TIGHTROPE ACROMIOCLAVICULAR (AC) DEVICE, TITANIUM, STAINLESS STEEL , MODELS AR-2257, AR-TBD

K052776 · Arthrex, Inc. · HTN · Dec 13, 2005 · Orthopedic

Device Facts

Record IDK052776
Device NameTIGHTROPE ACROMIOCLAVICULAR (AC) DEVICE, TITANIUM, STAINLESS STEEL , MODELS AR-2257, AR-TBD
ApplicantArthrex, Inc.
Product CodeHTN · Orthopedic
Decision DateDec 13, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The TightRope™ Acromioclavicular (AC) Device is intended as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting. Specifically, the Arthrex TightRope Acromioclavicular (AC) Device is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions.

Device Story

TightRope AC Device consists of two metal buttons (stainless steel or titanium) pre-threaded with FiberWire suture; includes pull-through sutures. Device functions as mechanical fixation system for bone fragments or ligament disruptions. Used by surgeons in clinical/OR settings to provide stabilization during healing process. Surgeon implants buttons and secures sutures to approximate bone or ligament structures. Provides alternative to traditional screw fixation; benefits patient by facilitating stabilization in areas where screws are contraindicated or as adjunct to other fixation systems.

Clinical Evidence

No clinical data provided; substantial equivalence based on design and intended use comparison to predicate devices.

Technological Characteristics

Materials: Stainless steel or titanium buttons, FiberWire suture. Form factor: Two-button system with pre-threaded suture. Energy source: None (mechanical). Sterilization: Not specified.

Indications for Use

Indicated for patients requiring fixation during healing of syndesmotic trauma, including acromioclavicular separations caused by coracoclavicular ligament disruptions, and as an adjunct for metaphyseal/periarticular small bone fragment fracture repair where screws are not indicated.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEC 1 3 2005 K05-2776 page 192 ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS TightRope™ Acromioclavicular (AC) Device | NAME OF SPONSOR: | Arthrex, Inc.<br>1370 Creekside Boulevard<br>Naples, Florida 34108-1945 | |------------------|---------------------------------------------------------------------------------------------------------------------| | 510(K) CONTACT: | Sally Foust, RAC<br>Regulatory Affairs Project Manager<br>Telephone: (239) 643-5553 ext. 125<br>FAX: (239) 598-5539 | TRADE NAME: Device TightRope™ Acromioclavicular (AC) COMMON NAME: Button/Suture ## DEVICE PRODUCT CODE/CLASSIFICATION: HTN: Single/multiple component metallic bone fixation appliances and accessories: 21 CFR 888.3030 ### PREDICATE DEVICES K043248: TightRope Syndesmosis Device (Arthrex, Inc.) K023294: Bosworth Coraco-Clavicular Screw and Washer (Howmedica Osteonics) K041356: Tenodesis Family (Arthrex, Inc.) K003227: Bio-Absorbable Corkscrew (Arthrex, Inc.) pre-1976, Kirschner (K) Wire (Synthes) # DEVICE DESCRIPTION AND INTENDED USE The TightRope™ Acromioclavicular (AC) Device is designed as two differently sized metal buttons, both stainless steel or both titanium, and FiberWire ™ suture. The buttons are pre-threaded with FiberWire suture, looped twice through the buttonholes. Pull-through FiberWire sutures are also looped through each button. The TightRope™ Acromioclavicular (AC) Device is intended as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting. {1}------------------------------------------------ Specifically, the Arthrex TightRope Acromioclavicular (AC) Device is intended to Opocilion in the hithing the healing process following a syndesmotic trauma, such provide intation of acromioclavicular separations due to coracoclavicular ligament disruptions. # SUBSTANTIALLY EQUIVALENCE Arthrex has determined that the TightRope™ Acromioclavicular (AC) Device is substantially equivalent to the predicate devices where basic features and intended uses are the same. Any design differences between the Arthrex TightRope™ Acromioclavicular (AC) Device when compared to predicate devices used in the standard medical practice for the treatment of acromioclavicular separations due to coracoclavicular disruptions are considered minor and do not raise any questions concerning safety and effectiveness. Any differences have been found to have no apparent effect on the performance, function, or intended use of the device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three heads, representing the department's focus on health, services, and people. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle. The seal is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Sally Foust, RAC Regulatory Affairs Project Manager Arthrex, Inc. 1370 Creekside Boulevard Naples, Florida 34108-1945 DEC 1 3 2005 Re: K052776 Trade/Device Name: TightRope™ Acromioclavicular (AC) Device Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HTN Dated: September 28, 2005 Received: October 3, 2005 Dear Ms. Foust: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 – Sally Foust, RAC This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N. Mellon Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Indications for Use 510(k) Number (if known): _ K05276 Device Name TightRope™ Acromioclavicular (AC) Device ### Indications for Use: The TightRope™ Acromioclavicular (AC) Device is intended as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting. Specifically, the Arthrex TightRope Acromioclavicular (AC) Device is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions. Prescription Use (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) …………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………………… Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of signature on Sion-Gill) Division of General, Restorative, and Neurological Devices 510(k) Number_152336
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...