In general, these devices are intended for use in primary or revision hip arthroplasty.
Device Story
Stryker Orthopaedics Hip Systems comprise modular components including femoral stems, acetabular shells, liners, femoral heads, augments, screws, and fixation accessories (cables, grips). Devices are implanted by orthopedic surgeons in clinical settings during primary or revision total/hemi-hip arthroplasty. Components are press-fit or cemented into the femur and acetabulum to restore joint function, biomechanics, and stability. This submission updates MR conditional labeling parameters based on new RF-heating testing per FDA guidance; no changes to device design, materials, or manufacturing. Output is a reconstructed hip joint providing structural support and articulation for patients with disabling joint disease or fractures.
Clinical Evidence
No clinical data was required or provided. Substantial equivalence is supported by non-clinical bench testing, specifically RF-related heating effects in an MR environment conducted in accordance with the FDA guidance document 'Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment' (2021).
Technological Characteristics
Modular hip arthroplasty components including femoral stems, acetabular shells, liners, and fixation accessories. Materials, dimensions, and design remain unchanged from previously cleared predicates. Devices are MR conditional. No software or electronic components.
Indications for Use
Indicated for patients requiring primary or revision total or hemi-hip arthroplasty due to noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, post-traumatic arthritis, correction of functional deformity, nonunion, femoral neck/trochanteric fractures, or revision of failed previous procedures. Specific components indicated for dislocation risks, high-risk primary/revision patients (history of dislocation, bone loss, soft tissue laxity, neuromuscular disease, intraoperative instability), or acetabular deficiencies. Cementless or cemented use depends on specific component design.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Stryker Orthopaedics Hip Systems Labeling Update (K153345)
K200823 — Zimmer, Inc. Hip Joint Prostheses MR Labeling I · Zimmer, Inc. · Jul 9, 2020
K211176 — Smith & Nephew Hip Systems · Smith & Nephew, Inc. · Jul 1, 2022
K250989 — Stryker and Serf hip devices · Howmedica Osteonics Corp (Dba Stryker Orthopaedics) · Jul 25, 2025
K153345 — Stryker Orthopaedics Hip Systems Labeling Update · Stryker Orthopaedics · May 25, 2016
K240783 — Anthology Hip Stems; CPCS Hip Stems; Femoral Heads; R3 Acetabular Liners; Spectron Hip Stems; Synergy Hip Stems · Smith & Nephew, Inc. · Oct 21, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
December 21, 2023,
Howmedica Osteonics Corp. dba Stryker Orthopaedics Julia Bally Official Correspondent 325 Corporate Drive Mahwah, New Jersey 07430
Re: K233498
Trade/Device Name: Stryker Orthopaedics Hip Systems Labeling Update Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH, KWZ, LZO, MEH, JDI, LWJ, KWL, JDG, KWY, MBL, HWC, JDQ, HRS, LRN. LYT. LZN Dated: October 30, 2023 Received: October 31, 2023
Dear Julia Bally:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Christopher Ferreira -S
for
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K233498
#### Device Name
Stryker Orthopaedics Hip Systems Labeling Update
Indications for Use (Describe)
ADM/MDM X3 Inserts, MDM Acetabular Inserts, MDM Acetabular Liners
The indications for use for total hip arthroplasty include:
- 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- 2. Rheumatoid arthritis:
- 3. Correction of functional deformity;
4. Revision procedures where other treatments or devices have failed; and.
5. Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are
unmanageable using other techniques.
6. Dislocation risks
MDM Liners are intended for cementless use only.
Universal Cement Restrictor, OmniFit Distal Cement Spacer
For cement spacers, mid-shaft restrictors and Cement Plugs:
· In cemented hip arthroplasty, when the cement spacer, restrictor and/or plug is thought to be advantageous.
#### Artisan Bone Plug
These bone plugs are intended to be placed in the femoral canal prior to the introduction of bone cemented hip procedure. The plug is placed distally to the femoral stem to help allow cement pressurization and to help prevent cement migration further down the femoral canal.
Trident® II Acetabular System (Trident II Cups (Clusterhold HA, Tritanium Clusterhole, Tritanium Multihole, Tritanium Solidback), 6.5mm Low Profile Hex Screw, Hex Dome Hole Plug
· Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
· Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
#### When used with MDM Liners
· Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
• Dislocation risks
When used with Constrained Liner:
• The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
The Trident® II Acetabular Shells are indicated for cementless use only.
