Choice Spine Lumbar Spacer System (Sabre™, Shark™, Hornet™, Harpoon™), Choice Spine VEO™ Lateral Access & Interbody Fusion System, Choice Spine Interbody Fusion System

K153107 · Choicespine, LP · MAX · May 19, 2016 · Orthopedic

Device Facts

Record IDK153107
Device NameChoice Spine Lumbar Spacer System (Sabre™, Shark™, Hornet™, Harpoon™), Choice Spine VEO™ Lateral Access & Interbody Fusion System, Choice Spine Interbody Fusion System
ApplicantChoicespine, LP
Product CodeMAX · Orthopedic
Decision DateMay 19, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Choice Spine Lumbar Spacer System (Sabre, Shark, Hornet, Harpoon): The Choice Spine Lumbar Spacer System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1 in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. When used as a vertebral body replacement device, the Choice Spine Intervertebral Body Device is intended for use in the thoracic and lumbar spine, from T1 to L5, for the replacement of a collapsed or unstable vertebral body resulting from a tumor or traumatic injury. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. Choice Spine VEO™ Lateral Access & Interbody Fusion System: The Choice Spine VEO™ Lateral Access & Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The Choice Spine VEO™ Lateral Access & Interbody Fusion System is designed to be used with autogenous graft and supplemental spinal fixation that is cleared for use in the lumbar spine. Choice Spine Interbody Fusion System: The Choice Spine Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The Choice Spine Interbody Fusion System is designed to be used with autogenous bone graft and a supplemental spinal fixation system that is cleared for use in the lumbar spine.

Device Story

Spinal interbody fusion and vertebral body replacement implants; families include Sabre, Shark, Hornet, Harpoon, VEO, and Interbody Fusion System. Implants feature hollow centers for bone graft placement; rectangular shapes; various heights/lengths/angulations. Delivered via posterior, transforaminal, lateral, anterior, or anterolateral surgical approaches. Used by orthopedic or neurosurgeons in clinical settings to stabilize spine, facilitate fusion, or replace collapsed/unstable vertebral bodies. Implants provide structural support; used with supplemental fixation and autograft. Benefit: stabilization of spinal segments, pain reduction, and restoration of vertebral height.

Clinical Evidence

No clinical data. Bench testing only; engineering analysis confirmed that the worst-case design for each implant family remains unchanged by the use of titanium-alloy versus PEEK.

