K950099 · Buckman Co., Inc. · MNH · Jan 16, 1997 · Orthopedic
Device Facts
Record ID
K950099
Device Name
SYNERGY(TM) TI POSTERIOR SPINAL SYSTEM
Applicant
Buckman Co., Inc.
Product Code
MNH · Orthopedic
Decision Date
Jan 16, 1997
Decision
SN
Submission Type
Traditional
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The SYNERGY™ Posterior Spinal System implants are intended to be used as a temporary construct that assists normal healing and are not intended to replace normal body structures. They are intended to stabilize the spinal operative site during the development of a solid fusion with bone graft, and are intended to be removed after the development of a solid fusion mass.
Device Story
The SYNERGY™ Posterior Spinal System is a temporary, posterior spinal fixation construct designed to stabilize the spine during bone graft fusion. The system comprises stainless steel or titanium screws, hooks, rods, and connectors. Surgeons implant the device via a posterior approach to the thoracic, lumbar, and sacral spine. The device provides mechanical stabilization; it does not replace body structures and must be removed after solid fusion occurs. The system is intended for use by experienced surgeons in a hospital setting. Clinical benefit is derived from stabilizing the spine to facilitate fusion in patients with specific deformities or degenerative conditions. The device is not intended for use as the sole means of spinal support; it requires concurrent bone grafting. Failure to achieve fusion may lead to implant fatigue, bending, or fracture.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by mechanical testing, engineering analysis, and comparison of design, materials, and indications to legally marketed predicate devices.
Technological Characteristics
Constructs made of 22-13-5 stainless steel (ASTM F-1314), 316 LVM stainless steel (ASTM F-138-86), 6Al-4V ELI titanium alloy (ASTM F-136-92), and commercially pure titanium (ASTM F-67-89). Instruments made of 17-4, 420, 440, 455 series stainless steel (ASTM F-899-84, A276-91). Components include pedicle screws, hooks, rods, and connectors. Sterilization via gravity steam (250°F/121°C for 60 minutes).
Indications for Use
Indicated for patients with severe spondylolisthesis (Grades 3-4) at L5-S1 using pedicle screws with autogenous bone graft. Also indicated for degenerative disc disease, idiopathic scoliosis, kyphotic/lordotic deformities, paralytic scoliosis, neuromuscular scoliosis, vertebral fracture/dislocation, tumors, spondylolisthesis, stenosis, pseudarthrosis, and failed previous fusion. Pedicle screw use limited to L5-S1; other components for thoracic/lumbar use (T1-Sacrum). 4.75mm rod constructs restricted to patients ≤32 kg. Contraindicated in patients with recent infection, morbid obesity, mental illness, substance abuse, fever, pregnancy, metal sensitivity, severe osteopenia, or inadequate anatomy for fixation.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
K081952 — SYNERGY SPINAL SYSTEM (6.35 POLARIS) · Biomet Spine · Oct 7, 2008
K973836 — SYNERGY 6.35MM SS ROD WITH SCREWS FOR ANTERIOR USE, PART OF THE CROSS MEDICAL-SYNERGY ANTERIOR SPINAL SYSTEM · Cross Medical Products, Inc. · Mar 5, 1998
K992369 — SYNERGY INTEGRAL SCREWS, MODELS, 2161,2162,2163,2164,2165,2166,2167,2168,2169,2210,2211,2212,2215,2216,2217,2220,2221,22 · Interpore Cross Intl. · Oct 13, 1999
K964254 — TITANIUM LIBERTY POSTERIOR SPINAL SYSTEM · Sofamor Danek USA,Inc. · Apr 21, 1997
Submission Summary (Full Text)
{0}
01/03/97 FRI 17:13 FAX 614 718 0515 CROSS MEDICAL 037
# K950099 JAN 16 1997
## 510(k)
### Summary of Safety and Effectiveness Information
**Regulatory Authority:**
Safe Medical Devices Act of 1990, 21 CFR 807.92
**Company Name/Contract:**
**Company:**
CROSS® Medical Products, Inc.