{4}------------------------------------------------
Restoration® Modular Hip System
The Restoration® Modular Hip System is indicated for use in:
· Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis
- · Correction of functional deformity;
- Revisions procedure where other treatments or devices have failed; and
• Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques
Additional indications specific to the Restoration Modular Hip System
The Restoration® Modular Hip System is intended to be used for primary and revision total hip arthroplasty, as well as in the presence of severe proximal bone loss. These femoral stems are designed to be press fit into the proximal femur.
Accolade II Femoral Stems, Anato Femoral Stem, Secur-Fit Advanced
The indications for use for total hip arthroplasty with stems include:
- · noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- · correction of functional deformity;
- · revision procedures where other treatments or devices have failed; and,
· nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of the Femoral Stem with compatible Howmedica Osteonics Constrained Liners:
· When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
The Stems are intended for cementless use only and are intended for total and hemiarthroplasty.
#### TRIDENT AND TRITANIUM® ACETABULAR COMPONENTS
• Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
- · Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
The HOWMEDICA OSTEONICS TRIDENT and TRITANIUM Acetabular Shells are intended for cementless use only. Dome hole plug is indicated for cementless hip arthroplasty, when an acetabular shell plug is thought to be advantageous
#### Acetabular Dome Hole Plug
The Dome Hole Plug is an optional device which is available to seal the Howmedica Osteonics Acetabular Shell components during cemented or cementless applications of the Howmedica Osteonics Done Hole Plug is threaded into the dome hole of the shell.
Indications
• In cemented or cementless hip arthroplasty, when an acetabular shell plug is thought to be advantageous.
## DALL-MILES® SYSTEMS
{5}------------------------------------------------
The DALL-MILES System is indicated for reattachment of the trochanter in any hip procedure using the trochanteric osteotomy (total or partial) approach.
The DALL-MILES Mini Cleat is indicated for vertical reattachment of the trochanter in any situation where the surgeon feels that the trochanter is at risk for detachment.
The Mini Cleat is intended for use with the DALL-MILES System for trochanteric reattachment only.
The DALL-MILES Cables and Cable Sleeves are indicated for trochanteric reattachment and trauma surgery of the hig; to stabilize bone graft material; and for supplementary cerclage fixation with plates and screws for fracture fixation.
The DALL-MILES Trochanteric Grips and Grip Plates are indicated for use in the greater trochanter due to trochanteric fracture or osteotomy with intramedullary fixation as the primary device.
The DALL-MILES Trochanteric Grip Plate is additionally indicated for use in the greater trochanter due to extended trochanteric osteotomies.
Exeter® V40™ Hip System (includes Orthinox V40 Femoral heads)
The Exeter® V40™ Femoral Hip System is indicated for:
· noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
• rheumatoid arthritis;
- correction of functional deformity;
- · revision procedures where other treatments or devices have failed; and,
· treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The Exeter® V40™ Femoral Stem Hip System is intended for use in total or hemi hip replacement. It is intended for cemented use only.
Exeter Centralizer, EXETER Intramedullary Plug
1. noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- 2. rheumatoid arthritis;
- 3. correction of functional deformity;
- 4. revision procedures where other treatments or devices have failed: and,
5. treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The Exeter Centralizer is intended to be used to centralize the femoral stem within the intramedullary canal. The Exeter Centralizer is intended to be used with bone cement.
The Exeter Intramedullary Bone Plug is intended to be used to restrict the migration of bone cement down the femoral canal and permit cement pressurization during total hip arthroplasty. The Exeter Intramedullary Bone Plug is intended to be used with bone cement.
#### Exeter X3 RimFit Cups
The indications for use for total hip arthroplasty include:
1. Painful, disabling joint disease of the hip from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
2. Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
3. Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
4. Where bone stock is of poor quality or inadequate for other reconstructive techniques, such as cementless fixation, as indicated by deficiencies of the acetabulum.
{6}------------------------------------------------
Restoration GAP II Acetabular Shell
Painful, disabling joint disease of the hip resultive arthritis, rheumatoid arthritis, post- traumatic arthritis or late stage avascular necrosis.
Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
Where bone stock is of poor quality or inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Segmental and/or cavitary acetabular defects which make it difficult to restore normal hip biomechanics or to reconstitute the normal structural continuity and integrity of the acetabulum, using standard total hip replacement acetabular components and procedures.
Gap Screws, Torx Screws, and Osteolock Bone Screws
HOWMEDICA OSTEONICS Torx Cancellous Bone Screws are intended for supplemental fixation of associated HOWMEDICA OSTEONICS cementless Acetabular Shells.
