Intended Use

Cobalt™ HV Bone Cement is indicated for use as bone cement in arthroplastic procedures of the hip, knee and other joints to fix plastic and metal prosthetic parts to living bone when reconstruction is necessary because of osteoarthritis, rheumatoid arthritis, traumatic arthritis, avascular necrosis, nonunion of fractures of the neck of the femur, sickle cell anemia osteoporosis, secondary severe joint destruction following trauma or other conditions (also for fixation of unstable fractures in metastatic malignancies), and revision of previous arthroplasty procedures.

Device Story

Cobalt™ HV Bone Cement is a two-component, radiopaque, rapidly setting PMMA bone cement. Components include a powder (methylmethacrylate-methylacrylate copolymer, benzoyl peroxide, zirconium dioxide) and a liquid (methylmethacrylate monomer, N,N-dimethyl-p-toluidine, hydroquinone). Upon mixing, the accelerator initiates free radical polymerization, creating polymer chains that harden within 15 minutes. Used by surgeons in clinical settings to fix plastic and metal prosthetic parts to living bone during joint arthroplasty or revision procedures. The device provides mechanical fixation, enabling reconstruction of joints damaged by disease or trauma. Benefit includes stable prosthetic attachment to bone.

Clinical Evidence

No clinical testing was performed. Substantial equivalence was established through comparative in vitro bench testing in accordance with the Class II Special Controls Guidance for PMMA Bone Cement.

Technological Characteristics

PMMA bone cement consisting of a powder component (copolymer, benzoyl peroxide, zirconium dioxide) and a liquid component (methylmethacrylate monomer, N,N-dimethyl-p-toluidine, hydroquinone). Radiopaque. Polymerization-based setting mechanism. Tested per Class II Special Controls Guidance for PMMA Bone Cement.

Indications for Use

Indicated for patients undergoing hip, knee, or other joint arthroplasty requiring fixation of prosthetic components to bone due to osteoarthritis, rheumatoid arthritis, traumatic arthritis, avascular necrosis, nonunion of femoral neck fractures, sickle cell anemia, osteoporosis, post-traumatic joint destruction, metastatic malignancy-related unstable fractures, or revision arthroplasty.

Regulatory Classification

Identification

Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K051496 AUG 4 - 2005 ## 510(k) Summary # Cobalt™ HV Bone Cement Applicant/Sponsor: Biomet Manufacturing Corp. P.O. Box 587 Warsaw, Indiana 46581-0587 Contact Person: Lonnie Witham Telephone: (574) 267-6639 Fax: (574) 372-1683 Proprietary Name: Cobalt™ HV Bone Cement Common Name: Bone Cement # Legally Marketed Devices To Which Substantial Equivalence Is Claimed: Device: Palacos® R Bone Cement Approved by: PMA (P810020, S001, S002, S003, S004) reclassified into Class II Manufacturer: Biomet Inc.; 56 East Bell Drive; Warsaw, IN 46582 Predicate Device: Generation 4® Bone Cement Cleared by: 510(k) Notification (K993836) Manufacturer: Biomet Inc.; 56 East Bell Drive; Warsaw, IN 46582 (Relevant to packaging and sterilization processes cleared for this device) #### Device Description: Cobalt™ Bone Cement provides two separate, pre-measured sterilized components which when mixed form radiopaque rapidly setting bone cement. #### Materials used: 40 g powder component (copolymer): - Methylmethacrylate-methylacrylate copolymer containing FD&C Blue No. 2 . Aluminum Lake - Benzoyl peroxide, hydrous 75% . - Zirconium dioxide . 20 ml of liquid component (monomer): - Methylmethacrylate (stabilized with hydroquinone) . - . N.N-dimethyl-p-toluidine Methymethacrylate monomer is the primary constituent of the liquid component. In much smaller quantities are the accelerator, N, N-dimethyl-p-toluidine, and the stabilizer, hydroquinone, both are typical constituents of PMMA bone cement. {1}------------------------------------------------ When the powder and liquid components are mixed, the accelerator speeds the generation of free radicals and the stabilizer in the liquid reacts with many of the early free radicals, but is soon consumed. Free radicals can then initiate formation of polymer chains. Polymerization proceeds slowly over the first few minutes. Polymer chains at the surface of the powder beads mingle with monomer and newly formed polymer chains, while smaller beads may dissolve completely. The cement temperature rises as set time of the cement approaches. Polymerization is essentially complete and the bone cement is hard within 15 minutes. ### Intended Use / Indications for Use: Cobalt™ Bone Cement is indicated for use as bone cement in arthroplastic procedures of the hip, knee and other joints to fix plastic and metal prosthetic parts to living bone when reconstruction is necessary because of osteoarthritis, rheumatoid arthritis, traumatic arthritis, avascular necrosis, nonunion of fractures of the femur, sickle cell anemia osteoporosis, secondary severe joint destruction following trauma or other conditions (also for fixation of unstable fractures in metastatic malignancies), and revision of previous arthroplasty procedures. ### Summary of the Technological Characteristics: Extensive in-vitro testing was performed in accordance with Class II Special Controls Guidance: Polymethylmethacrvlate (PMMA) Bone Cement: Guidance for Industry & FDA-July 17, 2002 to demonstrate the equivalence of Cobalt™ Bone Cement to Palacos® R Bone Cement. Test results showed that the technological characteristics (mechanical, chemical, physical & handling properties) of the two cements are substantially the same. ## Non-Clinical: The substantial equivalence of Cobalt™ Bone Cement to Palacos® R Bone Cement was determined by performing comparative in vitro testing and then comparing the results. The results showed that Cobalt™ Bone Cement possesses mechanical, chemical, physical and handling characteristics necessary to fulfill its intended use. In summary, Cobalt™ Bone Cement is substantially equivalent to Palacos® R Bone Cement for its primary intended use of fixation of prosthetic components as described in the device labeling. No clinical testing was performed. Clinical Testing: No clinical testing was required. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus symbol. The logo is black and white. 9200 Corpor Rockville M Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Lonnie Witham Biomet Manufacturing Corp. P.O. Box 587 Warsaw, Indiana 6581-0587 Re: K051496 Trade/Device Name: Cobalt™ HV Bone Cement Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: II Product Code: LOD Dated: June 3, 2005 Received: June 6, 2005 Dear Mr. Witham: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. AUG 4 - 2005 If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 -- Mr. Lonnie Witham This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Stupt Rhodes Mark Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## STATEMENT OF INDICATIONS FOR USE 510(k) Number Device Name: Cobalt™ HV Bone Cement ### Indications for Use: Cobalt™ HV Bone Cement is indicated for use as bone cement in arthroplastic procedures of the hip, knee and other joints to fix plastic and metal prosthetic parts to living bone when reconstruction is necessary because of osteoarthritis, rheumatoid arthritis, traumatic arthritis, avascular necrosis, nonunion of fractures of the neck of the femur, sickle cell anemia osteoporosis, secondary severe joint destruction following trauma or other conditions (also for fixation of unstable fractures in metastatic malignancies), and revision of previous arthroplasty procedures. Prescription Use X (Per 21 CFR 801.109) .. OR Over-The-Counter-Use (Optional Format 1-2-96) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE) Hyst Rurda (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number_________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...