5161-A Blazer Memorial Parkway
Dublin, Ohio 43017-1339
Phone: (614)718-0530
FAX: (614)718-0540
Medical Device Establishment
Registration Number: 1526354
**Contact:**
Regulatory Affairs Department
CROSS® Medical Products, Inc.
5161-A Blazer Memorial Parkway
Dublin, Ohio 43017-1339
Phone: (614)718-0530
FAX: (614)718-0540
**Device Name:**
**Trade Name:**
SYNERGY™ Posterior Spinal System
**Common Name:**
Posterior Spine Implants
Universal Spine System
**Classification Name:**
Appliance, Fixation, Spinal Interlaminal
Spondylolisthesis Spinal Fixation Device System
**Establishment Registration:**
**Number:** 1526354
{1}
01/03/97 FRI 17:13 FAX 614 718 0515
CROSS MEDICAL
038
## Classification:
The Orthopedic and Rehabilitation Devices Panel assigned the unique device classification code 87KWP (Interlaminal uses) to this device system. The published physical description of these devices is in 21 CFR, 888.3050. In addition, they are presently Class II medical devices. Class II medical devices are subject to Performance Standards. This device is also categorized under 87MNH (Spondylolisthesis Spinal Fixation), which is unclassified.
## Performance Standards:
Performance Standards applicable to the SYNERGY™ Posterior Spinal System have not been published by FDA. CROSS® Medical Products, Inc. produces this device according to the regulations and standards that are appropriate to the risk that Class II devices reasonably present. Voluntary performance standards, such as materials certifications, in-house SOP's and/or ASTM Standards are utilized as appropriate.
## Substantially Equivalent Devices(s):
1. TSRH® Spinal System
Danek Group
K901005, K901587, K904243,
K914306, K914966, K922520, K923242
2. PWB Lumbosacral Spinal System
CROSS® Medical
K920116, K924381
3. Harrington Spine System
Zimmer, Inc.
Preamendment Device
4. Danek/Sofamor Group
Dyna-Lok Spine System
K922617, K922767, K923239, K923241
## Device Description:
**Background:** This document summarizes safety and effectiveness information about the SYNERGY™ Posterior Spinal System contained in the 510(k) Premarket Notification submitted to FDA in support of its substantial equivalence.
**Purpose:** The SYNERGY™ Posterior Spinal System implants are intended to be used as a temporary construct that assists normal healing and are not intended to replace normal body structures.
{2}
01/03/97 FRI 17:14 FAX 614 718 0515
CROSS MEDICAL
039
They are intended to stabilize the spinal operative site during the development of a solid fusion with bone graft, and are intended to be removed after the development of a solid fusion mass.
The implantable portions of the SYNERGY™ Posterior Spinal System are made from surgical implant grade stainless steel, commercially pure titanium, and titanium alloy meeting ASTM specifications.
In the stainless steel system, the screws and set screws are made from 22-13-5 stainless steel (ASTM F-1314) and the balance of the implants are made from 316 LVM stainless steel (F-138-86, grade 2). Stainless steel alloy of these types are particularly well suited for use as surgical implants. In addition to excellent strength, fatigue and corrosion resistance characteristics, it has a long history of use in the human body as an implant material.
In the titanium system the screws, nuts, and set screws are made from 6Al-4V ELI titanium alloy (ASTM F-136-92). The rods are made from 6Al-4V ELI titanium alloy and commercially pure titanium (ASTM F-67-89, Grade 2). Both of these titanium types have excellent strength, fatigue, and corrosion resistance characteristics. Titanium devices are also useful as an alternative to devices made from stainless steel for selected patients with reactions or sensitivity to stainless steel implants. The surface finish of the titanium implants is tiodized type II, type III, and/or satin finished.
CAUTION: Mixing of dissimilar metals can accelerate the corrosion process. The titanium and stainless steel components of the SYNERGY™ Posterior Spinal System should not be used together. No components of the CROSS® Medical - SYNERGY™ Posterior Spinal System should be used with the components from any other system or manufacturer.
