The DRACO Locking Screws are intended to be used in combination with DRACO Locking plates for adult and pediatric patients as indicated for small bone fracture fixations for use include fixation of fractures, osteotomies, nonunions of the clavicle, scapula, olecranon, radius, metacarpals, metatarsals, Hallux Valgus osteotomy corrections, middle hand and middle foot bones, particular in osteopenic bone. The DRACO Compression Screws are indicated for fracture fixation of various bones, including, osteotomies in the foot (as Hallux Valgus treatment) or hand, foot or ankle surgery, fixation of bone fragments in long bones or small bone fractures. The size of the chosen screw should be adapted to the specific indication. The DRACO MetaFuse L Plates and the DRACO MetaFuse BLP Plates are used for adult and pediatric patients as indicated for small bone fracture fixations for use include fixation of fractures, ostectornies, non-unions of the clavicle, scapula, olecranon, radius, ulna, metacarpals, Hallux Valgus osteotomy corrections, middle hand and middle foot bones, particular in osteopenic bone. The DRACO MetaFuse MTP Plates, DuoTakú Plates, Open Wedge Plates, Osteoplate Plates, Plantar Plates, Calcaneus Offset Locking Plates can be used for adult and pediatric patients. Indications for use include fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the hand, feet, wrist, and toes.
Device Story
System of metallic bone fixation appliances including locking screws, compression screws, and various plates (MetaFuse, Takú, DuoTakú, Open Wedge, Osteoplate, Plantar, Calcaneus Offset). Used by orthopedic surgeons in clinical settings for internal fixation of small bone fractures, osteotomies, and joint fusions. Devices provide mechanical stabilization of bone fragments to facilitate healing. Screws and plates are implanted surgically; intended to remain in situ during bone union. Benefits include structural support for bone segments, particularly in osteopenic bone, aiding patient recovery and mobility.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Metallic bone fixation appliances. Materials include medical-grade metals typical for orthopedic fixation. Design includes locking and compression screws and various anatomical plate configurations. Mechanical fixation principle.
Indications for Use
Indicated for adult and pediatric patients requiring fixation of fractures, osteotomies, nonunions, joint fusion, or reconstruction of small bones (hand, feet, wrist, toes, clavicle, scapula, olecranon, radius, ulna, metacarpals, metatarsals). Includes Hallux Valgus corrections and use in osteopenic bone.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Related Devices
K232652 — Anatomical Plating System (various) · Tyber Medical, LLC · Dec 13, 2023
K190391 — TDM Plate and Screw Systems · Tdm Co., Ltd. · Nov 15, 2019
K222194 — Baby Gorilla®/Gorilla® Plating System · Paragon 28, Inc. · Aug 19, 2022
K131186 — SYNTHES 2.7MM AND 3.5MM VARIABLE ANGLE LCP MIDFOOT/HINDFOOT SYSTEM · Synthes (USA) Products, LLC · Oct 8, 2013
K221465 — Baby Gorilla/Gorilla Plating System · Paragon 28, Inc. · Jun 17, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 24, 2017
HNM Total Recon, LLC % Craig Pagan Consultant D2D Development, LLC 1135 W NASA Blvd. Melbourne, Florida 32901
Re: K163486
Trade/Device Name: DRACO Locking Screws;
DRACO Compression Screws; DRACO MetaFuse L Plates and the DRACO MetaFuse BLP Plates; DRACO MetaFuse MTP Plates, Takú Plates, DuoTakú Plates, Open Wedge Plates, Osteoplate Plates, Plantar Plates, Calcaneus Offset Locking Plates Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And
Accessories Regulatory Class: Class II Product Code: KTT, HWC, HRS, HTY Dated: June 20, 2017
Received: June 23, 2017
Dear Craig Pagan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K163486
Device Name DRACO Locking Screws
#### Indications for Use (Describe)
The DRACO Locking Screws are intended to be used in combination with DRACO Locking plates for adult and pediatric patients as indicated for small bone fracture fixations for use include fixation of fractures, osteotomies, nonunions of the clavicle, scapula, olecranon, radius, metacarpals, metatarsals, Hallux Valgus osteotomy corrections, middle hand and middle foot bones, particular in osteopenic bone.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known) K163486
Device Name DRACO Compression Screws
#### Indications for Use (Describe)
The DRACO Compression Screws are indicated for fracture fixation of various bones, including, osteotomies in the foot (as Hallux Valgus treatment) or hand, foot or ankle surgery, fixation of bone fragments in long bones or small bone fractures. The size of the chosen screw should be adapted to the specific indication.
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known) K163486
Device Name
DRACO MetaFuse L Plates and the DRACO MetaFuse BLP Plates
#### Indications for Use (Describe)
The DRACO MetaFuse L Plates and the DRACO MetaFuse BLP Plates are used for adult and pediatric patients as indicated for small bone fracture fixations for use include fixation of fractures, ostectornies, non-unions of the clavicle, scapula, olecranon, radius, ulna, metacarpals, Hallux Valgus osteotomy corrections, middle hand and middle foot bones, particular in osteopenic bone.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known) K163486
#### Device Name
DRACO MetaFuse MTP Plates, Takú Plates, Open Wedge Plates, Osteoplate Plates, Plantar Plates, Calcaneus Offset Locking Plates
#### Indications for Use (Describe)
The DRACO MetaFuse MTP Plates, DuoTakú Plates, Open Wedge Plates, Osteoplate Plates, Plantar Plates, Calcaneus Offset Locking Plates can be used for adult and pediatric patients. Indications for use include fixation of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the hand, feet, wrist, and toes.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.