The Digitalis Spacer MCP implants are intended for replacement of the metacarpophalangeal joint of the hand which has been damaged by: - Osteoarthrosis; - Rheumatoid arthritis; - Post traumatic arthritis. The Digitalis Spacer PIP implants are intended for replacement of the proximal interphalangeal joint of the hand which has been damaged by: - Rheumatoid arthritis; - Osteoarthritis: - Ankylosed joints or those with limited range of motion which have not responded to conservative treatment; - Non-functional joint due to inadequate bony alignment and joint space which cannot be restored by soft tissue reconstruction alone; - Destroyed articular surface(s).
Device Story
Double-stemmed, constrained, silicone elastomer prosthesis; functions as joint spacer for MCP or PIP joints. Implanted surgically to replace damaged osteo-cartilaginous heads; stems fit into intramedullary canals of phalanges/metacarpals. Acts as hinge joint to restore joint space and alignment. Used by orthopedic surgeons in clinical settings. Benefits patients by restoring joint function and mobility in cases of arthritis or joint destruction.
Clinical Evidence
Bench testing only. Included static strength testing, fatigue strength testing per ASTM F1781 (2021), crack propagation analysis, wear particle analysis, and LAL testing per AAMI ST-72.
Technological Characteristics
Constrained, double-stemmed silicone elastomer prosthesis. Dimensions and geometry similar to predicate devices. Fatigue testing performed per ASTM F1781 (2021). Biocompatibility/endotoxin testing per AAMI ST-72.
Indications for Use
Indicated for patients requiring replacement of metacarpophalangeal (MCP) or proximal interphalangeal (PIP) joints of the hand due to osteoarthrosis, rheumatoid arthritis, post-traumatic arthritis, ankylosed joints, limited range of motion unresponsive to conservative treatment, non-functional joints, or destroyed articular surfaces.
Regulatory Classification
Identification
A finger joint polymer constrained prosthesis is a device intended to be implanted to replace a metacarpophalangeal or proximal interphalangeal (finger) joint. This generic type of device includes prostheses that consist of a single flexible across-the-joint component made from either a silicone elastomer or a combination pf polypropylene and polyester material. The flexible across-the-joint component may be covered with a silicone rubber sleeve.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
BRM Extremities Srl % Margeaux Rogers Director, Regulatory Affairs Mcra, LLC 803 7th Street NW Washington, District of Columbia 20001
Re: K220142
Trade/Device Name: BRM Digitalis Spacer Regulation Number: 21 CFR 888.3230 Regulation Name: Finger joint polymer constrained prosthesis Regulatory Class: Class II Product Code: KYJ Dated: March 9, 2023 Received: March 9, 2023
Dear Margeaux Rogers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
April 4, 2023
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Farzana Farzana Sharmin -S Sharmin -S Date: 2023.04.04
Farzana Sharmin, PhD Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220142
Device Name BRM Digitalis Spacer
Indications for Use (Describe)
The Digitalis Spacer MCP implants are intended for replacement of the metacarpophalangeal joint of the hand which has been damaged by:
- · Osteoarthrosis;
- · Rheumatoid arthritis;
- Post traumatic arthritis.
The Digitalis Spacer PIP implants are intended for replacement of the proximal interphalangeal joint of the hand which has been damaged by:
- · Rheumatoid arthritis;
- · Osteoarthritis:
- · Ankylosed joints or those with limited range of motion which have not responded to conservative treatment;
- · Non-functional joint due to inadequate bony alignment and joint space which cannot be restored by soft tissue reconstruction alone;
- · Destroyed articular surface(s).
