Anteralign Spinal System with Titan nanoLOCK Surface Technology
K214010 · Medtronic Sofamor Danek USA · MAX · Apr 12, 2022 · Orthopedic
Device Facts
Record ID
K214010
Device Name
Anteralign Spinal System with Titan nanoLOCK Surface Technology
Applicant
Medtronic Sofamor Danek USA
Product Code
MAX · Orthopedic
Decision Date
Apr 12, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Anteralign Spinal System with Titan nanoLOCK™ Surface Technology System interbody cages with macro-, micro-, and nano- roughened surface textures are intended to be used in spinal fusion procedures on skeletally mature patients with symptomatic Degenerative Disease (DDD, defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies), degenerative spondylolisthesis, at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion. These patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. These patients should have had six months of nonoperative treatment prior to treatment with this device. Additionally, the Anteralign Spinal System with Titan nanoLOCK™ Surface Technology can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The Anteralign Spinal System with Titan nanoLOCK™ Surface Technology is intended to be used with autograff and/or allogenic bone graft comprised of cancellous bone graft, and/or demineralized allograft bone with bone marrow aspirate or a combination thereof. The Anteralign TL Interbody must be used with a posterior supplemental internal spinal fixation cleared for use in the lumbar spine. Miniplate and bone screw components are provided as an option for the TL interbody for the lumbosacral levels obligue or lateral above the bifurcation (12-L5) of the vascular structures, Indications of spinal instrumentation systems should be understood by the surgeon. The Anteralian Spinal System with Titan nanoLOCK™ Surface Technology TL interbody may be implanted via a minimally invasive OLIF or minimally invasive or open DLIF approach. The Arteralian™ LS interbody cage as a stand-alone device or in conjunction with supplemental fixation. The Anteralian LS™ interbody fusion Device may be inserted via minimally invasive or open anterior or oblique approach at one or two contiguous levels from L2 to S1. These approaches include anterior and oblique. When used as a stand-alone device, the Anteralign™ LS cage must be used with 3 screws with devices that have standard lordosis (≤16 degrees). If the physician chooses to use less than 3 screws or none of the provided screws, additional supplemental fixation in the lumbar spine must be used to augment stability. When used in patients as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity conditions, additional supplemental fixation) must be used. Additionally, cages with lordosis angles greater than 16° are intended to be used with supplemental fixation (e.g. facet screws or posterior fixation)
Device Story
Anteralign Spinal System consists of additive-manufactured titanium interbody cages (TL and LS), miniplates, and bone screws; designed for lumbar interbody fusion (L2-S1). Cages feature Titan nanoLOCK™ surface technology (macro/micro/nano-roughened) to improve bone fixation. TL cages require posterior supplemental fixation; LS cages may be used standalone (with 3 screws if lordosis ≤16°) or with supplemental fixation. Implantation via minimally invasive or open OLIF/DLIF/ALIF approaches. Central cavity packed with autograft/allograft bone graft. Surgeon uses device to provide structural support and correction during fusion; honeycomb windows and open voids facilitate bone ingrowth. Benefits include stabilization of vertebral segments and promotion of fusion in degenerative conditions.
Clinical Evidence
Bench testing only. Mechanical evaluation performed per FDA guidance for spinal systems, including ASTM F2052, F2077, F2119, F2182, F2213, F2267, and expulsion testing. Results demonstrate equivalent mechanical strength to predicate devices.
