CHOICE SPINE LUMBAR SPACER SYSTEM (SABRE, SHARK, HORNET, HARPOON)
Applicant
Choicespine, LP
Product Code
MAX · Orthopedic
Decision Date
Jul 16, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used as an intervertebral body fusion device, Choice Spine Lumbar Spacers are indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation (e.g., ORIA Claris, ORIA Spinal Clip System, etc.) and with autograft to facilitate fusion. When used as a vertebral body replacement device, Choice Spine Lumbar Spacers are intended for use in the thoracic and lumbar spine, from T1 to L5, for the replacement of a collapsed or unstable vertebral body resulting from a tumor or traumatic injury. The device system is designed for use with supplemental fixation (e.g., ORIA Claris, ORIA Spinal Clip System, etc.) and with autograft to facilitate fusion.
Device Story
Spinal spacer system (Sabre, Shark, Hornet, Harpoon) provides structural support for intervertebral fusion or vertebral body replacement. Rectangular, hollow-center implants accommodate autograft bone to facilitate fusion. Used in thoracic and lumbar spine; requires supplemental fixation (e.g., ORIA Claris/Spinal Clip). Implanted by surgeons during spinal procedures. Tantalum pins provide radiographic visualization. Benefits include stabilization of collapsed/unstable vertebral bodies or discogenic back pain relief.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical performance evaluated via static axial compression, bending, torsion, and dynamic axial compression/torsion testing per ASTM F2077-11 and ASTM F2267-04 standards.
Technological Characteristics
Materials: PEEK OPTIMA LT1 (ASTM F2026) for spacers; 17-4 Stainless Steel (ASTM F899) for trials. Tantalum marking pins (ASTM F560). Rectangular, hollow-center geometry with varying heights/lengths/lordosis. Non-active, implantable device.
Indications for Use
Indicated for skeletally mature patients requiring lumbar intervertebral fusion (L2-S1) for degenerative disc disease with up to Grade I spondylolisthesis after 6 months of failed non-operative treatment, or for thoracic/lumbar (T1-L5) vertebral body replacement following tumor or trauma.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
K140142 page 1 of 2
## JUL 1 6 2014
510(k) Summary for the 9 mm Wide Choice Spine Lumbar Spacer System
Date Prepared: July 14, 2014
Submitter Contact: Choice Spine, LP. 400 Erin Drive, Knoxville, TN 37919 Tel: 865.246.3333 Fax: 865.246.3334
Regulatory Contact: Kim Finch, Manager of Regulatory Affairs kfinch@choicespine.net
Device Name: Spinal Vertebral Body Replacement Device Intervertebral Fusion Device with Bone Graft, Jumbar
TradeName: Choice Spine Lumbar Spacer System (Sabre, Shark, Hornet, Harpoon)
Product Class: Class II
Requlation Number 888.3080 Intervertebral Body Fusion Device
888.3060 Spinal Intervertebral body fixation orthosis
Product Codes: MAX, MQP
Panel Code: 87
Indications for Use:
When used as an intervertebral body fusion device, Choice Spine lumbar Spacers are indicated for intervertebral body fusion of the lumbar spine, from L 2 - S 1 , in skeletally mature patients who have had six months of nonoperative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation systems such as (ORIA Claris, ORIA Spinal Clip System, etc.).
When used as a vertebral body replacement device, Choice Spine lumbar Spacers are intended for use in the thoracic and lumbar spine, from T1 to L5 , for the replacement of a collapsed or unstable vertebral body resulting from a turnor or traumatic injury. The device system is designed for use with supplemental fixation (ORIA Claris, ORIA Spinal Clip System, etc.) and with autograft bone to facilitate fusion.
{1}------------------------------------------------
K140142 page 2 of 2
Device Descriptions:
The 9mm width spacer system (SABRE, SHARK, HORNET, HARPOON) has a rectangular shape, a hollow center for placement of autograft bone, and a smooth anterior surface (nose). It is available in a variety of heights, length, and lordosis combinations with and without medial-lateral curvature to accommodate many different anatomic requirements.
#### Materials:
The 9mm width spacers are manufactured from PEEK OPTIMA LT1(Invibio) per ASTM F2026. The spacer includes tantalum marking pins per ASTM F560 for radiographic assistance. These technological characteristics and materials are identical to the predicate devices.
The 9mm width trials are manufactured from 17-4 Stainless Steel per ASTM F899 and are biocompatible.
### Predicate Device(s):
K2M, LLC Aleutian Spacer System K051454, K130699 Innovasis, Inc., Box Peek VBR System K062151 Choice Spine Lumbar Spacer System, K073669
#### Performance Standards:
Performance testing was completed by an independent laboratory following ASTM F2077-11 and ASTM F2267-04. The tests included static axial compression bending, static torsion, dynamic axial compression bending and dynamic torsion testing. The test methods and results are substantially equivalent to those of the predicate devices.
#### Conclusion:
The 9 mm wide spacer has the same technological characteristics, intended use, comparable performance data, and is manufactured using the same processes as the predicate devices. The sterilization method and parameters remain the same. In conclusion, the 9 mm wide devices are substantially equivalent to the listed predicates and do not raise new questions of safety or effectiveness.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 16, 2014
Choice Spine, LP Ms. Kim Finch Manager of Regulatory Affairs 400 Erin Drive Knoxville, Tennessee 37919
Re: K140142
Trade/Device Name: Choice Spine Lumbar Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, MQP Dated: June 13, 2014 Received: June 16, 2014
Dear Ms. Finch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Kim Finch
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use Statement K140142
The Company's Indications for Use Statement for Choice Spine Lumbar Spacer System is provided.
Indications for Use:
When used as an intervertebral body fusion device, Choice Spine Lumbar Spacers are indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device system is designed for use with supplemental fixation (e.g., ORIA Claris, ORIA Spinal Clip System, etc.) and with autograft to facilitate fusion.
When used as a vertebral body replacement device, Choice Spine Lumbar Spacers are intended for use in the thoracic and lumbar spine, from T1 to L5, for the replacement of a collapsed or unstable vertebral body resulting from a tumor or traumatic injury. The device system is designed for use with supplemental fixation (e.g., ORIA Claris, ORIA Spinal Clip System, etc.) and with autograft to facilitate fusion.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
n E. Dmitrie Division of
Concurrence of CDRH, Office of Device Evaluation (ODE)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.