The surgical instruments used to assist in the implantation of the system are made from 17-4, 420, 440 and 455 series stainless steels (ASTM F-899-84 and A276-91). Established medical grade plastics (Ultem and Radel) are used to construct the handles, cases, etc.
The SYNERGY™ Posterior Spinal System is made up of the following basic components:
The CROSS ® Medical - SYNERGY™ Posterior Spinal System posterior application components are grouped as follows:
1. INTEGRAL™ Open, Closed, and Reduction screws, and angled closed, iliac, and variable locking sacral screws with variable locking seats, hex nuts, and set screws. Only the INTEGRAL™ Open, Closed, Reduction, and Variable Locking screws are intended for pedicle fixation. (See INDICATIONS)
2. Open and closed spinal hooks with sliders, c-rings, and set screws.
3. Adjustable and fixed transverse connectors with set screws.
{3}
01/13/97 MON 16:44 FAX 614 718 05.5
006
4. Closed and Axial rod connectors with set screws.
5. Lateral connectors with set screws.
6. Rods.
7. Instruments
8. Sterilization case(s).
Note: While the variable locking screws and some fasteners (nuts and set screws) are used for both the 6.35 mm and 4.75 mm rod sizes, the remaining components are designed for specific rod diameters.
Note: The 5.5 and 8.0 mm diameter INTEGRAL™ Open and Reduction Screws, as well as the Angled Closed Screws, and Iliac Screws are only for use with the 6.35 mm rod system.
## INDICATIONS:
The SYNERGY™ Posterior Spinal System, when using the screws as pedicle screws, is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar-first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the screws fixed or attached to the lumbar and sacral spine; and (d) who are having the device removed after the development of a solid fusion mass.
Except for situations where screws are attached to the pedicles of the lumbar and sacral spine via a posterior approach for the treatment of severe spondylolisthesis (Grade 3 and Grade 4) at the L5-S1 vertebral joint, the specific indications for the SYNERGY™ Posterior Spinal System are:
1. Degenerative Disc Disease (as defined by chronic back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies).
2. Idiopathic scoliosis.
3. Kyphotic deformities of the spine.
4. Paralytic scoliosis and/or pelvic obliquity treated in combination with anterior instrumentation.
5. Lordotic deformities of the spine.
6. Neuromuscular scoliosis associated with pelvic obliquity.
7. Vertebral fracture or dislocation.
8. Tumors.
9. Spondylolisthesis.
10. Stenosis.
11. Pseudarthrosis.
12. Unsuccessful previous attempts at spinal fusion.
{4}
01/03/97 FRI 17:15 FAX 614 718 0515
CROSS MEDICAL
041
For these indications, the SYNERGY™ screws and lateral connectors are intended for sacral/iliac attachment only. All of the SYNERGY™ hooks and transverse connectors are intended for posterior thoracic and/or lumbar use only. As a whole, the levels of use for the SYNERGY™ Posterior Spinal System are T1 to the Sacrum. The levels of screw fixation are L3 to ilium.
Note: The 4.75 mm diameter rod/hook constructs are intended only for patients weighing 70 pounds (32 kg) or less.
## CONTRAINDICATIONS:
Contraindications include, but are not limited to:
1. History of recent infection; systemic, spinal or localized.
2. Morbid obesity.
3. Mental illness.
4. Alcoholism or drug abuse.
5. Fever or leukocytosis.
6. Pregnancy.
7. Metal sensitivity/allergies to the implant materials.
8. Severe osteopenia.
9. Presence of congenital abnormalities, vague spinal anatomy, tumors, or any other condition which prevents secure implant screw fixation and/or decreases the useful life of the device. Any condition where the device will interfere with anatomical structures or physiological performance, including inadequate tissue coverage over the operational site.
10. For pedicle screw cases, missing or congenitally deformed pedicles of the fifth lumbar (L5) vertebrae.
11. Patients unwilling or unable to follow post-operative care instructions.
12. Any circumstances not described under the heading INDICATIONS.
## WARNINGS AND PRECAUTIONS:
### Warnings:
- When used as a pedicle screw system, this device system is intended only for Grade 3 or 4 spondylolisthesis at the fifth lumbar-first sacral (L5-S1) vertebral joint.