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="text-decoration: underline;"><b> Prescription Use (Part 21 CFR 801 Subpart D) </b></span> | □ Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary – K220142
| Manufacturer: | BRM Extremities | |
|---------------------|-----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| | Registered Office: | Via Lorenzo Mascheroni 29<br>20145 Milano (MI) - Italy |
| | Headquarter: | Via Papa Giovanni XXIII, 9<br>23862 Civate (LC) - Italy |
| Contact: | Margeaux Rogers<br>Senior Director, Regulatory Affairs<br>MCRA, LLC<br>Phone: 202.552.5805<br>Email: mrogers@mcra.com | |
| Prepared By: | MCRA, LLC<br>803 7th Street NW<br>Washington, DC 20001<br>Phone: 202.552.5800 | |
| Date Prepared: | March 21, 2023 | |
| Device Trade Name: | BRM Digitalis Spacer | |
| Device Common Name: | Prosthesis, Finger, Constrained Polymer | |
| Classification: | 21 CFR 888.3230 – Finger joint polymer constrained prosthesi<br>Class II | |
| Product Codes: | KYJ | |
# Indications for Use:
The Digitalis Spacer MCP implants are intended for replacement of the metacarpophalangeal joint of the hand which has been damaged by:
- Osteoarthrosis;
- Rheumatoid arthritis;
- Post traumatic arthritis. ●
The Digitalis Spacer PIP implants are intended for replacement of the proximal interphalangeal joint of the hand which has been damaged by:
- Rheumatoid arthritis; •
- Osteoarthritis;
{4}------------------------------------------------
- Ankylosed joints or those with limited range of motion which have not responded to conservative treatment;
- . Non-functional joint due to inadequate bony alignment and joint space which cannot be restored by soft tissue reconstruction alone;
- Destroved articular surface(s). .
# Device Description:
The Digitalis Spacer is a double-stemmed, constrained, silicone prosthesis with a hinge joint. The subject device is available in two options, one intended for use in the metacarpophalangeal (MCP) and the other for use in the proximal interphalangeal joints (PIP). The implant MCP option is intended to be implanted to replace the osteo-cartilaginous heads of the metacarpophalangeal joint an act as a joint spacer between the resected head of the metacarpal at the distal head and the base of the proximal phalanx. The implant PIP option is intended to be implanted to replace the osteo-cartilaginous heads of the proximal interphalangeal joint and act as a joint spacer between the head of proximal phalanx and base of the medial phalanx.
# Primary Predicate Device:
The Digitalis Spacer is substantially equivalent to the primary predicate device Ascension Silicone MCP Spacers (K022892) with respect to intended use and indications for use, similar dimensions, and geometry.
# Additional Predicate Devices:
The Digitalis Spacer is substantially equivalent to the additional predicate devices Ascension PIP Spacers (K082231) and Stryker Silicone PIP & MCP (K870200 & K931588) with respect to intended use and indications for use, similar dimensions, and geometry.
# Reference Device:
The Digitalis Spacer is similar to the reference device NewPrim System (K191966) with respect to material.
# Discussion of Non-Clinical Testing/Performance Data:
Mechanical testing included evaluation of static strength of the silicone material as well as fatigue strength testing of the worst-case construct.
The BRM Digitalis Spacer is identical to the reference device in terms of material. Based on similarities in material and intended use, static strength testing performed on sections of the reference device components is applicable to the subject device; the static tensile strength of the BRM Digitalis Spacer is expected to exceed anticipated physiologic loading.
{5}------------------------------------------------
Fatigue testing performed on the subject device which incorporated both high and low cycle activities representative of intended physiologic use was performed according to ASTM F1781 (2021). Additionally, crack propagation was specifically analyzed as well as wear particle analysis. Additionally, the Digitalis Spacer is in compliance with LAL testing requirements for orthopedic devices per AAMI ST-72 testing.
# Substantial Equivalence Conclusions:
The Digitalis Spacer and the legally marketed predicate device have the same intended use and indications for use, similar dimensions, geometry and materials. The stems of the devices are fit into the intramedullary canals of the metacarpophalangeal and proximal interphalangeal joints. The devices are constrained and made of silicone elastomer.
Conclusions drawn from the nonclinical testing demonstrate that the BRM Digitalis Spacer is as safe, as effective, and performs as well as the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.