K171181 · Microport Orthopedics, Inc. · LZO · Aug 28, 2017 · Orthopedic
Device Facts
Record ID
K171181
Device Name
PROCOTYL® PRIME E-CLASS XLPE Liner
Applicant
Microport Orthopedics, Inc.
Product Code
LZO · Orthopedic
Decision Date
Aug 28, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The PROCOTYL® PRIME E-CLASS™ XLPE Liner is intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients. This device is indicated for the following conditions: 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankyloses, protrusion acetabuli, and painful hip dysplasia; 2) inflammatory degenerative joint disease such as rheumatoid arthritis; 3) correction of functional deformity; and, 4) revision procedures where other treatments or devices have failed. Shells with BIOFOAM® metal foam coating are intended only for uncemented arthroplasty.
Device Story
Acetabular liner component for total hip arthroplasty; used with PROCOTYL® PRIME Acetabular Cup System. Manufactured from vitamin E blended highly crosslinked ultra-high molecular weight polyethylene (VEXLPE). Available in standard, lipped, or face-changing lateralized configurations; inner diameters 22mm to 44mm. Implanted by orthopedic surgeons in clinical settings. Functions as bearing surface in hip prosthesis; reduces pain and improves joint function. Mechanical integrity verified via fatigue, push-out, lever-out, torque-out, and wear testing. Biocompatibility confirmed via extractables, cytotoxicity, irritation, and sensitization studies.
Clinical Evidence
No clinical data provided. Evidence based on non-clinical bench testing: material properties, bacterial endotoxins, push-out/lever-out/torque-out lock testing, frictional torque, axial cyclic loading, impingement testing, wear rate/debris analysis, and biocompatibility (extractables, cytotoxicity, irritation, sensitization).
Technological Characteristics
Material: Vitamin E blended highly crosslinked ultra-high molecular weight polyethylene (VEXLPE) per ASTM F2695-12. Form factor: Acetabular liner (standard, lipped, lateralized). Dimensions: 22mm-44mm ID. Mechanical interface: Locking mechanism for acetabular shells. Sterilization: Not specified.
Indications for Use
Indicated for skeletally mature patients requiring total hip arthroplasty for non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankyloses, protrusion acetabuli, painful hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity correction, or revision procedures.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
K173583 — Exactech® Novation® and AcuMatch® E-HXL Acetabular Liners · Exactech, Inc. · Mar 12, 2018
K070399 — 100KGY E-POLY ACETABULAR LINERS-ADDITIONAL PROFILES · Biomet Manufacturing, Inc. · May 4, 2007
K130365 — X-ALT HIGHLY CROSS LINKED VE ACETABULAR LINER · Encore Medical L.P. · Sep 23, 2013
K130652 — VITALITE E CROSS-LINKED POLYETHYLENE ACETABULAR INSERT · Consensus Orthopedics, Inc. · Dec 17, 2013
K181598 — Prime E-CLASS XLPE Liner · Microport Orthopedics, Inc. · Sep 14, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
MicroPort Orthopedics, Inc. Ms. Sarah Stroupe Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002
August 13, 2019
Re: K171181
Trade/Device Name: PROCOTYL® PRIME E-CLASS™ XLPE Liner Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, OQI, LPH, OQG Dated: July 28, 2017 Received: August 1, 2017
Dear Ms. Stroupe:
This letter corrects our substantially equivalent letter of August 28, 2017.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent for the indications for use stated in the enclosure to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Vesa Vuniqi Date: 2019.08.13 -S - - -15:02:35 -04'00'
Vesa Vuniqi Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K171181
Device Name PROCOTYL® PRIME E-CLASS™ XLPE Liner
#### Indications for Use (Describe)
PROCOTYL® PRIME hip system is intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.
Indications for Use
1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankyloses, protrusion acetabuli. and painful hip dysplasia:
2) inflammatory degenerative joint disease such as rheumatoid arthritis;
3) correction of functional deformity; and,
4) revision procedures where other treatments or devices have failed.
Shells with BIOFOAM® metal foam coating are intended only for uncemented arthroplasty.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners, inside of which is a white circle containing a red diamond shape. To the right of the square is the word "MicroPort" in blue, with the word "Orthopedics" in red underneath. Below that is the phrase "Full Function, Faster" in blue.