· HOWMEDICA OSTEONICS RESTORATION GAP Plate Screws are intended for fixation of the dome and illac plates of the associated HOWMEDICA OSTEONICS RESTORATION GAP Acetabular Shell, TRIDENT TRITANIUM Hemispherical Multihole Acetabular Shells, restoration Acetabular Augments, and Restoration Anatomic shells.
Insignia Hip Stem
Hip Arthroplasty Indications:
• Painful, disabling joint disease of the hip resulting from: noninflammatory degenerative joint disease (including osteoarthritis or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
· Revision of previous unsuccessful femoral head replacement, hip arthroplasty or other procedure.
· Correction of functional deformity
· Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of Insignia Hip Stems with compatible Howmedica Osteonics Constrained Liners: · When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
Additional indication specific to use of Insignia Hip Stems with compatible ADM and MDM Acetabular Components: • When the stem is to be used with compatible Howmedica Osteonics ADM and MDM Acetabular Components, the device is indicated for Dislocation risks
Insignia Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures.
#### Omnifit HFX Femoral Stems
For use as a Bipolar Hip Replacement:
- Femoral head/neck fractures or non-unions.
- · Aseptic necrosis of the femoral head.
· Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
Other Considerations:
· Pathological conditions or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
• Salvage of failed total hip arthroplasty.
{7}------------------------------------------------
· Femoral neck fractures.
For use as a Total Hip Replacement:
• Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
· Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Clinical circumstances which require an altered femoral resection level due to a proximal fracture, bone loss or calcar lysis.
Omnifit EON Cemented Femoral Stems
For use as a Bipolar Hip Replacement:
· Femoral head/neck fractures or non-unions.
- · Aseptic necrosis of the femoral head.
· Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion. Other Considerations:
· Pathological conditions or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
• Salvage of failed total hip arthroplasty.
- For use as a Total Hip Replacement:
• Painful, disabling joint disease of the hip resulting from: degenerative arthritis, post-traumatic arthritis or late stage avascular necrosis.
- · Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
Indications for use as a Total Hip Replacement include:
· noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- · correction of functional deformity;
- · revision procedures where other treatments or devices have failed; and,
· treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Restoration Anatomic Shell
• Painful, disabling joint disease of the hip from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
· Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
When used with MDM Liners
· Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
• Dislocation risks
When used with Constrained Liner:
• The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
{8}------------------------------------------------
Trident Constrained Acetabular Insert/ Constrained Acetabular Insert
The Constrained Acetabular Insert is intended for use as a component of a total hip prosthesis in primary or revision patients at a high risk of hip dislocation due to a history of dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.
Trident X3, Trident Crossfire, and Trident X3/Crossfire Elevated Rim Acetabular Liners
· Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
· Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
## UHR Bipolar
- · Femoral head/neck fractures or non-unions.
- · Aseptic necrosis of the femoral head.
· Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
Other Considerations:
· Pathological conditions or age considerations which indicate a more conservative acetabular procedure and an avoidance
of the use of bone cement in the acetabulum.
· Salvage of failed total hip arthroplasty.
Unitrax V40 Head, V40 Adapter Sleeve and C-Taper Adapter Sleeve
The HOWMEDICA OSTEONICS Unitrax Endoprosthesis, and the V40 Modular Adaptor sleeves are used as a hemiarthroplasty device for the following indications: femoral neck fractures, idiopathic avascular necrosis, and nonunions. The C-taper sleeves are intended for use as a Hemi-Hip Replacement with the following indications: femoral head/neck fractures or non-unions, aseptic necrosis of the femoral head/neck and oster- and post traumatic arthritis. The patient's acetabular bone stock must be adequate to support articulation with the ad of the endoprosthesis.
BIOLOX Delta Ceramic Heads (V40, C-Taper and Universal and C-Taper to Universal Taper Adapter Sleeve) The femoral heads are intended fixation to their mating hip stems, and can be used in cemented or cementless hip arthroplasty procedures.
For Use as a Total Hip Replacement:
· Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
- · Revision of previous cup arthroplasty or other procedures
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head.
- · Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
· Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
{9}------------------------------------------------
Alumina C-Taper and V40 Ceramic Heads
For Use as a Total Hip Replacement:
· Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
· Revision of previous cup arthroplasty or other procedures
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
Additional indications for the Alumina C-Taper Ceramic Heads
For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions.
- · Aseptic necrosis of the femoral head.
- · Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
· Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an
- avoidance of the use of bone cement in the acetabulum.
