K010081 · Exactech, Inc. · KWY · Feb 5, 2001 · Orthopedic
Device Facts
Record ID
K010081
Device Name
ACUMATCH L-SERIES UNIPOLAR ENDOPROSTHESIS
Applicant
Exactech, Inc.
Product Code
KWY · Orthopedic
Decision Date
Feb 5, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AcuMatch L-Series Unipolar Endoprosthesis was designed to address the more conservative clinical indications for joint replacement surgery. The L-Series Unipolar modular head is used in hemiarthroplasty of the hip. If total joint replacement is required at a future time, the procedure can be accomplished without removing the femoral stem.
Device Story
Modular unipolar hip endoprosthesis; consists of truncated spherical head and interchangeable neck sleeves. Used in hemiarthroplasty; allows future conversion to total hip replacement without femoral stem removal. Surgeon selects components based on patient requirements; requires specialized orthopedic surgical training. Device provides structural replacement for femoral head; restores joint function in degenerative or traumatic conditions. Benefits include conservative surgical approach and modularity for sizing.
Clinical Evidence
Bench testing only. Evidence includes Finite Element Analysis (FEA) and fatigue testing of production quality parts to verify performance under anticipated in vivo loading conditions. No clinical data presented.
Technological Characteristics
Modular unipolar endoprosthesis; cobalt chrome construction per ASTM F1537-94. Includes 15 head sizes (38-62 mm) and four offset options (-5, +0, +5, +10 mm). Supplied sterile via Gamma radiation (25-37 kGy, SAL 10^-6). Packaging includes PETG trays, Tyvek lids, and LD45 foam inserts.
Indications for Use
Indicated for skeletally mature individuals undergoing primary hip replacement for osteoarthritis, osteonecrosis, rheumatoid arthritis, post-traumatic degenerative problems, or femoral head replacement following femoral neck fracture. Contraindicated in patients with active infection, insufficient bone stock, neuromuscular disorders preventing hip control, or conditions where weight, age, or activity level suggest early failure. Also contraindicated for patients with acetabular degenerative changes or pelvic fractures.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis is a two-part device intended to be implanted to replace the head and neck of the femur. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a snap-fit acetabular component made of an alloy, such as cobalt-chromium-molybdenum, and ultra-high molecular weight polyethylene. This generic type of device may be fixed to the bone with bone cement (§ 888.3027) or implanted by impaction.
{0}------------------------------------------------
2320 NW 66TH COUR GAINESVILLE, FL 3265
352-377-1140
FAX 352-378-2617
Exactech® AcuMatch™ Integrated Hip System L-Series Unipolar Endoprosthesis
01008
FEB - 5 2001
Sponsor:
# 510(k) Summary of Safety and Effectiveness Special 510(k)
Exactech® Inc. 2320 N.W. 66th Court Gainesville, Florida 32653
(352) - 377 - 1140 Phone: (352) - 378 - 2617 Fax:
FDA Establishment Number 1038671
Gary J. Miller, Ph.D. Contact: Executive V.P. of Research and Development
Date:
01)08 DI
Section 4 Page 1 of 4
{1}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness Special 510(k)
### Classifications / Proprietary Names:
| Classification Name: | Prosthesis, Hip, Hemi-, Femoral, Metal |
|-----------------------|------------------------------------------------------------------------|
| Product Code: | KWL |
| C.F.R. Section: | 888.3360 |
| Device Class: | II |
| Classification Panel: | Orthopedic |
| Classification Name: | Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented or Uncemented |
| Product Code: | KWY |
| C.F.R. Section: | 888.3390 |
| Device Class: | II |
| Classification Panel: | Orthopedic |
Trade / Proprietary Model Names: AcuMatch L-Series Unipolar Endoprosthesis
### Legally Marketed Devices for Substantial Equivalence Comparison:
The Exactech AcuMatch L-Series Endoprosthesis is made of similar materials and is of a similar design to other legally marketed femoral components, most notably the Exactech Opteon Endoprosthesis:
| Model | Manufacturer | 510(k) Number |
|--------------------------------|----------------|---------------|
| Opteon Unipolar Endoprosthesis | Exactech, Inc. | K960538 |
| Unitrax Unipolar System | Howmedica | ---------- |
| Unipolar System | Smith & Nephew | ---------- |
| Bio-Moore II | BioMet | ---------- |
The proposed AcuMatch L-Series Unipolar Endoprosthesis has the same basic external design features as our predicate Exactech Opteon Endoprosthesis. These features include a material composition of cobalt chrome and identical off set sizing options. The main difference between the two designs is that the predicate Opteon model is a one-piece component and the L-Series model is a modular device with interchangeable neck segments.
ﺀ ﺃﻧﻪ ﻳ
{2}------------------------------------------------
## 510(k) Summary of Safety and Effectiveness Special 510(k)
In addition, the proposed L-Series Unipolar Endoprosthesis provides features similar to models marketed by Smith & Nephew, Howmedica and Biomet.
