K232425 · Orthosoft Inc. (D/B/A) Zimmer Cas) · QHE · Dec 12, 2023 · Orthopedic
Device Facts
Record ID
K232425
Device Name
Signature™ ONE System
Applicant
Orthosoft Inc. (D/B/A) Zimmer Cas)
Product Code
QHE · Orthopedic
Decision Date
Dec 12, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Indications for Use
The Signature™ ONE System is indicated, based on patient-specific radiological images with identifiable placement anatomical landmarks, to assist in pre-operative planning and/or intra-operative guiding of surgical instruments for shoulder replacement surgical procedures on patients not precluded from being radiologically scanned. The Signature™ ONE System is designed for use on a skeletally mature patient population. The targeted population has the same characteristics as the population that is suitable for the implants compatible with the Signature™ ONE System. The Signature™ ONE System is to be used with the glenoid components of the following shoulder implant systems in accordance with their indications: Zimmer® Trabecular Metal Reverse Plus® Shoulder. Comprehensive® Total Shoulder System, Comprehensive® Reverse Shoulder System, Comprehensive® Reverse Augmented Baseplates and Alliance® Glenoid System. The Signature™ ONE System pre-operative planning is also compatible with the humeral components of the following shoulder implant systems in accordance with their indications: Comprehensive® Total Shoulder System, Comprehensive® Reverse Shoulder System, and Identity™ Shoulder System. The Signature™ ONE System Guides and bone models are intended for single use only.
Device Story
System assists in pre-operative planning and intra-operative guidance for shoulder arthroplasty. Inputs: patient-specific radiological images. Process: software (Signature™ ONE Planner) enables surgeon to plan implant placement; additive manufacturing (selective laser sintering) produces patient-specific polyamide (nylon) guides and bone models based on plan. Output: physical guides/models used intra-operatively by surgeons to transfer plan to scapula/humerus. Used in clinical settings; guides/models are single-use and sterilized at hospital. Benefits: assists in accurate reproduction of pre-operative plan during surgery.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including performance testing, engineering analysis, usability engineering, and software verification/validation.
Technological Characteristics
Materials: Polyamide (nylon). Manufacturing: Additive manufacturing (selective laser sintering). Components: Patient-specific guides and bone models. Connectivity: Zimmer Biomet Portal (MDDS) for user interaction. Software: Planning and manufacturing applications. Sterilization: Hospital-based.
Indications for Use
Indicated for skeletally mature patients undergoing shoulder replacement surgery (anatomic or reverse) who are eligible for radiological scanning. Used with specific Zimmer Biomet glenoid and humeral implant systems.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Tornier Perform™ Patient-Matched Primary Reversed Glenoid and BLUEPRINT™ Patient Specific Instrumentation (K211359)
Submission Summary (Full Text)
{0}------------------------------------------------
December 12, 2023
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, the FDA logo is displayed in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" written in a clear, sans-serif font. The overall design is clean and professional, reflecting the organization's role in public health and safety.
Orthosoft Inc. (d/b/a Zimmer CAS) Nilam Dave Regulatory Affairs Specialist 75 Queen Street Suite 3300 Montreal, QC H3C 2N6 Canada
# Re: K232425
Trade/Device Name: Signature™ ONE System Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: QHE, KWT, KWS, PHX, MBF, HSD Dated: November 15, 2023 Received: November 16, 2023
Dear Nilam Dave:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Farzana Farzana Sharmin -S Date: 2023.12.12 Sharmin -S 16:22:45 -05'00'
Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232425
Device Name Signature™ ONE System
#### Indications for Use (Describe)
The Signature™ ONE System is indicated, based on patient-specific radiological images with identifiable placement anatomical landmarks, to assist in pre-operative planning and/or intra-operative guiding of surgical instruments for shoulder replacement surgical procedures on patients not precluded from being radiologically scanned.