· Salvage of failed total hip arthroplasty
Co Cr Femoral Heads (C-Taper and V40 Taper, LFIT and non-LFIT)
· Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatic arthritis or late stage avascular necrosis.
· Revision of previous cup arthroplasty or other procedures
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
Accolade C Femoral Stems, Accolade HFx Femoral Stems
Indications for use as a Total Hip Replacement include:
· noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- · correction of functional deformity;
- · revision procedures where other treatments or devices have failed; and,
· treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Accolade TMZF and Accolade TMZF Plus Femoral Stems
The subject hip stem is a single-use device intended for use in total hip replacement. It is intended for the reconstruction of the head and neck of the femoral joint. This hip stem is intended for primary reconstruction of the proximal femur or revision total hip arthroplasty. This device is intended for use with any currently available Howmedica Osteonics acetabular component and V40TM femoral heads that can be mated with a TMZF 5° 40'trunnion. Indications:
· Cementless primary hip surgery in cases of non-inflammatory degenerative joint disease including osteoarthritis.
avascular necrosis, rheumatoid arthritis, and correction of functional deformity.
• Treatment of nonunion, and femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
· Revision procedures where other treatments or devices have failed.
{10}------------------------------------------------
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{11}------------------------------------------------
## 510(k) Summary
| Sponsor | Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, NJ 07430 |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Julia Bally<br>Regulatory Affairs Specialist<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>+1 201-831-6265 |
| Alternate Contact | Karen Ariemma<br>Regulatory Strategy, Director<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>201-831-5718 |
| Date Prepared: | October, 30, 2023 |
| Proprietary Name: | Stryker Orthopaedics Hip Systems Labeling Update |
| Common Name: | Artificial Hip Replacement Components -Acetabular and Femoral |
## Classification Name:
Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis 21 CFR \$888.3358
Hip joint metal/polymer constrained cemented or uncemented prosthesis 21 CFR §888.3310
Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis 21 CFR §888.3353
Hip joint metal/polymer semi-constrained cemented prosthesis 21 CFR §888.3350
Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis 21 CFR §888.3360
Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis 21 CFR នុវិទ្យាល័យ
Smooth or threaded metallic bone fixation fastener 21 CFR §888.3040
Bone fixation cerclage 21 CFR §888.3010
{12}------------------------------------------------
Single/multiple component metallic bone fixation appliances and accessories 21 CFR §888.3030
## Product Codes:
LPH - prosthesis, hip, semi-constrained, metal/polymer, porous uncemented
KWZ - prosthesis, hip, constrained, cemented or uncemented, metal/polymer
LZO - prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented
MEH - prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous
JDI - prosthesis, hip, semi-constrained, metal/polymer, cemented
LWJ - prosthesis, hip, semi-constrained, metal/polymer, uncemented
KWL - prosthesis, hip, hemi-, femoral, metal
JDG - prosthesis, hip, femoral component, cemented, metal
KWY - prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented
MAY - prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish
MBL-prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous
HWC - screw, fixation, bone
JDQ - cerclage, fixation
HRS - plate, fixation bone
LRN - wire, surgical
LYT – fixation accessory
LZN - cement obturator
{13}------------------------------------------------
# Legally Marketed Device to Which Substantial Equivalence is Claimed:
The primary predicate submission number is K153345, Stryker Orthopaedics Hip Systems Labeling Update.