All the components have a spherical design, are made of cobalt chrome and supplied sterile.
#### Device Description:
The AcuMatch L-Series Unipolar is a modular endoprosthesis consisting of a truncated The Tical head and interchangeable neck sleeves. The device is fully machined from sphortour nead and filtering to ASTM F1537-94. The size offerings include 15 head sizes (38 mm - 62 mm) with four available offset options (-5, +0, +5 and +10).
Selection of the Acumatch Unipolar component is made by the surgeon in relationship to the requirements of the patient. The surgeon should become thoroughly familiar with the rechnique of implayation of the prosthesis by: 1) appropriate reading of the literature and 2) training in the operative skills and techniques required for hip arthroplasty surgeries.
### INTENDED USE
The AcuMatch L-Series Unipolar Endoprosthesis was designed to address the more conservative clinical indications for joint replacement surgery. The L-Series Unipolar modular head is used in hemiarthroplasty of the hip. If total joint replacement is required at a future time, the procedure can be accomplished without removing the femoral stem.
#### INDICATIONS
The AcuMatch L-Series Unipolar Endoprosthesis is indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. It is also indicated for use in replacement of the femoral head following femoral neck fracture.
### CONTRAINDICATIONS
The AcuMatch L-Series Unipolar Endoprosthesis is contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip joint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure of the system. The L-Series Unipolar is also contraindicated for use in patients with evidence of degenerative changes in the acetabulum and/or pelvic fractures.
> Section 4 Page 3 of 4
{3}------------------------------------------------
### 510(k) Summary of Safety and Effectiveness Special 510(k)
Packaging Information:
| Material | Composition |
|---------------------|-----------------------------------|
| Inner / Outer Trays | PETG - 0.040" thickness |
| Tray Lids | Spun-Bonded Olefin - Tyvek® |
| Inserts | Medium grade LD45 Foam |
| Box | Heavy weight cardboard |
| Outer Shrink-Wrap | Clear, Light-Weight Plastic |
| Shipping Cartons | Heavy-weight Corrugated Cardboard |
Utilization and implantation instructions are included in the package insert provided with each product. The name, size, dimension, material, lot, serial number and sterility status are indicated on the labeling.
Sterilization Specifications:
Method: Gamma radiation (Cobalt 60 source) Dose: 25 - 37 kGy Sterility Assurance Level (SAL): 10-6
#### Performance Data Summary:
Functional testing and engineering analysis were conducted to verify that the implant performance would be adequate for anticipated in vivo loading. Based on Finite Element Analysis (FEA), fatigue testing of production quality parts, and material certifications we conclude that the Exactech AcuMatch L-Series Unipolar Endoprosthesis is substantially equivalent to our predicate Exactech Opteon Unipolar device (#K960538).
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the text "Public Health Service" in a bold, sans-serif font. The text is arranged on a single line, with each word clearly legible. The overall impression is clean and straightforward, suggesting an official or institutional context.
Image /page/4/Picture/2 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's name in a circular arrangement around the edge. In the center of the seal is a stylized emblem consisting of three parallel, curved lines that resemble a bird in flight or a symbolic representation of human services.
FEB - 5 2001
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Lisa Simpson Regulatory Representative Exactech, Inc. 2320 N.W. 66th Court Gainesville, Florida 32653
Re: K010081
Trade Name: Exactech® AcuMatch™ L-series Unipolar Endoprosthesis Regulatory Class: II Product Code: KWY Dated: January 8, 2001 Received: January 10, 2001
Dear Ms. Simpson:
We have reviewed your Section 510(k) notification of intent to market the device referenced we have reviewed your became is substantially equivalent (for the indications for use above and we nave determined the worketed in interstate commerce prior to May 28, 1976, the stated in the encrosure) to device Amendments, or to devices that have been reclassified in enacificin cate of the Modical Dories Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general therefore, mainer the act include requirements for annual registration, listing of devices, Controlls of the rice free heeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (1 remance ripproval), in they or says of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to mspotwarity also 2007 regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket I notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{5}------------------------------------------------
### Ms. Lisa Simpson
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
L. Mark N. Milliken
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use
510(k) Number:
K01008/
Device Name:
Exactech® AcuMatch™ Integrated Hip System L-Series Unipolar Endoprosthesis
#### INDICATIONS
The AcuMatch L-Series Unipolar Endoprosthesis is indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, osteonecrosis. rheumatoid arthritis and/or post-traumatic degenerative problems. It is also indicated for use in replacement of the femoral head following femoral neck fracture.
#### CONTRAINDICATIONS
The AcuMatch L-Series Unipolar Endoprosthesis is contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip ioint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure of the system. The L-Series Unipolar is also contraindicated for use in patients with evidence of degenerative changes in the acetabulum and/or pelvic fractures.
Concurrence of CDRH, Office of Device Evalu (Division Sign-Off) Division of General Restorative Devices 510(k) Number Prescription Use or Over the Counter Use 人
Please do not write below this line - use another page if needed.
Section 3 Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.