FIXION UNIPOLAR MODULAR HEMI-HIP SYSTEM (FIXION MH)

K030972 · Disc-O-Tech Medical Technologies, Ltd. · KWL · May 30, 2003 · Orthopedic

Device Facts

Record IDK030972
Device NameFIXION UNIPOLAR MODULAR HEMI-HIP SYSTEM (FIXION MH)
ApplicantDisc-O-Tech Medical Technologies, Ltd.
Product CodeKWL · Orthopedic
Decision DateMay 30, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3360
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Fixion® MH System is intended for cemented or non-cemented use as a hemi-hip replacement. It is indicated for use in cases of: Femoral head and/or neck fractures or non-unions; Aseptic necrosis of the femoral head and/or neck; Osteo-, Rheumatoid-, and/or Post-traumatic arthritis of the hip, with minimal acetabular involvement.

Device Story

Modular hemi-hip prosthesis; includes expandable femoral stem and unipolar head ball. Surgeon uses inflation device (pump) to deliver saline into stem, causing expansion and mechanical abutment against medullary cavity; inflation pressure monitored via gauge. Used in OR for hip replacement. Provides structural support and articulation for hip joint; benefits patient by restoring joint function in cases of fracture, necrosis, or arthritis.

Clinical Evidence

Bench testing only. Performance evaluated per ISO 7206-1, ISO 7206-4, and ISO 7206-8 standards for endurance properties of stemmed femoral components. Compliance with ASTM F138-2000 for material specifications and ASTM F565-2000 for handling.

Technological Characteristics

Modular hemi-hip system; 316L stainless steel construction (ASTM F138). Expandable stem body; unipolar head ball. Mechanical fixation via saline-driven expansion. Dimensions per ISO 7206-1. Endurance testing per ISO 7206-4/8. Non-powered, mechanical device.

Indications for Use

Indicated for patients with femoral head/neck fractures, non-unions, aseptic necrosis, or hip arthritis (osteo-, rheumatoid, or post-traumatic) with minimal acetabular involvement requiring hemi-hip replacement.