- The screws of this device system are not intended for insertion into the pedicles to facilitate spinal fusions above the L5-S1 vertebral joint.
- Benefit of spinal fusions utilizing any pedicle screw fixation system has not been adequately established in patients with stable spines.
- Potential risks identified with the use of this device system, which may require additional surgery, include:
- device component fracture,
{5}
01/03/97 FRI 17:15 FAX 614 718 0515
CROSS MEDICAL
042
loss of fixation,
non-union,
fracture of the vertebra,
neurological injury, and
vascular or visceral injury.
See Warnings and Precautions, and Potential Adverse Effects sections of the package insert for a complete list of potential risks.
In general, the SYNERGY™ Posterior Spinal System should only be implanted by surgeons fully experienced in the use of such implants and the required specialized surgery techniques. Even with the use of spinal implants, a successful result in terms of pain, function, or fusion is not always achieved in every surgical case.
**CAUTION:** Federal law (USA) restricts these devices to sale by or on the order of a physician.
This device system is not intended to be the sole means of spinal support. Its use without a bone graft or in cases that develop into a non-union will not be successful. No spinal implant can withstand the loads of the body without maturation of a solid fusion mass, and in this case, bending, loosening or fracture of the implant will eventually occur.
The proper selection and compliance of the patient will greatly affect the results. Patients who smoke have been shown to have an increased incidence of non-union. These patients should be advised of this fact and warned of the consequences. Other poor candidates for spine fusion include obese, malnourished, poor muscle and bone quality, and nerve paralysis patients.
**Pre-Operative:**
1. Only those patients who meet the criteria of one or more of the INDICATIONS and do not have any conditions included in the CONTRAINDICATIONS should be selected for implantation.
2. The implant components should be handled and stored carefully, protected from any damage, including corrosive environments.
3. The surgeon must confirm that all necessary implants and instruments are on hand for the planned surgical construct. Components from other manufacturers should not be combined with components of the SYNERGY™ Posterior Spinal System.
4. All implants and instruments must be unpacked, inspected for damage, cleaned and sterilized prior to use in the operative field.
**Intra-Operative:**
1. The Instrumentation Technique Manual should be carefully followed.
{6}
01/03/97 FRI 17:16 FAX 614 718 0515 CROSS MEDICAL 043
2. Extreme caution should be used around the spinal cord and nerve roots, especially when inserting screws or hooks.
3. Breakage, slippage, misuse, or mishandling of the instruments or implant components, such as on sharp edges, may cause injury to the patient or operative personnel.
4. The implant must be handled and contoured carefully so as to avoid notching or scratching the surface.
5. Before closing the soft tissues, all of the nuts and set screws should be tightened firmly according to the operative technique. Recheck the tightness of all nuts and set screws after finishing to ensure that none loosened during tightening or manipulation of the other implants.
6. Explanted implants must never be reused.
## Post-Operative
1. The surgeon must consider removing the implant after healing, as the implants can loosen, fracture or corrode even after fusion has occurred. The risks and benefits of a second surgery must be carefully evaluated.
2. The patient must be adequately instructed regarding the risks and limitations of the implant, as well as post-operative care and rehabilitation.
3. The patient should be instructed in the proper use of crutches, canes, external braces or any other weight bearing or assist devices that may be required, and limit those physical activities which would place excessive stresses on the implants or cause delay of the healing process. The patient should also be instructed in the proper methods to ambulate, climb stairs, get in and out of bed, and perform activities of daily living, while minimizing rotational and bending stresses.
## POSSIBLE ADVERSE EFFECTS:
Pre-operatively, the patient should be made aware of the following possible adverse effects of spinal implant surgery. Additional surgery may be necessary to correct some of these effects.
1. Bending, loosening, or fracture of the implants or instruments.
2. Metal sensitivity to a foreign body, including possible tumor formation, auto-immune disease, metallosis and/or scarring.