| Femoral Stem Components | | | |
|-----------------------------------------------------|---------|--|--|
| Exeter Femoral Stems (V40 taper) | K173499 | | |
| SecurFit Max Femoral Stems | K153345 | | |
| SecurFit Max Plus Femoral Stems | K153345 | | |
| Accolade C Femoral Stems | K153345 | | |
| Accolade TMZF Femoral Stems | K153345 | | |
| Accolade TMZF Plus Femoral Stem | K153345 | | |
| Accolade HFx Femoral Stem | K153345 | | |
| Accolade II Femoral Stems | K153345 | | |
| SecurFit Advanced Femoral Stems | K153345 | | |
| Anato Femoral Stems | K153345 | | |
| OmniFit HFx Femoral Stems | K153345 | | |
| OmniFit EON Femoral Stems | K153345 | | |
| Insignia Hip Stem | K220731 | | |
| Restoration Modular Calcar Body | K212187 | | |
| Restoration Modular Cone Body | K212187 | | |
| Restoration Modular-Distal Bowed<br>Stem-Plasma | K212187 | | |
| Restoration Modular-Distal Stem-<br>Conical | K212187 | | |
| Restoration Modular-Distal Straight<br>Stem-Plasma° | K212187 | | |
| Restoration Modular 115mm Conical<br>Distal Stem | K220731 | | |
| Inserts/ Acetabular Liners | | | |
| ADM/MDM X3 Inserts (ETO),<br>MDM X3 Insert | K182468 | | |
| ADM/MDM X3 Inserts (GP) | K153345 | | |
| MDM Acetabular Liners | K153345 | | |
| Trident 0° Constrained Insert<br>(multiple styles) | K153345 | | |
| Trident 10° Constrained Insert | K153345 | | |
{14}------------------------------------------------
| Trident Crossfire Inserts (multiple sizes) | K153345 |
|-----------------------------------------------------------------------------------------------------------------------|---------|
| Trident X3 Inserts (GP) multiple sizes | K153345 |
| Trident X3 Acetabular Inserts (multiple styles) | K182468 |
| Acetabular Shells, Augments, Dome Hole Plug | |
| Trident Hemispherical Shells (multiple configurations) | K153345 |
| Tritanium Shells (multiple styles) | K153345 |
| Restoration Anatomic Acetabular Shell | K210893 |
| Trident Hemispherical Multi-Hole Shell | K153345 |
| Tritanium Multi-Hole Shell | K153345 |
| Trident HA PSL Solid Back Shell | K153345 |
| Trident HA PSL Cluster Shell | K153345 |
| Trident II Clusterhole HA Shell (multiple configurations) | K191358 |
| Trident II Tritanium Shells (multiple styles) | K191358 |
| Restoration Acetabular Augments | K153345 |
| Acetabular Dome Hole Plug | K220376 |
| Hex Dome Hole Plug | K191358 |
| GAP II | K222632 |
| Exeter X3 RimFit (GP) All Poly Cup | K213701 |
| Exeter X3 RimFit (ETO) All Poly Cup | K213701 |
| Femoral Heads and Adaptor Sleeves | |
| Biolox Delta Heads (Various sizes, styles) | K153345 |
| Adapter Sleeves (Various Syles) | K153345 |
| CoCR Heads (Various Styles/Sizes) | K153345 |
| C-taper Alumina Ceramic Heads | K153345 |
| V40 Orthinox Head | K173499 |
| Unitrax V40 Head (38, 40-56, 58, 61) | K153345 |
| Unitrax V40 Adapter Sleeve | |
| Unitrax C-Taper Adapter Sleeve | |
| UHR Bipolar | K222632 |
| Acessory Devices, Screws | |
| Torx Screws | K153345 |
| GAP Plate Screws | K153345 |
| Osteolock Screws | K153345 |
| UNIVERSAL CEMENT<br>RESTRICTOR | K220838 |
| 6.5mm Low Profile Hex Screw | K191358 |
| Artisan Bone Plug | K220838 |
| Dall-Miles Trochanteric grip, cables,<br>trochanteric grip plate, cable cleat<br>(multiple configurations and styles) | K202016 |
| Dall Miles Non-Beaded Cable | K202016 |
| Dall-Miles Beaded Cable and Sleeve<br>Set | K202016 |
| Dall-Miles Non-Beaded Cable and<br>Sleeve Set | K202016 |
| Dall Miles Cable Sleeve sets<br>(multiple styles/ configurations) | K202016 |
| Dall-Miles Cable Set Cable Grip | K202016 |
| Dall Miles Cable Sleeves | K202016 |
| Exeter Centralizer | K191414 |
| Exeter Intramedullary Plug | K191414 |
| Universal Distal Cement Spacer | K153345 |
{15}------------------------------------------------
## Device Description:
The devices covered by this submission are Stryker Total Hip components which include femoral stems, acetabular shells, liners, femoral heads, acetabular augments, acetabular bone screws, acetabular plugs, cables, trochanteric grips, cement restrictors, and distal femoral spacers. All devices are commercially available and have been cleared in prior 510(k) submissions.
All the subject devices have been cleared for MR conditional in previous 510(k)s. The purpose of this submission is to modify the MR conditional information in the instructions for use to update the parameters in which a patient who has the device can be safely scanned, per testing conducted accordance to updated FDA guidance. There have been no changes made to the devices requiring 510(k) clearance - only the MR conditional information in the instruction for use is being modified.
Intended Use: In general, these devices are intended for use in primary or revision hip arthroplasty.
