DEPUY DUPONT ORTHOPAEDICS FINGER JOINT PROSTHESIS

K970544 · Depuy, Inc. · KYJ · Sep 12, 1997 · Orthopedic

Device Facts

Record IDK970544
Device NameDEPUY DUPONT ORTHOPAEDICS FINGER JOINT PROSTHESIS
ApplicantDepuy, Inc.
Product CodeKYJ · Orthopedic
Decision DateSep 12, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3230
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The DePuy DuPont Finger Joint Prosthesis is indicated for cementless replacement of the metacarpophalangeal (MCP) joints of the finger where disabled by rheumatoid, degenerative or traumatic arthritis.

Device Story

Flexible, one-piece, hinged silicone elastomer implant; designed for cementless implantation across metacarpophalangeal (MCP) joint. Proximal and distal stems feature 30° angle to mimic relaxed hand position. Implanted by surgeons to replace damaged joints; restores joint function in patients with arthritis. Bench testing confirms durability through 10 million cycles of 90° flexion/extension.

Clinical Evidence

Bench testing only. Flexural fatigue testing performed for 10 million cycles at 90° flexion/extension, comparing DDO implant performance to Swanson and Sutter predicate devices.

Technological Characteristics

Silicone elastomer; one-piece hinged design; 30° angled proximal and distal stems; cementless fixation; dimensions/form factor designed for MCP joint replacement.

Indications for Use

Indicated for cementless replacement of metacarpophalangeal (MCP) joints in patients with rheumatoid, degenerative, or traumatic arthritis.

Regulatory Classification

Identification

A finger joint polymer constrained prosthesis is a device intended to be implanted to replace a metacarpophalangeal or proximal interphalangeal (finger) joint. This generic type of device includes prostheses that consist of a single flexible across-the-joint component made from either a silicone elastomer or a combination pf polypropylene and polyester material. The flexible across-the-joint component may be covered with a silicone rubber sleeve.

Predicate Devices

Submission Summary (Full Text)

{0} K970544 SEP 12 1997 SUMMARY OF SAFETY AND EFFECTIVENESS NAME OF FIRM: DePuy Inc. 700 Orthopaedic Drive Warsaw, Indiana 46581-0988 510(k) CONTACT: Cheryl Hastings Manager, Clinical Affairs TRADE NAME: DePuy DuPont Orthopaedics Finger Joint Prosthesis COMMON NAME: Finger Joint Prosthesis CLASSIFICATION: 888.3230 Finger joint, polymer, constrained prosthesis DEVICE PRODUCT CODE: 87 KYJ SUBSTANTIALLY EQUIVALENT DEVICES: ▲ Sutter Avanta MCP Joint Prosthesis ▲ Dow Corning Wright Swanson Finger Joint Implant DEVICE DESCRIPTION AND INTENDED USE: The DePuy DuPont Orthopaedics (DDO) Finger Joint Prosthesis is a flexible, one-piece, hinged silicone elastomer implant designed to be implanted across the metacarpophalangeal (MCP) joint. The proximal and distal stems of the prosthesis form a 30° angle which mimics the approximate position of the joint when the hand is relaxed. The DDO Finger Joint Prosthesis is indicated for cementless replacement of the metacarpophalangeal (MCP) joints of the finger where disabled by rheumatoid, degenerative or traumatic arthritis. BASIS OF SUBSTANTIAL EQUIVALENCE: The DDO Finger Joint Prosthesis is substantially equivalent to both the Dow Corning Wright Swanson Finger Joint Implant and the Sutter Avanta MCP Joint Prosthesis in that all of these implants are intended to replace the MCP joint when it is damaged due to rheumatoid, degenerative or traumatic arthritis. All of these implants are manufactured from silicone elastomer and all have a hinge design. The Swanson and Sutter implants are both designed as "straight" hinges while the DDO implant incorporates a 30° angle. Flexural fatigue testing shows that the DDO Finger Joint implants sustain the same or less damage than the Swanson or Sutter implants when cycled through a total of 90° of flexion/extension for 10 million cycles. {1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 12 1997 Ms. Cheryl Hastings Manager, Clinical Affairs DePuy, Inc. P.O. Box 988 700 Orthopaedic Drive Warsaw, Indiana 46581-0988 Re: K970544 Trade Name: DePuy DuPont Orthopaedics Finger Joint Prosthesis Regulatory Class: II Product Code: KYJ Dated: June 13, 1997 Received: June 16, 1997 Dear Ms. Hastings: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2} Page 2 - Ms. Cheryl Hastings This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} 510(k) Number (if known) K970544 Device Name DePuy DuPont Finger Joint Prosthesis Indications for Use: The DePuy DuPont Finger Joint Prosthesis is indicated for cementless replacement of the metacarpophalangeal (MCP) joints of the finger where disabled by rheumatoid, degenerative or traumatic arthritis. Concurrence of CDRH, Office of Device Evaluation ![img-1.jpeg](img-1.jpeg) Prescription Use ☑ OR Over-The Counter Use ☐ (Per 21 CFR 801.109)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...