BIOKNOTLESS AND LUPINE PLUS ANCHORS

K062170 · Depuy Mitek, A Johnson & Johnson Company · JDR · Aug 22, 2006 · Orthopedic

Device Facts

Record IDK062170
Device NameBIOKNOTLESS AND LUPINE PLUS ANCHORS
ApplicantDepuy Mitek, A Johnson & Johnson Company
Product CodeJDR · Orthopedic
Decision DateAug 22, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Bioknotless Plus Anchor with is indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization as follows: OPEN PROCEDURES SHOULDER 1. Bankart repair 2. SLAP lesion repair 3. Rotator cuff repair 4a. Capsule shift/capsulo-labral reconstruction, at the anterior glenoid rim site 4b. Capsule shift/capsulo-labral reconstruction at the lesser tuberosity of the humerus 5. Biceps tenodesis 6. Acromio-clavicular separation 7. Deltoid repair ELBOW 1. Biceps tendon reattachment 2. Tennis elbow repair ANKLE 1. Achilles tendon repair/reconstruction 2. Lateral stabilization 3. Medial stabilization at the medial talus site Foot: Hallux Valgus reconstruction 4. Midfoot reconstruction KNEE 1. Medial collateral ligament repair 2. Lateral collateral ligament repair 3. Joint capsule closure to anterior proximal tibia 4. Posterior oblique ligament or joint capsule to tibia repair 5. Extra capsular reconstruction / ITB tenodesis 6. Patellar ligament and tendon avulsion repairs. ARTHROSCOPIC PROCEDURES SHOULDER 1. Bankart repair 2. SLAP lesion repair 3. Rotator cuff repair 4. Capsule shift repair (glenoid rim) The Lupines Plus The Lupine Loop Anchor System is indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization as follows: OPEN PROCEDURES SHOULDER 1. Bankart repair 2. SLAP lesion repair 3. Rotator cuff repair 4a. Capsule shift/capsulo-labral reconstruction, at the anterior glenoid rim site 4b. Capsule shift/capsulo-labral reconstruction at the lesser tuberosity of the humerus 5. Biceps tenodesis 6. Acromio-clavicular separation 7. Deltoid repair ELBOW 1. Biceps tendon reattachment 2. Tennis elbow repair ANKLE 1. Achilles tendon repair/reconstruction 2. Lateral stabilization 3. Medial stabilization at the medial talus site Foot: Hallux Valgus reconstruction 4. Midfoot reconstruction KNEE 1. Medial collateral ligament repair 2. Lateral collateral ligament repair 3. Joint capsule closure to anterior proximal tibia 4. Posterior oblique ligament or joint capsule to tibia repair 5. Extra capsular reconstruction / ITB tenodesis 6. Patellar ligament and tendon avulsion repairs. ARTHROSCOPIC PROCEDURES SHOULDER 1. Bankart repair 2. SLAP lesion repair 3. Rotator cuff repair 4. Capsule shift repair (glenoid rim)

Device Story

Preloaded, absorbable disposable suture anchor/inserter assembly; used for soft tissue-to-bone fixation in shoulder, knee, ankle, foot, and elbow. Anchor constructed of molded Poly (L-lactide) (PLA) polymer; identical in design/configuration to predicate anchors. Sold with Ethibond, Panacryl, or Orthocord sutures. Used by surgeons in OR settings for open or arthroscopic procedures. Device provides mechanical fixation of soft tissue to bone; facilitates healing during postoperative immobilization period.

Clinical Evidence

Bench testing only.

Technological Characteristics

Absorbable suture anchor; material: Poly (L-lactide) (PLA) polymer. Single-piece anchor design. Preloaded with suture (Ethibond, Panacryl, or Orthocord). Disposable inserter assembly.

