REPROCESSED EXTERNAL FIXATION DEVICE

K023714 · Alliance Medical Corp. · KTT · Feb 3, 2003 · Orthopedic

Device Facts

Record IDK023714
Device NameREPROCESSED EXTERNAL FIXATION DEVICE
ApplicantAlliance Medical Corp.
Product CodeKTT · Orthopedic
Decision DateFeb 3, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

External Fixation Devices are intended to be used for the fixation of supracondylar, or condylar fractures of the femur; for fusion of a joint; for surgical procedures that involve cutting the bone, for fixation of bone fractures; bone reconstruction; as a guide pin for insertion of other implants; or may be implanted through the skin so that a pulling force or traction may be applied to the skeletal system; and others may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Device Story

Reprocessed external fixation devices consist of frames, clamps, rods, couplings, pins, posts, fasteners, wire fixations, bolts, washers, nuts, hinges, and sockets. Used in orthopedic surgery for bone fracture management, reconstruction, and correction. Devices are implanted through skin to provide skeletal stability or traction. Reprocessed by Alliance Medical Corporation to restore original function. Used by orthopedic surgeons in clinical/surgical settings. Benefit includes stabilization of bone segments to facilitate healing and deformity correction.

Clinical Evidence

Bench testing only. Validation of reprocessing and functional testing performed to demonstrate that reprocessed devices meet original performance specifications.

Technological Characteristics

Metallic alloys, plastics, and composites. Mechanical fixation system comprising frames, clamps, rods, and fasteners. No changes to materials, design, or dimensions compared to predicate devices. Sterilization via validated reprocessing protocols.

Indications for Use

Indicated for patients requiring external skeletal fixation for treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction, limb lengthening, and correction of bony or soft tissue deformities and defects.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K02371Y ## 510(k) SUMMARY | Submitter: | Alliance Medical Corporation<br>10232 South 51st Street<br>Phoenix, Arizona 85044 | | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------| | Contact: | Don Selvey<br>Vice President, Regulatory Affairs & Quality Assurance<br>(480) 763-5300 (o)<br>(480) 763-5310 (f) | | | Date of preparation: | August 29, 2002 | | | Name of device: | Trade/Proprietary Name: Reprocessed External Fixation<br>Devices<br>Common or Usual Name: External Fixation Devices, Fixation<br>Appliance, Single/Multiple Component<br>Classification Name: Single/Multiple Component Metallic Bone<br>Fixation Appliances and Accessories and Smooth or Threaded<br>Metallic Bone Fixation Fastener | | | Predicate device(s): | Legally marketed external fixation devices under various 510(k)<br>premarket notifications. | | | | K802814<br>K831576<br>K944092<br>K955848 | Orthofix® Axial External Fixation System<br>Orthofix®<br>Additional Accessories for the Orthofix® System<br>Orthofix® Modulsystem | | Device description: | External fixation devices are specially designed frames, clamps,<br>rods, rod-to-rod couplings, pins, posts, fasteners, wire fixations,<br>fixation bolts, washers, nuts, hinges, sockets, connecting bars<br>and screws used for the management of bone fractures and<br>reconstructive, as well as corrective, orthopedic surgery.<br>Materials used include metal alloys, plastic and composites.<br>These materials are chosen to address a wide range of<br>fractures and applications as well as to allow for the appropriate<br>amount of rigidity and stability. | | | Intended use: | External Fixation Devices are intended to be used for the<br>fixation of supracondylar, or condylar fractures of the femur; for<br>fusion of a joint; for surgical procedures that involve cutting the<br>bone, for fixation of bone fractures; bone reconstruction; as a<br>guide pin for insertion of other implants; or may be implanted<br>through the skin so that a pulling force or traction may be<br>applied to the skeletal system; and others may be used for<br>fixation of bone fractures, for bone reconstructions, as a guide<br>pin for insertion of other implants, or it may be implanted<br>through the skin so that a pulling force (traction) may be applied | | . {1}------------------------------------------------ to the skeletal system. - Reprocessed external fixation devices are indicated for use in Indications statement: patients requiring external skeletal fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities. fracture revision, bone reconstruction procedures, limb lenathening, correction of bony or soft tissue deformities and segmental bony or soft tissue defects. - Technological The design, materials, and intended use of the Reprocessed External Fixation Devices are identical to the predicate devices. characteristics: The mechanism of action of the Reprocessed External Fixation Device is identical to the predicate devices in that the same standard mechanical design, materials, shapes and sizes are utilized. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. - Bench and laboratory testing was conducted to demonstrate Performance data: performance (safety and effectiveness) of the Reprocessed External Fixation Devices. - Validation of reprocessing . - Function Testing . Performance testing demonstrates that Reprocessed External Fixation Devices perform as originally intended. - Conclusion: Alliance Medical Corporation concludes that the modified device (the Reprocessed External Fixation Device) is safe, effective and substantially equivalent to the predicate devices, as described herein. {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Don Selvey Vice President Regulatory Affairs and Quality Assurance Alliance Medical Corporation 10232 South 51st Street Phoenix, Arizona 85044 FEB 0 3 2003 Re: K023714 Trade Name: Reprocessed External Fixation Devices Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT, KTW, JEC Dated: November 1, 2002 Received: November 5, 2002 Dear Mr. Selvey: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Mr. Don Selvey This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Sincerely yours, Mark A. Melbern Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use Statement ll. 510(k) Number (if known): KO23714 Device Name: Alliance Medical Corporation Reprocessed External Fixation Devices Indications for Use: Reprocessed external fixation devices are indicated for use in patients requiring external skeletal fixation and treatment of fractures, osteotomy, arthrodesis, correction of deformities, fracture revision, bone reconstruction procedures, limb lengthening, correction of bony or soft tissue deformities and segmental bony or soft tissue defects. Concurrence of CDRH, Office of Device Evaluation (ODE) Mark N. Milliken Division Sign-Off) Division of General. Restorative and Neurological Devic 510(k) Number K0237 Prescription Use (per 21 CFR 801.109) or Over-the-Counter Use _ Alliance Medical Corporation Reprocessed External Fixation Devices Traditional 510(k)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...