Technological Characteristics

Materials: PEEK-Optima (ASTM F2026), Tantalum markers (ASTM F560), Ti-6Al-4V ELI (ASTM F136), stainless steel, aluminum, Radel R. Form factor: Rectangular spacers with hollow centers. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1) or vertebral body replacement (T1-L5) for tumor/trauma. DDD defined as discogenic back pain with degeneration confirmed by history/radiographs; may include Grade 1 spondylolisthesis or retrolisthesis. Requires 6 months non-operative treatment. Contraindications: none stated.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing right, arranged in a way that suggests unity and connection. The profiles are black and are set against a white background. Encircling the profiles are the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 19, 2016 Choice Spine, LP Ms. Kim Finch Manager of Regulatory Affairs 400 Erin Drive Knoxville, Tennessee 37919 Re: K153107 Trade/Device Name: Choice Spine Lumbar Spacer System (Sabre™, Shark™, Hornet™, Harpoon™), Choice Spine VEO™ Lateral Access & Interbody Fusion System, Choice Spine Interbody Fusion System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, MQP Dated: April 22, 2016 Received: April 25, 2016 Dear Ms. Finch: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing {1}------------------------------------------------ (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K153107 #### Device Name Choice Spine Lumbar Spacer System (Sabre™, Shark™, Hornet™, Harpoon™) Choice Spine VEOTM Lateral Access & Interbody Fusion System Choice Spine Interbody Fusion System #### Indications for Use (Describe) Choice Spine Lumbar Spacer System (Sabre, Shark, Hornet, Harpoon): The Choice Spine Lumbar Spacer System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1 in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. When used as a vertebral body replacement device, the Choice Spine Intervertebral Body Device is intended for use in the thoracic and lumbar spine, from T1 to L5, for the replacement of a collapsed or unstable vertebral body resulting from a tumor or traumatic injury. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. ### Choice Spine VEOTM Lateral Access & Interbody Fusion System: The Choice Spine VEO™ Lateral Access & Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The Choice Spine VEO™ Lateral Access & Interbody Fusion System is designed to be used with autogenous graft and supplemental spinal fixation that is cleared for use in the lumbar spine. #### Choice Spine Interbody Fusion System: The Choice Spine Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The Choice Spine Interbody Fusion System is designed to be used with autogenous bone graft and a supplemental spinal fixation system that is cleared for use in the lumbar spine. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary | Date | May 17, 2016 | |----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sponsor: | Choice Spine, LP<br>400 Erin Drive<br>Knoxville, TN 37919 | | Phone: | 865-246-3333 | | Fax: | 865-246-3334 | | Contact Person: | Kim Finch, Manager of Regulatory Affairs | | Proposed<br>Proprietary Trade<br>Name: | Choice Spine Lumbar Spacer System (Sabre™, Shark™, Hornet™, Harpoon™)<br>Choice Spine VEO™ Lateral Access & Interbody Fusion System<br>Choice Spine Interbody Fusion System | | Product Class: | Class II | | Classification<br>Name: | Choice Spine Lumbar Spacer System (Sabre™, Shark™, Hornet™, Harpoon™):<br>888.3080- Spinal Intervertebral Body Fusion Device 888.3060- Spinal Intervertebral Body Fixation Orthosis Choice Spine VEO™ Lateral Access & Interbody Fusion System:<br>888.3080- Spinal Intervertebral Body Fusion Device Choice Spine Interbody Fusion System:<br>888.3080- Spinal Intervertebral Body Fusion Device | | Device Product<br>Code: | Choice Spine Lumbar Spacer System (Sabre™, Shark™, Hornet™, Harpoon™):<br>MAX, MQP Choice Spine VEO™ Lateral Access & Interbody Fusion System:<br>MAX Choice Spine Interbody Fusion System:<br>MAX | | Purpose of<br>Submission: | The purpose of this submission is to gain clearance for additional devices<br>manufactured from titanium-alloy. The previously cleared system are Choice<br>Spine Lumbar Spacer System, K140142 and K073669, Choice Spine VEO™<br>Lateral Access & Interbody Fusion System (Formerly TranS1 Lateral Interbody<br>Fusion System) K100210 and K123997, and Choice Spine Interbody Fusion | TranS1 systems were acquired by Choice Spine in 2015. {5}------------------------------------------------ Device Description: This submission includes three lumbar implant families. The device description for each family is listed below. Choice Spine Lumbar Spacer System (Sabre, Shark, Hornet, Harpoon): The Choice Spine Lumbar Spacer System consists of interbody fusion devices (specifically, SABRE, SHARK™, HORNET™ & HARPOON™ Lumbar Spacers) comprised of polyetheretherketone (PEEK - Optima®) with tantalum markers (ASTM F2026 and ASTM F560) or Ti-6AI-4V ELI (ASTM F136). The spacers have a basic rectangular shape, a hollow center for placement of bone graft and a smooth bullet shaped anterior surface. They are available in an assortment of height, length and anteroposterior angulation combinations to accommodate many different anatomic requirements. These implants are delivered via a posterior or transforaminal approach. Choice Spine VEO™ Lateral Access & Interbody Fusion System: The Choice Spine VEO™ Lateral Access & Interbody Fusion System is a multicomponent system including instrumentation made of stainless steel. Aluminum. and Radel R and implants made of tantalum (ASTM F560) and PEEK Optima® (ASTM F2026) or Ti-6Al-4V ELI (ASTM F136). The implants are delivered via lateral approach. #### Choice Spine Interbody Fusion System: The Choice Spine Interbody Fusion System is a family of implants intended to aid in spinal fixation of the lumbar spine. This system includes implants made of Zeniva ZA-500 PEEK (ASTM F2026) with Tantalum markers (ASTM F560) or Ti-6Al-4V ELI (ASTM F136), which are delivered via anterior or anterolateral approach. #### Intended Use: Choice Spine Lumbar Spacer System (Sabre, Shark, Hornet, Harpoon): The Choice Spine Lumbar Spacer System is indicated for intervertebral body fusion of the lumbar spine, from L2 to S1 in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiquous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. When used as a vertebral body replacement device, the Choice Spine Intervertebral Body Device is intended for use in the thoracic and lumbar spine, from T1 to L5, for the replacement of a collapsed or unstable vertebral body resulting from a tumor or traumatic injury. The device system is designed for use with supplemental fixation and with autograft to facilitate fusion. #### Choice Spine VEO™ Lateral Access & Interbody Fusion System: The Choice Spine VEO™ Lateral Access & Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with {6}------------------------------------------------ degenerative disc disease (DDD) at one or two contiquous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondvlolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The Choice Spine VEO™ Lateral Access & Interbody Fusion System is designed to be used with autogenous graft and supplemental spinal fixation that is cleared for use in the lumbar spine. Choice Spine Interbody Fusion System: The Choice Spine Interbody Fusion System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. The Choice Spine Interbody Fusion System is designed to be used with autogenous bone graft and a supplemental spinal fixation system that is cleared for use in the lumbar spine. Comparison of The implants included in this submission are equivalent to the primary predicate Technological Alphatec Battalion Universal Spacer System (K143740), and additional predicates Characteristics: are Choice Spine Lumbar Spacer System (K140142 and K073669), Choice Spine Interbody Fusion System (K120991), and the Choice Spine VEO™ Lateral Access & Interbody Fusion Device System (K100210 and K123997). Non-Clinical Testing: The worst case design for each family does not change if the implant was manufactured from Titanium instead of PEEK. Therefore, no new non-clinical testing is needed. Conclusion: The subject implants are substantially equivalent in intended use, anatomic location, indications, method of stabilization, and material of manufacture.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...