The Signature™ ONE System is designed for use on a skeletally mature patient population. The targeted population has the same characteristics as the population that is suitable for the implants compatible with the Signature™ ONE System.
The Signature™ ONE System is to be used with the glenoid components of the following shoulder implant systems in accordance with their indications: Zimmer® Trabecular Metal Reverse Plus® Shoulder. Comprehensive® Total Shoulder System, Comprehensive® Reverse Shoulder System, Comprehensive® Reverse Augmented Baseplates and Alliance® Glenoid System.
The Signature™ ONE System pre-operative planning is also compatible with the humeral components of the following shoulder implant systems in accordance with their indications: Comprehensive® Total Shoulder System, Comprehensive® Reverse Shoulder System, and Identity™ Shoulder System.
The Signature™ ONE System Guides and bone models are intended for single use only.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> |
|----------------------------------------------|---------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span></span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Signature™ ONE System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on September 13, 2019.
| Sponsor: | Orthosoft Inc. (d/b/a Zimmer CAS)<br>75 Queen St., Suite 3300<br>Montreal, QC, CANADA H3C 2N6<br>Establishment Registration Number: 9617840 | |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Contact Person: | Nilam Dave<br>Regulatory Affairs Specialist<br>Telephone: +1 (206) 638-8947 | |
| Preparation Date: | 09 AUG 2023 | |
| Subject Device: | Trade Name: Signature™ ONE System<br>Common Name: Signature™ ONE System, ONE Planner<br>Shoulder | |
| | Classification Name:<br>• QHE- Shoulder joint metal/polymer semi-constrained<br>cemented prosthesis (21 CFR 888.3660) | |
| | Additional Product Codes:<br>• KWT- Shoulder joint metal/polymer non-<br>constrained cemented prosthesis (21 CFR 888.3650)<br>• KWS- Shoulder joint metal/polymer semi-<br>constrained cemented prosthesis(21 CFR 888.3660)<br>• PHX- Shoulder joint metal/polymer semi-<br>constrained cemented prosthesis (21 CFR 888.3660)<br>• MBF- Shoulder joint metal/polymer/metal<br>nonconstrained or semi-constrained porous-coated<br>uncemented prosthesis (21 CFR 888.3670)<br>• HSD- Shoulder joint humeral (hemi-shoulder)<br>metallic uncemented prosthesis (21 CFR 888.3690) | |
{4}------------------------------------------------
#### Predicate Device:
| K212560<br>(primary) | Signature™ ONE System | Zimmer CAS |
|------------------------|------------------------------------------------------------------------------------------------------------------------|---------------|
| K211359<br>(secondary) | Tornier Perform™ Patient-<br>Matched Primary<br>Reversed Glenoid and<br>BLUEPRINT™ Patient<br>Specific Instrumentation | Tornier, Inc. |
# Purpose and Device Description:
The Signature™ ONE System is developed to assist in preoperative planning (using the Signature™ ONE Planner) and to accurately transfer a pre-operative plan to orthopedic surgical procedures (using the Signature™ ONE Guides and bone model) if desired in skeletally mature individuals for Total Shoulder Arthroplasty. Both anatomic and reverse (TSA and RSA respectively) approaches are supported.
The Signature ONE Guides and Bone Models are designed and manufactured of polyamide (nylon) using additive manufacturing selective laser sintering (SLS), based on the approved/finalized pre-surgical plan and shipped prior to surgery. The guides and bone models are provided nonsterile and sterilized at the hospital. They are used intraoperatively to assist the surgeon in reproducing the plan on the scapula. The Signature ONE System surgical technique remains close to the conventional shoulder arthroplasty workflow.
The Signature™ ONE System uses a Non-Device Medical Device Data System (MDDS) called the Zimmer Biomet Portal for the interaction with external users (i.e. imaging technician and the surgeon). The internal users (i.e. the Zimmer Biomet operators) use manufacturing software applications to prepare the patient cases for the surgeon.