Indicated for skeletally mature patients with symptomatic DDD, degenerative spondylolisthesis, or spinal stenosis at 1-2 contiguous levels (L2-S1) requiring interbody fusion; includes patients with up to Grade 1 Spondylolisthesis or retrolisthesis. Also indicated as adjunct for multilevel degenerative scoliosis and sagittal deformity. Requires 6 months nonoperative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Anteralign™ Spinal System with Titan nanoLOCK™ Surface Technology (K212524)
K212524 — Anteralign Spinal System with Titan nanoLOCK Surface Technology · Medtronic Sofamor Danek USA, Inc. · Dec 8, 2021
K231241 — NanoHive Medical Lumbar Interbody System · NanoHive Medical, LLC · Jul 12, 2023
K193320 — KMTI Tesera SA Anterior Lumbar Interbody Fusion (ALIF) System · Kyocera Medical Technologies, Inc. · Feb 4, 2020
K234087 — restor3d TIDAL Lumbar Interbody Fusion System · Restor3D · Jan 22, 2024
K240126 — ProAM ALIF System · Pro Surgical, Inc. · May 3, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
April 12, 2022
Medtronic Sofamor Danek USA Elizabeth Hamilton Senior Regulatory Affairs Specialist 1800 Pvramid Place Memphis, Tennessee 38132
Re: K214010
Trade/Device Name: Anteralign™ Spinal System with Titan nanoLOCK™ Surface Technology Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: MAX, OVD Dated: March 14, 2022 Received: March 15, 2022
Dear Elizabeth Hamilton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K214010
#### Device Name
Anteralign™ Spinal System with Titan nanoLOCK™ Surface Technology
Indications for Use (Describe)
Anteralign Spinal System with Titan nanoLOCK™ Surface Technology System interbody cages with macro-, micro-, and nano- roughened surface textures are intended to be used in spinal fusion procedures on skeletally mature patients with symptomatic Degenerative Disease (DDD, defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies), degenerative spondylolisthesis, at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion. These patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. These patients should have had six months of nonoperative treatment prior to treatment with this device. Additionally, the Anteralign Spinal System with Titan nanoLOCK™ Surface Technology can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The Anteralign Spinal System with Titan nanoLOCK™Surface Technology is intended to be used with autograff and/or allogenic bone graft comprised of cancellous bone graft, and/or demineralized allograft bone with bone marrow aspirate or a combination thereof.
The Anteralign TL Interbody must be used with a posterior supplemental internal spinal fixation cleared for use in the lumbar spine.
Miniplate and bone screw components are provided as an option for the TL interbody for the lumbosacral levels obligue or lateral above the bifurcation (12-L5) of the vascular structures, Indications of spinal instrumentation systems should be understood by the surgeon
The Anteralian Spinal System with Titan nanoLOCK™ Surface Technology TL interbody may be implanted via a minimally invasive OLIF or minimally invasive or open DLIF approach.
The Arteralian™ LS interbody cage as a stand-alone device or in conjunction with supplemental fixation. The Anteralian LS™ interbody fusion Device may be inserted via minimally invasive or open anterior or oblique approach at one or two contiguous levels from L2 to S1. These approaches include anterior and oblique. When used as a stand-alone device, the Anteralign™ LS cage must be used with 3 screws with devices that have standard lordosis (\$16 degrees). If the physician chooses to use less than 3 screws or none of the provided screws, additional supplemental fixation in the lumbar spine must be used to augment stability. When used in patients as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity conditions, additional supplemental fixation) must be used. Additionally, cages with lordosis angles greater than 16° are intended to be used with supplemental fixation (e.g. facet screws or posterior fixation)
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
#### MEDTRONIC Sofamor Danek USA, Inc.
Anteralign™ LS
December 21, 2021
| I. | Submitter | Medtronic Sofamor Danek, USA Inc.<br>1800 Pyramid Place<br>Memphis, Tennessee 38132 |
|------|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact: | Elizabeth Hamilton<br>Senior Regulatory Affairs Specialist<br>Email: elizabeth.c.hamilton@medtronic.com |
| | Date Prepared: | December 20, 2021 |
| II. | Subject Device | |
| | Name of Device: | Anteralign™ Spinal System with Titan nanoLOCK™ Surface<br>Technology |
| | Product Codes: | MAX, OVD |
| | Common name: | Interbody Cages, Instruments |
| | Classification Name: | Intervertebral Body Fusion Device with Bone Graft, Lumbar<br>(21 CFR 888.3080)<br>Intervertebral Body Fusion with intrinsic screws (21 CFR<br>888.3080) |
| | Classification | Class II |
| III. | Predicate Devices: | Primary Predicate: Anteralign™ Spinal System with Titan<br>nanoLOCK™ Surface Technology (K212524, S.E. 12/08/2021)<br>Additional Predicate 2: The Endoskeleton™ TAS Interbody<br>System (K211258, S.E. 05/26/2021)<br>Additional Predicate 3: The Endoskeleton™ TAS Hyperlordotic |
{5}------------------------------------------------
Additional Predicate 4: Artic-XL 3D Ti Spinal System with Tionic Technology (K171689, S.E. 10/05/17) Additional Predicate 5: Capstone PEEK (K073291, S.E. 04/24/2008, K123027, S.E. 07/25/2013) Additional Predicate 6: LT Cage™ PEEK Lumbar Tapered Fusion Device (P9700155022, S.E. 09/10/2003) Additional Predicate 7: Crescent PEEK, (K094025, S.E. 04/26/2010)
Additional Predicate 8: Capstone Spinal System PEEK, (K073291 S.E. 04/28/2013 & K123027, S.E. 07/25/2013)
The predicate devices were not subjected to any Recall. IV. Description: The Anteralign™ Spinal System with Titan nanoLOCK™ Surface Technology consists of TL and LS interbody cages, mini plates, and bone screws.