Indications for Use

Indicated for soft tissue to bone fixation in shoulder, knee, ankle, foot, and elbow procedures requiring postoperative immobilization. Applicable to both open and arthroscopic surgical approaches.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K062170 page 142 # 510(k) SUMMARY AUG 2 2 2006 # Bioknotless Plus Anchor / Lupine Plus Anchor | Submitter's Name and<br>Address: | DePuy Mitek<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767 | |----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Kristine Christo<br>Senior Regulatory Affairs Specialist<br>DePuy Mitek<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767<br>Telephone: 508-828-3359<br>Facsimile: 508-977-6955<br>e-mail: KChristo@Dpyus.jnj.com | | Name of Medical Device | Classification Name: Screw, Fixation, Bone Staple<br>Common/Usual Name: Appliance for reconstruction of soft<br>tissue to bone<br>Proprietary Name: Bioknotless Anchor / Lupine Anchor | | Substantial Equivalence | Bioknotless Plus Anchor is substantially equivalent to:<br>Bioknotless Anchor, K002639, manufactured by DePuy Mitek.<br><br>Lupine Plus Anchor is substantially equivalent to:<br>Lupine Anchor, K050771, manufactured by DePuy Mitek | | Device Classification | Bone anchors/screws are classified by the FDA as Class II Medical<br>Devices under the generic category of Single/Multiple component<br>metallic bone fixation appliances and accessories.<br><br>Bioknotless Anchor / Lupine Anchor Systems carry FDA product code<br>MAI and is classified as single / multiple component metallic bone<br>fixation appliances and accessories soft tissue fastener under 21 CFR<br>888.3030. | | Device Description | Bioknotless Plus Anchor / Lupine Plus are a preloaded, absorbable | Bioknotless Anchor Plus / Lupine Anchor Plus {1}------------------------------------------------ K062170 disposable suture anchors/ inserters assembly for soft tissue repair to bone in the shoulder, knee, ankle, foot, and elbow. The absorbable polylactic acid (PLA) anchor is an identical anchor as that of the Bionotless Anchor / Lupine Anchor in design and configuration. The absorbable anchor is a one piece suture anchor constructed of molded Poly (L-lactide) polymer. The anchor systems may be sold with Ethibond Suture (NDA 17-804 and 17-809), Panacryl Suture (K964345), or Orthocord Suture (K040004 and K043298). The Bioknotless Anchor is indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization as follows: Shoulder, knee, ankle, foot, elbow The Lupine Anchor is indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization as follows: OPEN PROCEDURES SHOULDER: 1. Bankart repair,2. SLAP lesion repair 3. Rotator cuff repair, 4a. Capsule shift/capsulo-labral reconstruction, at the anterior glenoid rim site, 4b. Capsule shift/capsulo-labral reconstruction at the lesser tuberosity of the humerus, 5. Biceps tenodesis, 6. Acromio-clavicular separation, 7. Deltoid repair ELBOW: 1. Biceps tendon reattachment, 2. Tennis elbow repair ANKLE: 1. Achilles tendon repair/reconstruction, 2. Lateral stabilization 3. Medial stabilization at the medial talus site, Foot: Hallux Valgus reconstruction, 4. Midfoot reconstruction KNEE: 1. Medial collateral ligament repair, 2. Lateral collateral ligament repair 3. Joint capsule closure to anterior proximal tibia., 4. Posterior oblique ligament or joint capsule to tibia repair, 5. Extra capsular reconstruction / ITB tenodesis, 6. Patellar ligament and tendon avulsion repairs. ARTHROSCOPIC PROCEDURES SHOULDER:1. Bankart repair, 2. SLAP lesion repair, 3. Rotator cuff repair, 4. Capsule shift repair (glenoid rim) Safety and Performance The determination of substantial equivalence for this device was based on a detailed device description, and conformance to consensus and voluntary standards. Bench testing was performed demonstrating that the Bioknotless Anchor Plus / Lupine Anchor Plus met predetermined acceptance criteria. > Based on the indications for use, technological characteristics, and comparison to predicate devices, the Bioknotless Anchor Plus and Lupjne Anchor Plus has been shown to be substantially equivalent