K193533 · Biomet, Inc. · JDS · Feb 20, 2020 · Orthopedic
Device Facts
Record ID
K193533
Device Name
WIN Flexible Nail System
Applicant
Biomet, Inc.
Product Code
JDS · Orthopedic
Decision Date
Feb 20, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Intended Use
WIN Nails (for use in pediatric patients) are to be used for treatment of long-bone fractures including non-comminuted and comminuted mid-shaft fractures, subtrochanteric fractures, combination fractures of the shaft and neck, intertrochanteric fractures, combination intertrochanteric and subtrochanteric fractures.
Device Story
WIN Flexible Nail System consists of titanium alloy rods for internal fixation of pediatric long-bone fractures. Device implanted by orthopedic surgeons via surgical insertion into bone to provide osteosynthesis. System includes various nail sizes differentiated by surface treatment. Device serves as mechanical support for bone healing. No electronic, software, or algorithmic components.
Clinical Evidence
No clinical data provided. No non-clinical testing provided.
Technological Characteristics
Metallic bone fixation appliance; material: titanium alloy; form factor: flexible nails; surface treatment for size differentiation; sterilization: vendor-provided.
Indications for Use
Indicated for pediatric patients with long-bone fractures, including mid-shaft (comminuted/non-comminuted), subtrochanteric, distal third, intertrochanteric, and combined shaft/neck/intertrochanteric fractures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
K981529 — TITANIUM INTRAMEDULLARY NAIL, TITANIUM LOCKING SCREW · Smith & Nephew, Inc. · Jul 9, 1998
K233150 — NET BRAND Osteosynthesis Nailing System · Narang Medical , Ltd. · Feb 20, 2024
K081097 — ORTHOPEDIATRICS PEDIFLEX FLEXIBLE NAIL SYSTEM · Orthopediatrics, LLC · Jun 3, 2008
K240061 — TRIGEN MAX Tibial Nail System; INTERTAN MAX Hip Fracture Nail System · Smith & Nephew, Inc. · Sep 20, 2024
K251362 — PediFlex Flexible Nail System · OrthoPediatrics Corp. · Jun 24, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
February 20, 2020
Biomet Inc. Dominique Dompor Regulatory Affairs Specialist 56 East Bell Drive, PO Box 587 WARSAW IN 46581
Re: K193533
Trade/Device Name: WIN Flexible Nail System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: JDS Dated: December 19, 2019 Received: December 20, 2019
Dear Dominique Dompor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Michael Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193533
Device Name WIN Flexible Nail System
# Indications for Use (Describe)
WIN Flexible Nail System
WIN Nails (for use in pediatric patients) are to be used for treatment of long-bone fractures including non-comminuted and comminuted mid-shaft fractures, subtrochanteric fractures, combination fractures of the shaft and neck, intertrochanteric fractures, combination intertrochanteric and subtrochanteric fractures.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K193533
### 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the WIN Flexible Nail System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on September 13, 2019.
| Sponsor: | Biomet Inc.<br>56 East Bell Drive<br>PO Box 587<br>Warsaw, IN 46581<br>Establishment Registration Number:<br>1825034 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person: | Dominique Dompor<br>Regulatory Affairs Specialist<br>Telephone: (877) 652-0830 ext. 169 |
| Secondary Contact Person: | Stacy Hutchens<br>Scientific Affairs Manager<br>Telephone: (574) 371-9671 |
| Date: | December 19, 2019 |
| Subject Device: | Trade Name: WIN Flexible Nail System<br>Common Name: Nail, Fixation, Bone |
| | Classification Name:<br>Single/multiple component metallic bone<br>fixation appliances and accessories. |
| | • JDS - Nail, Fixation, Bone<br>(21 CFR §888.3030) |
### Predicate Device(s):
| Primary Predicate | 510(k) Number |
|-------------------------|---------------|
| Pediatric Fixation Rods | K000764 |
| Alternate Predicate | 510(k) Number |
| 2mm Pediatric Flex Nail | K022531 |
Purpose and Device Description:
The subject WIN Flexible Nail System is made of titanium alloy, which has been continuously made commercially available since the 1950s. This device is to
{4}------------------------------------------------
Intended Use and Indications for Use:
Summary of Technological Characteristics:
Summary of Performance Data: (Nonclinical and/or Clinical)
be implanted by insertion into long bones for fixation of fractures.
The purpose of this submission is:
- . To further identify and clarify cumulative changes made to the system since original clearance:
- o An extension to the WIN Flexible Nail product line offering;
- Added surface treatment to o aid in visual differentiation of nail sizes:
- A change to the o manufacturing location;
- o A change to the sterilization vendor site;
- . To update labeling in order to bring the Instructions for Use up to current practices.
This device is intended for osteosynthesis.
The WIN Flexible Nail System is to be used for treatment of long-bone fractures including non-comminuted and comminuted mid-shaft fractures, subtrochanteric fractures, distal third fractures, combination fractures of the shaft and neck, intertrochanteric fractures, combination intertrochanteric and subtrochanteric fractures.
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Intended Use: Identical to predicate
- Indications for Use: Identical to ● predicate
- Materials: Identical to predicate ●
- Design Features: Similar to predicate
- Sterilization: Identical to predicate
### Non-Clinical Tests:
- None provided .
### Clinical Tests:
{5}------------------------------------------------
## Substantial Equivalence Conclusion:
- None provided
The subject WIN Flexible Nail System has identical intended use and indications for use as the predicate devices. The subject device has similar technological characteristics to the predicates, and the information provided herein demonstrates that:
- . any differences do not raise new questions of safety and effectiveness; and
- the subject devices are at least as . safe and effective as the legally marketed predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.