OPTERTRAK LOGIC PSC TIBIAL INSERT

K110547 · Exactech Sarasota · JWH · Apr 14, 2011 · Orthopedic

Device Facts

Record IDK110547
Device NameOPTERTRAK LOGIC PSC TIBIAL INSERT
ApplicantExactech Sarasota
Product CodeJWH · Orthopedic
Decision DateApr 14, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The OPTETRAK Logic Total Knee System is indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present. In the USA, the OPTETRAK Logic Total Knee System is indicated for cemented use only.

Device Story

Optetrak Logic PSC Tibial Insert is a modification of the existing Optetrak Logic Total Knee System. Device features a central spine with increased medial/lateral thickness compared to the predicate, providing 2°-3° additional varus/valgus constraint for patients with minor ligament laxity. Used in orthopedic surgery for total knee replacement; implanted by surgeons. Device functions as a semi-constrained component within the knee joint, articulating with femoral components and tibial trays. Benefits include improved stability for patients requiring additional constraint during primary or revision knee arthroplasty.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including engineering drawing comparisons and cadaver laboratory assessments evaluating spine shear strength, component compatibility, and varus/valgus/internal/external constraint.

Technological Characteristics

Semi-constrained cemented knee prosthesis. Materials and manufacturing processes identical to predicate. Features modified central spine geometry for increased varus/valgus constraint (2°-3°). Compatible with existing Optetrak Logic tibial trays and femoral components.

Indications for Use

Indicated for skeletally mature individuals undergoing primary total knee replacement for osteoarthritis, osteonecrosis, rheumatoid arthritis, or post-traumatic degenerative conditions; also indicated for revision of failed reconstructions with sufficient bone stock and soft tissue integrity. Cemented use only.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KI10547 ### Exactech® Optetrak® Logic™ PSC Tibial Insert Special 510(k) - 510(k) Summary of Safety and Effectiveness APR 1 4 2011 #### Exactech® Inc. Sponsor: 2320 N.W. 66th Court Gainesville, FL 32653 Phone: (352) 327-4762 Fax: (352) 378-2617 FDA Establishment Number 1038671 - Contact: Patrick Hughes Regulatory Affairs Specialist Date: April 8, 2011 ### Trade or Proprietary or Model Name(s): Exactech® Optetrak® Logic™ PSC Tibial Insert ### Common Name: Cemented Total Knee Prosthesis #### Classification Name: 21 CFR 888.3560 - Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis. Class II. Product Code: JWH - prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer. ### Information on devices to which substantial equivalence is claimed: | | 510(k) Number | Trade or Proprietary or Model Name | Manufacturer | |--|---------------|------------------------------------|---------------| | | K093360 | Optetrak Logic Total Knee System | Exactech, Inc | {1}------------------------------------------------ ### Exactech® Optetrak® Logic™ PSC Tibial Insert Special 510(k) - 510(k) Summary of Safety and Effectiveness ### Indications for Use: The OPTETRAK Logic Total Knee System is indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present. In the USA, the OPTETRAK Logic Total Knee System is indicated for cemented use only. ### Device Description: The proposed Optetrak Logic PSC Tibial Insert devices represent modifications to existing Optetrak Logic Total Knee System components (K093360). Both proposed and predicate devices are made from the same materials using the same processes and are compatible with the same Optetrak Logic tibial trays and femoral components. The proposed Logic PSC devices combine the articulating geometry and locking features of predicate Optetrak Logic posterior stabilized inserts with a central spine that is slightly thicker in the medial/lateral dimensions, intended to provide surgeons with an option for treating patients with minor ligament laxity by introducing 2°-3° more varus/valgus constraint than the predicate. The proposed and predicate devices have the same intended use and basic fundamental scientific technology and share the following similarities: - . the same indications for use - . similar design features - . the same shelf life - are packaged and sterilized using the same materials and processes. . {2}------------------------------------------------ ## Exactech® Optetrak® Logic™ PSC Tibial Insert Special 510(k) - 510(k) Summary of Safety and Effectiveness ### Performance Data Table 1 shows performance data provided, cited, or referenced in this submission to support a conclusion of substantial equivalence: | Evaluation | Activities Performed | |-----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Logic PSC tibial spine shear strength | Review of predicate designs for clinical performance and resistance to shear forces Engineering drawing comparison of Logic PSC spine dimensions to predicate designs with successful clinical performance | | Logic PSC/mating femoral component compatibility | Cadaver laboratory assessment Review of predicate designs for clinical performance Engineering drawing comparison of Logic PSC dimensions to predicate designs that successfully mate with the same components | | Logic PSC varus/valgus constraint | Cadaver laboratory assessment Engineering drawing comparison of Logic PSC dimensions to predicate designs that successfully mate with the same components | | Logic PSC internal/external constraint | Cadaver laboratory assessment Engineering drawing comparison of Logic PSC dimensions to predicate designs that successfully mate with the same components | | Logic PSC tibial spine/mating femoral component clearance | Review of predicate designs for clinical performance Engineering drawing comparison of Logic PSC spine dimensions to predicate designs with successful clinical performance | ### Table 1: Optetrak Logic PSC Tibial Insert Performance Data #### Substantial Equivalence Conclusion: Results of engineering studies referenced in this 510(k) submission demonstrate the proposed Optetrak Logic PSC Tibial Insert devices are substantially equivalent to cleared predicate Optetrak Logic posterior stabilized tibial insert devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized caduceus, a symbol often associated with healthcare, consisting of a staff with a serpent entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Exactech, Inc. % Mr. Patrick Hughes Regulatory Affairs Specialist 2320 N.W. 66" Court Gainesville, Florida 32653 APR 1 4 2011 Re: K110547 Trade/Device Name: Exactech Optetrak Logic PSC Tibial Insert Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: II Product Code: JWH Dated: February 23, 2011 Received: March 24, 2011 Dear Mr. Hughes: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ #### Page 2 - Mr. Patrick Hughes device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Sincerely, yours, for Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Exactech® Optetrak® Logic™ PSC Tibial Insert Special 510(k) - Indications for Use 510(k) Number: KI10547 # Device Name: Exactech® Optetrak® Logic™ PSC Tibial Insert ### INDICATIONS The OPTETRAK Logic Total Knee System is indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present. In the USA, the OPTETRAK Logic Total Knee System is indicated for cemented use only. Prescription Use X (Part 21 CFR 801 Subpart D) and/or Over-The-Counter Use (21 CFR 807 Subpart C) Please do not write below this line - use another page if needed. Concurrence of CDRH, Office of Device Evaluation (ODE) for M. Mielkeom (Division Sign-Off) (Division Sign-On) Division of Surgical, Orthopedic, Divisionomative Devices 510(k) Number K110547
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...