> Anteralign™ Spinal System TL and LS interbody cages are additive manufactured titanium cages available in various heights, widths, lengths, and lordotic angles to accommodate patient anatomy. The TL cage is rectangular shaped whereas the LS cage is oval shaped. The interbodies are inserted between two lumbar vertebral bodies to give support and correction during lumbar interbody fusion surgeries. The interbodies have a central cavity that allows them to be packed with autogenous bone and/or allograft bone graft comprised of cancellous and/or corticocancellous bone graft, and/or demineralized allograft bone with bone marrow aspirate.
> The Anteralign™ Spinal System TL interbody cages are provided sterile and are intended to be used with supplemental fixation cleared for use in lumbar spine (L2-S1).
{6}------------------------------------------------
The TL interbody may be implanted via a minimally invasive OLIF or minimally invasive or open DLIF approach. Mini plates and screws are provided as options for antimigration of the Anteralign TL interbody. The miniplate is additively manufactured from titanium powder with a machined-wrought titanium bolt. The miniplate may be positioned either laterally or obliquely and oriented in either cephalad or caudal direction on the TL cage. The bone screw, which is manufactured from wrought titanium, is then placed through the miniplate intrinsic screw hole. Miniplates and bone screws are offered in different sizes and are provided sterile. Miniplates are only to be used with the TL interbody.
The Anteralign™ LS interbody cage may be used as a standalone device or in conjunction with supplemental fixation. When used as a standalone device, the Anteralign LS interbody cage is intended to be used with three screws and must have standard lordosis (≤16 degrees). The Anteralign™ LS interbody cage is intended to be used in the lumbar sacral region between L2 and S1 and may be implanted via open or minimally invasive procedures for OLIF 51 or ALIF approaches.
The interbody designs incorporate honeycomb windows and an open void to allow bone growth through the implant. The interbody device is treated with Titan Surface Technology, where Titan nanoLOCK™ Surface Technology (MMN) is designed to improve fixation to the adjacent bone. The Titan nanoLOCK™ Surface Technology provides a microscopicroughened surface with nano-scale features. The Titan nanoLOCK™ Surface Technology is specifically engineered to have nano-textured features at a nanometer (10-9) level, which have demonstrated the ability to elicit an endogenous cellular
{7}------------------------------------------------
and biochemical response attributed to these nanotextured features in vitro. The Titan nanoLOCK™ Surface Technology demonstrates the elements to be considered a Nanotechnology as outlined in the FDA Nanotechnology Guidance.
Stainless steel and titanium implants are not compatible. They must not be used together in a construct.
Anteralign Spinal System with Titan nanoLOCK™ Surface Indications for use: Technology System interbody cages with macro-, micro-, and nano- roughened surface textured features are intended to be used in spinal fusion procedures on skeletally mature patients with symptomatic Degenerative Disc Disease (DDD, defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies), degenerative spondylolisthesis, and/or spinal stenosis, at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion. These patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. These patients should have had six months of nonoperative treatment prior to treatment with this device. Additionally, the Anteralign Spinal System with Titan nanoLOCK™ Surface Technology can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
V.
The Anteralign Spinal System with Titan nanoLOCK™Surface Technology is intended to be used with autograft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, and/or demineralized allograft bone with bone marrow aspirate or a combination thereof.
{8}------------------------------------------------
The Anteralign TL Interbody must be used with a posterior supplemental internal spinal fixation cleared for use in the lumbar spine.