# 510(k) Summary
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Substantial Equivalence for the PROCOTYL® PRIME E-CLASS™ XLPE Liner.
| Submitted by: | MicroPort Orthopedics Inc.<br>5677 Airline Rd, Arlington TN, 38002<br>Phone: 866-872-0211<br>Fax: 855-446-2247 |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Date: | August 25, 2017 |
| Contact Person: | Sarah Evonne Stroupe<br>Regulatory Affairs Specialist II |
| Proprietary Name: | PROCOTYL® PRIME E-CLASS™ XLPE Liner |
| Common Name: | Acetabular Liner |
| Classification Name and Reference: | 21 CFR 888.3353 LZO/OQI<br>Hip joint metal/ceramic/polymer semi constrained<br>cemented or nonporous, uncemented prosthesis<br>Class II |
| Classification Name and Reference: | 21 CFR 888.3358 LPH/OQG<br>Hip joint metal/polymer/metal semi-constrained<br>porous-coated uncemented prosthesis<br>Class II |
| Subject Product Code and Panel Code: | Orthopedics/87/LZO/LPH/OQG/OQI |
| Primary Predicate Device: | PROCOTYL® PRIME Acetabular<br>Cup<br>System (K170444) |
| Additional Predicate Devices: | LINEAGE® Acetabular System (K002149 and<br>K052026) |
| Additional Predicate Devices: | DYNASTY® Acetabular System (K082924,<br>K061547, and K070785) |
| Reference Device: | StelKast's EXp Acetabular Liner (K094035, Models<br>SC3342 Non-Hooded 28mm, SC3343 Hooded<br>28mm) |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for MicroPort Orthopedics. The logo features a blue square with rounded corners, inside of which is a white circle containing a red four-pointed star. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. Below that is the text "Full Function, Faster" in blue.
# DEVICE INFORMATION
#### A. Intended Use
The PROCOTYL® PRIME E-CLASS™ XLPE Liner is intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients. This device is indicated for the following conditions:
- 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankyloses, protrusion acetabuli, and painful hip dysplasia;
- 2) inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3) correction of functional deformity; and,
- 4) revision procedures where other treatments or devices have failed.
Shells with BIOFOAM® metal foam coating are intended only for uncemented arthroplasty.
## B. Device Description
The PROCOTYL® PRIME E-CLASS™ XLPE Liner is an additional liner option for MicroPort Orthopedics' PROCOTYL® PRIME Acetabular Cup System (K170444). The design features are summarized below:
- Manufactured from E-CLASS™ vitamin E blended highly crosslinked ultra high . molecular weight polyethylene (VEXLPE) conforming to ASTM F2695-12
- Available in Standard, Lipped, or Face-Changing Lateralized Configurations .
- Inner Diameter sizes 22mm to 44mm o
## C. Substantial Equivalence Information
The design features and materials of the subject devices are substantially equivalent to those of the predicate and reference devices. The indications for use are identical to the predicate and reference devices. The fundamental scientific technology of the modified devices has not changed relative to the predicate and reference devices. The safety and effectiveness of the subject devices are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.
## D. Nonclinical Testing
The following nonclinical testing was performed to evaluate the E-CLASS™ material and design of the finished product.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for MicroPort Orthopedics. The logo features a blue square with rounded corners, inside of which is a white circle containing a red diamond shape. To the right of the square is the text "MicroPort" in blue, with "Orthopedics" in red below it. Underneath the company name is the tagline "Full Function, Faster" in a smaller, blue font.
The subject Acetabular Liner underwent material properties testing to compare its properties to those of currently marketed UHMWPE.
The subject Acetabular Liner was evaluated for bacterial endotoxins and found to be less than the USP endotoxin limit of 20 EU/device.
The subject Acetabular Liner underwent pre- and post-fatigue push-out, lever-out, and torque-out testing to evaluate its lock detail.
The subject Acetabular Liner underwent frictional torque testing to evaluate properties under pinch-loading conditions.