{16}------------------------------------------------
## Indications:
#### ADM/MDM X3 Inserts, MDM Acetabular Inserts, MDM Acetabular Liners
The indications for use for total hip arthroplasty include:
- 1. Noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- 2. Rheumatoid arthritis;
- 3. Correction of functional deformity;
- 4. Revision procedures where other treatments or devices have failed; and,
5. Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head
involvement that are unmanageable using other techniques.
6. Dislocation risks
MDM Liners are intended for cementless use only.
#### Universal Cement Restrictor, OmniFit Distal Cement Spacer
For cement spacers, mid-shaft restrictors and Cement Plugs:
· In cemented hip arthroplasty, when the cement spacer, restrictor and/or plug is thought to be advantageous.
#### Artisan Bone Plug
These bone plugs are intended to be placed in the femoral canal prior to the introduction of bone cement in a cemented hip procedure. The plug is placed distally to the femoral stem to help allow cement pressurization and to help prevent cement migration further down the femoral canal.
Trident® II Acetabular System (Trident II Cups (Clusterhole HA, PLS Clusterhold HA, Tritanium Clusterhole, Tritanium Multihole, Tritanium Solidback), 6.5mm Low Profile Hex Screw, Hex Dome Hole Plug
· Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
• Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
{17}------------------------------------------------
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
When used with MDM Liners
· Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
- Dislocation risks
When used with Constrained Liner:
· The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability. The Trident® II Acetabular Shells are indicated for cementless use only.
#### Restoration® Modular Hip System
The Restoration® Modular Hip System is indicated for use in:
- Noninflammatory degenerative joint disease, including osteoarthritis and avascular ● necrosis;
- Rheumatoid arthritis ●
- Correction of functional deformity;
- Revisions procedure where other treatments or devices have failed; and ●
- . Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques
Additional indications specific to the Restoration Modular Hip System
The Restoration® Modular Hip System is intended to be used for primary and revision total hip arthroplasty, as well as in the presence of severe proximal bone loss. These femoral stems are designed to be press fit into the proximal femur.
#### Accolade II Femoral Stems, Anato Femoral Stem, Secur-Fit Advanced
The indications for use for total hip arthroplasty with stems include:
- . noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and,
- nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with ● head involvement that are unmanageable using other techniques.
{18}------------------------------------------------
Additional indication specific to use of the Femoral Stem with compatible Howmedica Osteonics Constrained Liners:
- . When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
The Stems are intended for cementless use only and are intended for total and hemiarthroplasty.
## TRIDENT AND TRITANIUM® ACETABULAR COMPONENTS
· Painful, disabling joint disease of the hip resulting from: degenerative arthritis,
rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- · Revision of previous unsuccessful femoral head replacement, cup arthroplasty or
other procedure.
- · Clinical management problems where arthrodesis or alternative reconstructive
techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or is inadequate for other reconstructive techniques
as indicated by deficiencies of the acetabulum.
The HOWMEDICA OSTEONICS TRIDENT and TRITANIUM Acetabular Shells are intended
for cementless use only.
Dome hole plug is indicated for cementless hip arthroplasty, when an acetabular
shell plug is thought to be advantageous
## Acetabular Dome Hole Plug
The Dome Hole Plug is an optional device which is available to seal the Howmedica Osteonics Acetabular Shell components during cementless applications of the acetabular cup. The Howmedica Osteonics Done Hole Plug is threaded into the dome hole of the shell.
Indications
In cemented or cementless hip arthroplasty, when an acetabular shell plug is thought to . be advantageous.
## DALL-MILES® SYSTEMS
{19}------------------------------------------------
The DALL-MILES System is indicated for reattachment of the trochanter in any hip procedure using the trochanteric osteotomy (total or partial) approach.
The DALL-MILES Mini Cleat is indicated for vertical reattachment or reinforcement of the trochanter in any situation where the surgeon feels that the trochanter is at risk for detachment.
The Mini Cleat is intended for use with the DALL-MILES System for trochanteric reattachment only.
The DALL-MILES Cables and Cable Sleeves are indicated for trochanteric reattachment and trauma surgery of the hip; to stabilize bone graft material; and for supplementary cerclage fixation with plates and screws for fracture fixation.
The DALL-MILES Trochanteric Grips and Grip Plates are indicated for use in the fixation of the greater trochanter due to trochanteric fracture or osteotomy with intramedullary fixation as the primary device.
The DALL-MILES Trochanteric Grip Plate is additionally indicated for use in the fixation of the greater trochanter due to extended trochanteric osteotomies.
