Exactech Optetrak Logic Enhanced Assembly

K152170 · Exactech, Inc. · JWH · Sep 3, 2015 · Orthopedic

Device Facts

Record IDK152170
Device NameExactech Optetrak Logic Enhanced Assembly
ApplicantExactech, Inc.
Product CodeJWH · Orthopedic
Decision DateSep 3, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Intended Use

The OPTETRAK Comprehensive Knee Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present. In the USA, the OPTETRAK Comprehensive Knee Systems are indicated for cemented use only, except for the OPTETRAK Logic PS and CR Porous Femoral Components, which are indicated for cemented or cementless use. The Optetrak Logic CC Total Knee System is indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present. The Optetrak Logic CC Total Knee System is indicated for cemented use only.

Device Story

Exactech Optetrak Logic Enhanced Assembly consists of modified tibial inserts and trays for total knee arthroplasty. Device features dimensional modifications to enhance intraoperative assembly of components. Used by orthopedic surgeons in clinical settings during primary or revision total knee replacement surgery. Prosthesis replaces damaged joint surfaces to restore function and reduce pain. Substantial equivalence established via mechanical testing (micromotion and shear) comparing modified components to previously cleared Optetrak Logic systems.

Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence supported by mechanical engineering studies, specifically micromotion and shear testing, demonstrating performance equivalent to predicate devices.

Technological Characteristics

Semi-constrained, cemented, polymer/metal/polymer knee prosthesis. Modifications involve dimensional changes to tibial inserts and trays to facilitate assembly. Materials and design principles remain consistent with previously cleared Optetrak Logic systems.

Indications for Use

Indicated for skeletally mature individuals requiring primary total knee replacement or revision due to osteoarthritis, osteonecrosis, rheumatoid arthritis, or post-traumatic degenerative conditions, provided sufficient bone stock and soft tissue integrity exist.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized human figure with three faces in profile, representing the department's focus on health and human well-being. The logo is simple, using only black and white. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 3, 2015 Exactech, Incorporated Mr. Patrick Hughes Senior Regulatory Affairs Specialist 2320 Northwest 66th Court Gainesville, Florida 32653 Re: K152170 Trade/Device Name: Exactech Optetrak Logic Enhanced Assembly Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: July 31, 2015 Received: August 4, 2015 Dear Mr. Hughes, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related {1}------------------------------------------------ adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Mark N. Melkerson -S Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152170 Device Name Exactech Optetrak Logic Enhanced Assembly #### Indications for Use (Describe) Optetrak Logic Total Knee System: The OPTETRAK Comprehensive Knee Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, theumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revious reconstructions where sufficient bone stock and soft tissue integrity are present. In the USA, the OPTETRAK Comprehensive Knee Systems are indicated for cemented use only, except for the OPTETRAK Logic PS and CR Porous Femoral Components, which are indicated for cemented or cementless use. #### Optetrak Logic CC Total Knee System: The OPTETRAK Logic CC Total Knee System is indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, theumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revious reconstructions where sufficient bone stock and soft tissue integrity are present. The OPTETRAK Logic CC Total Knee System is indicated for cemented use only. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | |----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------| |----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Exactech® Optetrak Logic® Enhanced Assembly Special 510(k) - 510(k) Summary of Safety and Effectiveness | Sponsor: | Exactech, Inc.<br>2320 N.W. 66th Court<br>Gainesville, FL 32653 | |----------|-----------------------------------------------------------------| | Phone: | (352) 377-1140 | | Fax: | (352) 378-2617 | | | FDA Establishment Number 1038671 | | Contact: | Patrick Hughes<br>Senior Regulatory Affairs Specialist | | Date: | July 31, 2015 | ## Trade or Proprietary or Model Name(s): Exactech® Optetrak Logic® Total Knee System & Exactech® Optetrak Logic® CC Total Knee System #### Common Name: Cemented Total Knee Prosthesis #### Classification Name: Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer #### Information on devices to which substantial equivalence is claimed: | 510(k) Number | Trade or Proprietary Model Name | Manufacturer | |---------------|-------------------------------------------|---------------| | K093360 | Optetrak Logic PS | | | K111400 | Optetrak Logic CR, Sizes 1-5 | | | K121307 | Optetrak Logic CR, Sizes 0 & 6 | | | K110547 | Optetrak Logic PSC | Exactech, Inc | | K123342 | Optetrak Logic CRC | | | K132161 | Optetrak Logic 17mm & 19mm Tibial Inserts | | | K150890 | Optetrak Logic CC | | | K101981 | Optetrak Logic Combo Trays | | #### Indications for Use: The OPTETRAK Comprehensive Knee Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present. In the USA, the OPTETRAK Comprehensive Knee Systems are indicated for cemented use only, {4}------------------------------------------------ # Exactech® Optetrak Logic® Enhanced Assembly Special 510(k) - 510(k) Summary of Safety and Effectiveness except for the OPTETRAK Logic PS and CR Porous Femoral Components, which are indicated for cemented or cementless use. The Optetrak Logic CC Total Knee System is indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present. The Optetrak Logic CC Total Knee System is indicated for cemented use only. ## Device Description: This submission proposes dimensional modifications to features common to all Optetrak Logic tibial inserts, Optetrak Logic trapezoidal tibial trays, and Optetrak Logic Combo tibial trays. These modifications are designed to enhance intraoperative assembly. ## Testing Description: This submission includes references to the following mechanical testing: - . Micromotion testing - Shear testing ## Substantial Equivalence Conclusion: Results of engineering studies referenced in this 510(k) submission demonstrate proposed Optetrak Logic and Optetrak Logic CC devices are substantially equivalent to cited cleared predicate Optetrak Logic devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...