The subject Acetabular Liner complete long term axial cyclic loading to evaluate the subject Acetabular Liner's mechanical integrity and fatigue resistance properties.
The subject Acetabular Liner underwent impingement testing to evaluate its mechanical integrity during lip impingement in a fatigue loading profile.
The subject Acetabular Liner underwent wear testing to determine the wear rate. A wear particle analysis was performed on the wear debris.
Additionally, a range of motion study was performed for the PROCOTYL® PRIME Acetabular Cup System, originally submitted as part of K170444.
## E. Clinical Testing
Clinical data was not provided for the subject devices.
## F. Biocompatibility
Both bench and animal studies were performed to evaluate the biocompatibility of the E-CLASS™ material.
The E-CLASS™ material underwent an extractables study to determine it's the irritant potential of the material and to compare this potential to currently marketed Vitamin E UHMWPE.
The E-CLASS™ material underwent cytotoxicity testing, performed using cell monolayers, to determine the irritant potential of the material.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners, inside of which is a white circle containing a red diamond shape. To the right of the square is the word "MicroPort" in a large, bold, blue font. Below that is the word "Orthopedics" in a smaller, red font, and below that is the phrase "Full Function, Faster" in a smaller, blue font.
The E-CLASS™ material underwent irritation and sensitization studies, performed on animal specimens, to determine the irritant and allergenic potential of the material.
No biocompatibility testing was performed for the associated instrumentation, as all associated instruments have been previously cleared.
## G. Component and Accessory Compatibility
The PROCOTYL® PRIME E-CLASS™ XLPE Liner may be used with PROCOTYL® PRIME Acetabular Cup System Acetabular Shells and Instrumentation (K170444).
The PROCOTYL® PRIME E-CLASS™ XLPE Liners may be used with previously cleared MicroPort Femoral Heads.
Table 1 shows the compatibility of the subject Acetabular Liner with previously cleared MicroPort Orthopedics products.
| 510(k) | Intended Combinations |
|---------------------------------|--------------------------------------------------------|
| with Femoral Heads (up to 44mm) | |
| K893685 | Ceramic-Polyethylene with ID 28-36mm** |
| K920593 | Ceramic-Polyethylene, OD 28mm |
| K925512 | Ceramic-Polyethylene, OD 28mm |
| K932222 | Metal-Polyethylene, OD 28mm XXL |
| K002149 | Metal-Polyethylene with ID 22-32mm |
| K021349* | Metal-Polyethylene with ID 38-56mm |
| K004043* | Metal-Polyethylene with ID 28-36mm |
| K051348* | Metal-Polyethylene with ID 38-56mm |
| K072656 | Ceramic-Polyethylene with ID 38-46mm with neck sleeves |
| K130376 | Ceramic-Polyethylene with ID 32-40mm |
| K140043 | Ceramic-Polyethylene with ID 28mm |
| with Acetabular Shells | |
| K170444 | Quad and Solid Acetabular Shells |
| with Instrumentation | |
| K170444 | Acetabular Liner Trials |
#### Table 1: Subject Acetabular Liners Compatibility
*Metal femoral heads in K021349, K004043 and K051348 were originally cleared for use with metal-metal bearings, and later cleared for compatibility with DYNASTY® A-CLASS® polyethylene liners in K070785.
**36mm Forte Ceramic heads were originally cleared for use in PMA P030027, and later cleared for compatibility with PROCOTYL® L / O Acetabular System in K142119.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners, inside of which is a white circle containing a red star-like shape. To the right of the square is the word "MicroPort" in blue, with the word "Orthopedics" in red underneath. Below that is the phrase "Full Function. Faster."
## H. Conclusion
The indications for use and fundamental scientific technology of the PROCOTYL® PRIME E-CLASS™ XLPE Liner are similar to the predicate and reference devices. The materials of the subject device are considered to be substantially equivalent to the predicate and reference devices. The design features, materials information, predicate and subject testing, and analysis provided in the Master File support the substantial equivalence of the PROCOTYL® PRIME E-CLASS™ XLPE Liner.
MicroPort Orthopedics BV Hoogoorddreef 5, 1101 BA Amsterdam, The Netherlands +31 20 545 01 00
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.