3. Skin or muscle sensitivity in patients with inadequate tissue coverage over the operative site, which may result in skin breakdown and/or wound complications.
4. Nonunion or delayed union.
5. Infection.
{7}
01/03/97 FRI 17:17 FAX 614 718 0515 CROSS MEDICAL 044
6. Nerve or vascular damage due to surgical trauma, including loss of neurological function, dural tears, radiculopathy, paralysis and cerebral spinal fluid leakage.
7. Gastrointestinal, urological, and/or reproductive system compromise, including sterility, impotency and/or loss of consortium.
8. Pain or discomfort.
9. Bone loss due to resorption or stress shielding, or bone fracture at, above or below the level of surgery.
10. Hemorrhage of blood vessels and/or hematomas.
11. Misalignment of anatomical structures, including loss of proper spinal curvature, correction, reduction and/or height.
12. Bursitis.
13. Bone graft donor site pain.
14. Inability to resume activities of normal daily living.
15. Death.
NOTE: The SYNERGY™ Posterior Spinal System Instrumentation Technique Manual should be carefully followed. It supplies important additional information on proper usage of the implants and instruments.
## PACKAGING:
All packages containing implants or instruments should be sealed and intact upon receipt. The product should not be used and should be immediately returned to CROSS² Medical Products, Inc., if the package or product is damaged.
## Sterilization/Resterilization:
All instruments and implants are supplied Non-Sterile. Non-Sterile instruments and implants are packaged in "clean only" condition. The labeling of the implants & instruments clearly indicates their sterility status. All packaging materials must be removed prior to sterilization. High temperature steam sterilization should be used, with the following cycle having been laboratory validated:
| Method: | Steam |
| --- | --- |
| Cycle: | Gravity |
| Temperature: | 250°F (121°C) |
| Exposure Time: | 60 minutes |
Note: It is recommended to dry and/or cool the parts to prevent condensation after the steam cycle.
{8}
01/03/97 FRI 17:17 FAX 614 718 0515
CROSS MEDICAL
045
Only sterile products should be used in the operative field. The recommended sterilization cycle is based on HIMA/AORN protocols. Other sterilization methods and cycles may also be suitable. However, individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques.
## INSTRUMENTATION:
Device specific instrumentation is necessary for the insertion and anchoring of the SYNERGY™ Posterior Spinal System. However, much of the surgical procedure is done with general orthopedic instrumentation of the appropriate size and type. SYNERGY™ Posterior Spinal System instruments are made from stainless steel meeting ASTM F899-84 and A276-91 standards.
## PRODUCT COMPLAINTS:
Any dissatisfaction with the product quality, labeling or performance should be reported to CROSS® Medical Products, Inc. immediately by the customer or user. Furthermore, if any of the implants "malfunction" (i.e. do not meet any of their performance specifications or otherwise do not perform as intended) and may have caused or contributed to the death or serious injury of the patient, CROSS® Medical Products, Inc. should be notified immediately by telephone, fax, or written correspondence.
When filing a complaint, the name, part number and lot number of the part should be provided along with the name and address of the person filing the complaint. In addition, the nature of the complaint should be clearly communicated along with a notification of whether a written report from CROSS® Medical Products, Inc. is requested.
## COMPARISON WITH PREDICATE DEVICE(S):
The SYNERGY™ Posterior Spinal System has similar design concept, materials, features, intended use and surgical approach as existing legally marketed posterior spine system such as the TSRH®, Harrington, Dyna-Lok, and PWB Lumbosacral Systems. All of these systems are used to treat similar or the same conditions, and all have essentially the same cautions and contraindications for use.
## CONCLUSION:
Based on the basic design concept, the use of well known materials, feature comparisons, mechanical testing, indications for use, surgical approach, preproduction quality assurance planning and engineering analysis CROSS® Medical Products, Inc. believes that sufficient evidence exists to reasonably conclude that this device is substantially equivalent to existing legally marketed posterior spine systems.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.