Exeter® V40TM Hip System (includes Orthinox V40 Femoral heads)
The Exeter® V40™ Femoral Hip System is indicated for:
- . noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis; ●
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and, ●
- . treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The Exeter® V40™ Femoral Stem Hip System is intended for use in total or hemi hip replacement. It is intended for cemented use only.
## Exeter Centralizer, EXETER Intramedullary Plug
1. noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- 2. rheumatoid arthritis:
- 3. correction of functional deformity:
4. revision procedures where other treatments or devices have failed: and,
{20}------------------------------------------------
5. treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The Exeter Centralizer is intended to be used to centralize the femoral stem within the intramedullary canal. The Exeter Centralizer is intended to be used with bone cement.
The Exeter Intramedullary Bone Plug is intended to be used to restrict the migration of bone cement down the femoral canal and permit cement pressurization during total hip arthroplasty. The Exeter Intramedullary Bone Plug is intended to be used with bone cement.
#### Exeter X3 RimFit Cups
The indications for use for total hip arthroplasty include:
1. Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis.
post-traumatic arthritis or late stage avascular necrosis.
2. Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
3. Clinical management problems where arthrodesis or alternative reconstructive techniques are less
likely to achieve satisfactory results.
4. Where bone stock is of poor quality or inadequate for other reconstructive techniques, such as cementless fixation, as indicated by deficiencies of the acetabulum.
The EXETER X3 RimFit Cup is intended for cemented use only.
#### Restoration GAP II Acetabular Shell
Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post- traumatic arthritis or late stage avascular necrosis.
Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
Where bone stock is of poor quality or inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Segmental and/or cavitary acetabular defects which make it difficult to restore normal hip biomechanics or to reconstitute the normal structural continuity and integrity of the acetabulum, using standard total hip replacement acetabular components and procedures.
{21}------------------------------------------------
## HOWMEDICA OSTEONICS Torx Cancellous Bone Screws are intended for
supplemental fixation of associated HOWMEDICA OSTEONICS cementless Acetabular
Shells.
• HOWMEDICA OSTEONICS RESTORATION GAP Plate Screws are intended for
fixation of the dome and iliac plates of the associated HOWMEDICA OSTEONICS
RESTORATION GAP Acetabular Shell, TRIDENT TRITANIUM Hemispherical Multihole
Acetabular Shells, restoration Acetabular Augments, and Restoration Anatomic shells.
#### Insignia Hip Stem
Hip Arthroplasty Indications:
Painful, disabling joint disease of the hip resulting from: noninflammatory degenerative . joint disease (including osteoarthritis or avascular necrosis), rheumatoid arthritis or posttraumatic arthritis.
Revision of previous unsuccessful femoral head replacement, hip arthroplasty or other . procedure.
Correction of functional deformity .
. Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of Insignia Hip Stems with compatible Howmedica Osteonics Constrained Liners:
- . When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
Additional indication specific to use of Insignia Hip Stems with compatible ADM and MDM Acetabular Components:
- When the stem is to be used with compatible Howmedica Osteonics ADM and MDM . Acetabular Components, the device is indicated for Dislocation risks
Insignia Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty procedures.
#### Omnifit HFX Femoral Stems
{22}------------------------------------------------
For use as a Bipolar Hip Replacement:
- . Femoral head/neck fractures or non-unions.
- . Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular ● involvement or distortion.
#### Other Considerations:
- Pathological conditions or age considerations which indicate a more conservative acetabular ● procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty.
- · Femoral neck fractures.
For use as a Total Hip Replacement:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are ● less likely to achieve satisfactory results.
- Clinical circumstances which require an altered femoral resection level due to a proximal ● fracture, bone loss or calcar lysis.
#### Omnifit EON Cemented Femoral Stems
For use as a Bipolar Hip Replacement:
- . Femoral head/neck fractures or non-unions.
- . Aseptic necrosis of the femoral head.
- . Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
Other Considerations:
- Pathological conditions or age considerations which indicate a more conservative acetabular ● procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty. ●
For use as a Total Hip Replacement:
- . Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- . Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
Indications for use as a Total Hip Replacement include:
{23}------------------------------------------------
- noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis:
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and,
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur ● with head involvement that are unmanageable using other techniques.
### Restoration Anatomic Shell
· Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
· Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
• Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
When used with MDM Liners
· Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
• Dislocation risks
When used with Constrained Liner:
• The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
The Restoration® Anatomic Shell is indicated for cementless use only.
#### Trident Constrained Acetabular Insert/ Constrained Acetabular Insert
The Constrained Acetabular Insert is intended for use as a component of a total
hip prosthesis in primary or revision patients at a high risk of hip dislocation due to a
history of dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or
intraoperative instability.