Regulatory Classification

Identification

A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY 3 0 2003 Disc-O-Tech Medical Technologies, Ltd. Fixion® Unipolar Modular Hemi-Hip System # 510(K) Summary # KO30972 Disc-O-Tech Medical Technologies Ltd. Fixion® Unipolar Modular Hemi-Hip System (Fixion® MH) -10/16 mm stem body; 12 mm stem neck; 316L stainless steel head ball. ### Company Name and Address Disc-O-Tech Medical Technologies Ltd. 3 Hasadnaot St. Herzelia 46728 Israel #### Submitter's Name and Contact Person Hila Wachsler-Avrahami Disc-O-Tech Medical Technologies Ltd. 3 Hasadnaot St. Herzelia 46728 Israel Tel .: + 972 9 9511511 Fax .: + 972 9 9548939 #### Date Prepared March 26, 2003 #### Trade/Proprietary Name Fixion® Unipolar Modular Hemi-Hip System (Fixion® MH) #### Classification Name Prosthesis, Hip, Hemi-, Femoral, Metal 21 CFR § 888.3360 Class II {1}------------------------------------------------ # Predicate Devices - 1. Fixion® Unipolar Modular Hemi-Hip System (K014072) by Disc-O-Tech. - 2. Fixion Interlocking Proximal Femur Intramedullary Nailing System (K010988, K012967) by Disc-O-Tech. ## Performance Standards The following standards were used: - 1. ISO 7206-1 (1995): Implants for Surgery Partial and Total Hip Joint Prostheses - Part 1: Classification and Description of Dimensions. - 2. ISO 7206-4 (Draft, 1999): Implants for Surgery Partial and Total Hip Joint Prostheses - Part 4: Determination of Endurance Properties of Stemmed Femoral Components. - 3. ISO 7206-8 (1995): Implants for Surgery Partial and Total Hip Joint Prostheses = Part 8: Endurance Performance of Stemmed Femoral Components with Application of Torsion. - 4. Guidance Document for Femoral Stem Prostheses (Draft), ORDB/DGRD/CDRH/FDA, August 1, 1995. - 5. Guidance Document for Testing Non-Articulating, "Mechanically Locked", Modular Implant Components (Draft), ORDB/DGRD/CDRH/FDA, May 1, 1995. - 6. ASTM F138-2000: Standard Specification for Stainless Steel Bar and Wire for Surgical Implants. - 7. ASTM F565-2000: Standard Practice for Care and Handling of Orthopedic Implants and Instruments. ## Intended Use The Fixion® MH System is intended for cemented or non-cemented use as a hemi-hip replacement. It is indicated for use in cases of: - > Femoral head and/or neck fractures or non-unions; - Aseptic necrosis of the femoral head and/or neck; A - Osteo-, Rheumatoid-, and/or Post-traumatic arthritis of the hip, with A minimal acetabular involvement. {2}------------------------------------------------ # System Description The Fixion® MH System is a modular hemi-hip system, which consist of the following main components: - Implants (stainless steel), including the stem and head ball. The stem is the 1. femoral diaphyseal component. It consists of an expandable stem body and a neck part. The unipolar head ball is the prosthesis hip component that articulates within the acetabulum. - The Instrumentation Set is a set of accessories, including pre-operative 2. tools, femoral canal preparation tools and accessories such as the stem insertion handle, inflation adapter, screwdriver for stem cap insertion and devices for implant removal. - The Inflation Device is a pump, which rotation of its handle delivers saline 3. into the stem. This action causes the expansion of the stem and its abutment to the bone medullary cavity. The pump pressure gauge indicates the inflation pressure. # Substantial Equivalence The Fixion® MH 10/16 stem (with a neck width of 12 mm) and the Fixion® MH 316L stainless steel head balls are substantially equivalent to the cleared stem and head balls of the Fixion® MH System (K014072): - They have the same intended use and indications for use; ું. - ું જ They incorporate the same basic design; - They have the same operating principles; જુન - ું જ They have the same fixation method; - ું જ They are packed and sterilized using the same materials and processes. The added components are made of the same materials as the cleared Fixion® MH stem. In addition, the dimensions of the added Fixion® MH stem body, neck and head balls are within the range of sizes of other cleared, marketed systems. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a caduceus, which is a symbol often associated with healthcare and medicine. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAY 3 0 2003 Ms. Hila Wachsler-Avrahami Regulatory Affairs Disc-O-Tech Medical Technologies Ltd. 3 Hasadnaot Street Herzelia Israel 46728 Re: K030972 Trade/Device Name: Fixion® Unipolar Modular Hemi-Hip System (Fixion® MH) Regulation Number: 21 CFR 888.3360 Regulation Name: Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis Regulatory Class: II Product Code: KWL Dated: April 29, 2003 Received: May 1, 2003 Dear Ms. Wachsler-Avrahami: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {4}------------------------------------------------ #### Page 2 - Ms. Lori Glastetter forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Mark A. Milliman Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indication for Use 510(K) Number (if known): _ Ko3 o972 # Device Name: Fixion® Unipolar Modular Hemi-Hip System (Fixion® MH) Indication for Use: The Fixion® MH System intended for cemented or noncemented use as a hemi-hip replacement. It is indicated for use in cases of: - Femoral head and/or neck fractures or non-unions; A - Aseptic necrosis of the femoral head and/or neck; A - 八 Osteo-, Rheumatoid-, and/or Post-traumatic arthritis of the hip, with minimal acetabular involvement. # (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use | OR | Over the Counter Use | |----------------------|----|----------------------| | (per 21 CFR 801.109) | | | *[Signature]* (Division Sign-Off) Division of General, Restorative and Neurological Devices | 510(k) Number | K030972 | |---------------|---------| |---------------|---------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...