LINEAGE ACETABULAR SYSTEM

K002149 · Wrightmedicaltechnologyinc · LPH · Aug 31, 2000 · Orthopedic

Device Facts

Record IDK002149
Device NameLINEAGE ACETABULAR SYSTEM
ApplicantWrightmedicaltechnologyinc
Product CodeLPH · Orthopedic
Decision DateAug 31, 2000
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia: 2. inflammatory degenerative joint disease such as rheumatoid arthritis; correction of functional deformity; 3. revision procedures where other treatments or devices have failed, and, treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

Device Story

LINEAGE™ Acetabular System is a modular hip prosthesis component set for uncemented total hip arthroplasty. System comprises hemispherical titanium alloy shells (with/without 14° peripheral rim flare) and UHMWPE acetabular liners. Shells feature CP titanium plasma spray or porous bead coatings; options include solid, quadrant, or multi-hole configurations with threaded apical hole plugs. Liners include 0° or 15° overhangs and optional 4+mm lateralized shift; internal geometry accommodates existing cobalt chrome or ceramic femoral heads via WMT12/14, Orthomet®, or A-line tapers. Liner locks into shell via 18° male taper and peripheral ring assembly. Device used by orthopedic surgeons in clinical/OR settings to replace diseased hip joint surfaces, aiming to reduce pain and restore function. No software or electronic components.

Clinical Evidence

No clinical data provided; substantial equivalence supported by bench testing and materials data.

Technological Characteristics

Materials: Titanium alloy shell, UHMWPE liner, CP titanium plasma spray or porous bead coating. Design: Hemispherical shell with 14° flare option; 0°/15° liner overhangs; 4+mm lateralized shift option. Interface: 18° male taper and peripheral ring locking mechanism. Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients undergoing total hip arthroplasty for non-inflammatory/inflammatory degenerative joint disease (e.g., osteoarthritis, rheumatoid arthritis, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), functional deformity correction, revision procedures, or treatment of nonunion, femoral neck, and trochanteric fractures of the proximal femur with head involvement.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a logo with the text "AUG 31 2000" above it. The logo is a stylized "W" shape. To the right of the logo, the text "WRIGHT MEDICAL TECHNOLOGY" is displayed, along with the address "5677 AIRLINE ARLINGTON, TN" and the phone number "901-867-99". Image /page/0/Picture/1 description: The image shows the number 000012. The number is printed in a bold, sans-serif font. The numbers are slightly distressed, giving them a vintage or worn look. The background is plain and white. # 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves as a Summary of Safety and Effectiveness for the use of the LINEAGE™ Acetabular System. Submitted By: Date: Contact Person: Proprietary Name: Common Name: Classification Name and Reference: Wright Medical Technology, Inc. ROAD 3800 July 11, 2000 Ehab M. Esmail Senior Regulatory Affairs Associate LINEAGETM Acetabular System Metal/ Polymer Acetabular Components 21 CFR 888.3358 Prosthesis, Hip, Semi-Constrained, metal/polymer, Uncemented - Class II Orthopedics/87/LPH ## Device Product Code and Panel Code: ## DEVICE INFORMATION ### A. INTENDED USE Indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: - 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia: - 2. inflammatory degenerative joint disease such as rheumatoid arthritis; - correction of functional deformity; 3. - revision procedures where other treatments or devices have failed, and, র্ব - treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. Image /page/0/Picture/25 description: The image shows the logo for Wright Medical Technology, Inc. The logo features a stylized "W" symbol above the word "WRIGHT" in a bold, sans-serif font. Below "WRIGHT" is the text "MEDICAL TECHNOLOGY, INC." in a smaller, sans-serif font. Image /page/0/Picture/26 description: The image shows a globe with a map of the world on it. The globe is mostly white and gray, with some darker areas representing landmasses. The map shows the continents of Asia, Europe, and Africa. The globe is sitting on a stand, and there is a black arm extending from the top of the globe. {1}------------------------------------------------ ## 000013 The LINEAGETM Acetabular System are single use components, intended for use in conjunction with associated ceramic or metal femoral heads as part of an uncemented total hip arthroplasty. #### B. DEVICE DESCRIPTION The LINEAGE™ Acetabular System consists of metal acetabular shells and UHMWPE acetabular liners. The LINEAGE™ Acetabular Shell will be available in a hemispherical titanium allov shell with and without a 14° peripheral rim flare. The shells will be coated with commercially pure titanium plasma spray or porous beads. Design features of the LINEAGE™ Acetabular Shell are summarized below: - · Total hemispherical design - · Hemispherical design with 14° flare - · Coated with CP Ti plasma spray or porous beads - · Solid, quadrant, and multi-hole options - · Threaded apical hole plug The LINEAGE™ Acetabular Liners will be available with a 0° and 15° overhangs with/without a 4+mm lateralized shift. The Liner's internal geometry will be intended to be used with our existing femoral heads manufactured from cobalt chrome or ceramic with WMT12/14 taper, Orthomet® taper and A-line taper. The LINEAGE™ Liner's external geometry will be designed to accept the LINEAGE™ Acetabular Shells. Design features of the LINEAGE™ Acetabular Liner are summarized below: - · 360° liner overhang positioning options - · Features an 18° male taper and peripheral ring assembly to lock the liner into the acetabular shell - · The lip of the 15° liner will be machined to 210° - · The inserts will be offered with 0°and 15° overhangs - · The inserts will also be offered with a 4+mm lateralized shift - · Internal diameter will have a 2mm chamfer to minimize impingement. The thinnest part of any UHMWPE articulating insert will be greater than 4 mm if attached to a metal or ceramic backing. #### C. SUBSTANTIAL EQUIVALENCE INFORMATION The intended use, material, type of interface, and design features of the LINEAGE™ Acetabular System are substantially equivalent to the competitive devices previously cleared for market. The safety and effectiveness of the LINEAGE™ Acetabular System are adequately supported by the substantial equivalence information, materials data, and testing results provided within the Premarket Notification. Image /page/1/Picture/23 description: The image shows the text "510(K) SUMMARY" in a bold, serif font. The text is arranged on two lines, with "510(K)" on the first line and "SUMMARY" on the second line. The text appears to be part of a document or report. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three swooping lines, resembling an abstract representation of a bird or a wave. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 31 2000 Mr. Ehab M. Esmail Senior Regulatory Affairs Associate Wright Medical Technology, Inc. 5677 Airline Road Arlington, Tennessee 38002 Re: K002149 Trade Name: Lineage™ Acetabular System Regulatory Class: II Product Code: LPH Dated: July 13, 2000 Received: July 17, 2000 Dear Mr. Esmail: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895, A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {3}------------------------------------------------ Page 2 - Mr. Ehab M. Esmail If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, vonne R. Vochner Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page ***_ of _*** 510(k) Number (if known): K002149 LINEAGE™ Acetabular System Device Name: Indications For Use: Indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following concitions: - l ) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; - 2) inflammatory degenerative joint disease such as rheumatoid arthritis; - 3) correction of functional deformity; - 4) revision procedures where other treatments or devices have failed; and, - treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur 5) with head involvement that are unmanageable using other techniques. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daine R. Lochner (Division Sign-Off) Division of General Restorative Devices 510(k) Number 4002147 Prescription Use h U (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...