{24}------------------------------------------------
## Trident X3, Trident Crossfire, and Trident X3/Crossfire Elevated Rim Acetabular Liners
- . Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other ● procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as ● indicated by deficiencies of the acetabulum.
## UHR Bipolar
- · Femoral head/neck fractures or non-unions.
- · Aseptic necrosis of the femoral head.
- · Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
#### Other Considerations:
· Pathological conditions or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty.
## Unitrax V40 Head, V40 Adapter Sleeve and C-Taper Adapter Sleeve
The HOWMEDICA OSTEONICS Unitrax Endoprosthesis, and the V40 Modular Adaptor sleeves are used as a hemiarthroplasty device for the following indications: femoral neck fractures, idiopathic avascular necrosis, and non-unions. The C-taper sleeves are intended for use as a Hemi-Hip Replacement with the following indications: femoral head/neck fractures or nonunions, aseptic necrosis of the femoral head/neck and osteo- and post traumatic arthritis. The patient's acetabular bone stock must be adequate to support articulation with the head of the endoprosthesis.
## BIOLOX Delta Ceramic Heads (V40, C-Taper and Universal and C-Taper to Universal Taper Adapter Sleeve)
The femoral heads are intended for mechanical fixation to their mating hip stems, and can be used in cemented or cementless hip arthroplasty procedures.
#### For Use as a Total Hip Replacement:
{25}------------------------------------------------
- . Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous cup arthroplasty or other procedures
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
## For Use as a Bipolar Hip Replacement
- . Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head.
- . Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
- . Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty ●
## Alumina C-Taper and V40 Ceramic Heads
For Use as a Total Hip Replacement:
- Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid ● arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous cup arthroplasty or other procedures
- Clinical management problems where arthrodesis or alternative reconstructive techniques are . less likely to achieve satisfactory results.
- . Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
## Additional indications for the Alumina C-Taper Ceramic Heads
## For Use as a Bipolar Hip Replacement
- . Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head.
- . Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
- . Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- . Salvage of failed total hip arthroplasty
## Co Cr Femoral Heads (C-Taper and V40 Taper, LFIT and non-LFIT)
- Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid ●
{26}------------------------------------------------
arthritis, post-traumatic arthritis or late stage avascular necrosis.
- . Revision of previous cup arthroplasty or other procedures
- . Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- . Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
#### Accolade C Femoral Stems, Accolade HFx Femoral Stems
Indications for use as a Total Hip Replacement include:
- . noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- . rheumatoid arthritis:
- correction of functional deformity; ●
- revision procedures where other treatments or devices have failed; and,
- . treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
#### Accolade TMZF and Accolade TMZF Plus Femoral Stems
The subject hip stem is a single-use device intended for use in total hip replacement. It is intended for the reconstruction of the head and neck of the femoral joint. This hip stem is intended for primary reconstruction of the proximal femur or revision total hip arthroplasty. This device is intended for use with any currently available Howmedica Osteonics acetabular component and V40™ femoral heads that can be mated with a TMZF 5° 40'trunnion.
#### Indications:
- Cementless primary hip surgery in cases of non-inflammatory degenerative joint disease . including osteoarthritis, avascular necrosis, rheumatoid arthritis, and correction of functional deformity.
- . Treatment of nonunion, and femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
- Revision procedures where other treatments or devices have failed.
#### Summary of Technological Characteristics:
There have been no changes requiring 510(k) clearance to the technological characteristics of the Stryker Total Hip systems as a result of the revision to the labeling. The subject devices have the same design and are manufactured from the same materials as the predicate devices.
{27}------------------------------------------------
### Non-Clinical Testing:
The Stryker Orthopedics Hip Systems have been previously evaluated and cleared for conditional use in a Magnetic Resonance Environment through non-clinical testing as outlined in Attachment CH3.08 - Performance Testing.
New testing has been performed to comprehensively assess the RF-related heating effects induced by the subject devices when implanted into bone, following the FDA guidance document, "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," dated May 20, 2021.
The labeling of the Stryker Orthopedics Hip Systems has been modified to provide the parameters under which a patient who has the device can be safely scanned.
Clinical Testing: Clinical testing was not required as a basis for substantial equivalence.
Conclusion: The Stryker Orthopaedics Hip System devices are substantially equivalent to the predicate device identified in this premarket notification.
Device comparison showed that the proposed device is substantially equivalent in intended use, materials and performance characteristics to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.