SYNTHES MINI EXTERNAL FIXATOR
Device Facts
| Record ID | K961350 |
|---|---|
| Device Name | SYNTHES MINI EXTERNAL FIXATOR |
| Applicant | Synthes (Usa) |
| Product Code | KTT · Orthopedic |
| Decision Date | Jun 28, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3030 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Synthes Mini External Fixator stabilizes and provides treatment for fractures of the hand and foot.
Device Story
External fixation system for hand and foot fractures; components include threaded Kirschner wires, wire-to-rod clamps, rod-to-rod clamps, rods, and angled connecting bar. Procedure: K-wires inserted into proximal/distal fracture fragments using soft tissue protector; clamps and rods attached to wires; fracture repositioned; nuts tightened to secure construct; K-wires cut. Used by surgeons in clinical settings to provide rigid stabilization during fracture healing.
Clinical Evidence
Bench testing only. Mechanical testing performed to compare stiffness and fixation strength against Howmedica Mini Hoffmann External Fixation System; results indicate Synthes device is at least equivalent to or stiffer than predicate.
Technological Characteristics
Components: Threaded Kirschner wires (stainless steel), clamps (stainless steel), angled connecting bar (stainless steel), rods (carbon fiber reinforced epoxy). Mechanical fixation principle. Non-sterile; requires sterilization by user.
Indications for Use
Indicated for patients requiring stabilization and treatment of fractures of the hand and foot.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Howmedica Mini Hoffmann External Fixation System
Related Devices
- K963618 — SYNTHES SIMPLE SMALL EXTERNAL FIXATOR (SSEF) · Synthes (Usa) · Nov 12, 1996
- K042019 — HOFFMANN II MICRO EXTERNAL FIXATION SYSTEM · Howmedica Osteonics Corp. · Sep 24, 2004
- K050048 — HOFFMAN II MICRO EXTERNAL FIXATION SYSTEM · Howmedica Osteonics Corp. · Mar 4, 2005
- K211112 — OIC External Fixation System · Orthopaedic Implant Company · May 5, 2021
- K092002 — AFGHAN EXTERNAL FIXATOR · Smith & Nephew, Inc. · Dec 10, 2009