to predicate devices under the Federal Food, Drug and Cosmetic Act. Indications for Use {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # AUG 2 2 2006 DePuy Mitek A Johnson & Johnson Company % Ms. Kristine Christo Senior Regulatory Affairs Specialist 325 Paramount Drive Raynham, Massachusetts Re: K062170 Trade/Device Name: Bioknotless Plus Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and acessories Regulatory Class: Class II Product Code: HWC, JDR, MAI Dated: July 28, 2006 Received: July 31, 2006 Dear Ms. Christo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Kristine Christo ﻟ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or 240-276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Paulare Buell Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ page 192 # INDICATIONS FOR USE 510(k) Number (if known): K062 | 70 #### Device Names: Bioknotless Plus Anchor Indications for Use: The Bioknotless Plus Anchor with is indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization as follows: ### OPEN PROCEDURES SHOULDER 1. Bankart repair - 2. SLAP lesion repair 3. Rotator cuff repair - 4a. Capsule shift/capsulo-labral reconstruction, at the anterior glenoid rim site - 4b. Capsule shift/capsulo-labral reconstruction at the lesser tuberosity of the humerus - 5. Biceps tenodesis 6. Acromio-clavicular separation 7. Deltoid repair #### ELBOW - 1. Biceps tendon reattachment - 2. Tennis elbow repair #### ANKLE - 1. Achilles tendon repair/reconstruction 2. Lateral stabilization 3. Medial stabilization at the medial talus site - Foot: Hallux Valgus reconstruction 4. Midfoot reconstruction #### KNEE - 1. Medial collateral ligament repair - 2. Lateral collateral ligament repair - 3. Joint capsule closure to anterior proximal tibia - 4. Posterior oblique ligament or joint capsule to tibia repair - 5. Extra capsular reconstruction / ITB tenodesis - 6. Patellar ligament and tendon avulsion repairs. #### ARTHROSCOPIC PROCEDURES SHOULDER - 1. Bankart repair - 2. SLAP lesion repair - 3. Rotator cuff repair - 4. Capsule shift repair (glenoid rim) Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Ey Distismon (0) Say Page 1 of Division of Ge and Neurological Devices 510(k) Number K062170 {5}------------------------------------------------ ### INDICATIONS FOR USE 510(k) Number (if known): Kobo2170 Device Names: Lupine Plus Anchor INDICATIONS FOR USE: The Lupines Plus The Lupine Loop Anchor System is indicated for use in soft tissue to bone fixation in association with adequate postoperative immobilization as follows: #### OPEN PROCEDURES SHOULDER - 1. Bankart repair 2. SLAP lesion repair 3. Rotator cuff repair 4a. Capsule shift/capsulo-labral reconstruction, at the anterior glenoid rim site 4b. Capsule shift/capsulo-labral reconstruction at the lesser tuberosity of the humerus 5. Biceps tenodesis 6. Acromio-clavicular separation 7. Deltoid repair #### ELBOW 1. Biceps tendon reattachment 2. Tennis elbow repair ANKLE 1. Achilles tendon repair/reconstruction 2. Lateral stabilization 3. Medial stabilization at the medial talus site Foot: Hallux Valgus reconstruction 4. Midfoot reconstruction KNEE 1. Medial collateral ligament repair 2. Lateral collateral ligament repair 3. Joint capsule closure to anterior proximal tibia 4. Posterior oblique ligament or joint capsule to tibia repair 5. Extra capsular reconstruction / ITB tenodesis 6. Patellar ligament and tendon avulsion repairs. ARTHROSCOPIC PROCEDURES SHOULDER 1. Bankart repair 2. SLAP lesion repair 2. SLAP lesion repair 3. Rotator cuff repair 3. Rotator cuff repair 4. Capsule shift repair (glenoid rim) | Prescription Use | √ | |-----------------------------|---| | (Part 21 CFR 801 Subpart D) | | AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of __________ | | <div> <p>(Division Sign-Off)</p> </div> | |--|-----------------------------------------| |--|-----------------------------------------| 510(k) Premarket Notification: Special Bioknotless Anchor Plus / Lupine Anchor Plus Division of General, Restorative, and Neurological Devices **510(k) Number** K062170
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...