The purpose of this submission is to introduce new hardware kits for the existing compatible Comprehensive Reverse Augment implants and to include the addition of pre-operative planning of humeral components. A Rotational Guide component will be available as part of these two kits. Modifications have been made to the software applications to accommodate the new guide ordering option. The overall manufacturing process, materials, sterilization methods, have not changed from the previous primary predicate and principal of operation
{5}------------------------------------------------
remains
similar.
The Signature™ ONE System is indicated, based on Indications for Use: patient-specific radiological images with identifiable placement anatomical landmarks, to assist in pre-operative planning and/or intra-operative guiding of surgical instruments for shoulder replacement surgical procedures on patients not precluded from being radiologically scanned.
> The Signature™ ONE System is designed for use on a skeletally mature patient population. The targeted population has the same characteristics as the population that is suitable for the implants compatible with the Signature™ ONE System.
> The Signature™ ONE System is to be used with the glenoid components of the following shoulder implant systems in accordance with their indications and contraindications: Zimmer® Trabecular Metal Reverse Plus® Shoulder, Comprehensive® Total Shoulder ---------------------------------------------------------------------------------------------------------------------------------------------------------------------System, Comprehensive® Reverse System, Shoulder Comprehensive® Reverse Augmented Baseplates and Alliance® Glenoid System.
> The Signature™ ONE System pre-operative planning is also compatible with the humeral components of the following shoulder implant systems in accordance with their indications and contraindications: Comprehensive® Shoulder System. Comprehensive® Total Reverse Shoulder System, and Identity™ Shoulder System.
> The Signature™ ONE System Guides and bone models are intended for single use only.
Differences in Indications for Use: The indications for use was updated to include compatibility with implant systems for humerus preoperative planning that was not present with primary predicate device K212560.
{6}------------------------------------------------
#### Summary of Technological Characteristics:
The rationale for substantial equivalence is based on consideration of the following characteristics:
- o Intended Use: Same as predicate device
- Indications for Use: Similar to predicate device with 0 new compatibility addition
- Materials: Same as predicate device
- Sterilization: Same as predicate device ●
- Design Features: The new Rotational Guide component is made available for use with the Comprehensive Reverse Augment workflow from the primary predicate device and provided in two new kits. Modifications have been made to the software applications and the overall workflow from the primary predicate device to incorporate humerus pre-operative planning.
# Summary of Performance Data (Nonclinical and/or Clinical):
The following performance data was provided in support of the substantial equivalence determination:
# Device Performance Testing:
Verification and Validation Testing for Signature™ ONE System was conducted with the following aspects:
- Performance Tests: Performance tests documented to ensure o the performance of the implemented features and verify related design inputs.
- Engineering Analysis: Analysis documented to ensure the o performance of the implemented features and verify related design inputs
- Usability Engineering: Performance of the system in regards o to human factors engineering.
- Validation: Validation performed to validate related user O needs, intended use and safety and effectiveness.
# Software Verification and Validation Testing
Software tests were conducted to satisfy the Basic Level of Documentation per requirements of Content of Premarket Submissions for Device Software Functions Guidance for Industry and Food and Drug Administration Staff issued on June 14, 2023 and IEC 62304 (Medical Device Software- Software Life Cycle
{7}------------------------------------------------
Processes). The testing demonstrates that the Signature™ ONE System does not raise any new issues of safety and effectiveness as compared to the predicate devices.
### Substantial Equivalence Conclusion
The proposed and predicate device have the same intended use and similar indications for use, technological characteristics and principle of operation with the difference due to compatibility addition of pre-operative planning of humeral components and new guide. The new Rotational Guide in proposed device is similar to guides provided in primary predicate device to assist in intraoperative guiding of surgical instruments. In summary, any differences between the subject and predicate devices do not raise different questions of safety and effectiveness and the proposed device is at least as safe and effective as the legally marketed predicate devices.