Miniplate and bone screw components are provided as an option for anti-migration for the TL interbody for the lumbosacral levels oblique or lateral above the bifurcation (L2-L5) of the vascular structures. Indications and contraindications of spinal instrumentation systems should be understood by the surgeon
The Anteralign Spinal System with Titan nanoLOCK™ Surface Technology TL interbody may be implanted via a minimally invasive OLIF or minimally invasive or open DLIF approach.
The Anteralign™ LS interbody cage may be used as a standalone device or in conjunction with supplemental fixation. The Anteralign LS™ interbody fusion Device may be inserted via minimally invasive or open anterior or oblique approach at one or two contiguous levels from L2 to S1. These approaches include anterior and oblique. When used as a stand-alone device, the Anteralign™ LS cage is must be used with 3 screws with devices that have standard lordosis (≤16 degrees). If the physician chooses to use less than 3 screws or none of the provided screws, additional supplemental fixation in the lumbar spine must be used to augment stability. When used in patients as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity conditions, additional supplemental fixation (e.g., posterior fixation) must be used. Additionally, cages with lordosis angles greater than 16° are intended to be used with supplemental fixation (e.g. facet screws or posterior fixation).
{9}------------------------------------------------
Comparison of The subject Anteralign™ LS implants have the same, and /or VI. Technological similar features as the following predicate devices: Characteristics with the Primary Predicate)- Anteralign™ Spinal System with .
Predicate Devices:
- Titan nanoLOCK™ Surface Technology (K212524, S.E. 12/8/21) is the primary predicate, because the Anteralign™ LS devices are incorporated into this Spinal System. It serves as a predicate for identical intended identical materials, identical use, use, a sterilization method, identical levels of implantation, identical fundamental scientific technology, identical manufacturing method, and similar indications.
- Additional Predicate 2- The Endoskeleton™ TAS Interbody System (K211258, S.E. 05/26/21) serves as a predicate for identical intended use, identical indications, identical method of sterilization, identical levels of implantation, identical fundamental scientific technology and similar overall design and sizing.
- Additional Predicate 3 -The Endoskeleton™ TAS Hyperlordotic with nanoLOCK™ (K211258, S.E. 05/26/21) serves as a predicate for identical intended use, identical indications, identical method of sterilization, identical levels of implantation, identical fundamental scientific technology and similar overall design and sizing.
- Additional Predicate 4 - Artic-XL 3D Ti Spinal System with Tionic Technology (K171689, S.E. 10/05/17) was chosen as a predicate due to identical intended use, identical materials, identical method of sterilization,
{10}------------------------------------------------
identical surgical approach, identical fundamental technology, similar sizing and similar indications.
- Additional Predicate 5 -Capstone Peek (K073291, S.E 4/24/08, K123027, S.E. 7/25/13) serves as a mechanical test predicate.
- Additional Predicate 6 LT Cage™ PEEK Lumbar . Tapered Fusion Device, P9700155022, S.E 9/10/2003 serves as a mechanical test predicate.
- . Additional Predicate 7- Crescent PEEK, (K094025, S.E.4/26/10) serves as a mechanical test predicate.
- Additional Predicate 8- Capstone Spinal System PEEK, (K073291 S.E 4/28/13 & K123027, S.E. 7/25/13) serves as a mechanical test predicate.
Both Subject and Predicate devices are based on the same technological characteristics of providing support and correction by surgically implanting fusion devices between lumbar or lumbosacral vertebral bodies
#### VII. Performance Data: Mechanical Testing:
In accordance with, "Guidance for Industry and FDA Staff – Spinal System 510(k)'s", Medtronic has evaluated the subject devices to demonstrate substantial equivalence to the predicate devices. Medtronic completed mechanical testing on the devices according to the following industry standards.
- •
- . ASTM F2052
{11}------------------------------------------------
- ASTM F2077 .
- ASTM F2119 ●
- ASTM F2182
- ASTM F2213
- ASTM F2267 ●
- . Expulsion Testing
This evaluation has demonstrated that the subject devices have equivalent mechanical strength in comparison to the predicate devices.
Based on the supporting information provided in this pre-VIII. Conclusion: market notification, the subject Anteralign™ LS devices are substantially equivalent to the identified predicates.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.