unsuccessful arthrodesis with either poor positioning or pain in the hip, or where low back pain or knee pain is becoming disabling
- Revision of an unsuccessful cemented or un-cemented hip replacement stem, . providing sufficient bone stock is present
- Revision of a previous unsuccessful femoral osteotomy, Girdlestone resection, cup . arthroplasty or hemi arthroplasty.
Page 2 of 3
Portland Orthopaedics Pty. Ltd. MARGRONTM Hip Replacement System Special 510(k) Notification
{2}------------------------------------------------
# (03244)
#### Substantial Equivalence:
The MARGRONTM Hip Replacement System is substantially equivalent to the current MARGRONTM Hip Replacement System (K992815) in that:
- the intended use is the same ●
- the same operating principles is used .
- the same final design configuration is incorporated �
- both are labeled as sterile, single patient use .
- both are packaged and labeled using the same materials and processes �
#### Summary of Testing:
All materials used in the fabrication of this MARGRON™ Hip Replacement System were evaluated with the original design through physical testing (fatigue, torque and corrosion testing) and biocompatibility testing. It has been shown to be substantially equivalent to the predicate device.
Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, rendered in black.
Public Health Service
JAN 2 0 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. David Sekel CEO Portland Orthopaedics Pty. Ltd Unit 3, 44 McCauley Strect Matraville, NSW, 2036 Australia
Re: K032641 : Ito22011 Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: II Product Code: MEH, LPH Dated: December 18, 2003 Received: December 22, 2003
Dear Mr. Sekel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can he found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2 - Mr. David Sekel
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), plcasc contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-frec number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Sincerely yours,
Mark N Milliman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### 510(k) Number (if known): _
Device Name:
MARGRONTM Hip Replacement System
#### Indications for Use:
The MARGRON™ Hip Replacement System is intended for primary or revision restruction, without bone cement, of the femoral portion of a severely disabled and/or very painful hip joint, where radiographic evidence of sufficient sound bone is present.
KO32641
The patient should be skeletally mature. The patient's condition should be due to one or more of the following:
- Osteoarthritis .
- Rheumatoid arthritis .
- Tumor conditions involving the upper third of the femur or the acetablum .
- Ankylosing spondylitis ●
- Psoriatic arthritis .
- Old osteomyelitis with a long infection-free period and a normal WBC, ESR and C-reactive protein .
- Non union of femoral neck fracture or avascular necrosis of the femoral head .
- Post-traumatic fracture/dislocation of the hip �
- Revision of an unsuccessful arthrodesis with either poor positioning or pain in the hip, or where . low back pain or knee pain is becoming disabling
- Revision of an unsuccessful cemented or un-cemented hip replacement stem, providing sufficient . bone stock is present
- Revision of a previous unsuccessful femoral ostcotomy, Girdlestone resection, cup arthroplasty or ◆ hemi arthroplasty.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | <span style="text-decoration: overline;">X</span> OR <span style="text-decoration: overline;"></span> Over-The-Counter Use |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
for
| | (Division Sign-Off) |
|--|----------------------------------------------------------------|
| | Division of General and Restorative Devices |
| | 510(k) Number <span style="text-decoration: overline;"></span> |
Mark N Millenson
Restorative
Medical Devices
K032641
Portland Orthopaedics Pty. Ltd. MARGRONTRA Hip Replacement System Special 510(k) Notification